MONITORING INTRA OPERATIF

Mindi Widayani
NRP 122.022.1115
FK UPN “Veteran” Jakarta
STANDAR MONITORING DASAR
UNTUK ANESTESI

WHO Guidelines for Safe Surgery 2009


Standards for basic intraoperative monitoring
( ASA : American Society of Anesthesiologists)

• Petugas anestesi yang berkualitas


harus hadir dalam ruangan
STANDAR I selama pelaksanaan anestesi
umum, anestesi regional dan
monitoring perawatan anestesi.

• Selama semua anestesi,


oksigenasi, ventilasi, sirkulasi dan
STANDAR II suhu pasien harus selalu terus
menerus dievaluasi.
MONITORING THE ANAESTHETIC EQUIPMENT

in addition to confirming that the patient is fit for surgery, the


anaesthesia team should complete the ABCDEs : an examination of the
- Airway equipment
- Breathing system (including oxygen and inhalational agents)
Oxygen Breathing Infusion
- suCtion Alarms
Supply
- Drugs and Devices Systems Devices
- Emergency medications, equipment and assistance to confirm
their availability and functioning.

Recommendations for Standards of Monitoring During Anaesthesia and Recovery 4th


Edition. The Association of Anaesthetists of Great Britain and Ireland. 2007.
Monitoring Devices

Induction and Maintenance of


Anaesthesia Recovery from Anaesthesia

• Pulse oximeter • Pulse oximeter


• Non invasive blood pressure monitor • Non-invasive blood pressure monitor
• Electrocardiograph • The following must also be
• Airway gases: oxygen, carbon dioxide immediately available
and vapour • Electrocardiograph
• Airway pressure • Nerve stimulator
• The following must also be available • Means of measuring temperature
• A nerve stimulator whenever a • Capnograph
muscle relaxant is used
• A means of measuring the patient’s
temperature

Recommendations for Standards of Monitoring During Anaesthesia and Recovery 4th


Edition. The Association of Anaesthetists of Great Britain and Ireland. 2007.
Monitoring Devices
Regional Techniques & Sedation
Additional Monitoring
for Operative Procedures
• Some patients will require • Patients must have appropriate
additional, mainly invasive, monitoring, including a
monitoring, e.g. vascular or minimum of the following
intracranial pressures, cardiac devices :
output, or biochemical • 1. Pulse oximeter
variables. • 2. Non-invasive blood pressure
• Specific devices designed to monitor
monitor loss of consciousness • 3. Electrocardiograph
using adaptations of either
surface EEG monitoring or
auditory evoked potentials
have become available.

Recommendations for Standards of Monitoring During Anaesthesia and Recovery 4th


Edition. The Association of Anaesthetists of Great Britain and Ireland. 2007.
1. MONITORING SISTEM KARDIOVASKULER

Routine monitoring Advanced monitoring

• Cardiac activity • Direct arterial blood


• Non-invasive blood pressure
pressure ( NIBP ) • Cardiac filling pressure
• Electrocardiography ( ECG ) monitor
• Central venous pressure
• Pulmonary capillary
wedge pressure

Tujuan :
Untuk menjamin fungsi yang adekuat dari sirkulasi pasien selama
anestesi.
Electrocardiography
Cardiac activity
Arrhythmia: Lead II
Myocardial ischemia
Electrolyte imbalance
Pacemaker function

Non-invasive blood pressure


Cuff: width 120-150 % limb
diameter, air bladder includes
more than halfway around limb
Manometer: aneroid, mercury
Detector: manual, automated
DIRECT ARTERIAL PRESSURE MONITOR

Indications
Continuous blood pressure monitor:
anticipated cardiovascular instability,
direct manipulation of cardiovascular Contraindications
system, inability to accurate Complications
- Local infection
measurement directly -Direct trauma: AV-fistula,
- Impaired blood
Aneurysm
circulation: Raynaud’s
Frequent arterial blood sampling: - Hematoma
phenomenon, DM
ABG, Acid-base / electrolyte / glucose - Infections
- Risks of thrombosis:
disturbance, Coagulopathies - Thrombosis
hyperlipidemia,
- Embolization
previous brachial
- Massive blood loss
artery cannulation
2. MONITORING SISTEM RESPIRASI

Ventilatory Oxygenation Machine and


monitoring monitoring Circuit
monitoring
Ventilatory monitoring
Clinical skills
Direct observation: rate, rhythm, volume of respiration
Auscultation: precordial, esophageal stethoscope
Palpation: reservoir bag movement

Monitoring devices
Spirometer
Airway pressure manometer
Circuit disconnection alarm
Ventilatory monitoring

Bila ETT atau laryngeal mask dimasukkan, posisi yang benar harus
diverifikasi dengan penilaian klinik dan identifikasi dari karbon
dioksida pada gas ekspirasi.

Analisa karbon dioksida end tidal yang kontinu, digunakan dari waktu
ke waktu pada pemakaian ETT atau laryngeal mask, sampai
ekstubasi/ pencabutan atau pemindahan ke tempat perawatan paska
operasi, harus menggunakan metode kuantitatif  kapnografi,
kapnometri atau spektroskopi mass.
Ventilatory monitoring
Capnometer (End-tidal CO2 analysis)
Relationship with PaCO2 : ETCO2 < PaCO2 Capnogram : normal curve
~ 3-6 mmHg mainly depends on dead space 1. Dead space air (no CO2)
ventilation 2. Mixed bronchus & alveolus air (CO2 upstroke)
3. Alveolus air (CO2 plateau)
Normal value 30 – 35 mmHg

Inspiration
3 ETCO2
2
1
Oxygenation monitoring

Clinical skills
Direct observation: impaired mental function, sympathetic
overactivities, appearance(+ cyanosis)
Auscultation: wheezing, crepitation

Monitoring devices
Arterial blood gas analysis
Percutaneous O2 measurement
Pulse oximeter Tujuan :
Untuk menjamin konsentrasi oksigen yang adekuat dalam
gas inspirasi dan darah selama semua anestesi.
Oxygenation monitoring

• Pulse oximeter
– Percent of oxyhemoglobin / total
hemoglobin
– Oxyhemoglobin absorp 940 nm.
– Deoxyhemoglobin absorp 660 nm.
– Caution: SpO2  PaO2
Oxygenation monitoring
• Pulse oximeter
– SpO2 correlates with PaO2 • Pulse oximeter artifacts
as in Oxygen-hemoglobin
dissociation curve – Abnormal hemoglobin: COHb, MetHb,
HbF
– SpO2 90 ~ PaO2 60 mmHg
– Dye: Methylene blue
(moderate hypoxemia)
– Anemia
75 ~ PaO2 40 mmHg
– Ambient light
(mixed venous oxygen sat.)
– Arterial saturation
50 ~ PaO2 27 mmHg
– Blood flow
(P50)
– Motion
– Nail polish
– Electro-cautery
3. MONITORING SUHU TUBUH

• Suhu tubuh pasien yang mengalami anestesi umum seharusnya diawasi.


Indikasi • Prosedur yang sangat singkat (kurang dari 15 menit) mungkin merupakan pengecualian dalam hal
ini.

• Diukur dengan thermistor atau thermocouple. Thermistor adalah semikonduktor yang resistensinya
menurun tanpa ada peringatan. Thermocouple adalah sirkuit 2 lempeng logam yang digabungkan
Teknik sehingga perbedaan potensial dihasilkan bila logam dalam suhu yang berbeda.
• Probe thermistor dan thermocouple sekali pakai tersedia untuk monitoring temperatur dari
membran timpani, rektum, nasofaring, esofagus, kandunh kemih dan kulit.

Komplikasi • Berhubungan dengan trauma yang disebabkan oleh probe.


4. VOLUME STATUS AND
RENAL MONITORING

Estimate blood loss


Urine output
Hemodynamic stability
Urine output
Kateterisasi kandung kemih adalah satu – satunya metode yang dapat dipercaya
untuk mengawasi keluaran urin.

Indikasi Pertimbangan klinis Kontraindikasi

• Pasien dengan gagal • Keuntungan tambahan • Kateterisasi kandung


jantung kongesti, gagal dengan menaruh kemih seharusnya
ginjal, penyakit hepar kateter foley adalah dilakukan dengan hati –
lanjut atau syok. kemampuan untuk hati pada pasien
• Prosedur bedah seperti memasukkan dengan resiko tinggi
operasi jantung, operasi thermistor pada ujung infeksi.
aorta atau renal, kateter jadi kandung
craniotomy, operasi kemih atau suhu inti
abdomen mayor, atau dapat dimonitor lebih
operasi dengan baik.
pergeseran cairan yang
banyak terjadi.
• Operasi yang lama dan
pemberian diuretik.
5. MONITORING NEUROMUSKULAR
PERIPHERAL NERVE STIMULATOR

• Monitoring fungsi neuromuskular dari semua pasien yang menerima obat neuromuskular
blok yang lama kerja panjang atau sedang.
Indikasi • Menilai paralisis selama induksi rapid sequence atau selama infus kontinyu dari obat lama
kerja pendek.

• Tidak ada kontraindikasi untuk monitoring neuromuskular, meskipun pada beberapa lokasi
Kontraindikasi mungkin akan menghalangi prosedur bedah.

• Stimulasi ulnar dari otot adductor pollicis dan saraf wajah untuk orbicularis oculi adalah
yang paling sering dimonitor. Karena inhibisi reseptor neuromuskuler yang harus dimonitor,
Teknik dan maka stimulasi langsung pada otot harus dihindari dengan meletakkan elektroda pada
Komplikasi daerah saraf dan tidak melebihi otot tersebut.
• Komplikasi stimulasi saraf terbatas pada iritasi kulit dan abrasi pada tempat pelekatan
elektroda.
DAFTAR PUSTAKA

WHO Guidelines for Safe Surgery. 2009.

Practice Advisory for Intraoperative Awareness and Brain Function Monitoring.


American Society of Anesthesiologists, Inc. Lippincott Williams & Wilkins, Inc.
2006.

Recommendations for Standards of Monitoring During Anaesthesia and


Recovery 4th Edition. The Association of Anaesthetists of Great Britain and
Ireland. 2007.