IEC 61000-4-30 Power Quality Measurement Methods Compliance Report

ENCORE SERIESTM 61000 System

Udin (declared voltages) 230V, 50 Hz & 120V, 60 Hz

DRANETZ BMI 1000 New Durham Road, Edison, New Jersey 08818-4019 USA

Technology International (Europe), Ltd. 60 Shrivenham Hundred Business Park Shrivenham, Swindon, Wiltshire SN6 8TY, England

61000 System 61000-4-30 Test Report

TABLE OF CONTENTS
61000-4-30 Test Results Summary ......................................................................3 1. TEST RESULTS ...............................................................................................4 1. TEST RESULTS ...............................................................................................4 1.1 MEASUREMENT AGGREGATION INTERVALS ....................................4 1.2 TIME CLOCK UNCERTAINITY ................................................................4 1.3 FLAGGING .............................................................................................4 1.4 FREQUENCY ........................................................................................4 1.5 VOLTAGE MAGNITUDE ........................................................................4 1.6 FLICKER ...................................................................................................5 61000-4-15 ....................................................................................................5 61000-4-30 ....................................................................................................5 1.7 DIPS , SWELLS, & INTERRUPTIONS ..................................................5 1.8 UNBALANCE..........................................................................................5 1.9 HARMONICS..........................................................................................6 1.10 INTERHARMONICS ...............................................................................6 1.11 MAINS SIGNALLING ...............................................................................6 1.12 UNDER/OVER DEVIATION ....................................................................6 2. POWER QUALITY TEST AND SPECIFICATIONS...........................................7 2.1 Range of Influence .................................................................................7 2.2 Accuracy of Parameters .........................................................................7 2.3 Test Standards .......................................................................................8 2.4 Other Standards ......................................................................................8 2.5 QUALITY SYSTEM ................................................................................8 3. Equipment Under Test ......................................................................................9 3.1 Test Configurations ................................................................................9 3.2 Equipment Under Test Constructional Details ...................................12 3.3 PRODUCT VARIANTS .........................................................................12 4 Test Equipment Used .....................................................................................13 4.1 Frequency, Flicker, Harmonics, Interharmonics, Voltage Magnitude ........13 4.2 Voltage Unbalance ....................................................................................14 4.3 Technical Statements ...............................................................................15 5 DETAILED TEST FILES and RESULTS..........................................................16 ANNEX A ........................................................................................................25 Certificates Of Compliance..............................................................................25 IEC 61000-4-30 Certificate .................................................................................26 ISO 9001 Certificate............................................................................................27 ISO 17025 Certificate..........................................................................................28 ANNEX B ........................................................................................................29 Test Instruments and Facilities........................................................................29 Test Instruments used for 61000-4-30 ................................................................30 Test Instruments used for additional verification .................................................30 Test Facility Photographs ...................................................................................31

Page 2 of 34

61000 System 61000-4-30 Test Report

1. TEST RESULTS

1.1

MEASUREMENT AGGREGATION INTERVALS


The test results are that the Encore Series 61000 System complies with the requirements of section 4.4 of 61000-4-30 for aggregation over time limits of cycle-by-cycle, 200msec, 3 second, 10 minute, and 2 hour intervals as required for the parameters of rms variations, harmonics, interharmonics, unbalance using sequence components, under/over deviation, flicker (Pst and Plt), and magnitude of supply, as appropriate.

1.2

TIME CLOCK UNCERTAINITY


The test results are that the Encore Series 61000 System complies with the requirements of section 4.6 of 61000-4-30 for free-running clock of 1 second/day or +/- 1 cycle at 50/60Hz when synchronized with GPS option.

1.3

FLAGGING
The test results are that the Encore Series 61000 System complies with the requirement of section 4.7 of 61000-4-30 for flagging intervals where rms variations have occurred through the use of the time stamped, detailed characterized event lists

1.4

FREQUENCY
The test results are that the Encore Series 61000 System measurements complies with section 5.1 of 61000-4-30 for the specified measurement accuracy of 0.01 Hz. Also, during the frequency variation tests did not have any influence on Flicker, Harmonics, Interharmonics, Unbalance, or Voltage Magnitude measurements during the evaluation.

1.5

VOLTAGE MAGNITUDE
The test results are that the Encore Series 61000 System measurements complies with section 5.2 of 61000-4-30 for the specified measurement accuracy of 0.1% of the declared voltage for Udin 230 V, 50 hz and Udin 120 V, 60 hz. Also, during the voltage variation tests did not have any influence on Flicker, Frequency, Harmonics, Interharmonics, or Unbalance measurements during the evaluation

Page 4 of 34

61000 System 61000-4-30 Test Report

1.6 FLICKER
61000-4-15
The test results are that the Encore Series 61000 System measurements complies with section 5.3 of 61000-4-30 for the specified measurement accuracy of 0.05 pst. Also, during the flicker variation tests did not have any influence on Frequency, Harmonics, Interharmonics, Unbalance, or Voltage Magnitude measurements during the evaluation.

61000-4-30
The test results are that the Encore Series 61000 System measurements complies with section 5.3 of 61000-4-30 for the specified measurement accuracy of 0.1 Pst for a known Pst of 1. This then translates to 10% for Pst levels above 1, for example a Pst of 20 would allow a tolerance of 2 Pst. Also, during the flicker variation tests did not have any influence on Frequency, Harmonics, Interharmonics, Unbalance, or Voltage Magnitude measurements during the evaluation.

1.7 DIPS , SWELLS, & INTERRUPTIONS


The test results are that the Encore Series 61000 System measurements complies with sections 5.4 and 5.5 of 61000-4-30 for the specified hysteresis of 2% and detection within a 0.5 cycle for fixed nominal, reporting the required magnitude and duration parameters.

1.8

UNBALANCE
The test results are that the Encore Series 61000 System measurements complies with section 5.7 of 61000-4-30 for the specified measurement accuracy of 0.15% of the negative sequence divided by positive sequence for a known unbalance of 1%. This then translates to 0.15% for each 1% unbalance levels above 1%, for example an unbalance of 5% would allow a tolerance of 0.75%. Also, during the voltage unbalance variation tests did not have any influence on Flicker, Frequency, Harmonics, or Interharmonics measurements during the evaluation.

Page 5 of 34

61000 System 61000-4-30 Test Report

1.9

HARMONICS
61000-4-7 The test results are that the Encore Series 61000 System measurements complies with section 5.8 of 61000-4-30 for the specified measurement accuracy of 5% maximum allowable error for harmonic signals greater than 1%. Also, during the harmonics variation tests did not have any influence on Frequency, Flicker, Interharmonics, Unbalance, or Voltage Magnitude measurements during the evaluation.

1.10 INTERHARMONICS
61000-4-7 The test results are that the Encore Series 61000 System measurements complies with section 5.9 of 61000-4-30 for the specified measurement accuracy of 5% maximum allowable error for harmonic signals greater than 1%. Also, during the interharmonics variation tests did not have any influence on Frequency, Flicker, Harmonics, Unbalance, or Voltage Magnitude measurements during the evaluation.

1.11 MAINS SIGNALLING


The test results are that the Encore Series 61000 System measurements complies with section 5.10 of 61000-4-30 for the specified measurement for mains signaling frequencies under 3KHz using the 5 Hz bins from the calculations of the harmonics/interharmonics.

1.12 UNDER/OVER DEVIATION


The test results are that the Encore Series 61000 System measurements complies with section 5.12 for the specified measurement in section 5.12 of 61000-4-30 for over/under deviations based on the nominal voltage over the aggregation intervals of 200msec and 10 minutes.

Page 6 of 34

61000 System 61000-4-30 Test Report

2. POWER QUALITY TEST AND SPECIFICATIONS


2.1 Range of Influence
The range of influence for each parameter per EN61000-4-30 table 1 is applicable for Class A performance. For the equipment under test Class A performance is claimed for the test results. Frequency - 42.5 Hz - 57.5 Hz for 50 Hz Systems and 51 Hz - 69 Hz for 60 Hz Systems Voltage Magnitude - 0% -200 % of the declared voltage. Flicker - 0% to 20% pst Unbalance- 0% to 5% Harmonics - twice the values of IEC 61000-2-4, Class 3 Interharmonics - twice the values of IEC 61000-2-4, Class 3 Dips and Swells - 2% hysteresis

2.2

Accuracy of Parameters
The accuracy of each parameter is defined as specified by the applicable standard(s) for a given parameter. Frequency - EN 61000-4-30 specifies within 0.5 Hz Voltage Magnitude - EN 61000-4-30 specifies 1 % Flicker - EN 61000-4-15 specifies 0.1 pst and for EN61000-4-30 specifies < 0.5 state 1 and 0.1 for states 2 and 3. Unbalance- EN 61000-4-30 specifies 0.5 % Harmonics - EN 61000-4-30 speciifes 3 % Interharmonics - EN 61000-4-30 specifies 0.5% Dips and Swells - EN 61000-4-30 specifies 2%

Page 7 of 34

61000 System 61000-4-30 Test Report

2.3

Test Standards
Harmonics IEC 61000-4-7, 2003
General Guide on Harmonics and Interharmonics measurements and Instrumentation, for power supply sytems and equipment connected thereto

Flicker

IEC-61000-4-15, 2002

Flickermeter Functional and design specifications

Power Quality
Power Quality Measurement Methods

IEC 61000-4-30, 2003

2.4

Other Standards
Additional standards are referenced here that are not part of 61000-4-30 but are included in this section in reference to the Dranetz-BMI quality system.

ISO 17025 - 2005


General Requirements for the competence of testing and calibration laboratories.

ISO 9001 - 2000


Quality Management Systems Requirements

2.5

QUALITY SYSTEM
Dranetz-BMI operates a quality system that is fully documented, in terms of its procedures. The company has been granted and maintained continuous ISO 9001 Certification by Quality Management Institute (a division of CSA) since 1994. Also, the company has been granted and maintained continuous ISO 17025 Certification by Technology International (Europe), Ltd. since 2004. A copy of these certificates can be found in Annex A of this report.

Page 8 of 34

61000 System 61000-4-30 Test Report

3. Equipment Under Test


3.1 Test Configurations
There were two test configurations of the Encore Series 61000 System with two unit samples of each configuration that were under test and are detailed as follows. Configuration #1 Serial Number Module Slot 1 61MZP Module Serial Number 61MZPAA206 Cal Date
Feb 16, 2006 Jan 26, 2006

Cal Due
Feb 16, 2007 Jan 26, 2007 Apr 19, 2007 Mar 6, 2007

Encore Series 61000 Mainframe#1 5536 Voltage Module Encore Series 61000 Mainframe#2 5536 Voltage Module

61PQBA153

5536XA344 61PQBA237

---------61MZP

61MZPBA015 Apr 19,


2006

5536BA062

----------

Mar 6, 2006

Figure 1 Front View, Encore Series 61000 System

Page 9 of 34

61000 System 61000-4-30 Test Report

Figure 2 - Rear View, Configuration #1

Figure 3 5536 Voltage Module used with Configuration #1

Page 10 of 34

61000 System 61000-4-30 Test Report

Configuration #2

Mainframe Serial Number 61PQBA070

Module Slot 1 61MVS

Module Serial Number

Cal Date
2006

Cal Due
Jan 20, 2007 Jan 20, 2007

Encore Series 61000 Mainframe#3 Encore Series 61000 Mainframe#4

61MVSAA075 Jan 20,

61PQBA071

61MVS

61MVSAA074 Jan 20,


2006

Figure 4 - Rear View, Configuration #2

Page 11 of 34

61000 System 61000-4-30 Test Report

3.2

Equipment Under Test Constructional Details


Product Family The product family of compliance for Dranetz-BMI Models 61000PQ, 61000VCM, and 61MVS are claimed under the basis of similarity of construction and functionality and are described as follows. Series 61000PQ - Power Quality Monitor instrument capable of measuring power quality parameters in nominal frequency range of 50/60 Hz and including impulse transients. Series 61000VCM - Power Quality Monitor same as 61000PQ measures parameters in nominal frequency range of 50/60 but does not measure transients.

3.3

PRODUCT VARIANTS
The product models and variant model numbers are briefly described below. Similarities of Construction Construction variants of models 61000PQ and 61000VCM are functional derived from the same platform and compliance is claimed by similarities of design and functionality. Also, the noted constructional differences do not have an impact on the test results.

Page 12 of 34

61000 System 61000-4-30 Test Report

4 Test Equipment Used


4.1 Frequency, Flicker, Harmonics, Interharmonics, Voltage Magnitude
A Fluke 6100A was utilized to perform frequency, flicker, harmonics, interharmonics, and voltage magnitude . Equipment
Fluke Model 6100A Fluke Model 6100A

Serial Number
876149616 86094282

Calibration Date
Oct 26, 2005 Dec 21, 2005

Calibration Due
Oct 26. 2006 Dec 21, 2006

Reference
61000-4-30

Description Test Programs Fluke 6100A programmed parameters

Revision Level
1.0

Page 13 of 34

61000 System 61000-4-30 Test Report

4.2 Voltage Unbalance


A 3 phase test fixture consisting of the following equipment listed in the table were utilized to perform voltage unbalance test specified per IEC 61000-4-30.
Equipment Wavetek Model 75 (Phase A) Wavetek Model 75 (Phase B) Wavetek Model 75 (Phase C) Bogen Amplifier Model MT125C (Phase A) Bogen Amplifier Model MT125C (Phase B) Bogen Amplifier Model MT125C (Phase C) HP Digital Meter Model 3458A Serial Number B7260657 B92021737 B90040873 Asset# 083 Calibration Date Not Applicable Not Applicable Not Applicable Not Applicable Calibration Due Not Applicable Not Applicable Not Applicable Not Applicable

Asset# 513

Not Applicable

Not Applicable

Asset# 494

Not Applicable

Not Applicable

2823A05911

Mar 10, 2006

Mar 31, 2007

Page 14 of 34

61000 System 61000-4-30 Test Report

4.3 Technical Statements


Dranetz-BMI has been a leader in the market of Power parameter measurements since being first established in 1963. Later evolving into Power Quality Monitoring and Measurements by introducing its first Power Disturbance Analyzer in 1975. For over thirty years the company's reputation has long been established at the forefront of Power Measuring and Monitoring techniques and considered by many as the leader and competent in this area of expertise. To establish the validity of our test approach and test results we have employed the services of
Technology International (Europe), Ltd. 60 Shrivenham Hundred Business Park Shrivenham, Swindon, Wiltshire SN6 8TY, England

Since 1992, Technology International (Europe) Ltd. was appointed by the UK Department of Trade and Industry, and notified to the EU Commission, as a Competent Body for the EMC Directive, and later appointed as a Notified Body for the Low Voltage Directive, Machinery Directive and Radio and Telecommunication Terminal Equipment Directive. Dranetz-BMI contracted the services of Technology International Ltd. as an independent third party to competently audit the requirements of IEC 61000-4-30 , verify our testing methods and test results. Technology International Ltd. has audited our test process and as a competent body issued Dranetz-BMI certification to ISO 17025 to competently perform the tests required by IEC 61000-4-30. Additionally, Technology International Ltd. verified and audited the testing process on site, verified the test results, and issued a certificate of compliance to the requirements of the IEC 61000-4-30 standard: 2003, Class A for the Encore Series 61000 System. A copy of the certificate of compliance is included in Annex A of this report.

Page 15 of 34

61000 System 61000-4-30 Test Report

5 DETAILED TEST FILES and RESULTS


Frequency - 230V, 50 Hz
Applied Waveform freq42_1.bat freq46_1.bat freq50_1.bat freq53_1.bat freq57_1.bat freq42_2.bat freq46_2.bat freq50_2.bat freq53_2.bat freq57_2.bat freq42_3.bat freq46_3.bat freq50_3.bat freq53_3.bat freq57_3.bat Frequency 42.5 Hz 46.75 Hz 50 Hz 53.75 Hz 57.5 Hz 42.5 Hz 46.75 Hz 50 Hz 53.75 Hz 57.5 Hz 42.5 Hz 46.75 Hz 50 Hz 53.75 Hz 57.5 Hz Test State1 Test State1 Test State1 Test State1 Test State1 Test State2 Test State2 Test State2 Test State2 Test State2 Test State2 Test State2 Test State2 Test State2 Test State2 Comments
Allowable Measurement Accuracy 0.01 Hz

5536 Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Results 61MVS Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance

Frequency - 120V, 60 Hz
Applied Waveform freq51_1.bat freq55_1.bat Freq60_1.bat Freq64_1.bat Freq69_1.bat Freq51_2.bat Freq55_2.bat Freq60_2.bat Freq64_2.bat Freq69_2.bat Freq51_3.bat Freq55_3.bat Freq60_3.bat Freq64_3.bat Freq69_3.bat Frequency 51 Hz 55.5 Hz 60 Hz 64.5 Hz 69 Hz 51 Hz 55.5 Hz 60 Hz 64.5 Hz 69 Hz 51 Hz 55.5 Hz 60 Hz 64.5 Hz 69 Hz Test State1 Test State1 Test State1 Test State1 Test State1 Test State2 Test State2 Test State2 Test State2 Test State2 Test State3 Test State3 Test State3 Test State3 Test State3 Comments
Allowable Measurement Accuracy 0.01 Hz

5536 Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Results 61MVS Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance

Page 16 of 34

61000 System 61000-4-30 Test Report

Flicker 61000-4-15
Applied Waveform

230V, 50 Hz
Flicker Changes Per Minute (CPM) 1 2 7 39 110 1620 4000 4800 7 Pst Comments
Allowable Measurement Accuracy pst 0.05

5536 Pass

Results 61MVS Pass

50f_1.bat 50f_2.bat 50f_7.bat 50f_39.bat 50f_110.bat 50f_1620.bat 50f_4000.bat Not Required 50f_7p3.bat

1 1 1 1 1 1 1 1 3

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance Not Required within tolerance

Pass Pass Pass Pass Pass Pass Pass Pass Pass

Pass Pass Pass Pass Pass Pass Pass Pass Pass

Flicker 61000-4-15
Applied Waveform

120V, 60 Hz
Flicker Changes Per Minute (CPM) 1 2 7 39 110 1620 4000 7 Pst Comments
Allowable Measurement Accuracy pst 0.05

5536 Pass

Results 61MVS Pass

60f_1.bat 60f_2.bat 60f_7.bat 60f_39.bat 60f_110.bat 60f_1620.bat Not Required 60f_7p3.bat

1 1 1 1 1 1 1 3

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance Not Required within tolerance

Pass Pass Pass Pass Pass Pass Pass Pass

Pass Pass Pass Pass Pass Pass Pass Pass

Page 17 of 34

61000 System 61000-4-30 Test Report

Flicker 61000-4-30 Table 3


Applied Waveform Test State Flicker

230V, 50 Hz
Comments Results

50f1_0.bat 50f1_5.bat 50f1_10.bat 50f1_15.bat 50f1_20.bat 50f2_0.bat 50f2_5.bat 50f2_10.bat 50f2_15.bat 50f2_20.bat 50f3_0.bat 50f3_5.bat 50f3_10.bat 50f3_15.bat 50f3_20.bat

State 1 State 1 State 1 State 1 State 1 State 2 State 2 State 2 State 2 State 2 State 3 State 3 State 3 State 3 State 3

Changes Per Minute (CPM) 39 39 39 110 110 39 39 39 110 110 39 39 39 110 110

Pst

Allowable Measurement Accuracy

5536 Pass

61MVS Pass

0 5 10 15 20 0 5 10 15 20 0 5 10 15 20

within tolerance pst = < 0.5 within tolerance pst = 0.5 within tolerance pst = 1.0 within tolerance pst = 1.5 within tolerance pst = 2.0 within tolerance pst = < 0.5 within tolerance pst = 0.5 within tolerance pst = 1.0 within tolerance pst = 1.5 within tolerance pst = 2.0 within tolerance pst = < 0.5 within tolerance pst = 0.5 within tolerance pst = 1.0 within tolerance pst = 1.5 within tolerance pst = 2.0

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Flicker 61000-4-30 Table 3


Applied Waveform Test State Flicker

120V, 60 Hz
Comments Results

60f1_0.bat 60f1_5.bat 60f1_10.bat 60f1_15.bat 60f1_20.bat 60f2_0.bat 60f2_5.bat 60f2_10.bat 60f2_15.bat 60f2_20.bat 60f3_0.bat 60f3_5.bat 60f3_10.bat 60f3_15.bat 60f3_20.bat

State 1 State 1 State 1 State 1 State 1 State 2 State 2 State 2 State 2 State 2 State 3 State 3 State 3 State 3 State 3

Changes Per Minute (CPM) 39 39 39 110 110 39 39 39 110 110 39 39 39 110 110

Pst

Allowable Measurement Accuracy

5536 Pass

61MVS Pass

0 5 10 15 20 0 5 10 15 20 0 5 10 15 20

within tolerance pst = < 0.5 within tolerance pst = 0.5 within tolerance pst = 1.0 within tolerance pst = 1.5 within tolerance pst = 2.0 within tolerance pst = < 0.5 within tolerance pst = 0.5 within tolerance pst = 1.0 within tolerance pst = 1.5 within tolerance pst = 2.0 within tolerance pst = < 0.5 within tolerance pst = 0.5 within tolerance pst = 1.0 within tolerance pst = 1.5 within tolerance pst = 2.0

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Page 18 of 34

61000 System 61000-4-30 Test Report

Harmonics - 230V, 50 Hz
Applied Waveform Test State Harmonics Comments Results

50h1_0.bat 50h1_25.bat 50h1_50.bat 50h1_75.bat 50h1_100.bat 50h2_0.bat 50h2_25.bat 50h2_50.bat 50h2_75.bat 50h2_100.bat 50h3_0.bat 50h3_25.bat 50h3_50.bat 50h3_75.bat 50h3_100.bat

State 1 State 1 State 1 State 1 State 1 State 2 State 2 State 2 State 2 State 2 State 3 State 3 State 3 State 3 State 3

Twice the values of 61000-2-4 Class 3 0% 25% 50% 75% 100% 0% 25% 50% 75% 100% 0% 25% 50% 75% 100%

Allowable Measurement Accuracy 5%

5536 Pass

61MVS Pass

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Harmonics - 120V, 60 Hz
Applied Waveform Test State Harmonics Comments Results

60h1_0.bat 60h1_25.bat 60h1_50.bat 60h1_75.bat 60h1_100.bat 60h2_0.bat 60h2_25.bat 60h2_50.bat 60h2_75.bat 60h2_100.bat 60h3_0.bat 60h3_25.bat 60h3_50.bat 60h3_75.bat 60f3_20.bat

State 1 State 1 State 1 State 1 State 1 State 2 State 2 State 2 State 2 State 2 State 3 State 3 State 3 State 3 State 3

Twice the values of 61000-2-4 Class 3 0% 25% 50% 75% 100% 0% 25% 50% 75% 100% 0% 25% 50% 75% 100%

Allowable Measurement Accuracy 5%

5536 Pass

61MVS Pass

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Page 19 of 34

61000 System 61000-4-30 Test Report

Interharmonics - 230V, 50 Hz
Applied Waveform Test State Harmonics Comments Results

50i1_0.bat 50i1_25.bat 50i1_50.bat 50i1_75.bat 50i1_100.bat 50i2_0.bat 50i2_25.bat 50i2_50.bat 50i2_75.bat 50i2_100.bat 50i3_0.bat 50i3_25.bat 50i3_50.bat 50i3_75.bat 50i3_100.bat

State 1 State 1 State 1 State 1 State 1 State 2 State 2 State 2 State 2 State 2 State 3 State 3 State 3 State 3 State 3

Twice the values of 61000-2-4 Class 3 0% 25% 50% 75% 100% 0% 25% 50% 75% 100% 0% 25% 50% 75% 100%

Allowable Measurement Accuracy 5%

5536 Pass

61MVS Pass

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Interharmonics - 120V, 60 Hz
Applied Waveform Test State Harmonics Comments Results

60i1_0.bat 60i1_25.bat 60i1_50.bat 60i1_75.bat 60i1_100.bat 60i2_0.bat 60i2_25.bat 60i2_50.bat 60i2_75.bat 60i2_100.bat 60i3_0.bat 60i3_25.bat 60i3_50.bat 60i3_75.bat 60i3_100.bat

State 1 State 1 State 1 State 1 State 1 State 2 State 2 State 2 State 2 State 2 State 3 State 3 State 3 State 3 State 3

Twice the values of 61000-2-4 Class 3 0% 25% 50% 75% 100% 0% 25% 50% 75% 100% 0% 25% 50% 75% 100%

Allowable Measurement Accuracy 5%

5536 Pass

61MVS Pass

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Page 20 of 34

61000 System 61000-4-30 Test Report

Unbalance 230V, 50Hz


Test State Unbalance Comments 5536 Pass Pass Pass Pass Pass Pass Pass Results 61MVS Pass Pass Pass Pass Pass Pass Pass

State 1 State 1 State 1 State 1 State 1 State 2 (100%) State 2 (200%) State 3 (100%)

Percent unbalance of declared voltage 1% 2% 3% 4% 5% 0.73 of declared voltage - Phase A 0.80 of declared voltage - Phase B 0.87 of declared voltage - Phase C 0.73 of declared voltage - Phase A 0.80 of declared voltage - Phase B 0.87 of declared voltage - Phase C 1.52 of declared voltage - Phase A 1.40 of declared voltage - Phase B 1.28 of declared voltage - Phase C

Allowable Measurement Accuracy 5%

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance

Pass within tolerance within tolerance Pass

Pass

Pass

120V, 60Hz
Test State Unbalance Comments 5536 Pass Pass Pass Pass Pass Pass Pass Results 61MVS Pass Pass Pass Pass Pass Pass Pass

State 1 State 1 State 1 State 1 State 1 State 2 (100%) State 2 (200%) State 3 (100%) State 3 (200%)

Percent unbalance of declared voltage 1% 2% 3% 4% 5% 0.73 of declared voltage - Phase A 0.80 of declared voltage - Phase B 0.87 of declared voltage - Phase C 0.73 of declared voltage - Phase A 0.80 of declared voltage - Phase B 0.87 of declared voltage - Phase C 1.52 of declared voltage - Phase A 1.40 of declared voltage - Phase B 1.28 of declared voltage - Phase C 1.52 of declared voltage - Phase A 1.40 of declared voltage - Phase B 1.28 of declared voltage - Phase C

Allowable Measurement Accuracy 5%

within tolerance within tolerance within tolerance within tolerance within tolerance within tolerance

Pass within tolerance within tolerance Pass

Pass

Pass

within tolerance

Pass

Pass

Page 21 of 34

61000 System 61000-4-30 Test Report

Voltage Magnitude 230V, 50 Hz


Applied Waveform Test State Voltage Magnitude Comments Results

Allowable Measurement Accuracy 0.1%

5536 Pass Pass Pass Pass Pass Pass N/A

61MVS Pass Pass Pass Pass Pass Pass N/A

50v1_25.bat 50v1_50.bat 50v1_75.bat 50v1_100.bat 50v1_200.bat N/A

State 1 State 1 State 1 State 1 State 1 State 2

N/A

State 3

25% 50% 75% 100% 200% Determined by flicker, unbalance, harmonics, & interharmonics Determined by flicker, unbalance, harmonics, & interharmonics

within tolerance within tolerance within tolerance within tolerance within tolerance Refer to test results for flicker, unbalance, harmonics, & interharmonics Refer to test results for flicker, unbalance, harmonics, & interharmonics

N/A

N/A

120V, 60 Hz
Applied Waveform Test State Voltage Magnitude Comments Results

Allowable Measurement Accuracy 0.1%

5536 Pass Pass Pass Pass Pass Pass N/A

61MVS Pass Pass Pass Pass Pass Pass N/A

60v1_25.bat 60v1_50.bat 60v1_75.bat 60v1_100.bat 60v1_200.bat N/A

State 1 State 1 State 1 State 1 State 1 State 2

N/A

State 3

25% 50% 75% 100% 200% Determined by flicker, unbalance, harmonics, & interharmonics Determined by flicker, unbalance, harmonics, & interharmonics

within tolerance within tolerance within tolerance within tolerance within tolerance Refer to test results for flicker, unbalance, harmonics, & interharmonics Refer to test results for flicker, unbalance, harmonics, & interharmonics

N/A

N/A

Page 22 of 34

61000 System 61000-4-30 Test Report

Dips & Swells 230 V, 50 Hz


Applied Waveform Test Percent amplitude of declared voltage 88% 89% 91% 88% 89% 91% 88% 89% 91% 88% 89% 91% 109% 111% 112% 109% 111% 112% 109% 111% 112% 109% 111% 112% Comments Results

50_88_1.bat 50_89_1.bat 50_91_1.bat 50_88_1.bat 50_89_1.bat 50_91_1.bat 50_88_10.bat 50_89_10.bat 50_91_10.bat 50_88_50.bat 50_89_50.bat 50_91_50.bat 50_09_1.bat 50_11_1.bat 50_12_1.bat 50_09_2.bat 50_11_2.bat 50_12_2.bat 50_09_10.bat 50_11_10.bat 50_12_10.bat 50_09_50.bat 50_11_50.bat 50_12_50.bat

Dip 1 cycle Dip 1 cycle Dip 1 cycle Dip 2 cycles Dip 2 cycles Dip 2 cycles Dip 10 cycles Dip 10 cycles Dip 10 cycles Dip 50 cycles Dip 50 cycles Dip 50 cycles Swell 1 cycle Swell 1 cycle Swell 1 cycle Swell 2 cycles Swell 2 cycles Swell 2 cycles Swell 10 cycles Swell 10 cycles Swell 10 cycles Swell 50 cycles Swell 50 cycles Swell 50 cycles

Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded

61MVS Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Page 23 of 34

61000 System 61000-4-30 Test Report

120 V, 60 Hz
Applied Waveform Test Percent amplitude of declared voltage 88% 89% 91% 88% 89% 91% 88% 89% 91% 88% 89% 91% 109% 110% 111% 109% 110% 111% 109% 110% 111% 109% 110% 111% Comments Results

60_88_1.bat 60_89_1.bat 60_91_1.bat 60_88_1.bat 60_89_1.bat 60_91_1.bat 60_88_10.bat 60_89_10.bat 60_91_10.bat 60_88_60.bat 60_89_60.bat 60_91_60.bat 60_09_1.bat 60_11_1.bat 60_12_1.bat 60_09_2.bat 60_11_2.bat 60_12_2.bat 60_09_10.bat 60_11_10.bat 60_12_10.bat 60_09_60.bat 60_11_60.bat 60_12_60.bat

Dip 1 cycle Dip 1 cycle Dip 1 cycle Dip 2 cycles Dip 2 cycles Dip 2 cycles Dip 10 cycles Dip 10 cycles Dip 10 cycles Dip 60 cycles Dip 60 cycles Dip 60 cycles Swell 1 cycle Swell 1 cycle Swell 1 cycle Swell 2 cycles Swell 2 cycles Swell 2 cycles Swell 10 cycles Swell 10 cycles Swell 10 cycles Swell 60 cycles Swell 60 cycles Swell 60 cycles

Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded No Event Event Recorded Event Recorded

61MVS Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass

Page 24 of 34

61000 System 61000-4-30 Test Report

ANNEX A

Certificates Of Compliance

IEC 61000-4-30 ISO 9001

ISO 17025

Page 25 of 34

61000 System 61000-4-30 Test Report

IEC 61000-4-30 Certificate

Page 26 of 34

61000 System 61000-4-30 Test Report

ISO 9001 Certificate

Page 27 of 34

61000 System 61000-4-30 Test Report

ISO 17025 Certificate

Page 28 of 34

61000 System 61000-4-30 Test Report

ANNEX B Test Instruments and Facilities

Test Instruments used for IEC 61000-4-30

Test Instruments used for additional verification

Page 29 of 34

61000 System 61000-4-30 Test Report

Test Instruments used for 61000-4-30


Test equipment used for IEC 61000-4-30 has been noted accordingly in this report and documented appropriately.

Test Instruments used for additional verification


The following is a summary of additional test equipment that Dranetz-BMI utilizes to meet or exceed additional compliance tests or criteria not required by 61000-4-30. Although, these instruments were not used for this test it is included to demonstrate Dranetz-BMIs commitment and excellence in meeting or exceeding applicable standards and measurement methods for electrical instrumentation.
Description Electrostatic Discharge Fast Transients / Burst Generator Surge Generator Mainframe Surge Network Ring Wave Surge Network Bi-Wave EmissionsSpectrum Analyzer EmissionsAntenna EmissionsLine Impedance Stabilization Network Emissions Software Program EmissionsRF Shield Room Insulation Tester Hipot Temperature & Humidity Chambers Manufacturer Keytek Schaffner Keytek Keytek Keytek Wayne Kerr Model Number Series 2000 NSG 1025 711A P1 P3 SSA1000A Serial Number 8702041 52393337 8507388 8507393 8507394 9700001627 Standard reference 61000-4-2 61000-4-4 61000-4-5 61000-4-5 61000-4-5 EN 55011

Wayne Kerr Farnell

1JELPA30102 LSN30B

9700001413 000268

EN 55011 EN 55011

Wayne Kerr

EASY 1

N/A

EN 55011

Lindgren Hipotronics Tenney

34-1/1-0 100 Series Benchmaster

3754 1310-1196 SB11421-7 SB11421-21

EN 55011 61010-1 N/A

Page 30 of 34

61000 System 61000-4-30 Test Report

Test Facility Photographs


Electrostatic Discharge

Electrical Fast Transients

Page 31 of 34

61000 System 61000-4-30 Test Report

Surge Network

Emissions Spectrum Analyzer & Computer Setup

Page 32 of 34

61000 System 61000-4-30 Test Report

Emissions Shielded Room

Page 33 of 34

61000 System 61000-4-30 Test Report

Insulation Tester

Temperature & Humidity Chamber

Page 34 of 34