Feedback Control of Type 1 Diabetes: Sandip D. Kulkarni
Feedback Control of Type 1 Diabetes: Sandip D. Kulkarni
Feedback Control of Type 1 Diabetes: Sandip D. Kulkarni
Sandip D. Kulkarni
By Sandip Kulkarni 1
Type 1 Diabetes Mellitus
Hyperglycemia: Kidney failure, heart
disease, blindness, brain damage
180 mg/dl
Target
Blood
Glucose (BG)
range
Overcom-
70 mg/dl pensation
Hypoglycemia: Neurological
60 mg/dl damage, unconsciousness,
seizures, death Severe
Hypoglycemia
By Sandip Kulkarni 2
Glucose Sensing
Self Monitoring of Blood Continuous Glucose
Glucose (SMBG) Monitoring (CGM) Device
Pricks into
the
abdominal
Finger prick Noisy signals skin
Provides estimates of blood Delay
glucose (BG) only when the Sensor faults (drift, dropout)
user samples Requires calibration with SMBG
Sensor performance variability
Requires a control solution
(+/- 30% or more observed**)
Inaccurate measurement (+/- FDA-approved only for use with
20% observed*) SMBG
By Sandip Kulkarni 3
Insulin Delivery Mechanisms
Continuous Subcutaneous Insulin Infusion (CSII)
User entry:
By Sandip Kulkarni 6
Linearized discrete-time model
u(k)
insulin infusion GLUCOSE
x(k+1)=Ax(k)+Bu(k) y(k)=Cx(k)
+Gc(k) +v(k)
SENSOR
d(k)
meal
States
G glucose
X Insulin sensitivity
I sc1
Insulin in Intradermal
I
sc 2 Compartments
x
I p Insulin in Plasma
Gsc Infused Glucose
Q
1
Stomach Compartments C=(100000
Q2 By Sandip Kulkarni
00) 7
What happens today...
Meals,
Exercise
INSULIN
PUMP
bolus
basal
Open-loop
Subject
therapy
GLUCOSE
SENSOR
By Sandip Kulkarni 8
Control Block Diagram
MPC
bolus
basal + Safety
Mod.
Subject
GLUCOSE
SENSOR
By Sandip Kulkarni 9
18 Hour Closed Loop Control Experiment in
Hospital
4 PM: 30 min 7 PM: manual 4.4u pre-meal Overnight
220 2.2
exercise
180 1.8
160 1.6
140 1.4
12:00 14:00 16:00 18:00 20:00 22:00 00:00 02:00 04:00 06:00 08:00
Finger Sticks CGM Insulin injection
By Sandip Kulkarni 10