Neutron Activation Analysis
Neutron Activation Analysis
Gamma Detector
The prepared sample and standard sample are placed in a detector one at a time. The detector system counts and records gamma radiation emissions for a period of time.
Time varies, but is usually in the range of 5 minutes to an hour.
Specialized software analyzes radiation peaks. Peak data is correlated to specific elements for identification and quantification.
Electrons > autoradiography Gammas > NAA Similar story for many (not all) elements
Gamma-ray spectrometer
gamma ray
+
e h e h eh Voltage supply
Germanium crystal
counts
display energy
11
activity
short 1/2-life
short
irradiation time
long
activity
short 1/2-life
long 1/2-life
short
delay time
long
12
Advantages of NAA
Simultaneous analysis for many elements Chemical preparation rarely necessary Non-destructive testing Reliable quantitative results Very high sensitivity (for some elements)
13
2 3 5 23 22 12 4 2
73
19 elements not even listed! Bottom line great analysis techniqueIF youve got the right problem
14
Gamma-ray spectrum showing several short-lived elements measured in a sample of pottery irradiated for 5 seconds, decayed for 25 minutes, and counted for 12 minutes with an HPGe detector.
Gamma Spectra from a Sample of Pottery Irradiated for 5 Seconds, Decayed for 25 Minutes, and Counted for 12 Minutes with an HPGe Detector.
Gamma Spectra from a Sample of Pottery Irradiated for 24 Hours, Decayed for 9 Days, and Counted for 30 Minutes on a HPGe Detector.