Product Information Xforce P, Xforce HP and Xforce K load cells

Xforce Family

Xforce HP, Fmax 50 N

Patented Xforce load cells exclusive at Zwick Roell Xforce load cells are only available from Zwick Roell. These high accuracy load cells are used for all Zwick load frames. Applicational range Xforce load cells are ideal for tensile, compression and flexural / bending testing as well as through-zero cyclic testing. Parasitic forces All Xforce load cells are highly insensitivie to parasitic forces (transverse forces, bending moments, torque moments). Construction type and configuration of the load cells All Xforce load cells are based on the rotation-symmetric or axis-symmetric principle and are therefore highly insensitive to transverse forces. The low overall height allows shorter test arrangements to be achieved with, for example, reduced bending issues. The construction type provides a high operating force, very small measurement distances and high stiffness. Simple mechanical pin system All load cells have an accurately fitted play-free connection pin so that grips and test fixtures can be quickly and firmly aligned on the machine test axis. Self identifying sensor plugs The smart load cells have a unique electronic identification system stored on internal EEPROM: testXpert II testing software automatically identifies sensor type and properties no calibration required at sensor change

force and travel limits automatically imported sensor overloads plus date stored in EEPROM Second connection bolt for two test areas Xforce K load cells are the only load cells which can be used for two test areas because of double sided connection studs. Faster load cell changeover by option connection via mounting stud If several load cells are used or if there is to be a frequent changing of load cells, we recommend the option connection via mounting stud. This offers flexibility and significant time savings. The load cell cables will not be strained by screwing/ unscrewing when changing load cells. Reference positions for various test set-ups are automatically achieved (with the threaded mounting the reference position changes depending on the amount of threads used).

Xforce K with a second connection

PI 716 2.0212

Xforce quick-change system

Product Information Xforce P, Xforce HP and Xforce K load cells


System calibration Before delivery Zwick Roell calibrates each load cell together with the testing machine, its drive system and the measurement and control electronics as one complete system. This ensures that our customers can carry out tests to the highest possible accuracy as soon as the machine is commissioned on site. Calibration and accuracy according to ISO 7500-1 The Xforce load cells far exceed the requirements of the Standards. Xforce load cells fulfill all five criteria of the accuracy classes according to ISO 7500-1, over a very wide measurement range. (see fig. 1) In connection with testControl the Xforce HP and Xforce K load cells typically offer a better linearity (relative accuracy) than is required by ISO 7500. The linearity error (relative accuracy) is less than 1% of the measured value even at 0.1% of the load cell capacity, it is less than 0.25% from 0.4% of load cells capacity. (see fig. 2) This very wide measurement range often avoids the need for a second load cell, thereby saving the cost of a second load cell and subsequent annual calibration costs. The entire measurement range is available even with large preloads caused by heavy test fixtures or specimen grips. The load cell can be used to the full capacity even when fixtures or grips represent 50 % of the load cell capacity. Over load security, force limits The load cells are extremely robust, and have a high operating force (up to 150% of their maximum capacity), so the need for overload protection can often be avoided (e.g. preloaded spring packages, mechanical bearings or guides for transverse force absorption). Xforce load cells are very robust and can withstand loads up to 300 % of their capacity without mechanical damage and up to 150% without zero shift. Xforce HP load cells in the 5 - 100 N range have integrated overload protection and are therefore protected against greater overloads (tensile forces, compressive forces, bending moments etc.). The unique use of both soft and hard limit switches within the testing machine control system allows the load cells and test fixtures to be easily protected against crosshead overtravel (only integrated in Zwick Roell machines and software). Load limits can be configured in testXpert II to automatically switch off the test system at predefined levels, thereby protecting the load cell.
Fulfillment of all 5 criteria according to ISO 7500-1

ISO 7500: Accuracy grade 0.5


Accuracy < 0.5% Repeatability < 0.5% Reversibility < 0.75% Zero error < 0.05% ISO 7500: Accuracy Resolution < 0.25% Accuracy < 0.5% Repeatability < 0.5% Reversibility < 0.75% Zero error < 0.05% Resolution < 0.25%

Xforce P grade 0.5


fulfill all 5 criterias from 2% of FN

Xforce Xforce HP fulfill all P 5 criterias fulfill all Xforce K from 1% of FN 5 criterias
from 2% of FN 5 criterias from 1% of FN

0.2% 0.4%

Measurement range Xforce HP 100% of FN ... fulfill all 1%

Xforce K

Measurement range ISO 7500: Accuracy grade 1 0.2% < 1% ... 100% of FN 1% Accuracy 0.4% Repeatability < 1% Reversibility < 1,5% Zero error < 0,1% ISO 7500: Accuracy grade 1 Resolution < 0,5% Accuracy < 1% Repeatability < 1% Xforce P Fulfill all 5 criterias Reversibility < 1,5% from 0.4% of FN Zero error < 0,1% Resolution < 0,5% Xforce HP Fulfill all 5 criterias Xforce from 0.2% of F criterias Xforce K P Fulfill all 5 N from 0.4% of FN 0.2% Xforce 0.4%

HP Xforce K

Measurement range 1% all 5 criterias100% of FN ... Fulfill


from 0.2% of FN

Fig. 1 Accuracy grade 1: Xforce HP and XforceMeasurement range K load cells fulfill the 0.2% 0.4% 1% requirements of all five criteria from 0.2% of... 100% of FN the capacity, Xforce P load cells from 0.4% of the capacity. Accuracy grade 0.5: Xforce load cells fulfill the requirements of all five criteria from 1% rsp. 2% of the capacity.

Xforce HP typical accuracy < 0.25% Xforce to from 0.4% of F Linearity according K ISO 7500-1 N Xforce HP typical accuracy < 1% Xforce HP typical accuracy < 0.25% from 0.1% of FN Xforce K from 0.4% of FN Xforce K
0.1% 0.4% Xforce HP Measurement range 1% ... typical accuracy < 1%100% of FN from 0.1% of FN Measurement range ... 100% of FN

Xforce K
0.1% 0.4%

1%

Fig. 2 The linearity error (relative accuracy) is < 0.25% in the range from 0.4 ... 100% of the capacity. The measurement range which achieves a relative accuracy < 1%, typically starts at 0.1% of the rated capacity.

Xforce P (up to 100 N)

Xforce HP (up to 100 N)

Xforce P Xforce HP (200 N - 250 kN) (200 N - 10 kN) Multiple bending beam 150% of FN 150% of FN 300% of FN 100% of FN 150% of FN 150% of FN 300% of FN 100% of FN 150% of FN 150% of FN 300% of FN 100% of FN

Xforce K (10 - 250 kN) Bending ring

150% of FN 150% of FN 300% of FN 100% of FN 150% of FN 150% of FN 1.000 N (2 200% to 4.000% of FN(2 (200 N) 0.07% of Fact / mm 0.1% of FN / mm +10...+60 C -30...+60 C 0.005% of FN per K 0.005% of FN per K 2mV/V IP 42 3.5 m 2mV/V IP 54 3.5 m 2mV/V IP 54 3.5 m 0.01% of FN per K 0.004% of FN per K 0.01% of FN per K 0.0025% of FN per K 0.2% of FN / mm +10...+60 C -30...+60 C 0.2% of FN / mm +10...+60 C -30...+60 C 0.25% of Fact / mm 0.07% of Fact / mm 0.015% of Fact / mm 0.005% of FN / mm +10...+60 C -30...+60 C 0.001% of FN per K 0.004% of FN per K

0.25% of Fact / mm

0.2% of FN / mm +10...+60 C -30...+60 C

0.01% of FN per K

Design Force limits/ranges Operating force Limit force Force at break Limiting transverse load at mounting stud Parasitic influences Influence of bending moment Influence of torque Temperature range Storage temperature Temperature influence on the zero-signal TK0 Temperature influence on the characteristic value TKC Further characteristics Nominal characteristic Protection class Cable length

0.01% of FN per K

2mV/V IP 42 3.5 m

(1

Product Information Xforce P, Xforce HP and Xforce K load cells

(2

HP/K version recommended for applications in which bending moments or overloads in the tensile/compression direction may occur. Via integrated mechanical support frame.

Explanations regarding parasitic influences and force measurement in general can be found in Application Note AI00108 (Force measurement with Xforce load cells), which we will be happy to make available on request.

Xforce P ( up to 100 N)

Xforce HP

(1

(up to 100 N)

Maximum capacity FN [kN] Maximum capacity FN [lbf] Item number Item number for ProLine (2 Accuracy grade 1 from (0.4% of FN) Accuracy grade 0.5 from (2% of FN) Connection Dia. of mounting stud [mm] Limiting bending moment [Nm] (3 Limiting torque [Nm] 10 2.25 060253 063920 0.04 N 0.2 N M28x1.5 8 1.0 10 20 4.50 060256 063921 0.08 N 0.4 N M28x1.5 8 1.30 10 50 11.24 060257 063922 0.2 N 1N M28x1.5 8 2.0 10

5 1.12 057091 063919 0.02 N 0.1 N M28x1.5 8 0.7 10

100 22.48 060258 063923 0.4 N 2N M28x1.5 8 3.0 10

5 1.12 063924 063927 0.01 N 0.05 N M28x1.5 8 0.7 10

10 2.25 063925 063929 0.02 N 0.1 N M28x1.5 8 1.0 10

20 4.50 063926 063930 0.04 N 0.2 N M28x1.5 8 1.30 10

50 11.24 060259 063932 0.1 N 0.5 N M28x1.5 8 6 10

100 22.48 060260 063933 0.2 N 1N M28x1.5 8 6 10

(1

(2

HP version recommended for applications in which bending moments or overloads in the tensile/compression direction may occur. Only in combination with a ProLine load frame; please see note regarding this. (3 Max. bending moments Mb for load cells not loaded in the measurement direction. If the load cell is simultaneously loaded to its nominal load the values must be halved.

Product Information Xforce P, Xforce HP and Xforce K load cells


Xforce P (200 N - 10 kN) Maximum capacity FN [kN] Maximum capacity FN [lbf] Item number Item number for ProLine (2 Accuracy grade 1 from (0.4% of FN) Accuracy grade 0.5 from (2% of FN) Connection via Dia. of mounting stud [mm] Limiting bending moment [Nm] (3 (4 Limiting torque [Nm] (3 Xforce P (20 - 250 kN) Maximum capacity FN [kN] Maximum capacity FN [lbf] Item number Item number for ProLine (2 Accuracy grade 1 from (0.4% of FN) Accuracy grade 0.5 from (2% of FN) Connection via Dia. of mounting stud [mm] Limiting bending moment [Nm] (3 (4 Limiting torque [Nm] (3 Xforce HP (6 (200 N - 10 kN) Maximum capacity FN [kN] Maximum capacity FN [lbf] Item number Item number for ProLine (2 Accuracy grade 1 from (0.2% of FN) Accuracy grade 0.5 from (1% of FN) Connection via Dia. of mounting stud [mm] Limiting bending moment [Nm] (3 (4 Limiting torque [Nm] (3 Xforce K (6 (10 - 250 kN) Maximum capacity FN [kN] Maximum capacity FN [lbf] Item number Item number for ProLine (2 Accuracy grade 1 from (0.2% of FN) Accuracy grade 0.5 from (1% of FN) Connection via Dia. of mounting stud [mm] Limiting bending moment [Nm] (4 Limiting torque [Nm]
(1 (2

0.2 45 011563 018542 0.8 N 4N M28x1.5 8 3 (2) 14 (5)

0.5 112 011562 018540 2N 10 N M28x1.5 8 7 (5) 35 (7)

0.5 112 057993 058423 2N 10 N M28x1.5 20 7 (5) 35 (7)

1 225 011560 018539 4N 20 N M28x1.5 20 (1 17 (15) 50 (17)

2.5 562 011558 018538 10 N 50 N M28x1.5 20 34 (30) 80 (17)

5 1124 011556 018537 20 N 100 N M28x1.5 20 58 (50) 130 (17)

10 2248 017955 40 N 200 N flange 1 20 115 (80) 200 (17)

10 2248 011554 018536 40 N 200 N M28x1.5 20 115 (80) 200 (17

20 4496 017907 019242 80 N 400 N flange 1 36 460 (250) 1500 (250) 0.2 45 011571 018548 0.4 N 2N M28x1.5 8 3 (2) 14 (5) 10 2248 325944 20 N 100 N flange 1 20 500 500

30 6744 017908 019246 120 N 600 N flange 1 36 500 (250) 1800 (250)

50 11240 017909 019248 200 N 1000 N flange 1 36 650 (250) 3000 (250)

100 22481 017910 019254 400 N 2000 N flange 2 60 4500 (3500) 10000 (6500) 2.5 562 011568 018545 5N 25 N M28x1.5 20 34 (30) 80 (17)

150 33721 017911 600 N 3000 N flange 2 60 5000 (4000) 12500 (5800)

250 56202 017912 1000 N 5000 N flange 2 60 6000 (5000) 15000 (11000) 10 2248 011565 018543 20 N 100 N M28x1.5 20 115 (80) 200 (17) 250 56202 318930 500 N 2500 N flange 2 60 30000 55000

0.5 112 011570 018547 1N 5N M28x1.5 8 7 (5) 35 (7)

0.5 112 057991 058424 1N 5N M28x1.5 20 7 (5) 35 (7)

1 225 011569 018546 2N 10 N M28x1.5 20 (1 17 (15) 50 (17)

5 1124 011566 018544 10 N 50 N M28x1.5 20 58 (50) 130 (17)

10 2248 017953 018554 20 N 100 N flange 1 20 115 (80) 200 (17) 150 33721 320304 300 N 1500 N flange 2 60 8000 20000

(5

20 4496 318936 325222 40 N 200 N flange 1 36 600 500

30 6744 325642 325644 60 N 300 N flange 1 36 700 500

50 11240 318934 325223 100 N 500 N flange 1 36 1100 1800

100 22481 318932 325328 200 N 1000 N flange 2 60 4800 10000

Note: with Xforce load cells the mounting stud diameter has been changed from 8 to 20 mm for the 1kN load cell! Only in combination with a ProLine load frame; please see note regarding this. (3 The values relating to the limit moments for the connecting system are in brackets. If these are exceeded, re-calibration will be necessary. (4 Max. bending moments Mb for load cells (LC) not loaded in the measurement direction. If the LC is simultaneously loaded to its nominal load the values must be halved. (5 Already included: Option 2. Mounting studs for use of the load cell in two test areas. (6 HP/K version recommended for applications in which bending moments or overloads in the tensile/compression direction may occur.

All data at ambient temperature.