Professional Documents
Culture Documents
Electrical and Electronic Measuring Instruments
Electrical and Electronic Measuring Instruments
Saswat Nayak
Roll No – 1707395
Semester – IV
Branch/Section – EEE (05)
2018-19
• Working Principle
• Suggested Improvements
Clamp meter
SOURCE : WIKIPEDIA
7
6. Energy meter
• Origin • Working Principle
Thomas Edison at first worked on a direct current (DC) The driving system has two electromagnets. The upper one is called
electromechanical meter with a direct reading register, but instead the shunt electromagnet, and the lower one is called series
developed an electrochemical metering system, which used an
electromagnet. The series electromagnet is excited by the load
electrolytic cell to totalize current consumption. At periodic intervals
the plates were removed and weighed, and the customer billed. The current flow through the current coil. The coil of the shunt
electrochemical meter was labor-intensive to read and not well electromagnet is directly connected with the supply and hence
received by customers. carry the current proportional to the shunt voltage. This coil is
• Suggested Improvements called the pressure coil. The center limb of the magnet has the
copper band. These bands are adjustable. The main function of the
The Energy Meter can be improved if the following limitations are copper band is to align the flux produced by the shunt magnet in
taken care of such a way that it is exactly perpendicular to the supplied voltage.
The first major device to measure in-circuit ESR was based on Carl W. The ESR Meter can be improved if the following limitations are
Vette's "US Patent #4216424: Method and apparatus for testing
taken care of
electrolytic capacitors“, under the Creative Electronics brand. The
Creative Electronics ESR meter was the primary device many used An ESR meter does not measure the capacitance of a capacitor
for the duration of the patent. The patent expired in 1998, when A faulty short-circuited capacitor will incorrectly be identified by an
many other companies entered the market. ESR meter as having ideally low ESR
ESR may depend upon operating ; a capacitor which has excessive ESR
Additional patents extended the original work, including John G. at operating temperature and voltage may test as good if measured
Bachman's 2001 "US Patent #6677764: System for protecting
electronic test equipment from charged capacitors" cold and unpowered
An ESR meter can be damaged by connection to a capacitor with
significant voltage across it, either because of residual stored charge
• Working Principle or in a live circuit
When an ESR meter is used as a milli ohmmeter, any significant
Most ESR meters work by discharging a real electrolytic capacitor inductance present between the test probes will make measurements
and passing an electric current through it for a short time, too short meaningless.
for it to charge appreciably. This will produce a voltage across the
device equal to the product of the current and the ESR plus a
negligible contribution from a small charge in the capacitor; this
voltage is measured and its value divided by the current (shown in
ohms or milliohms on a digital display or by the position of a
pointer on a scale. The process is repeated tens or hundreds of
thousands of times a second.
ESR Meter
SOURCE : AMAZON 9
8. Static Meter
• Origin • Suggested Improvements
The Static Meter can be improved if the following limitations are taken
A static meter includes a field head which develops an electrical
care of
charge representative of the strength of an electrostatic field at the
location of the field head. It was invented by Carlton M. Slough in
1983, Texas. The LCD display depends on a battery or external power source.
In case of fluctuations or transients, it can record an error.
• Working Principle Warming of the meter during its use can change its properties leading
to errors in measured value.
The electrical charge is converted to an AC signal which is then The A/D converter has a limitation on word length which can cause
provided to metering apparatus. The metering apparatus provides quantization noise giving rise to error in measured value.
an indication of the strength of the electrostatic field in accordance There is a voltage limitation. If it is increased beyond the limit, the
with the AC signal. meter will be damaged.
The digital nature makes it unsuitable for adjusting tuning circuits or
peaking tunable responses.
They are expensive due to high manufacturing cost.
Static Meter
SOURCE : WIKIPEDIA
10
9. Polarity Indicator
• Origin
Polarity Indicator was invented to indicate direction of current and
polarity of voltage by Kaisling William in 1916. It related in general to
a device for visualizing the direction of the current flowing in a
circuit and is particularly adapted for use in connection with the
ignition and lighting circuits of an automobiles or the like.
• Working Principle
The opamp 741 is the core of the polarity indicator circuit. Its non-
inverting input is used to test the points for polarity. It has a gain of
around 150 which enables it to test low voltage levels. The test
result is displayed through the two LED’s D1 and D2. D1 lights up by
positive polarity and D2 lights up by negative polarity.
Polarity Indicator
SOURCE : Electroschematics
11
10. Flux meter
• Origin • Suggested Improvements
The flux meter is the advanced form of the ballistic galvanometer
which has certain advantages like the meter has low controlling The Flux Meter can be improved if the following limitations are taken
torque and heavy electromagnetic damping and is used for care of
measuring the flux of the permanent magnet.
The flux linking with the coil is varied by either removing it from the
magnetic field or by reversing the field of the magnet. The change
of the flux induces the electromotive force in the coil. This emf
induces the current in the search coil and send it through the flux
meter. Because of the current, the pointer of the flux meter
deflects, and their deflection is directly proportional to the change
in the value of flux linkages.
As, the variation of the flux linkages reduces, coil stop moving
because of their high electromagnetic damping. The high
electromagnetic damping is because of the low resistance circuit
between the flux meter and the search coil.
Flux meter
SOURCE : Circuit Globe
12
11. Microwave Power Meter
• Origin • Working Principle
When Sigurd and Russell Varian, the inventors of the klystron Usually a microwave power meter will consist of a measuring head
microwave power tube in the late 1930s, were in the early which contains the actual power sensing element, connected via a
experimental stages of their klystron cavity, the detection diodes of
cable to the meter proper, which displays the power reading. The
the day were not adequate for those microwave frequencies. The
story is told that Russell cleverly drilled a small hole at the head may be referred to as a power sensor or mount. Different
appropriate position in the klystron cavity wall, and positioned a power sensors can be used for different frequencies or power
fluorescent screen alongside. This technique was adequate to reveal levels. Several modern power sensor heads contain electronics to
whether the cavity was in oscillation and to give a gross indication of create a digital output and can be plugged via USB into a PC which
power level changes as various drive conditions were adjusted. acts as the power meter.
Early measurements of high power system signals were • Suggested Improvements
accomplished by arranging to absorb the bulk of the system power
into some sort of termination and measuring the heat buildup versus
The use of directional couplers to extend the range of low-level
time.
power meters is a well established technique.
Historically the means of operation in most power sensor and meter A cascade method would allow calibration of low power meter
combinations was that the sensor would convert the microwave coupler combinations for measurement up to megawatt range.
power into an analogue voltage which would be read by the meter
and converted into a power reading.
13
12. Phase meter
• Origin • Suggested Improvements
The earliest meter was Samual Gardiner’s (USA) lamphour meter The Phase Meter can be improved if the following limitations are taken
patented in 1872. It measured the time during which energy was care of
supplied to the load, as all the lamps connected to this meter were
controlled by one switch. Subdividing lighting circuits became
practical with the introduction of Edison’s light bulb, and this meter
became obsolete.
Incorrect magnitudes of fluxes.
Thomas Alva Edison’s 'electric meter' patented in 1881 (USA patent Incorrect phase angles.
No. 251,545) used the electrochemical effect of current. Edison later
added a counting mechanism to aid meter reading. Change in strength of brake magnet
There were other electrolytic meters, like the German Siemens- Change in disc resistance
Shuckert hydrogen meter and the Schott&Gen. Jena mercury meter. Abnormal friction of moving parts
• Working Principle
One of the earliest reliable sources used for audio test was the first A weakness of this method of measuring distortion is that extraneous
product made by Hewlett-Packard in 1939, the HP200A audio signals such as noise, AC line pickup, etc. corrupt the measurement
oscillator. The clever and inexpensive design of the HP200A allowed causing the indicated distortion to be higher than actual.
testers to generate very high quality, low distortion sine waves that
These effects can be reduced by adding a broadband filter between
could be used for testing. This was followed by the company's
introduction of the HP320A and HP320B Distortion Analyzers in the output of the analyzer and the AC voltmeter. For good results the
1941. filter should be of at least second but preferably fourth order on the
low and high cut-off frequencies.
• Working Principle Before adding a filter the AC voltmeter reading should be noted with
no signal applied to the electronics under test (notch switch in the
It is such a device which measures detects the frequencies which UNNOTCHED position).
are responsible for producing harmonic distortion in our signal. A low-pass filter can also be added to the output of the sine wave
generator to reduce the level of harmonics applied to the amplifier
So once the frequencies are detected, the distortion meter now under test. The filter should have a cut-off frequency somewhat
acts such that it works to eliminate the effect of such frequencies. higher than 1 kHz and should be of at least second order.
For this purpose it has the option to set certain predefined
parameters, so that any frequencies which lie outside these set of
parameters are eliminated and not allowed to pass along with the
original signal. The original signal can then be produced at the
output in whatever way it is required.
The power factor meter has the moving system called pointer
which is in equilibrium with the two opposing forces. Thus, the
pointer of the power factor meter remains at the same position
which is occupied by it at the time of disconnection.
• Origin
Synchroscope was invented for indicating the frequency and phase
relations between the alternating current outputs of two generators
by Jr Frederick W Hornbruch in 1940.
• Working Principle
17
16. Magnetometer
• Origin
The first magnetometer capable of measuring the absolute magnetic
intensity was invented by Carl Friedrich Gauss in 1833 and notable
developments in the 19th century included the Hall effect, which is
still widely used.
• Working Principle
• Suggested Improvements
Ammeter
20
SOURCE : WIKIPEDIA
19. Voltmeter
• Origin
• Suggested Improvements
The original principals behind voltmeters were established by a
Danish physicist named Hans Christian Oersted (1777-1851) in 1820, Improvised design to simplify the construction, cheapen the cost of
when he discovered that an electric current in a wire produced a manufacture, and facilitate the adjustment and repair of the instrument :
magnetic field around it.
almost all types of voltmeters are founded off of the models in My improved volt meter comprises a removable needle of soft iron
which the pointers were attached to moving coils, which were placed between the poles of a permanent magnet, and enclosed in a
developed by French physicist Jacques-Arsene d'Arsonval(1851- shell or tube of copper, to prevent excessive oscillations. A mirror
1940) in 1882. placed under the index enables the readings to be made with greater
accuracy.
• Working Principle
Voltmeter
SOURCE : WIKIANSWERS
21
20. Magnetic Flow Meter
• Origin
• Suggested Improvements
In 1832, Michael Faraday (1791 – 1867) assembled a large scale
open channel magmeter and attempted to use this to measure
the flow of water passing under London’s Waterloo Bridge. Lining in the context of electromagnetic flow meters refers to
In 1915, the Americans M.W. Smith and Joseph Slepian filed a applying resin or other alternative materials to the inner walls of a
patent for “A device to measure the speed of a boat by means of detector's measurement tube (metal pipe). Lining is important in
magnetohydrodynamics.”
order to efficiently detect generated very low electromotive force
In 1930 the same idea was adapted to closed conduits by the levels and ensure corrosion and abrasion resistance to the fluids
Briton E.J. Williams.
being measured.
In 1952 the Dutch company Tobi-Meter introduced the first An improved perivascular electromagnetic flowmeter
commercial magmeter.
Improved magnetic field for an electromagnetic flowmeter with point
In 1962 the British scientist J.A. Shercliff published the “Theory of electrodes
electromagnetic flow-measurement.”
Concept of a next-generation electromagnetic phase-shift flowmeter
• Working Principle for liquid metals
Improved compact electromagnetic liquid flow meter
A magnetic field is applied to the metering tube, which results
in a potential difference proportional to the flow velocity
perpendicular to the flux lines. The physical principle at work
is electromagnetic induction. The magnetic flow meter requires
a conducting fluid, for example, water that contains ions, and
an electrical insulating pipe surface, for example, a rubber-lined
steel tube.
• Origin
Varmeter was invented for measuring the reactive volt amperes of
single phase alternating current circuits by Harry M Witherow of
General Electric Co in 1940.
• Working Principle
The current which passes through the current coil is the load
current. The load current has a phase difference of 90° concerning
that of the supply voltage, and it is given by the equation shown
below.
Varmeter
SOURCE : Circuit Globe
23
22. Permeameter
• Origin
The name was first applied by Silvanus P. Thompson to an apparatus
devised by himself in 1890, which indicates the mechanical force
required to detach one end of the sample, arranged as the core of a
straight electromagnet, from an iron yoke of special form; when this
force is known, the permeability can be easily calculated.
• Working Principle
Permeameter
SOURCE : WEB
• Suggested Improvements
24
23. Coulometer
• Origin
Karl Fischer titration is a classic titration method in analytical
chemistry that uses coulometric or volumetric titration to determine
trace amounts of water in a sample. It was invented in 1935 by the
German chemist Karl Fischer.
It used an instrument of chemical analysis that determines the
amount of a substance released in electrolysis by measurement of
the quantity of electricity used, called Coulometer
• Working Principle
25
24. Barometer
• Origin
Although Evangelista Torricelli is universally credited with inventing
the barometer in 1643, historical documentation also suggests
Gasparo Berti, an Italian mathematician and astronomer,
unintentionally built a water barometer sometime between 1640
and 1643.
• Working Principle
• Suggested Improvements
• Origin
In 1880, Pierre and Jacques Curie published an experimental
demonstration connecting mechanical stress and surface charge on a
crystal. This phenomenon became known as the piezoelectric effect.
Closely related to this phenomenon is the Curie point, named for the
physicist Pierre Curie, which is the temperature above which
piezoelectric material loses spontaneous polarization of its atoms.
• Working Principle
27