QA Tyyi Tyyii Yti Rtrtrtuutyytrty Ytyi Total

You might also like

Download as ppt, pdf, or txt
Download as ppt, pdf, or txt
You are on page 1of 32

Quality is a Lousy Idea-

If it’s Only an Idea


Quality Assurance vs.
Quality Control
Quality Assurance Quality Control
An overall A series of
management plan to analytical
guarantee the measurements used
integrity of data to assess the
(The “system”) quality of the
analytical data
(The “tools”)
True Value vs. Measured
Value
True Value Measured Value
The known, The result of an
accepted value of individual’s
a quantifiable measurement of a
property quantifiable
property
Accuracy vs. Precision
Accuracy Precision
How well a How well a series
measurement of measurements
agrees with an agree with each
accepted value other
Accuracy vs. Precision
Systematic vs.
Random Errors
Systematic Error Random Errors
Avoidable error due Unavoidable errors
to controllable that are always
variables in a present in any
measurement. measurement.
Impossible to
eliminate
Quality Control Measures
• Standards and Calibration
• Blanks
• Recovery Studies
• Precision and Accuracy Studies
• Method Detection Limits
• NJQLs
Standards and Calibration
• Prepared vs. Purchased Standard
• Signals: Peak Area, Beer’s Law
• Calibration Curves
• Continuing Calibration Checks
• Internal Standards
• Performance Testing.
Calibration Curves
Graphical representation of the
relationship between:

• The concentration of the analyte


and
• The analytical signal
Calibration Curve for DDT
500
6
Peak area x 10

400
y = 9.3005x + 4.3313
300

200 2
R = 0.9989
100

0
0 10 20 30 40 50 60

Parts per trillion DDT


Continuing Calibration
Verification
• Many methods don’t require that
daily calibration curves are
prepared
• A “calibration verification” is
analyzed with each batch of samples
Sample Batch
• 10 - 20 samples (method defined)
or less
• Same matrix
• Same sample prep and analysis
• Contains a full set of
QC samples
Internal Standards
• A compound chemically similar to
the analyte
• Not expected to be present in the
sample
• Cannot interfere in the analysis
• Added to the calibration standards
and to the samples in identical
amounts.
Internal Standards
• Refines the calibration process
• Analytical signals for calibration
standards are compared to those
for internal standards
• Eliminates differences in random
and systematic errors between
samples and standards
Performance Testing
Blind samples submitted to
laboratories ?
?

Labs must periodically ?


analyze with acceptable
results in order to maintain
accreditation
Blanks, Blanks, Blanks
• Laboratory Reagent Blanks
• Instrument Blanks

• Field Reagent Blanks

• Trip Blanks
Laboratory Reagent Blanks
• Contains every reagent used in the
analysis
• Is subjected to all analytical
procedures
• Must give signal below detection
limit
• Most methods require one with
every batch
Instrument Blank
• A clean sample (e.g., distilled water)
processed through the instrumental
steps of the measurement process; used
to determine instrument
contamination.
Field Reagent Blanks
• Prepared in the lab, taken to the
field

• Opened at the sampling site,


exposed to sampling equipment,
returned to the lab.
Trip Blanks
• Prepared in the lab, taken to the
field
• Not opened
• Returned to the lab

• Not always required in EPA methods


Recovery Studies
• Matrix Spikes
• Laboratory Control Samples

• Surrogates .
Matrix Spikes
• Sample spiked with a known
amount of analyte
• Subjected to all sample prep and
analytical procedures
• Determines the effect of the matrix
on analyte recovery
• Normally one per batch
Laboratory Control Sample
• Analyte spiked into reagent water
• Subjected to all sample prep and
analytical procedures
Laboratory Control Sample
Also known as:
• Laboratory Fortified Blank (LFB)
• Quality Control Sample (QCS)
Surrogates
• Similar to an internal standard
• Added to all analytical samples,
and to all QC samples to monitor
method performance, usually
during sample prep
• Methods often have specific
surrogate recovery criteria
• Most common in Organic methods
Quality Control Measures
• Standards and Calibration
• Blanks
• Recovery Studies
• Precision and Accuracy Studies
• Method Detection Limits
• NJQLs
Precision and Accuracy
• Required for initial certification
and annually thereafter
• A series of four laboratory control
samples
• Must meet accuracy (recovery) and
precision (standard deviation)
requirements, often in method
Precision and Accuracy
• Required with a change in
instrumentation or personnel
• Specific to the analyst
• Other names include:
P&A, DOC, IDOC
Method Detection Limit
“The minimum concentration of a
substance that can be measured and
reported with 99% confidence that the
analyte concentration is greater than
zero”
N.J.A.C 7:18 - 1.7
Method Detection Limit
• MDLs are determined according to
40 CFR, part 136, Appendix B
• Seven replicate laboratory control
samples, analyzed for precision
• Multiply standard deviation by
3.14 (Student’s t- value)
Method Detection Limit
• Must be performed initially for
certification
• Must meet criteria specified in
method
• Must be performed with change in
instrumentation or test method
• Annually with ELCP
New Jersey
Quantitation Limits
(NJQLs)
• The minimum concentration of an
analyte that can be quantified with
statistical confidence
• 5 x MDL, for the NJ Lab
Certification Program

You might also like