Neofuge 15

High Speed Refrigerated Bench-top


Centrifuge

User Manual
(Version20100408)

Heal Force Bio-Meditech Holdings Group


Preface
Without the permission of Heal Force, no party is allowed to reprint or diffuse any
contents about this manual, including photographs and audio-video products.

The operator may copy some chapters of this manual only for interior use, such as
chapters on how to manage incidents. These chapters have been clearly marked in the
contents of this manual.

If any damage to the instruments caused by users due to not obedience by the
“Operating Environment of Instruments”, Heal Force will not be responsible for this.

Heal Force has the right to modify the operating manual in any time without any proper
notices.

Operating environment of instruments

In order to extend the machine’s service life, we shall pay attention to the
following factors that may damage the machine:
- Chemical action;
- Effects of environments, including natural ultraviolet radiation
- Erosion and abrasion of protective cover parts and other safety parts
* Used at home;
* Altitude: ≤2000 m
* Ambient temperature: +5℃~+35℃;
* Relative humidity: ≤80%;
* Operating range of voltage: AC230 V,50/60 Hz;
* Adequate ventilation facilities must be installed inside;
* There are no vibration and airflow to influence the performance around;
* There are no electric dust and volatile gas and corrosive gas to exist in the ambient air.

Notice
* Please read this manual carefully if you use this instrument for the first time!
* Neofuge 15 High Speed Bench-top Centrifuge should be operated only by the trained
and authorized staff;
* The maintenance of the instrument should be carried out by the agents authorized by
Heal Force or Heal Force;
* The following materials should be forbidden:
- Flammable and explosive substances;
- Strong chemical materials;
- Tonic or radioactive substances, or pathogenic microorganism;
* The system of Neofuge 15 High Speed Bench-top Centrifuge can be maintained only by
qualified servicemen using appropriate tools.
* If any problem not mentioned in this manual occurs, please contact with agents
authorized by Heal Force or Heal Force for proper processing methods.
* Please use accessories provided by Heal Force. If other accessories are used, Heal Force
should not be responsible for bad consequences. However, the user may apply to Heal
Force to check whether the accessories are accordance with the requirements of Heal
Force.
* Neofuge 15 High Speed Bench-top Bench-top Centrifuge should be checked and
maintained at stated intervals.

Warning sign

Attention: Please read Operating and Technical Instructions


carefully before using!

Attention: High voltage! Danger!

Signal Ground!

DO not lift by the front panel!


The label of caution in the following accessories must be stuck on the
centrifuge that the user can see it easily.
Table of Contents

I. Safety clauses ................................................................................................................... 5

1.1 Caution of operation.................................................................................... 6


1.2 Relevant reference standards:.................................................................................. 6
II. Introduction of Neofuge 15 High Speed Bench-top Centrifuge............................... 7
2.1 Appearance of Neofuge 15 High Speed Bench-top Centrifuge.............................. 7
2.2 General ....................................................................................................................... 8
2.3 Device structure ......................................................................................................... 8
2.4 Safety protection device ............................................................................................ 9
2.6 Settling the centrifuge..................................................................................... 10
2.7 Configuration......................................................................................................... 10
III. Available optional accessories ................................................................................. 11
A List of Rotor Accessories (test tube rack) .............................................................. 11
IV. Preparations before using ....................................................................................... 13
4.1 Transportation and installation............................................................................ 13
4.2 Proper place for the centrifuge............................................................................. 13
4.3 Fixed placement of equipment ............................................................................. 13
4.4 Proper connection of electric supply.................................................................... 14
V. Operating instructions ................................................................................................ 14
5.1 Control panel and display interface..................................................................... 14
5.2 Starting ................................................................................................................... 15
5.3 Opening .................................................................................................................. 16
5.4 Closing .................................................................................................................... 16
5.5 Installation of rotor ............................................................................................... 16
5.6 Maintenance of lock sleeve and supporting shaft of rotor................................. 18
5.7 The calculation of rotor load ................................................................................ 18
5.8 Samples filling into the centrifugal containers ................................................... 19
5.9 Safety use of rotor.................................................................................................. 19
5.10 Parameters setting ............................................................................................... 19
VI. Maintenance ............................................................................................................. 24
6.1 Cleaning ............................................................................................................. 24
6.2 Maintenances by HEAL FORCE .................................................................. 25
6.3 Warranty (contract maintenance) qualification .......................................... 25
VII. Trouble shooting ................................................................................................... 26
7.1 To open lid in case of emergency .......................................................................... 26
7.2 Warning message of trouble............................................................................. 26
7.3 Safety protection.................................................................................................... 29
VIII. Technical data ....................................................................................................... 30
I. Safety clauses
Neofuge 15 High Speed Bench-top Centrifuge is designed according to the present
technical standards:
Machinery Industrial Standards of the People’s Republic of China JB/T 5519-91
High-speed Refrigeration Centrifuge,
Safety standards:
GB/9706.1-1995 Medical Electric Equipment, Part 1:Universal safety requirements,
GB4793.1-1995 Measurement, controlling the laboratory to use the security request for
the electric equipment First part: Current requirements,
GB4793.7-2001 Measurement, controlling the laboratory to use the security of the
electric equipment Use the special request for the centrifuge in the laboratory ,
GB191-2000 Graphic presentation sign of warehousing and transportation of the
packaging,
GB/T14710-1993 The electric equipment environment of medical use requires and test
method,
GB9969.1-1998 Industrial products operation instructions General provisions, so it
must be used according to design requirements;
When using Neofuge 15 High Speed Bench-top Centrifuge according to the following
improper or inappropriate methods, equipment may be destroyed or men be injured:
z To use not accordance with the design requirements;
z The operator and maintainer have not been trained;
z The user has made improper modification to the design without authorization;
z Not notice or understand regulations for safety use.

Persons who use or maintain Neofuge 15 High Speed Bench-top Centrifuge


should read and understand related usage and safety regulations in this
Operating and Technical Instructions.
The following regulations should be obeyed to avoid accidents:

Operating and Technical Instructions is the integral part of the Neofuge 15 High Speed
Bench-top Centrifuge, which should be put beside “Neofuge 15 High Speed Bench-top
Centrifuge” for the reference of operators.
Neofuge 15 High Speed Bench-top Centrifuge is designed to separate biologic cells, organism and
bacteria. The density of the sample won’t exceed 1.2g/cm3 at the maximum speed.
When the Neofuge 15 High Speed Bench-top Centrifuge is in operation (when the
sample is being separated or rotors are running), operators and dangerous substances should
be absent within 30 cm of the centrifuge. The ventilation hole of the centrifuge can not be
clogged by substances.
If the following safety precautions are not obeyed when using Neofuge 15 High Speed
Bench-top Centrifuge, operators or other persons may be hurt or the centrifuge and samples
to be separated may be damaged:
z The centrifuge is not corrosion proof or implosion proof, so it must be used in the
environment free of corrosion or implosion;
z The following materials should be forbidden:
- Flammable and explosive substances;
- Strong chemical materials;
- Tonic or radioactive substances, or pathogenic microorganism;
z It is necessary to take effective sealing measures when performing segregation of
poisonous substances and pathogenic microorganism cell. Furthermore, effective
sterilizing measures should be taken after the application. For detail, please see relevant
descriptions in Maintenance.
z Corrosive substances may do damage to materials inside the centrifuge and the rotors, so
they should be put inside the sealed containers while separating.

1.1 Caution of operation


z Before running (when separating samples), appropriate rotors must be fixed firmly;
z In operation (when the rotor is running) or the centrifuge has stopped (the rotor is still
running), opening the lid by hand is strongly forbidden;
z Accessories of Neofuge 15 High Speed Bench-top Centrifuge must be provided by Heal
Force. Some common parts, such as glass and plastic containers for separating, should
also be approved to be qualified products and in be accordance with the requirements of
the highest speed and maximum centrifugal force of related rotors;
z Do not use the centrifuge or separate samples is forbidden when the lid is open;
z Don’t turn on the power switch of the centrifuge equipment when disassembling the
front panel of the centrifuge;
z Changing mechanical parts and electronic devices should be carried out by related
persons appointed by Heal Force;
z When using the centrifuge, the user must use proper rotors according to load. Overload is
forbidden;
z The rotor should be checked regularly. If the rotor is eroded or damaged, it should not be
used any more;
z Maintenance should be carried out according to “Clean and sterilize” periodically.

1.2 Relevant reference standards:


z National standards: Machinery Industrial Standards of the People’s Republic of China
JB/T 5519-91 High Speed Refrigeration Centrifuge;
z GB/9706.1 Medical Electric Equipment, Part 1:Universal safety requirements;
z GB4793.1-1995 Measurement, controlling the laboratory to use the security request for
the electric equipment First part: Current requirements;
z GB4793.7-2001 Measurement, controlling the laboratory to use the security of the
electric equipment Use the special request for the centrifuge in the laboratory;
z GB191-2000 Graphic presentation sign of warehousing and transportation of the
packaging;
z GB/T14710-1993 The electric equipment environment of medical use requires and test
method;
z GB9969.1-1998 Industrial products operation instructions General provisions;
z Meet enterprise standards: Q/TEU C42-2005 Neofuge 15 High Speed Bench-top
Centrifuge.
II. Introduction of Neofuge 15 High Speed Bench-top
Centrifuge
2.1 Appearance of Neofuge 15 High Speed Bench-top Centrifuge

Figure 1: Front and side view of Neofuge 15 High Speed Bench-top Centrifuge

Figure 2: Back side view of Neofuge 15 High Speed Bench-top Centrifuge

Description of figure one and figure two:


1. Front cover casing 2. Lid 3. Top observation hole at the top 4. Cover casing at the
top 5 Foot 6. Control panel and display window 7. Front ventilation hole 8.
Power jack/switch
2.2 General
Neofuge 15 High Speed Bench-top Centrifuge with multi-functional protection
introduces advanced technology such as speed adjustment via frequency control,
microcomputer control, touch button, digital setup, digital LCD. Six kinds of rotors are
available (For details, see “Table 1: Type of rotor and technical parameters”). This
centrifuge is used to separate biologic cells, organism and bacteria.

2.3 Device structure


This device is composed of many components, such as lid system, chamber system,
drive system, rotor system, base system, electric power system, control system, display
system and warning system.
2.3.1 Lid system is composed of the lid, lid hinge, gas springs, lid lock, lid warning, and
opening hole and so on. The lid hinge is at the back of the machine frame and the lid lock is
in front of the machine frame. Only when the lid lock is locked firmly, may the centrifuge run;
otherwise the warning system will run (The lights is on and the buzzer sounds) and the device
can’t run.
To open the lids, you only need to press “ ” (opening key) on the control panel. When
the lid rise to certain height, the lid hinge and springs can support the lid.
If power cut occurs or the opening key is useless, the sample can be taken out by the
following steps: pry out guy rope button, which connects with the guy rope, from emergency
lid opening hole at the bottom. Pull out it, and then the lid can be opened by hand.
When rotors are running and power supply is on, to
open lid by hand is forbidden!

2.3.2 The chamber system is composed of the internal stainless steel chamber, rubber seal
rings. The chamber may provide a working condition of steady temperature.
2.3.3 This device adopts frequency conversion electric machinery to drive the rotor with load
sample. It applies cone connection and which is useful for connection of rotors and rotary
shafts.
2.3.4 Rotor system is composed of various kinds of angle rotors and horizontal rotors (For
details, see Table 1: Type of rotor and technical parameter), and centrifugal tube etc. The
rotor rotates at a certain speed to create a relative centrifugal force field to separate the
samples.
As the centrifugal force is several thousand times more than g (acceleration of earth gravity),
it is important to use and maintain the rotor safely and carefully!
2.3.5 The base system is composed of the machine frame, protection steel ring, motherboard,
top cover and rubber supporting legs and so on.
2.3.6 The power supply system includes power supply jack/switch, power switch, which is
used to supply electricity.
2.3.7 The control system includes identification of the rotor code, setting of the rotary speed
and centrifugal force, running time, temperature, choice of accelerating, control of display
system and warning system of the complete machine and so on. The control system is
composed of the main control panel, checking console for speed, package board of sensors
(identification of rotors and checking of balance quality of rotors). To open the cover in front
of the device is strictly forbidden!
2.3.8 The display system is composed of the display board and PVC keyboard panel (control
panel). It is the interactive media. It can show every setting parameter synchronously, and can
track the actual change. In addition, it can display various kinds of troubles and warnings.
2.3.9 The warning system consists of over speed, electrical motor overheat, lid failure and
temperature control failure. When the troubles mentioned above occur, the system will give
an alarm. At this moment, the red warning light on the control panel glimmers, the buzzer
sounds the alarm. The machine is unable to start (starting forbidden). The machine stops
automatically, until the trouble is treated. Then the machine can be restarted. Note: you may
press “ESC” on the control panel to cancel the warning tone sounded by the buzzer.

2.4 Safety protection device


Neofuge 15 High Speed Bench-top Centrifuge has a series of safety protection devices:
z The machine frame is made of steel plate and protection steel ring is made of steel
casting. The internal bladder of the inner chamber is stainless. The front frame is made of
plastics;
z The structure of the lid is explosion proof. There is a observation window in the center of
the top, the front locking device.
Only when is the centrifuge on and does the rotor stop, it is allowed to press “ ” to
open the lid. Only when is the centrifuge lid locked, you may start the centrifuge to work!
z Overspeed stop
z The warning alarm is given when the rotary speed exceeds the setting speed of 200 r/min.
The rotor stops automatically if the speed exceeds the maximum rated speed of 250r/min.
Wait until the rotor stops completely, then open the lid. Restart when the trouble is
solved;
z Opening forbidden in operation
“ ”is pressed by accident in operation, the device won’t respond;
z Opening in case of emergency
If the electricity is off or there is something wrong with the rotor abruptly, you may open
the lid by hand when it is useless to open it by pressing (see “Trouble shooting”).

Don’t play with safety protection device at will!


2.5 Technical specifications
2.5.1 Technical data
Rated/maximum speed 15,000/16,000min(fully loaded)
Maximum capacity 4×100 g
Maximum centrifugal acceleration 24,900×g (maximum rated
speed)
Rated Power 470VA
Drive system Frequency conversion electric
machinery
Noise ≤65dB(A)
2.5.2 Technical parameters
Speed control range 300~16,000r/min
Speed control precision ±1%(2000~16000r/min)
Time setting ranges 1min~9h and 59mins
2.6 Settling the centrifuge
2.6.1 This device should be put on the swing-out table-board of the enough rigidity, and be
free from vibration and direct irradiation of the heat source and sunlight.
2.6.2 There should be space of 10cm~15cm around the device for the purpose of
ventilation.
2.6.3 After mounting, to adjust level and make the four shores support at the table-board.
2.6.4 Range of application for operational power supply: AC230V, 50/60 Hz.

Grounding for the device should be made, and the


ground line of the device should be connected to that of
the electric network!
Man-made power cut is forbidden during the rotor’s operation
period. Otherwise it will cause damage to control circuit.
2.7 Configuration
The accessories are prepared with the device .
(For details, see “Packing list”).

You should check the contents when receiving products


and unpacking. If there is something wrong, please
contact to Heal Force or agents authorized by HealForce.
III. Available optional accessories
Our company will provide users with rotors of various specifications which are
corresponding to Neofuge 15 High Speed Bench-top Centrifuge. You can select your desired
rotors at your discretion when purchasing the Neofuge 15 High Speed Bench-top Centrifuge
(for detail, please see Table 1-Type of Rotors and Technical Parameters).
If you want these optional accessories, please consult the manufacturer or Heal Force or
agents authorized by Heal Force for technical indicators and purchase No..

Table 1: Type of rotor and technical parameters


Angle of Maximum
Rotor Rated Maximum
the angle centrifugal
Type of rotor numbe speed capacity
rotor acceleration
r (r/min) (ml)
(°) (×g)
Angle rotor #1501 45
16,000 24×1.5/2.0 24,900
F15.15.02
Angle rotor #1505 40
13,000 24×1.5 16,060
F15.13.01
Angle rotor #1506 40
13,000 32×0.2 16,060
F15.13.002
Angle rotor #1503 45
8,500 6×50 10,016
F15.08.50
Angle rotor #1504 45
8,500 6×50 10,016
F15.08.50F
Horizontal rotor S15.04.100 #1502 4,000 4×100 2,612

A List of Rotor Accessories (test tube rack)


Rema
No. Name Order No. Qty Color Specification
rk
1501rotor-1*0.5ml test Optio
1 ADP17.02-001 24 Yellow Ø8*32
tube rack nal
1501rotor-1*0.2ml test Optio
2 ADP17.02-002 24 Blue Ø6.5*22
tube rack nal
1502 rotor-test tube rack Optio
3 SP15.04.100.1-00 4 Grey Ø12*104
10x1.5/2ml nal
1502rotor-test tube rack Optio
4 SP15.04.100.5-00 4 Green Ø12*104
3x5/7ml nal
1502 rotor-test tube rack Orang Ø12*104 Optio
5 SP15.04.100.7-00 4
5x7ml e nal
1502 rotor-test tube rack Ø17*104 Optio
6 SP15.04.100.15-00 4 White
3x15ml nal
1502 rotor-test tube rack SP15.04.100.C15-1 Ø17*120 Optio
7 4 Green
C15ml 0 nal
1502 rotor-test tube rack Orang Ø 25*115 Optio
8 SP15.04.100.25-10 4
25ml e nal
1502 rotor-test tube rack Ø36*110 Optio
9 SP15.04.100.50-10 4 blue
50ml nal
1502 rotor-test tube rack SP15.04.100.C50-1 Ø29*116 Optio
10 4 Yellow
C50ml 0 nal
11 1502 rotor-test tube rack SP15.04.100.S50-1 4 White Ø29*114 Optio
S50ml 0 nal
6 Optio
12 Green Ø29*115
1x50ml test tube rack ADP15.94-001 nal
6 Orang Optio
13 Ø25.5*96
1x38ml test tube rack ADP15.94-002 e nal
6 Optio
14 Brown Ø18*114
1x16ml test tube rack ADP15.94-003 nal
6 Optio
15 Yellow Ø16*112
2x12ml test tube rack ADP15.94-004 nal
6 Optio
16 Grey Ø13*104
3x6.5ml test tube rack ADP15.94-005 nal
6 Optio
17 White Ø11*85
4x3.5ml test tube rack ADP15.94-006 nal
4x1.5ml_Micro test tube 6 Optio
18 Black Ø11*(>41)
rack ADP15.94-007 nal

Rotor figures:

Figure 3: Rotor number~#1501

Figure 4: Rotor number~#1502


Figure 5: Rotor number~#1503

IV. Preparations before using


4.1 Transportation and installation
Neofuge 15 High Speed Bench-top Centrifuge is transported in special-purpose wooden case with
protection materials for buffering inside. After package is unpacked, the protection materials for buffering
should be taken out.

The net weight of the device is about 58 Kg, so it is required to


lift the device on both sides (the left and right) to keep its balance.
The cover of the electric apparatus in the front should not be
pressed!
Be straight and no vibration when conveying!

Please use special packing cases in long-distance transporting. Be straight! Handle with
care!

4.2 Proper place for the centrifuge


Neofuge 15 High Speed Bench-top Centrifuge is for indoor use, and the place should meet
the following:
z A safety clearance of 30cm must be maintained around the centrifuge. When running,
there should be no maleficent substance or related person;
z The adapter or table should be solid to avoid vibration; if mobile racks or trolly are used,
they should be with locking devices to ensure the safe operation of the centrifuge;
z If the centrifuge is placed beside the wall or in the corner, there should be space of 10cm
between the back and the wall and 15cm between the left side and right side and the wall
to ensure good ventilation and heat emission.
z The centrifuge should be put far away from windows to avoid direct irradiation of the
heat source and the sunlight.
z The four supporting legs of the centrifuge should stand evenly on the table-board, which
is swing-out.
z The centrifuge should be put in the clean room with the constant ambient temperature
between 5℃ and 35℃ and ambient humidity less than 80%.

4.3 Fixed placement of equipment


After the centrifuge has been placed, don’t move it frequently. After being moved, be
sure the centrifuge is swing-out, four supporting legs of which stand evenly on the
table-board. There should be no vibration.

To avoid damage to rotor axis, please take out the rotor


when moving the centrifuge.

4.4 Proper connection of electric supply


A power socket shall be provided for power line of centrifuge separately. The earthing of
the supply socket connected to the centrifuge must be in good condition. The power wires of
the centrifuge should be in accordance with the safety standards of the nation and local area.
And the voltage and frequency of the power supply should be in accordance with the
requirements stated in the technical description or specifications labeled on the nameplate of
the centrifuge. The power jack/switch is at the back side of the machine on its left hand.
Please make use of the power line attached with the machine, and connect it properly with
the power socket of the machine before switching it into conduction with network power. “|”
stands for off and “O” for break.
As the rotor is packed inside the centrifugal cavity,
please do not press “►” and “ ” when the machine is
switched on for the first time.

V. Operating instructions
5.1 Control panel and display interface

Speed/F Time

S/F

Figure 6:Sketch map of control panel


Description of figure 6:Brief Introduction to Control Panel
1. Data display window 2. Value increment/decrement key 3. Alarm indicator lamp 4. Key of
opening 5. “Quick run” mode
6. Key of stopping 7. Start-up key 8. Rotary speed/centrifugal force switch key
4 3

Type Time Info


2
0000
1
00:00:00
0 0 0 0 0 rpm
Speed 0 0 0 0 0 rcf 1

Stand By
5 6
Figure 7:Sketch map of “Main interface”

Description of figure 7:Brief Introduction to Main Interface


1. Accelerating and decelerating curve indication zone: The number indicates the serial No.
of the accelerating and decelerating curve;
2. Warning indication area of lid lock and running status: “ ” for lock status, “ ”for
opening status, “ ”for stopping status when the arrow is fixed, “ ”for running status
when the arrow cycles. “ ” for buzzer warning is off, “ ” for buzzer warning is on;
3. Time indication area: When the rotor is normal operation, the remaining running time
will be displayed. When the “■”is pressed or when the duration of runs is due, the
value of “00:00:00”will be displayed. The running time will be displayed when the
“►►”is pressed for temporary running;
4. Type of current rotor: The serial No. of current is “0000” when the rotor is absent, and
“9999” when the rotor can not be recognized.
5. Current rotary speed/ centrifugal force: display current and actual rotary speed value of
the rotor, “rpm”, ”00000” for no run; display current and actual centrifugal force (g) of
the rotor, “rcf” ,“00000” for no run;
6. Set indication area of parameter and running status: show “stand by” when the machine
is ready to work; show “Running” when the machine runs properly. Show relevant preset
value when the rotating speed, accelerating and decelerating curve are set. Show failure
reasons when warning.

5.2 Starting
Insert one end of the power wire into the supply socket at the bottom of the left side, the
other end connected to the independent socket for civil power supply. The voltage of supply
power applied is AC230V and frequency is 50/60Hz. After connecting, turn on the power
switch at the bottom of the left side, the buzzer chirps and the icon “Heal Force” is displayed
on display window. At the same time, self-check is carried out. When the self-check has been
completed, the main interface appears (see Fig. 7). Now is for the following operation. As the
rotor is packed inside the centrifugal cavity, please do not press “►” and “ ” when the
machine is switched on for the first time.
5.3 Opening
If the electricity is on and rotor doesn’t run, you may press “ ” (key of opening) on
the control panel to open the lid, the data display window appears “Lift lid”. Under the air
spring action, the lid will open automatically and spring to a certain height, and then it needs
to manually lift and open the lid. The inner chamber will appear in front of the user. The icon
on the main interface will show the tapered end “unlocked”. At this point, it is applicable to
take out the rotor and package if the machine is switched on for the first time.
Notice: The lid will spring to a certain height when it is opened. Therefore, heads or
other objects should not stay at the place above the lid. It is dangerous!
When the lid can’t open automatically, you may take out things inside the chamber by
hand if necessary, see “Trouble shooting”.

5.4 Closing
Press the lid down until front side lid hooks on the lid skip over the lid bolt and a ringing
sound is heard. The bottom of the lid hooks will touch travel switch. At this time, the icon on
the main interface shows that the tapered end is closed and the lid has been locked firmly.

While closing the lid, press the lid properly, do not push
powerfully in order not to damage the lid lock!
5.5 Installation of rotor

Inappropriate types of rotors and centrifuge tubes may


influence the performance, or even damage the
centrifuge.
Rotors used must be confirmed and designated by Heal Force to avoid troubles.
Various types and specification of rotors and centrifugal containers confirmed by Heal Force
are listed in this instruction. Rotors and containers of various specifications as designated in
this instruction are recommended (for details, see “Table 1: Types and specifications of rotors
and technical parameters”).
Please use special inner hexagon attached to the machine when installing rotors.

Steps to install rotors are as follows (see Fig. 8, 9, 10) :


z To turn on the electricity, wait until self-check is completed;
z To press “ ” to open the lid of the centrifuge, make sure the inner part of the chamber
is clean and is free of any foreign matters.
z To clean the external surface of rotor lock sleeve;
z Prepare the rotor you decide to use, hold it by both hands, aim the rotor centre at lock
sleeve, put it vertically on the bottom of the cone, remove two hands, and then slight
press the rotor by hand to ensure rotor base is placed on No.5 panel (shown on figure 8)
z Use special tools for assembling and disassembling of rotor (5mm inner hexagon ring
spanner) attached to the machine, screw on clockwise to fasten the fastening screws.
You should check the mounting position before using or
the rotor has been used some time. Screw the fastening
screw again if possible to ensure the rotor is mounted
firmly.
Figure 8:Diagram for rotor lock sleeve and supporting axis
Figure 9:Diagram for Installation of rotor

Figure 10:Sketch map after the completion of rotor installation

Description of figure8 , figure 9and figure 10:


1. Fastening screw 2. Lock sleeve 3. Electrical machine axis (cone) 4. Motor cover
5. Rotor supporting panel of the electrical machine axis
5.6 Maintenance of lock sleeve and supporting shaft of rotor
Steps to maintain lock sleeve and supporting shaft of rotor
z To turn on the electricity, wait until the self-check is completed;
z To press “ ” to open the lid of the centrifuge.
z To screw off the fastening screw (lock sleeve of the rotor) with special tools for
assembling and disassembling of rotors attached with the machine, then clean it.
Attention: the clockwise direction is for screwing and counterclockwise for loosening;
z To clean the cone section (see Fig. 9);
z To clear dust, impurity and remains separated before away from the centrifugal chamber
to make it clean and clear;
z To clean external surface of the lock sleeve of the rotor;
z To screw the fastening bolt into the screw of the cone section slowly with special tools
for assembling and disassembling rotors (inner hexagon spanner of 5 mm). Attention:
appropriate strength is permissible. The bolt is not allowed to touch the bottom,
otherwise the fastening bolt and cone section may be damaged and the rotor can’t be
mounted.

5.7 The calculation of rotor load


z Calculation of maximum load capacity
When the Neofuge 15 High Speed Centrifuge is running at high speed, there is enormous
centrifugal force; each rotor is designed to hold enough mechanical strength at the maximum
rated rotary speed-that is “safety factor”; However, the “safety factor” determines the rotor
load not to exceed it maximum rated load.
When separating samples, the total weight is not allowed to exceed the maximum rated
load of the rotor. Otherwise, you should reduce the weight of samples or calculate
permissible running speed (NPERM) to ensure the rotor load not to exceed its maximum rated
load.
Calculation method of permissible running speed (NPERM)is as follows:
NPERM=Nmax ×(maximum permissible load÷actual load)0.5
Nmax: Maximum rated rotary speed;
maximum permissible load:
actual load:

No overloading! The rotor may explode, thus pieces


produced may do damage to the centrifuge!

5.8 Samples filling into the centrifugal containers

Please carefully ensure the centrifugal container (centrifugal tube


etc.) to be in accordance with permissible maximum rated
acceleration (centrifugal force); please slow down for use if
possible.
When the centrifuge runs it is better to keep balance. It should not be interfered due to
vibration in separated area for samples. So it is necessary to fill the centrifugal container with
samples evenly. All samples should be placed in appropriate containers.

Take care of the operation life of centrifugal containers


used, especially the ones run under maximum
permissible load and velocity. Please change it for a new
one if the centrifugal container (plastic, glass container) is
damaged.

5.9 Safety use of rotor


5.9.1 Before the rotor runs, you should load samples and place tubes correctly.
5.9.2 When installing the centrifugal cup of the swing-out rotor, please check whether pintles
in the bearing are in the slots on both sides of the centrifugal cup and whether the four
centrifugal cups are flexible by hand.
5.9.3 The horizontal rotor is not allowed to experience long running at the critical speed of
2000r/min, otherwise, major vibration may occur and the working life will be shortened.
5.9.4 When changing rotors, please unscrew bolts in the counterclockwise direction to loosen
the rotor.

5.9.5 Do not start the machine when fastening screw is loosed.


5.9.6 When repeated running is needed for the centrifuge, you should check the fastening
screw after the rotor has run several times.
5.9.7 It is allowed to load samples into the centrifugal tubes symmetrically at different time.
Asymmetrical loading is strictly forbidden.

5.10 Parameters setting


Attention:
z You should install the rotor in the rotary shaft before setting parameters.
z This machine adopts touching buttons. You merely need to tightly touch related places
when setting parameters.
z You may set up the parameters again when mistakes occur in the operation.

5.10.1 Selection of accelerating and decelerating curve


This machine provides users with a selection function of curve lines, consisting of 9
different kinds of elevating speed curve lines.
z Switch on the power switch of the machine. Shows to enter the main interface when
initialization is completed.
z Press directly [▲] or [▼] button of the key [ ] on control panel one time to require
the preset numerical value and the numerical value won’t be changed when the machine
stops running. Continue to press it for more than one second to add or decrease preset
elevating speed curve line. At the same time, the lowest line on main interface shows
[↑↓] on the left side and preset elevating speed curve line on the right side. The
date will be stored into the system automatically and the interface returns to its
original status when setting is finished two seconds.

5.10.2 Setting of rotary speed:


Rotary speed range of Neofuge 15 High Speed Bench-top Centrifuge: The minimum
rotary speed: 300r/min, the maximum rotary speed: 16, 000 r/min.
Set rotary speed of rotors according to the following steps:
To see whether there is a dot before “00000 rpm” (as shown in figure 7), if there is no
dot, press again the “Speed/F” key to switch, and then press [▲] or [▼] one time to require
the preset numerical value and the numerical value won’t be changed during the periods of
stopping and running. Continue to press [▲] and [▼] on “Speed” icon of control panel (as
shown in figure 4) for more than one second to add or decrease preset rotary speed. At the
same time, the lowest line on main interface shows [↑↓] on the left side and preset rotary
speed on the right side. The date will be stored into the system automatically and the
interface returns to its original status when setting is finished two seconds.

5.10.3 Setting of the centrifugal force:


It is also feasible to change the setting of the rotary speed of Neofuge 15 High Speed
Bench-top Centrifuge through setting the numerical value of “centrifugal force”. Only when
you understand the type No. of the rotor may you convert the centrifugal force and rotary
speed regularly. The value of the centrifugal force and rotary speed will be displayed at the
same time.
So the value of the centrifugal force and rotary speed are displayed synchronously and
regularly only the centrifuge has run several seconds and the type No. of the rotor is
identified.

To set the centrifugal force as follows:


1. To see whether there is a dot before “00000 rcf ” on the bottom of the main interface
(as shown in figure 8), if there is no dot, press again the “Speed/F” key to switch to
“00000 rcf” , and then press [▲] or [▼] one time to require the preset numerical value.
2. Continue to press [▲] and [▼] to add up or decrease the value in “00000”
3. The date will be stored into the system automatically and the interface returns to its
original status when the value in “00000” reaches to your desired value two seconds.

Attention: Numerical values of centrifugal force set


should not be too low; otherwise rotary speed of some rotors may
be under 300r/min, leading to instability of devices!
5.10.4 Calculation on centrifugal force
Generally speaking, the relative centrifugal force is several thousand times more than the
G-force (g). It is the unit to measure the efficiency of various instruments to separate or
deposit things. The calculation of the centrifugal force is related to the centrifugal rational
speed and centrifugal radius. The method is as follows:
RCF=11.18 ×(n/1000)2× r
r:Centrifugal radius, unit: cm;
n:Centrifugal speed, unit: rpm (Rpm)
Attention:The numerical value of the maximum centrifugal force is related to the
maximum radius
The radius of the rotor and the shape of centrifugal contain are taken into account in
setting of “The numerical value of the centrifugal force “.

5.10.5 Setting of running time:


Setting range of running time of rotors of Neofuge 15 High Speed Bench-top
Centrifuge:1m~9h and 59m.
To press [▲] and [▼] on “Time” icon one time to require the preset numerical value and
the numerical value won’t be changed during the periods of stopping and running. Continue
to press it for more than one second to add or decrease preset running time. At the same time,
the lowest line on main interface shows [↑↓] on the left side and preset time on the right
side. The date will be stored into the system automatically and the interface returns
to its original status when setting is finished two seconds. The 3rd area of figure 5 shows
preset time.

5.10.6 Running of device


z The device can’t start until the users’ data have been set, the rotor has been installed,
samples to be separated placed correctly and the lid closed.
Attention:A lid lock icon on the main interface may tell you “Whether the lid of the device is
closed correctly”.
z Centrifugal work starts
1. To press the start key [►] on the control panel and the rotor is running evening at the
speed of 300r/min; At the same time, carry out rotor identification work.
2. After the rotor has been identified, the rotor accelerates according to preset elevating
speed curve line to preset rotary speed;
3. When press the start key [►] on the control panel, the running time is counting down
from “09:59”. When it displays zero, that is “00:00”, the centrifuge stops and the work is
finished;
4. In operation, the rotary speed and the centrifugal force (see Figure 7) rise separately from
“00000” and “00000” until they reach the setting “Rotary speed” and “Centrifugal
force”, and the centrifugal force of them is converted automatically according to the
rotor detected;
5. There should be warning prompts if the rotary speed and centrifugal force of the rotor
exceed the permissible maximum value or setting value; and if there are temperature
protection operation of the electric machinery, over-voltage operation of the compressor
and something wrong with the (EEROM, RAM) and so on, there should also be warning
prompts;
6. Opening of the lid is forbidden while the rotor is running.
Attention:For more details about concrete warning prompts, see “Trouble shooting”.
5.10.7 Imbalance detection
If the rotor doesn’t run smoothly and the vibration of the rotational shaft exceeds the
setting scope in operation, the device should stop at once and warning indication (normally
imbalance load imposed on the load) prompts appear. Open the lid after the running comes
to a stop. The device may start once more after the lid is opened to eliminate the trouble.

5.10.8 Modification of parameter during operation


The parameters except the accelerating curve and braking curve can be modified in
operation, including the rotating speed, running time, centrifugal force and so on.
After the parameter has been modified and the main interface menu appears, the newly
setting parameters will come into effect.

5.10.9 Termination of the centrifuge operation


There are two ways to terminate the centrifuge operation:
In normal operation, the running time has been set. So when the running time is out, the
centrifuge will stop automatically.
When the rotary speed of the rotor falls to zero, that is “RPM:00000” and “RCF:
00000”, you may press the opening key “ ” on the control panel and lift the lid slightly to
open the lid. The state of the lid lock displays the tapered end is open, then you can take out
the samples separated in the centrifuge.
In normal operation, you may press “■” on the control panel any time to terminate the
centrifuge on half way, and the centrifuge stops immediately. The countdown stops at the
same time. You can press “ ” to open the lid only the “RPM” and “RCF” (centrifugal force)
centrifugal force) displayed fall to zero. The state of the lid lock displays the tapered end is
open at the same time, and then you can take out the samples separated in the centrifuge.
Attention:If the centrifuge stops on half way, the running time which is updated to
“Setting time” will be “00:00” (see Figure 7), and remains to the next start.

5.10.10 Disassembling of rotors

Attention: When disassembling rotors, you should catch


them and lift by both hands. Vibration is forbidden.

Steps to disassemble rotors are as follows:


1. To stop the centrifuge;
2. To open the lid;
3. To take out the centrifugal container first if the horizontal rotor exists;
4. To unscrew the fastening screw in the counterclockwise direction with special tools
(torque spanner or inner hexagon spanner of 5mm) for disassembling. (no need to take
out)
5. To catch the rotor by both hands and vertically lift it. Vibration is forbidden;
6. You may take the sample separated, container and rotor from the fastening bolt (lock
sleeve of the rotor) if the sample separated is polluted or to prevent it to be polluted!
7. Then you may take the sample from the rotor in appropriate places where there are
preventive measures, for example:to put the sample separated, container and rotor in the
safe chest for biological operation and open the rotor to take the sample out after being
purified.

5.10.11 Alarm clearance


z Description of trouble message
The low line of the display screen shows all trouble messages, [ ] on the left side and
relative trouble messages on the right side. At the same time, the buzzer shall sound
and the warning red lights flash. Alarming sound will not vanish automatically, but it can be
eliminated by the operator. Press key [ ] to cancel the warning of the buzzer. Red lights
alarming will not cancel automatically until finally the trouble is solved.

Table 2:Table of trouble message prompt


Over Rotor Maximum Speed Current rotary speed of rotor exceeds max rated
rational speed
Over Rotor Set Speed Current rotary speed of rotor exceeds the setting
speed
Lid Open in Running Lid is opened when the device runs (centrifugal
operation)
Lid Open Failure Lid is not opened correctly
Speed Sensor Error Something wrong with speed sensor
I2c_BUS_Error_ I 2C warning
Main board Registration Error Main board CPU self-check register warning
Main board Ram Error_ Main board CPU self-check inner ram warning
Error Eprom_ EEPROM error
Motor Error Electrical machine failure
Inverter Protect Protection of frequency converter
Rotor Imbalance Imbalance warning
Keyboard Registration Error Keyboard CPU self-check register warning
Keyboard Ram Error Keyboard CPU self-check inner ram warning

5.10.12 Operation records


Please keep records of each operation (one for each device) to ensue the normal
operation of the equipment and prolong the service life and for the accumulation of operation
data of equipment to diagnose the trouble and the cause of the accident. The recommended
table of operation data is as follows:

Table 3 Table of operation records


Weight of
Program
Velocity
Rotary
speed

Date of Name of monotube Rotor Working


Time

Operator
test sample (including No. condition
sample)
VI. Maintenance
6.1 Cleaning

In the event that hazardous substances overflow or come into the equipment, user shall
performing cleaning work as required.
Users have the responsibility to clean and purify the centrifuge when dangerous
substance spills over into the machine. Users shall clean and purify it according
to the method recommended by the manufactures to ensure that the machine
won’t be damaged when using cleaning and purifying methods not
recommended by the manufactures. Inappropriate detergents and incorrect methods may do
harm to the centrifuge or spare parts inside; Cleaning and sterilization should be carried out
according to operating procedures described in the operating and technical instructions.

6.1.1 Implementation of cleaning and purifying


Turn off the power switch and draw out the power wire
before cleaning!

The centrifuge shell, rotor chamber, rotor and centrifugal containers used and so on
should be cleaned and maintained regularly (or determined by the service condition); thus
contamination on them can be cleared away to avoid damage to the spare parts used and
environments pollution.
If you are not clear which kind of should be selected for use while washing solvents
while cleaning the centrifuge, please contact Heal Force.
Organic solvents are forbidden, as they can’t decompose
grease in the bearings of the motor; Liquid, especially
organic solvents should be in contact with the supporting
shaft and ball bearings when being cleaned;
6.1.2 Implementation of high-pressure sterilization

Rotors used by this centrifuge can have high-pressure sterilization. Temperature:


127.5℃Absolute pressure: 250kPa, keep for at least 10 min. Please refer to instructions
offered by other manufactures on other separating containers.
Check whether it is permissible to sterilize rotors and
containers with high-pressure; the temperature and
time for high-pressure sterilization should be strictly controlled
according to different rotors and containers for separating;

6.1.3 Maintenance
6.1.3.1 Don’t contact the rotor with sharp objects. When being moved and disassembled, you
should take care not to damage the rotor, because scratches and scars may cause cracks in
operation.
6.1.3.2 Check whether there are spots, grooves, minute cracks in the rotor (especially the
bottom of the test tube) regularly. If any of them is found, please stop using them
immediately and contact Heal Force.
6.1.3.3 The rotor should be cleaned once every week in normal operation. Rinse the rotor
immediately if it is used to separate saline solution or other corrosive samples. If there are
liquid drops sprinkling on the rotor, blot it up and rinse partly immediately.
6.1.3.4 Please clean with sponges or cotton cloth dipped in neutral cleaning solution, then
wash the cleaning solution with distilled water. Don’t sprinkle or spray water on the rotor,
because it may make the rotor corrosive due to residual water. Invert the rotor to make it dry.
6.1.3.5 Inspection of rotor parts and safety consideration
6.1.3.6 Continuity of grounding protection.
6.1.3.7 Turn off the electricity and pull out the supply wire connected to the back wall of the
device before disassembling the cover casing in front of the device. Power operation is
forbidden to avoid the electric shock or damage to the device. Attention: This should be
operated only by trained maintainers of Heal Force!
6.1.3.8 Only spare parts provided by Heal Force can be used in this device.
6.1.3.9 Turn off the electricity when the centrifuge isn’t required.
6.1.3.10 Transportation and Storage
This machine belongs to the precision instrument. Water, vibration, leaning and inverted
placement are forbidden in the process of transportation and storage.

Frangible upwards damp-proof

6.2 Maintenances by HEAL FORCE


Neofuge 15 High Speed Bench-top Centrifuge produced by Heal Force is maintained by
professional maintainers once every year. Maintenance includes the following:
z To check electric devices (interior circuit);
z To check whether the centrifuge is in a stable position;
z To check electromagnetism lid locking system and other safety circuits;
z To check rotor locking device and supporting shaft (drive shaft);
z To check the performance of rotors and clean rotors by maintainers.
Heal Force will go to your house for maintenance not free of charge. Costs have been
determined in the sales contract. To change parts is free with the period of warranty;
otherwise users should pay for it.

6.3 Warranty (contract maintenance) qualification


The period of warranty (contract maintenance) begins from the day when users receive
products. If there is something wrong with our products, maintainers of Heal Force may go to
your house to repair them free of charge with the period of warranty (contract maintenance).
Users can change defective parts caused by its own quality, not by users, free of charge;
otherwise users should pay for it. The shell can’t enjoy the warranty
Warranty (contract maintenance) qualification
z Users operate according to the requirements of the operating and technical instructions;
z Installation, debugging, position changing, accessories adding, maintenance and so on
are carried by Heal Force or professional persons;
z Normal maintenance is carried out regularly according to methods determined in the
operating and technical instructions.
z Heal Force will repair damaged devices not free of charge if the damage is caused by
users because they operate them incorrectly.
Heal Force will treat any problem of the device quality (except damage caused due to
incorrect operation) free of charge within one years from the day when the device sells; you
should pay for the repair if the period of warranty is out.
VII. Trouble shooting
7.1 To open lid in case of emergency
In normal use, the function of automatic opening is forbidden when the electricity is off
or the operation to open the lid is disable by accident. However, things in the chamber must
be taken out. In this case, you may open the lid of the centrifuge by hand to take samples out.
Attention: this is for emergency only!
When the electricity is off and there is no brake,
the rotor will run a long time. Please wait!
Steps to open the lid in case of emergency are as follows:

7.1.1 To confirm all the rotors are not running (to observe from the eye hole at the top
of the lid);
7.1.2 To turn off electricity;
7.1.3 You may pull out the plastic plug at the lower part of the front wall of the chest with a
knife or screwdriver. You may pull the line connected to the back of the plastic plus to open
the lid, and then take out the sample separated in the centrifuge;
7.1.4 After it, you should put the line into the device and place the plug in its original place;

7.2 Warning message of trouble


The warning messages and treatment are listed in the following table. You may deal with
them according to prompts; if you can’t deal with them or the warning messages are not
listed, please contact maintainers of Heal Force at once.
Table of warning message of trouble (1)
Trouble Cause of trouble and treatment
No electricity, connection of the power supply is not good or abrupt
power cut
No display or no z To check whether the supply socket and the wire is in good
display abruptly condition, and the socket is live;
z Connection of the power switch not in good condition;
z Please contact maintainers of Heal Force if you can’ t solve it.
Cause of trouble:
The device stops 1. Exceeding the maximum rated speed of the rotor:
abruptly 2. If the rotary speed exceeds the rated speed of 250r/min, there will be
a warning alarm. At the moment, you should stop the device to reset
the rotary speed;
3. Exceeding the setting speed of the rotor:
4. Electric machinery of frequency conversion overheated or
compressor overheated:
5. Frequency conversion motor or compressor overheated:At this
moment, the electricity is cut off inside the device. The device stops;
6. Please check the electricity if there is no display on the keyboard
panel.
7. The voltage may be too low. Check the supply voltage.
1. The lid should be locked if the rotor is still running.
2. Manually open the door after the power is switched off, and then switch on the
power again to see if it is normal.
Lid locked 3. Safety-protection procedure activates when the power fails unexpectedly.
Operator can’t open the lid until 5 minutes after the power is switched on..
4. Please contact maintainers of Heal Force if you can’t solve the problem.
1. Exceeding the critical speed. Minor vibration is permissible.
2. Check whether the rotor is locked firmly.
3. Check symmetry of the load of rotors. Ensure the device is
swing-out.
Major vibration
4. Check the mounting of rotors.
5. Check the transmission shaft. Turn it by hand. If it can’t run
smoothly, there may be something wrong. Please contact maintainers
of Heal Force if you can’t solve the problem.
Digitron
1. No connection to the power supply. Check the distribution power.
indicator lamp
2. Two fuses of 6.3A on the PCB board are burnt out. Please ask for
on the control
professionals to change them.
panel off when
3. Please contact maintainers of Heal Force if you can’t solve the
the power switch
problem.
is turned on
Control panel Maybe it is interfered by the electric network. Turn off and wait for one
display abnormal minute. Turn on the device, and then everything will be all right.
Button pressed,
1. Check the electricity of the frequency conversion board.
no run for the
2. Electric control circuit failed. Change the electric control panel.
electric machine
1. Cut off electricity.
Burnt smoke 2. Whether the electric machine is burnt out.
smell from the 3. Whether the electric units are burnt out.
machine 4. Turn the transmission shaft by hand. If the shaft doesn’t rotate
smoothly, please ask professionals from the manufacturer for help.
Other troubles To be solved by professionals from the manufacturer.

Table of warning message of trouble (2)


Trouble messages Cause of trouble and treatment
Cause of trouble:Current rotary speed exceeds the maximum
rated speed more than 250rpm;
z Maybe it is overshoot. Return to the normal state
Over Rotor Max Speed
automatically several seconds later;
z Please contact maintainers of Heal Force if you can’t solve
it.
Cause of trouble:Current rotary speed exceeds the setting speed
Over Rotor Set Speed more than 200rpm;
z The acceleration may be overshoot. It will return to normal
state several seconds later;
Cause of trouble:Lid of the device is open or not closed
completely; rotor is running;
Lid Open When Run z The warning alarm will disappear when the rotor stops
completely;
z Close the lid again;
Cause of trouble:opening of the lid failed
z Close the device again, then start;
z When there is something wrong with the PCB control
Lid Open Fail circuit, please contact maintainers of Heal Force if you
can’t solve it
z Please open the lid by hand when you need to take
sample( please refer to 7.1)
Cause of trouble:speed sensor ineffective (no signal);
z Close the lid again, then start;
Speed Sensor Error z When there is something wrong with the PCB control
circuit, please contact maintainers of Heal Force if you
can’t solve it
Cause of trouble:I2C bus drive signal read-write error;
z Shut down, then start;
I2c_BUS_Error z When there is something wrong with the PCB control
circuit, please contact maintainers of Heal Force if you
can’t solve it
Cause of trouble: main board CPU internal register read-write
error;
Main board Registration z Shut down, then start;
Error z When there is something wrong with the PCB control
circuit, please contact maintainers of Heal Force if you
can’t solve it
Cause of trouble: main board CPU internal ram read-write error;
z Shut down, then start;
Main board Ram Error z When there is something wrong with the PCB control
circuit, please contact maintainers of Heal Force if you
can’t solve it
Cause of trouble:EEROM read-write error;
z Shut down, then start;
Error Eeprom z When there is something wrong with the PCB control
circuit, please contact maintainers of Heal Force if you
can’t solve it
Cause of trouble: imbalance of rotor load;
z Open the lid, take out the rotor, adjust rotor load to balance
status.
Rotor Imbalance
z When there is something wrong with the PCB control
circuit, please contact maintainers of Heal Force if you
can’t solve it
Keyboard Registration Cause of trouble: keyboard cpu internal register read-write
Error error;
z Shut down, then start;
z When there is something wrong with the PCB control
circuit, please contact maintainers of Heal Force if you
can’t solve it
Cause of trouble: keyboard cpu internal ram read-write error;
z Shut down, then start;
Keyboard Ram Error z When there is something wrong with the PCB control
circuit, please contact maintainers of Heal Force if you
can’t solve it

7.3 Safety protection


Protection function of automatic stopping
Note:All protection functions will come into effect once the electricity is on.
z Safety protection for “Overspeed”:If the rotary speed of the rotor is 250r/min
above the rated speed, the protection function of automatic stopping for “Over
max speed” will come into effect. The prompt message
“error_over_rotor_max_speed” appears and the buzzer alarms at the same time.
If the rotary speed of the rotor is 250r/min above the rated speed, the device will
stop automatically.
z When the running time is out, the device will stops automatically, and the
running time becomes “00:00:00”.
z Processing program in case of abnormity
If there is abnormal vibration, noise or peculiar smell, shut down the device, cut off
the electricity, take samples out and check. Please contact Heal Force if you can’t
solve it.

z Notice for technology support


Please inform the maintainer of Heal Force of the serial No. and version No. in
advance for convenient diagnosis when you need our help.
You may find the serial No. on the nameplate at the back of the device. You may find the
version No. of the software on the display when the machine is opened.
VIII. Technical data
Function/Parameter Technical indicators
z Used at home.
z There are no electric dust and volatile gas and corrosive
gas to exist in the ambient air; There are no vibration
Operating environment and airflow to influence the performance around
z Altitude:≤2000m;
z Relative humidity:≤80%;
Ambient temperature z +5℃~+35℃
Applied voltage z AC230V,50/60 Hz
Rated power z 470VA
Type of fuses and parameter z 317,125 in, AC250V,6.3A
Set time range z 1min~9h and 59min
z 15,000/16,000 RPM;
Rated/Maximum rotary
speed of rotor
(speed max)
Minimum rotary speed
z 300 RPM;
(speed min)
Maximum centrifugal z At the maximum rotary speed (speed max), the
force (RCFmax) maximum centrifugal force (RCFmax) is 24,900×g;
Ascending velocity z Max(highest) ascending velocity, less than 5min;
Descending velocity z Minimum (lowest) descending velocity, less than 5min;
Complete machine z Complete machine vibration of the centrifuge should not
vibration exceed 0.5g
Steady precision of
z ±1%(2000~16000r/min)
rotary speed
Time error of
z Time error of centrifuge:±1m per 30m.
centrifuge
Noise (At the
z Less than or equal to 65dB at the maximum speed;
maximum speed)
External dimensions
z 525mm(L)*390mm(W)* 330mm(H)
(Device)
Net Weight: z 58Kg
External
z 820mm(L)*720mm(W)* 770mm(H)
dimensions(Packaging)
Gross Weight z 103Kg
z Machinery Industrial Standards of the People’s Republic of
China JB/T 5519-91 High-speed Refrigeration Centrifuge;
z Safety standards: Designed according to GB/9706.1
Electric Equipment for Medical Use- Part 1:Universal safety
requirements;
Relevant standard that
z GB4793.1-1995 Measurement, controlling the laboratory to
quotes:
use the security request for the electric equipment First part:
Current requirements;
z GB4793.7-2001 Measurement, controlling the laboratory to
use the security of the electric equipment Use the special
request for the centrifuge in the laboratory;
z GB191-2000 Graphic presentation sign of warehousing and
transportation of the packaging;
z GB/T14710-1993 The electric equipment environment of
medical use requires and test method;
z GB9969.1-1998 Industrial products operation instructions
General provisions;
z Meets Enterprise standards: Q/TEUC42-2005 Neofuge 15
High Speed Bench-top Centrifuge.
Manufacturer
Shanghai Lishen Scientific Equipment Co., Ltd.
Address 6788 Songze Road, Qingpu District, Shanghai 201706, P. R. China

International Sales
Nison Instrument (Shanghai) Limited
Address Floor 15, No.2 Hua Shan Road, Shanghai 200040, P. R. China
Tel +86 21-6272 8646
Fax +86 21-6271 0529
E-mail export@healforce.com
Website www.healforce.com

Version 20100408
Information is subject to change and/or updating without notice.
Copyright © 2010 Heal Force. All rights reserved