Disinfection of Endoscope 2
Disinfection of Endoscope 2
Contamination
Preperation
Precleaning
after examination
1. Drying
2. Functional check
3. Maintanance
4. Storage
Cleaning
Flushing
Flushing
Disinfection
Pre-cleaning
Pre-clean
immediately after
the examination!
Care & Maintenance
Manual Cleaning and disinfection
Released Chemicals
Pre-cleaning
Immerse distal end in a
basis with cleaning
solution and alternately
actuate suction and
air/water valves
Disconnect
endoscope from
supply units
Pre-cleaning
Safely transport to
the separate
reprocessing room
(unclean side)
Leakage testing
Connect leakage tester, and
increase pressure until
the indicator is within the blue
range of the manometer
(approx. 160 mmHg)
Process documentation
Automated reprocessing
Process documentation
Features
Validated reprocessing
Standard luer lock connectors
All channels involved
16
Endoscopic Reprocessing
Further reprocessing connections:
Irrigation tube for Water-Jet Channel
(Luer Connection)
13991SWJ
13925xKS only
(Extend of delivery)
17
Endoscopic Reprocessing
Connection tubes on the handle 13925xKS
long Colonoscope
18
Endoscopic Reprocessing
Connection tubes (cleaning adaptors) on the
supply connector
19
Endoscopic Reprocessing
Connected scope in the Cleaning and
Disinfection machine
20
Endoscopic Reprocessing
Wassenburg WD440
Maintenance & storage
Carefully dry all channels
using medical-grade
compressed air (pressure
limited to 0.5 bar)
Check endoscope
functionality in
accordance with
instructions for use
Maintenance & storage