Monitoring Intra Operatif: Mindi Widayani NRP 122.022.1115 FK UPN "Veteran" Jakarta
Monitoring Intra Operatif: Mindi Widayani NRP 122.022.1115 FK UPN "Veteran" Jakarta
Mindi Widayani
NRP 122.022.1115
FK UPN “Veteran” Jakarta
STANDAR MONITORING DASAR
UNTUK ANESTESI
Tujuan :
Untuk menjamin fungsi yang adekuat dari sirkulasi pasien selama
anestesi.
Electrocardiography
Cardiac activity
Arrhythmia: Lead II
Myocardial ischemia
Electrolyte imbalance
Pacemaker function
Indications
Continuous blood pressure monitor:
anticipated cardiovascular instability,
direct manipulation of cardiovascular Contraindications
system, inability to accurate Complications
- Local infection
measurement directly -Direct trauma: AV-fistula,
- Impaired blood
Aneurysm
circulation: Raynaud’s
Frequent arterial blood sampling: - Hematoma
phenomenon, DM
ABG, Acid-base / electrolyte / glucose - Infections
- Risks of thrombosis:
disturbance, Coagulopathies - Thrombosis
hyperlipidemia,
- Embolization
previous brachial
- Massive blood loss
artery cannulation
2. MONITORING SISTEM RESPIRASI
Monitoring devices
Spirometer
Airway pressure manometer
Circuit disconnection alarm
Ventilatory monitoring
Bila ETT atau laryngeal mask dimasukkan, posisi yang benar harus
diverifikasi dengan penilaian klinik dan identifikasi dari karbon
dioksida pada gas ekspirasi.
Analisa karbon dioksida end tidal yang kontinu, digunakan dari waktu
ke waktu pada pemakaian ETT atau laryngeal mask, sampai
ekstubasi/ pencabutan atau pemindahan ke tempat perawatan paska
operasi, harus menggunakan metode kuantitatif kapnografi,
kapnometri atau spektroskopi mass.
Ventilatory monitoring
Capnometer (End-tidal CO2 analysis)
Relationship with PaCO2 : ETCO2 < PaCO2 Capnogram : normal curve
~ 3-6 mmHg mainly depends on dead space 1. Dead space air (no CO2)
ventilation 2. Mixed bronchus & alveolus air (CO2 upstroke)
3. Alveolus air (CO2 plateau)
Normal value 30 – 35 mmHg
Inspiration
3 ETCO2
2
1
Oxygenation monitoring
Clinical skills
Direct observation: impaired mental function, sympathetic
overactivities, appearance(+ cyanosis)
Auscultation: wheezing, crepitation
Monitoring devices
Arterial blood gas analysis
Percutaneous O2 measurement
Pulse oximeter Tujuan :
Untuk menjamin konsentrasi oksigen yang adekuat dalam
gas inspirasi dan darah selama semua anestesi.
Oxygenation monitoring
• Pulse oximeter
– Percent of oxyhemoglobin / total
hemoglobin
– Oxyhemoglobin absorp 940 nm.
– Deoxyhemoglobin absorp 660 nm.
– Caution: SpO2 PaO2
Oxygenation monitoring
• Pulse oximeter
– SpO2 correlates with PaO2 • Pulse oximeter artifacts
as in Oxygen-hemoglobin
dissociation curve – Abnormal hemoglobin: COHb, MetHb,
HbF
– SpO2 90 ~ PaO2 60 mmHg
– Dye: Methylene blue
(moderate hypoxemia)
– Anemia
75 ~ PaO2 40 mmHg
– Ambient light
(mixed venous oxygen sat.)
– Arterial saturation
50 ~ PaO2 27 mmHg
– Blood flow
(P50)
– Motion
– Nail polish
– Electro-cautery
3. MONITORING SUHU TUBUH
• Diukur dengan thermistor atau thermocouple. Thermistor adalah semikonduktor yang resistensinya
menurun tanpa ada peringatan. Thermocouple adalah sirkuit 2 lempeng logam yang digabungkan
Teknik sehingga perbedaan potensial dihasilkan bila logam dalam suhu yang berbeda.
• Probe thermistor dan thermocouple sekali pakai tersedia untuk monitoring temperatur dari
membran timpani, rektum, nasofaring, esofagus, kandunh kemih dan kulit.
• Monitoring fungsi neuromuskular dari semua pasien yang menerima obat neuromuskular
blok yang lama kerja panjang atau sedang.
Indikasi • Menilai paralisis selama induksi rapid sequence atau selama infus kontinyu dari obat lama
kerja pendek.
• Tidak ada kontraindikasi untuk monitoring neuromuskular, meskipun pada beberapa lokasi
Kontraindikasi mungkin akan menghalangi prosedur bedah.
• Stimulasi ulnar dari otot adductor pollicis dan saraf wajah untuk orbicularis oculi adalah
yang paling sering dimonitor. Karena inhibisi reseptor neuromuskuler yang harus dimonitor,
Teknik dan maka stimulasi langsung pada otot harus dihindari dengan meletakkan elektroda pada
Komplikasi daerah saraf dan tidak melebihi otot tersebut.
• Komplikasi stimulasi saraf terbatas pada iritasi kulit dan abrasi pada tempat pelekatan
elektroda.