SIREMOBIL Compact L
XP
Maintenance Instructions
System
© Siemens, 2009
© Siemens, 2009
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.
2009
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
Copyright
“© Siemens, 2009“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA
and/or Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.
0 Table of Contents
1 _______ General Information______________________________________________ 5
Prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Systems equipped with an external I.I. laser light localizer. . . . . . . . . . . . . . . . . . . . . 6
For systems equipped with an integrated I.I. laser light localizer . . . . . . . . . . . . . . . . 6
When an integrated laser light localizer is present (close to tube) . . . . . . . . . . . . . . . 6
Required tools, measurement devices, and auxiliary materials . . . . . . . . . . . . . . . . . . . . 7
Spare parts which may be needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Text emphasis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
General safety notes (in existing documents) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
General electrical safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Radiation safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Mechanical safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Safety note: Risk of infection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Integrated I.I. laser light localizer option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
I.I. laser light localizer option or single tank laser light localizer option . . . . . . . . . . . 13
Explanation of Abbreviations in the Maintenance Protocol . . . . . . . . . . . . . . . . . . . . . . 14
Maintenance interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Inspection of surroundings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Power sockets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Mechanical safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
External I.I. laser light localizer (if present) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Integrated I.I. laser light localizer (if present) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Integrated laser light localizer, close to tube (if present) . . . . . . . . . . . . . . . . . . . . . . 19
Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Voltage discharge rubber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Image quality (IQ) quick test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Checks in accordance with IEC 62353. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Operating values, inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
1.1 Prerequisites
The prerequisites described in Chapter 1 of the Troubleshooting Guide apply to mainte-
nance as well.
1.2.2 For systems equipped with an integrated I.I. laser light localizer
NOTE The items listed are detailed in the SPC (Spare Parts Catalog)
if there is no other reference. Exception: items marked with
an asterisk ( * ).
Bei einer Gefahr, die bei Nichtvermeidung zum Tod oder zu einer
schweren Körperverletzung führen kann.
1.6.7 I.I. laser light localizer option or single tank laser light localizer option
Abbrevia- Explanation
tion
SI Safety Inspection
SIE Electrical safety
SIM Mechanical safety
PM Preventive Maintenance
PMP Periodic preventive maintenance
PMA Preventive maintenance adjustments
PMF Preventive check of operating values and functions
Q System Quality, Image Quality
QIQ Image Quality
QSQ System quality
SW Software Maintenance
The work steps identified by these abbreviations are listed as check-off points in the main-
tenance protocol and should be checked off upon completion.
• Now unlock the emergency stop switch. The lifting column can be moved again.
3.2.4 Monitor
SIM Monitor(s)
• Check monitor(s) for damage.
• Check monitor(s) for proper attachment to the monitor trolley.
3.2.5 Labels
SIM Caution labels
• Ensure that all required caution labels are attached and in good condition.
- Replace any illegible labels.
SIM ID labels
• Ensure that all required ID labels are attached and in good condition.
- Replace any illegible labels.
3.4 Operation
SIE Check the operating functions.
• Check whether the operating icons and button labels are clearly legible and recogniz-
able.
• Check whether the light indicators of the control elements and displays are properly
functioning.
• Check all system operating functions.
X NOTE The IQ quick test is part of the safety inspection, but should
only be performed once maintenance has been completed.
See (Final result/quality inspection and maintenance / p. 32).
NOTE The checks in accordance with IEC 62353 are part of the
safety inspection but can only be performed at the end of
maintenance. See (Final Work Steps / p. 33).
4.1 Maintenance
PMP Cleaning the system
• Clean the entire system:
- Visible cables
- Exterior surfaces
- Contact surface of wheels
- Interior space
PMP System ventilation
• Clean the ventilation slots of the SIREMOBIL stand and monitor trolley.
PMP Memoskop ventilation
• Clean the ventilation slots on the MEMOSKOP.
• Check that the fan is functioning properly.
NOTE Perform this check with the lowest possible kV values so that
the front edges of the slot diaphragm blades are effectively
visualized.
PMP Upkeep
• Entire system: Touch up any paint damage as needed.
• Fill out the protocol "Record of checks in accordance with IEC 62353" completely and
then save it with the existing protocols in the system binder.