Download as pdf or txt
Download as pdf or txt
You are on page 1of 16

DATA SHEET

VENTILATION R

SERVO-i ADULT

HIGHLIGHTS
®
For adult and pediatric patients NAVA - improved synchrony and unique
monitoring capabilities
Intuitive user interface
For invasive as well as non invasive ventilation
Designed for cost-efficiency
®
The Open Lung Tool - Lung recruitment with
Flexible placement Dynamic Compliance and etCO2.
Modular – interchangeable plug-in units Available in MR and Transport editions for
Enables Heliox delivery continuity of ventilatory care
| Critical Care | SERVO-i ADULT | 3 |

SERVO-i VENTILATOR FAMILY


SERVO-i ADULT

SERVO-i Configurations UBE UEE


®
NIV NAVA
The SERVO-i Ventilator family consists of four configurations:
®
NAVA
Infant Heliox
Adult Nasal CPAP
UBE Universal Basic Edition NIV PC
UEE Universal Extended Edition NIV PS
Bi-Vent
Standard Configuration
Y Sensor Measuring
Options CO2 Analyzer
Not applicable Nebulizer
Alarm output connector
®
The SERVO-i Ventilator family addresses the very different Open Lung Tool
®
requirements of neonatal, pediatric and adult needs from a single Automode
ventilation platform. All four configurations are the same ventilator, SIMV (PRVC) + PS
equipped with different functions. They can be customized and PRVC
upgraded with different options for future needs. (SERVO-i Adult VS
and SERVO-i Infant can also be upgraded to SERVO-i Universal.) SIMV (VC) + PS
The same ventilator can be used at the bedside, during transport* VC
and in the MR-room* facilitating training, operation and SIMV (PC) + PS
maintenance, increasing efficiency and flexibility. PC
PS/CPAP
*An agreement with MAQUET must be signed, see conditions in All patient category software
the SERVO-i MR declaration (Order no. 66 71 670) and the KEY TO ABBREVIATIONS
Interhospital Transport Declaration (Order no. 66 64 721)
respectively. NAVA Neurally Adjusted Ventilatory Assist
NIV Non-invasive ventilation
SIMV Synchronized Intermittent Mandatory
Ventilation
PRVC Pressure Regulated Volume Control
VS Volume Support
VC Volume Control
PS Pressure Support
PC Pressure Control
CPAP Continuous Positive Airway Pressure
| 4 | SERVO-i ADULT | Critical Care |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

The system – General Non-operating conditions


The device complies with Impact: Peak acceleration: 15 g.
requirements of Medical Device Pulse duration: 6 ms.
Directive 93/42/EEC. Number of impacts: 1000.
Classification: Class I equipment. According Storage temperature: –25 to +60º C (–13 to 140º F)
to IEC/EN 60 601-1. Storage Relative Humidity: < 95%
Standards: IEC/EN 60 601-1 (Type B). Storage Atmospheric Pressure: 470 to 1060 hPa
IEC/EN 60 601-2-12.
EN 794-1.
IP classification IP 20 Power supply
Electromagnetic compatibility Power supply, automatic range 100 – 120 V AC ±10%,
(EMC): selection: 50 – 60 Hz, or
Emission: According to IEC/EN 60601-1-2 220 – 240 V AC ±10%,
(Edition 2:2001). 50 – 60 Hz.
Immunity- Extended test to According to IEC/EN 60601-1-2. Plug-in battery module:
30 v/m: Battery backup: Two battery modules are
The 'EMC Declaration, Information to the Responsible delivered with the ventilator. Up
Organization' is available from MAQUET. to six battery modules can be
included.
Patient range (all ventilation Weight 10 – 250 kg. Battery capacity: Rechargable, 12 V, 3.5 Ah each.
modes): Recharge time: Approximately 3 h/battery.
Battery backup time: At least 3 h, when using six
batteries.
Operating conditions
Operating temperature: +10 to +40ºC External 12 V DC: 12.0 V – 15.0 V DC, 10 A
Relative humidity: 15 to 95% non-condensing Max power consumption: At 100 – 120 V: 2 A, 190 VA,
Atmospheric pressure: 660 to 1060 hPa 140 W.
Lowest pressure in breathing –400 cmH2O At 220 – 240 V: 1 A, 190 VA,
system: 140 W.
| Critical Care | SERVO-i ADULT | 5 |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

The ventilator – General User interface


Dimensions: (See dimensional drawings Weight: Approximately 5 kg
page -) Attachment: Can be attached to the mobile
User Interface: W 355 x D 53 x H 295 mm cart, a table, rail or pole (15 –
Patient Unit: W 300 x D 205 x H 415 mm 30 mm diameter).

Weight: Approximately 20 kg (Patient


Screen
Unit 15 kg, User Interface 5 kg)
Type: TFT-LCD module
Method of triggering: Flow and pressure
Size: 31 cm (12.1”) diagonal
Max. operating pressure: Approximately 115 cmH2O
Viewing area: 246.0 x 184.5 mm
Bias flow: 2 l/min

Inspiratory channel
Gas supply
Pressure drop: Max. 6 cmH2O at a flow of 1 l/s
Inlet gas pressure Air/O2: 200 – 650 kPa / 2.0 – 6.5 bar /
29 – 94 PSI Internal compressible factor: Max. 0.1 ml/cmH2O
Connection standards available: AGA, DISS, NIST, or French Gas delivery system: Microprocessor controlled valve
standard. Inspiratory flow range: 0 to 3.3 l/s
Unavailable gas/loss of gas The flow from an unavailable
pressure: gas (air or O2) is automatically
compensated for so that the Expiratory channel
patient gets the preset volume Pressure drop: Max. 3 cmH2O at a flow of 1 l/s
and pressure. Internal compressible factor: Max. 0.1 ml/cmH2O
PEEP regulation: Microprocessor controlled valve
Patient system gas connectors Rise time, expiratory flow <12 ms for 10 – 90 % response
measurement: at flow of 0.05 – 3.2 l/s
Conical fittings: Male 22 mm / female 15 mm.
In accordance with ISO 5356-1. Expiratory flow range: 0 to 3.2 l/s
Gas exhaust port: Male 30 mm cone
| 6 | SERVO-i ADULT | Critical Care |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

Alarms Alarms
Airway pressure (upper): End-tidal CO2 (upper and lower 0.5 – 20%.
Invasive ventilation: 16 – 120 cmH2O limit): 4 – 100 mm Hg.
0.5 – 14 kPa.
Non Invasive Ventilation: 16 - 40 cmH2O
Leakage out of range in NIV: Yes. Described in the User’s
manual.
Expired minute volume 0.5 – 60 l/min
(Upper alarm limit): Technical: Yes. Described in the User’s
manual.
Expired Minute Volume 0.5 – 40 l/min
(Lower alarm limit): Autoset (alarm limits) Invasive ventilation, controlled
specification: modes only
No patient effort (Apnea) alarm 15 – 45 s
High airway pressure: Mean peak pressure
Automatic return to support mode on patient triggering
+10 cmH2O or at least
35 cmH2O.
No consistent patient effort: Yes, described in User's manual
Upper minute volume: Expiratory minute volume
Respiratory frequency: 1 – 160 breaths/min + 50%.
High end expiratory pressure: 0 – 55 cmH2O Lower minute volume: Expiratory minute volume
Low end expiratory pressure: 0 – 47 cmH2O. – 50%.
Note. Setting the alarm to 0 Upper respiratory frequency: Breathing frequency + 40%.
(zero) is equal to alarm off.
Lower respiratory frequency: Breathing frequency – 40%.
High continuous pressure: Set PEEP level + 15 cmH2O
exceeded for more than 15 High end expiratory pressure: Mean end expiratory pressure
seconds. + 5 cmH2O.
O2 concentration: Low end expiratory pressure: Mean end expiratory pressure
Set value ±5 vol% or 18 vol%
– 3 cmH2O.
Gas supply: Below 200 kPa / 2.0 bar /
Upper end tidal carbon End tidal carbon dioxide
29 PSI and
dioxide concentration (etCO2): concentration + 25%.
above 650 kPa / 6.5 bar / 94 PSI
Lower end tidal carbon End tidal carbon dioxide
Battery: Limited battery capacity:
dioxide concentration (etCO2): concentration – 25%.
10 min.
No battery capacity: less than
3 min.
Low battery voltage.
| Critical Care | SERVO-i ADULT | 7 |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

Ventilation Modes – Invasive ventilation Ventilation modes – Non-invasive Ventilation (optional)


Controlled ventilation: NIV PC
PC Optional
NIV PS
VC
PRVC Optional
Ventilation modes - NAVA (Optional)
Supported ventilation: NAVA Neurally Adjusted Ventilatory
VS Optional Assist via endotracheal tube or
PS/CPAP tracheostomy

Combined ventilation: NIV NAVA Neurally Adjusted Ventilatory


SIMV (VC) + PS: Assist via non-invasive patient
SIMV (PC) + PS: Comes with the corresponding interfaces
controlled ventilation mode.
(Optional)
Waveform and loop presentations
SIMV (PRVC) + PS Comes with the corresponding
controlled ventilation mode. Real time waveforms - up to 4 waveforms can be displayed
(Optional) simultaneously:
Bi-Vent Pressure controlled ventilation Pressure
on two independently adjustable Flow
levels, allowing unrestricted Volume
spontaneous breathing on both
CO2 Requires SERVO-i CO2 Analyzer
levels. (Optional)
option
Automode Control mode: VC <—> Support Edi Requires SERVO-i NAVA option
mode: VS
Control mode: PC <—> Support Loops:
mode: PS Volume / Pressure* *A reference loop and three
Control mode: PRVC <—> overlaying loops can be
Support mode: VS displayed.
(Optional) Flow / Volume* *Displayed simultaneously with
Open Lung Tool graphical
trends, if requested.
| 8 | SERVO-i ADULT | Critical Care |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

Displayed Trended Displayed Trended


Monitoring value value*
Monitoring value value*
End Expiratory Flow: Yes Yes
Breathing frequency: Yes Yes
Static Compliance: Yes Yes
Spontaneous breaths per No Yes
minute (RRsp): Dynamic Compliance: Yes Yes
Peak Airway Pressure: Yes Yes Inspiratory Resistance Yes Yes
Mean Airway Pressure: Yes Yes Expiratory Resistance Yes Yes
Pause Airway Pressure: Yes Yes Elastance Yes Yes
End Expiratory Pressure: Yes Yes Time Constant Yes No
CPAP Pressure: Yes Yes P0.1: Yes Yes
Inspired Tidal Volume: Yes Yes Work of Breathing patient: Yes Yes
Expired Tidal Volume: Yes Yes Work of Breathing ventilator Yes Yes
Inspired Minute Volume: Yes Yes Shallow Breathing Index (SBI) Yes Yes
Expired Minute Volume: Yes Yes Supply pressure (O2 and air): Yes No
Leakage fraction in NIV (%): Yes Yes Battery remaining time: Yes No
Ti/Ttot: Yes No Barometric pressure: Yes No
I:E ratio: Yes No
Total PEEP: Yes No Log function
Edi peak: Yes Yes Event log: Alarms.
Edi min: Yes Yes Ventilator settings.
O2 Concentration (measured): Yes Yes Apnea periods.
Immediate functions.
CO2 End tidal concentration Yes Yes
(etCO2): Service log: Technical alarms.
Test results.
CO2 Minute elimination (CO2): Yes Yes Preventive maintenance.
Tidal CO2 elimination (VTCO2): Yes Yes Service history.
MVe sp / MVe: Yes No Configuration log.
Spontaneous Exp. Minute Yes Yes
Volume (MVe sp):
*Stored trend values for up to 24 hours
| Critical Care | SERVO-i ADULT | 9 |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

Parameter settings: Setting range: Parameter: Setting range:


TInsp (s): 0.1 – 5
Parameter:
NAVA level (cmH2O/µV): 0 - 15
Inspiratory tidal volume (ml): 100 – 2000 Edi trigger sensitivity (µV): 0.1 – 2.0
Inspiratory minute volume 0.5 – 60 TPause (s): 0 – 1.5
(l/min): TPause (% of breath cycle time): 0 – 30
Apnea, time to alarm (s): 15 – 45 Flow trigger sensitivity level 0 – 100%
Automode Trigger timeout (s): 7 – 12 (fraction of bias flow):
PC/PS above PEEP (cmH2O) 0 – (120 - PEEP) Press. trigg sensitivity (cmH2O) -20 – 0
PC/PS above PEEP in NIV 0 – (32 - PEEP) Insp. rise time (% of breath 0 – 20
(cmH2O) cycle time):
PEEP (cmH2O) 0 – 50 Insp. rise time (s): 0 – 0.4
PEEP in NIV (cmH2O) 2 – 20 Insp. cycle off (% of peak flow): 1 – 70
CPAP pressure (cmH2O): – Insp. cycle off in NIV (% of peak 10 – 70
flow):
CMV frequency (breaths/min): 4 – 100
Nebulizer time (min): 5 – 30
SIMV frequency (breaths/min): 1 – 60
Breath cycle time, SIMV (s): 1 – 15
PHigh (cmH2O) (PEEP +1) – 50 Backup settings Setting range:
THigh (s): 0.2 – 10 Parameter:
TPEEP (s): 0.2 – 10
PS above PHigh (cmH2O): 0 – (120 - PHigh) Inspiratory tidal volume (ml): 100 - 2000
O2 concentration (%): 21 – 100 PC above PEEP (cmH2O) 5 – (120 - PEEP)
I:E ratio: 1:10 – 4:1 PC above PEEP in NIV 5 – (32 - PEEP)
CMV frequency (breaths/min): 4 – 100
I:E ratio: 1:10 – 4:1
TInsp (s): 0.1 – 5
| 10 | SERVO-i ADULT | Critical Care |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

Parameter: Setting range Non-invasive Ventilation (optional)


Oxygen breaths: 100% for 1 minute Max. leakage compensation level:
Start breath: Initiation of 1 breath (In SIMV Adult 65 l/min
mode initiation of 1 mandatory Leakage overrange detection: Automatic
breath)
Disconnect detection: Automatic
Pause hold: Insp. or exp (0 – 30 seconds)
Disconnect flow: Configurable
Alarm silence/reset: 2 minute silence and reset of
Low: 7.5 l/min
latched alarms
High: 40 l/min
Compliance compensation: On/Off
Disabled Deactivates disconnect
Automode (optional with Automode On/Off
detection
Universal Basic Edition):
Servo Ultra Nebulizer (optional): Nebulizer On/Off
Connect detection: Manual, or automatic via bias
Backup ventilation Backup On/Off flow

Suction Support
Pre oxygenation time: Max. 2 min
Post oxygenation time: Max. 1 min
Suction phase time: No maximum level
Adjustable oxygen level: 21 – 100 %

Saving of data
Recording of current waveform 20 seconds of data will be
and parameter values: recorded (10 seconds before
and 10 seconds after
activation).

Communication/Interface
Serial port: RS-232C - isolated. For data
communication via the
Communication Interface
Emulator (CIE).
Network connection (optional): MIB (Medical Information Bus)
monitor connection
Data transfer (optional) Via Ventilation Record Card
Screen picture transfer Via Ventilation Record Card
(optional)
| Critical Care | SERVO-i ADULT | 11 |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

Open Lung Tool (OLT) (optional with Universal Basic Edition) NAVA (optional)
Three simultaneous graphical 1. EIP and PEEP (End Standards: IEC/EN 60601-1
trends, presented Inspiratory Pressure and (Type CF, defibrillator proof)
breath-by-breath: Positive End Expiratory
pressure).
Size:
2. VTi and VTe (Inspiratory
Edi module: 154 x 90 x 21 mm
and Expiratory Tidal
volume). Edi catheter cable: Length: 2 m
3. C dyn i and VTCO2*
(Dynamic inspiratory Weight - Edi module: 0.25 kg
Compliance [= VTi/(EIP – PEEP)]
Power source - Edi module Powered from SERVO-i. <3 W
and Tidal CO2 elimination*).
supply voltage: at 12 V (normal operation)
* Requires SERVO-i CO2 Parameters: Edi waveform
Analyzer option. Edi leads waveforms
NAVA estimated pressure
Values stored breath by breath: Up to 21.600 breaths. waveform (Pest)
Cursor function: A cursor can be moved with the Edi catheters: 6 Fr, length 49 cm
main rotary dial or via the touch 6 Fr, length 50 cm
screen. When moving it along 8 Fr, length 100 cm
the graph, the numeric 8 Fr, length 125 cm
parameter values valid for that 12 Fr, length 125 cm
actual moment will be shown. 16 Fr, length 125 cm
Zoom function: Time resolution of the x-axis
can be selected in five different
steps.
Time marking: Hours and minutes (when values
are measured).
| 12 | SERVO-i ADULT | Critical Care |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

SERVO-i CO2 Analyzer (optional) SERVO-i CO2 Analyzer – Performance


Standard: EN 864, ISO 9918. Measuring method: Mainstream, dual-wavelength,
IEC/EN 60601-1 non-dispersive infrared.
(Type BF, defibrillator proof) Parameters: Capnogram.
CO2 End tidal concentration
(etCO2).
Size:
CO2 Minute elimination ( CO2).
CO2 Analyzer Module: 154 x 90 x 43 mm
Tidal CO2 elimination (VTCO2).
Sensor: 32.0 x 42.4 x 21.6 mm
Measuring range: 0 to 100 mm Hg CO2 partial
Weight: pressure.
CO2 Analyzer Module: 0.45 kg 0 to 13.3 kPa CO2 partial
Sensor: 18 g pressure.
0 to 13.2% CO2 volume (at a
Airway adapter: 10 g
barometric pressure of
1013 hPa).
Connectors and cables:
Step response time: <25 ms (10 to 90% step
CO2 Analyzer Module: 15-pole D-sub female response).
connector.
Warm-up time: 30 s to initial CO2 indication,
Sensor: 20-pole, 2.4 m cable. max. 5 min to full specification.
Oxygen concentration Automatic. Values supplied from
Power source: compensation: the SERVO-i Ventilator System.
CO2 Analyzer Module supply Powered from the SERVO-i. Barometric pressure Automatic. Values supplied from
voltage: compensation: the SERVO-i Ventilator System.
Power consumption: 8 W at 12 V, during warm up. Digitizing rate: 87 Hz
6.5 W at 12 V, during normal Airway adapter dead space: <5 ml
operation.
Sensor: Powered from the CO2 Analyzer
Module.
| Critical Care | SERVO-i ADULT | 13 |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

Y Sensor measuring (optional) Servo Ultra Nebulizer (optional)


Size: Weight: Approx. 125 g
Y Sensor Module: 154 x 90 x 43 mm Dimensions: H 105 mm x L 108 mm x
Y sensor adult: Length 84 mm W 60 mm
Connection cable length: 2.0 m
Weight: Nebulizer T-piece connections: Inlet/outlet: 22/15 mm
Y Sensor Module: 0.4 kg outer/inner diameter and 22 mm
inner diameter, ISO standard.
Y sensor adult: 10.5 g
Pediatric patient tubes:
Adapters 22/10 mm outer
Sensor material: Makrolon polycarbonate. diameter and 15/10 mm outer
Tubing: 2.0 m. Medical grade PVC. diameter.
Power source – Y Sensor Powered from the SERVO-i. Internal volume: 60 ml
Module supply voltage: <5 W at 12 V (normal operation). Ultrasonic generator frequency: 2.4 MHz
Particle size (water): Mass Median Diameter (MMD)
= approximately 4.0 µm,
Y sensor measuring – Performance
measured distally in
Measuring method: Fixed orifice, differential endotracheal tube 8 mm inner
pressure. diameter.
Parameters: Airway pressure. Output from nebulizer (water) – Min. 0.1 ml water/min at 0.1 l
Airway flow. minimum water flux: gas flow/s.
Inspiratory and expiratory Min. 0.3 ml water/min at 0.5 l
volumes. gas flow/s.
Measuring range: 2 to 180 l/min Min. 0.5 ml water/min at 1.0 l
Airway adapter dead space: < 9.0 ml gas flow/s.
Buffer liquid: Sterile water
Max. medication temperature: 55º C (131º F)
Volume, medication cup: Max. 10 ml
Noise level: Max. 50 dBA, measured at
0.3 m distance.

Aeroneb Nebulizer Systems (optional)


See separate data sheets.
| 14 | SERVO-i ADULT | Critical Care |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

SERVO-i Mobile Cart (optional) SERVO-i IV Pole (optional)


Weight: 20 kg Max load (total): 6 kg
Dimensions: H 1015 mm x L 640 mm x
W 560 mm
(see dimensional drawing Gas trolley (optional)
below) Max load: 2x10 kg bottles
Docking: Dockable to SERVO-i Mobile
Cart.
SERVO-i Drawer kit (optional) Dockable to a separate wall
Weight: 4.5 kg clamp.
Dimensions: H 240 mm x L 210 mm x
W 300 mm
Compressor Mini (optional)
See separate data sheet.
SERVO-i Holder (optional)
Weight: 3.5 kg
Dimensions: H 352 mm x L 247 mm x Alarm output connector (optional)
W 159 mm Connector: 4-pole Modular connector
(see dimensional drawing Ratings: Max 40 V DC, Max 500 mA,
below) Max 20 W

SERVO-i Shelf Base (optional) Service


Weight: 1.2 kg Regular maintenance: Once every 12 months or at
Dimensions: H 29 mm x L 205 mm x W 159 least after 5000 operating hours.
mm
(see dimensional drawing
below) Note
For inaccuracies and more detailed technical specifications please
refer to the User’s manual.
Gas cylinder restrainer (optional)
Max load: 2 x 5-liter bottles
ORDERING INFORMATION

SERVO-i, Ventilator and accessories: See separate information:


“SERVO-i, Version 5.0 — System Flow Chart” (Order no: 66 70
102).
| Critical Care | SERVO-i ADULT | 15 |

TECHNICAL SPECIFICATIONS
SERVO-i ADULT

Dimensional drawings

SERVO-i on Mobile Cart SERVO-i Holder

SERVO-i (patient unit) on SERVO-i Holder

SERVO-i Shelf Base


© Maquet Critical Care AB 2010. All rights reserved. • MAQUET reserves the right to modify the design and specifications contained herein without prior notice.
Order No. 66 70 107 • Printed in Sweden • 100601 • 10 English • Data Sheet

Maquet Critical Care AB


Röntgenvägen 2
SE-171 54 Solna, Sweden
Phone: +46 (0) 8 730 73 00
www.maquet.com
GETINGE GROUP is a leading global provider of products and
systems that contribute to quality enhancement and cost efficiency
within healthcare and life sciences. We operate under the three
brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh
focuses on patient mobility and wound management solutions.
For local contact: GETINGE provides solutions for infection control within healthcare
and contamination prevention within life sciences. MAQUET
Please visit our website specializes in solutions, therapies and products for surgical
www.maquet.com interventions and intensive care.

You might also like