User Manual
(Version20100408)
The operator may copy some chapters of this manual only for interior use, such as
chapters on how to manage incidents. These chapters have been clearly marked in the
contents of this manual.
If any damage to the instruments caused by users due to not obedience by the
“Operating Environment of Instruments”, Heal Force will not be responsible for this.
Heal Force has the right to modify the operating manual in any time without any proper
notices.
In order to extend the machine’s service life, we shall pay attention to the
following factors that may damage the machine:
- Chemical action;
- Effects of environments, including natural ultraviolet radiation
- Erosion and abrasion of protective cover parts and other safety parts
* Used at home;
* Altitude: ≤2000 m
* Ambient temperature: +5℃~+35℃;
* Relative humidity: ≤80%;
* Operating range of voltage: AC230 V,50/60 Hz;
* Adequate ventilation facilities must be installed inside;
* There are no vibration and airflow to influence the performance around;
* There are no electric dust and volatile gas and corrosive gas to exist in the ambient air.
Notice
* Please read this manual carefully if you use this instrument for the first time!
* Neofuge 15 High Speed Bench-top Centrifuge should be operated only by the trained
and authorized staff;
* The maintenance of the instrument should be carried out by the agents authorized by
Heal Force or Heal Force;
* The following materials should be forbidden:
- Flammable and explosive substances;
- Strong chemical materials;
- Tonic or radioactive substances, or pathogenic microorganism;
* The system of Neofuge 15 High Speed Bench-top Centrifuge can be maintained only by
qualified servicemen using appropriate tools.
* If any problem not mentioned in this manual occurs, please contact with agents
authorized by Heal Force or Heal Force for proper processing methods.
* Please use accessories provided by Heal Force. If other accessories are used, Heal Force
should not be responsible for bad consequences. However, the user may apply to Heal
Force to check whether the accessories are accordance with the requirements of Heal
Force.
* Neofuge 15 High Speed Bench-top Bench-top Centrifuge should be checked and
maintained at stated intervals.
Warning sign
Signal Ground!
Operating and Technical Instructions is the integral part of the Neofuge 15 High Speed
Bench-top Centrifuge, which should be put beside “Neofuge 15 High Speed Bench-top
Centrifuge” for the reference of operators.
Neofuge 15 High Speed Bench-top Centrifuge is designed to separate biologic cells, organism and
bacteria. The density of the sample won’t exceed 1.2g/cm3 at the maximum speed.
When the Neofuge 15 High Speed Bench-top Centrifuge is in operation (when the
sample is being separated or rotors are running), operators and dangerous substances should
be absent within 30 cm of the centrifuge. The ventilation hole of the centrifuge can not be
clogged by substances.
If the following safety precautions are not obeyed when using Neofuge 15 High Speed
Bench-top Centrifuge, operators or other persons may be hurt or the centrifuge and samples
to be separated may be damaged:
z The centrifuge is not corrosion proof or implosion proof, so it must be used in the
environment free of corrosion or implosion;
z The following materials should be forbidden:
- Flammable and explosive substances;
- Strong chemical materials;
- Tonic or radioactive substances, or pathogenic microorganism;
z It is necessary to take effective sealing measures when performing segregation of
poisonous substances and pathogenic microorganism cell. Furthermore, effective
sterilizing measures should be taken after the application. For detail, please see relevant
descriptions in Maintenance.
z Corrosive substances may do damage to materials inside the centrifuge and the rotors, so
they should be put inside the sealed containers while separating.
Figure 1: Front and side view of Neofuge 15 High Speed Bench-top Centrifuge
2.3.2 The chamber system is composed of the internal stainless steel chamber, rubber seal
rings. The chamber may provide a working condition of steady temperature.
2.3.3 This device adopts frequency conversion electric machinery to drive the rotor with load
sample. It applies cone connection and which is useful for connection of rotors and rotary
shafts.
2.3.4 Rotor system is composed of various kinds of angle rotors and horizontal rotors (For
details, see Table 1: Type of rotor and technical parameter), and centrifugal tube etc. The
rotor rotates at a certain speed to create a relative centrifugal force field to separate the
samples.
As the centrifugal force is several thousand times more than g (acceleration of earth gravity),
it is important to use and maintain the rotor safely and carefully!
2.3.5 The base system is composed of the machine frame, protection steel ring, motherboard,
top cover and rubber supporting legs and so on.
2.3.6 The power supply system includes power supply jack/switch, power switch, which is
used to supply electricity.
2.3.7 The control system includes identification of the rotor code, setting of the rotary speed
and centrifugal force, running time, temperature, choice of accelerating, control of display
system and warning system of the complete machine and so on. The control system is
composed of the main control panel, checking console for speed, package board of sensors
(identification of rotors and checking of balance quality of rotors). To open the cover in front
of the device is strictly forbidden!
2.3.8 The display system is composed of the display board and PVC keyboard panel (control
panel). It is the interactive media. It can show every setting parameter synchronously, and can
track the actual change. In addition, it can display various kinds of troubles and warnings.
2.3.9 The warning system consists of over speed, electrical motor overheat, lid failure and
temperature control failure. When the troubles mentioned above occur, the system will give
an alarm. At this moment, the red warning light on the control panel glimmers, the buzzer
sounds the alarm. The machine is unable to start (starting forbidden). The machine stops
automatically, until the trouble is treated. Then the machine can be restarted. Note: you may
press “ESC” on the control panel to cancel the warning tone sounded by the buzzer.
Rotor figures:
Please use special packing cases in long-distance transporting. Be straight! Handle with
care!
V. Operating instructions
5.1 Control panel and display interface
Speed/F Time
S/F
Stand By
5 6
Figure 7:Sketch map of “Main interface”
5.2 Starting
Insert one end of the power wire into the supply socket at the bottom of the left side, the
other end connected to the independent socket for civil power supply. The voltage of supply
power applied is AC230V and frequency is 50/60Hz. After connecting, turn on the power
switch at the bottom of the left side, the buzzer chirps and the icon “Heal Force” is displayed
on display window. At the same time, self-check is carried out. When the self-check has been
completed, the main interface appears (see Fig. 7). Now is for the following operation. As the
rotor is packed inside the centrifugal cavity, please do not press “►” and “ ” when the
machine is switched on for the first time.
5.3 Opening
If the electricity is on and rotor doesn’t run, you may press “ ” (key of opening) on
the control panel to open the lid, the data display window appears “Lift lid”. Under the air
spring action, the lid will open automatically and spring to a certain height, and then it needs
to manually lift and open the lid. The inner chamber will appear in front of the user. The icon
on the main interface will show the tapered end “unlocked”. At this point, it is applicable to
take out the rotor and package if the machine is switched on for the first time.
Notice: The lid will spring to a certain height when it is opened. Therefore, heads or
other objects should not stay at the place above the lid. It is dangerous!
When the lid can’t open automatically, you may take out things inside the chamber by
hand if necessary, see “Trouble shooting”.
5.4 Closing
Press the lid down until front side lid hooks on the lid skip over the lid bolt and a ringing
sound is heard. The bottom of the lid hooks will touch travel switch. At this time, the icon on
the main interface shows that the tapered end is closed and the lid has been locked firmly.
While closing the lid, press the lid properly, do not push
powerfully in order not to damage the lid lock!
5.5 Installation of rotor
Operator
test sample (including No. condition
sample)
VI. Maintenance
6.1 Cleaning
In the event that hazardous substances overflow or come into the equipment, user shall
performing cleaning work as required.
Users have the responsibility to clean and purify the centrifuge when dangerous
substance spills over into the machine. Users shall clean and purify it according
to the method recommended by the manufactures to ensure that the machine
won’t be damaged when using cleaning and purifying methods not
recommended by the manufactures. Inappropriate detergents and incorrect methods may do
harm to the centrifuge or spare parts inside; Cleaning and sterilization should be carried out
according to operating procedures described in the operating and technical instructions.
The centrifuge shell, rotor chamber, rotor and centrifugal containers used and so on
should be cleaned and maintained regularly (or determined by the service condition); thus
contamination on them can be cleared away to avoid damage to the spare parts used and
environments pollution.
If you are not clear which kind of should be selected for use while washing solvents
while cleaning the centrifuge, please contact Heal Force.
Organic solvents are forbidden, as they can’t decompose
grease in the bearings of the motor; Liquid, especially
organic solvents should be in contact with the supporting
shaft and ball bearings when being cleaned;
6.1.2 Implementation of high-pressure sterilization
6.1.3 Maintenance
6.1.3.1 Don’t contact the rotor with sharp objects. When being moved and disassembled, you
should take care not to damage the rotor, because scratches and scars may cause cracks in
operation.
6.1.3.2 Check whether there are spots, grooves, minute cracks in the rotor (especially the
bottom of the test tube) regularly. If any of them is found, please stop using them
immediately and contact Heal Force.
6.1.3.3 The rotor should be cleaned once every week in normal operation. Rinse the rotor
immediately if it is used to separate saline solution or other corrosive samples. If there are
liquid drops sprinkling on the rotor, blot it up and rinse partly immediately.
6.1.3.4 Please clean with sponges or cotton cloth dipped in neutral cleaning solution, then
wash the cleaning solution with distilled water. Don’t sprinkle or spray water on the rotor,
because it may make the rotor corrosive due to residual water. Invert the rotor to make it dry.
6.1.3.5 Inspection of rotor parts and safety consideration
6.1.3.6 Continuity of grounding protection.
6.1.3.7 Turn off the electricity and pull out the supply wire connected to the back wall of the
device before disassembling the cover casing in front of the device. Power operation is
forbidden to avoid the electric shock or damage to the device. Attention: This should be
operated only by trained maintainers of Heal Force!
6.1.3.8 Only spare parts provided by Heal Force can be used in this device.
6.1.3.9 Turn off the electricity when the centrifuge isn’t required.
6.1.3.10 Transportation and Storage
This machine belongs to the precision instrument. Water, vibration, leaning and inverted
placement are forbidden in the process of transportation and storage.
7.1.1 To confirm all the rotors are not running (to observe from the eye hole at the top
of the lid);
7.1.2 To turn off electricity;
7.1.3 You may pull out the plastic plug at the lower part of the front wall of the chest with a
knife or screwdriver. You may pull the line connected to the back of the plastic plus to open
the lid, and then take out the sample separated in the centrifuge;
7.1.4 After it, you should put the line into the device and place the plug in its original place;
International Sales
Nison Instrument (Shanghai) Limited
Address Floor 15, No.2 Hua Shan Road, Shanghai 200040, P. R. China
Tel +86 21-6272 8646
Fax +86 21-6271 0529
E-mail export@healforce.com
Website www.healforce.com
Version 20100408
Information is subject to change and/or updating without notice.
Copyright © 2010 Heal Force. All rights reserved