Professional Documents
Culture Documents
SERVOi_DATA_6670107_10_Datasheet_Adult_EN_ALL[1]
SERVOi_DATA_6670107_10_Datasheet_Adult_EN_ALL[1]
VENTILATION R
SERVO-i ADULT
HIGHLIGHTS
®
For adult and pediatric patients NAVA - improved synchrony and unique
monitoring capabilities
Intuitive user interface
For invasive as well as non invasive ventilation
Designed for cost-efficiency
®
The Open Lung Tool - Lung recruitment with
Flexible placement Dynamic Compliance and etCO2.
Modular – interchangeable plug-in units Available in MR and Transport editions for
Enables Heliox delivery continuity of ventilatory care
| Critical Care | SERVO-i ADULT | 3 |
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
Inspiratory channel
Gas supply
Pressure drop: Max. 6 cmH2O at a flow of 1 l/s
Inlet gas pressure Air/O2: 200 – 650 kPa / 2.0 – 6.5 bar /
29 – 94 PSI Internal compressible factor: Max. 0.1 ml/cmH2O
Connection standards available: AGA, DISS, NIST, or French Gas delivery system: Microprocessor controlled valve
standard. Inspiratory flow range: 0 to 3.3 l/s
Unavailable gas/loss of gas The flow from an unavailable
pressure: gas (air or O2) is automatically
compensated for so that the Expiratory channel
patient gets the preset volume Pressure drop: Max. 3 cmH2O at a flow of 1 l/s
and pressure. Internal compressible factor: Max. 0.1 ml/cmH2O
PEEP regulation: Microprocessor controlled valve
Patient system gas connectors Rise time, expiratory flow <12 ms for 10 – 90 % response
measurement: at flow of 0.05 – 3.2 l/s
Conical fittings: Male 22 mm / female 15 mm.
In accordance with ISO 5356-1. Expiratory flow range: 0 to 3.2 l/s
Gas exhaust port: Male 30 mm cone
| 6 | SERVO-i ADULT | Critical Care |
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
Alarms Alarms
Airway pressure (upper): End-tidal CO2 (upper and lower 0.5 – 20%.
Invasive ventilation: 16 – 120 cmH2O limit): 4 – 100 mm Hg.
0.5 – 14 kPa.
Non Invasive Ventilation: 16 - 40 cmH2O
Leakage out of range in NIV: Yes. Described in the User’s
manual.
Expired minute volume 0.5 – 60 l/min
(Upper alarm limit): Technical: Yes. Described in the User’s
manual.
Expired Minute Volume 0.5 – 40 l/min
(Lower alarm limit): Autoset (alarm limits) Invasive ventilation, controlled
specification: modes only
No patient effort (Apnea) alarm 15 – 45 s
High airway pressure: Mean peak pressure
Automatic return to support mode on patient triggering
+10 cmH2O or at least
35 cmH2O.
No consistent patient effort: Yes, described in User's manual
Upper minute volume: Expiratory minute volume
Respiratory frequency: 1 – 160 breaths/min + 50%.
High end expiratory pressure: 0 – 55 cmH2O Lower minute volume: Expiratory minute volume
Low end expiratory pressure: 0 – 47 cmH2O. – 50%.
Note. Setting the alarm to 0 Upper respiratory frequency: Breathing frequency + 40%.
(zero) is equal to alarm off.
Lower respiratory frequency: Breathing frequency – 40%.
High continuous pressure: Set PEEP level + 15 cmH2O
exceeded for more than 15 High end expiratory pressure: Mean end expiratory pressure
seconds. + 5 cmH2O.
O2 concentration: Low end expiratory pressure: Mean end expiratory pressure
Set value ±5 vol% or 18 vol%
– 3 cmH2O.
Gas supply: Below 200 kPa / 2.0 bar /
Upper end tidal carbon End tidal carbon dioxide
29 PSI and
dioxide concentration (etCO2): concentration + 25%.
above 650 kPa / 6.5 bar / 94 PSI
Lower end tidal carbon End tidal carbon dioxide
Battery: Limited battery capacity:
dioxide concentration (etCO2): concentration – 25%.
10 min.
No battery capacity: less than
3 min.
Low battery voltage.
| Critical Care | SERVO-i ADULT | 7 |
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
Suction Support
Pre oxygenation time: Max. 2 min
Post oxygenation time: Max. 1 min
Suction phase time: No maximum level
Adjustable oxygen level: 21 – 100 %
Saving of data
Recording of current waveform 20 seconds of data will be
and parameter values: recorded (10 seconds before
and 10 seconds after
activation).
Communication/Interface
Serial port: RS-232C - isolated. For data
communication via the
Communication Interface
Emulator (CIE).
Network connection (optional): MIB (Medical Information Bus)
monitor connection
Data transfer (optional) Via Ventilation Record Card
Screen picture transfer Via Ventilation Record Card
(optional)
| Critical Care | SERVO-i ADULT | 11 |
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
Open Lung Tool (OLT) (optional with Universal Basic Edition) NAVA (optional)
Three simultaneous graphical 1. EIP and PEEP (End Standards: IEC/EN 60601-1
trends, presented Inspiratory Pressure and (Type CF, defibrillator proof)
breath-by-breath: Positive End Expiratory
pressure).
Size:
2. VTi and VTe (Inspiratory
Edi module: 154 x 90 x 21 mm
and Expiratory Tidal
volume). Edi catheter cable: Length: 2 m
3. C dyn i and VTCO2*
(Dynamic inspiratory Weight - Edi module: 0.25 kg
Compliance [= VTi/(EIP – PEEP)]
Power source - Edi module Powered from SERVO-i. <3 W
and Tidal CO2 elimination*).
supply voltage: at 12 V (normal operation)
* Requires SERVO-i CO2 Parameters: Edi waveform
Analyzer option. Edi leads waveforms
NAVA estimated pressure
Values stored breath by breath: Up to 21.600 breaths. waveform (Pest)
Cursor function: A cursor can be moved with the Edi catheters: 6 Fr, length 49 cm
main rotary dial or via the touch 6 Fr, length 50 cm
screen. When moving it along 8 Fr, length 100 cm
the graph, the numeric 8 Fr, length 125 cm
parameter values valid for that 12 Fr, length 125 cm
actual moment will be shown. 16 Fr, length 125 cm
Zoom function: Time resolution of the x-axis
can be selected in five different
steps.
Time marking: Hours and minutes (when values
are measured).
| 12 | SERVO-i ADULT | Critical Care |
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
TECHNICAL SPECIFICATIONS
SERVO-i ADULT
Dimensional drawings