Professional Documents
Culture Documents
JKMR (TM 8) Kendali Mutu Akuarasi Dan Repro Timer Generator Sinar-X
JKMR (TM 8) Kendali Mutu Akuarasi Dan Repro Timer Generator Sinar-X
Materials
1.Digital Meter (1, 3 phase)
2.Spinning Top (1 phase)
1 phase
full-wave rectifier
1 phase
full-wave rectifier
1 , AC 50 Hz AC 50 Hz
Half-wave Rectifier
1 , AC 50 Hz AC 100 Hz
Full-wave Rectifier
1 , AC 50 Hz Constant
Constant Generator
AC
AC50
50Hz DC100
Hz AC 100Hz
Hz
Full-waveRectifier
Full-wave Rectifier
X-Ray
0.04 s
kV
1 phase
full-wave rectifier
(x ) -Population
( xi )
50 56 9
50 53 0 ( xi x ) 2
SD
50 50 9 n 1
50 54 1
20
50 52 1
5 1
x 53 20
SD 1.12
October 16, 2019 Jaminan Mutu Radiologi II 15
2.Exposure Time
Accuracy & Precision SD 1.12
Setting Reading
Time (ms) Time (ms) ( xi x ) 2
Ac = Accuracy
(x ) ( xi )
50 56 9
xx
50 53 0 Ac 100
50 50 9 x
50 54 1 53 50
50 52 1 100
50
x 53 20 6%
x 53 20 3%
October 16, 2019 Jaminan Mutu Radiologi II 18
2.Exposure Time
Dampak dari exposure time yang tidak akuat
X-Ray
0.04 s
kV
X-Ray
0.04 s
kV