Plasma Low Temp.

Sterilizer
Plasma Low Temp. Sterilizer

A. THE GRAPHICAL USER INTERFACE


PANEL (GUI Panel)

B. THE PRINTER

C. THE STERILIZATION CELL &


SLIDING DOOR

D. STERILIZATION AGENT STORAGE

E. SERVICE DOOR

F. ON/OFF SWITCH

G. STERILIZATION AGENT
CARTRIDGE

H. DATA CONNECTION & LOGGING


Data ports
a. RS232 Port
b. USB Port
c. Ethernet PORT
Plasma Low Temp. Sterilizer
Plasma Low Temp. Sterilizer
Plasma Low Temp. Sterilizer

Fast Program: Fast Program provides fast turnaround for instruments


which need surface sterilization such as telescopes and rechargeable batteries
for use in laparoscopic, orthopedic or ophthalmology procedures.
• Duration: 31min (29min with Trivac)
• H2O2 Sterilant: Non-Concentrated H2O2
• Loading: Max load 5kg (top tray only)

Standard Program: Standard program is designed to sterilize general surgical instruments with
flexible lumens (1x850mm and below) and rigid lumens (1x400mm double sided). S
tandard Program is designed to handle Single Channel Flexible Lumens/Endoscopes
(maximum 6 lumens per load).
• Duration: 48min (42min with Trivac)
• H2O2 Sterilant: Concentrated H2O2
• Loading: Max load 10.0kg (12.0kg for 160Lt)

Advanced Program: Advanced Program is designed to sterilize general surgical instruments


which has long rigid lumens (1x500mm double sided, and maximum 6 lumens per load).
• Duration: 57min (53min with Trivac)
• H2O2 Sterilant: Concentrated H2O2
• Loading: Max load 10.0kg (12.0kg for 160Lt)
Plasma Low Temp. Sterilizer