Professional Documents
Culture Documents
Benesphera Electrolyte Analyzer: Troubleshooting During Installation
Benesphera Electrolyte Analyzer: Troubleshooting During Installation
• Solution:
• Flush periodically with Daily cleaner.
• Avoid loose connections
• Insert all staged electrodes into the reference housing after
cleaning inner silicon seals individually and repeat above procedure
with air flush with 0.9 mm plain bold needle provided in the trouble
shooting kit and then connect the electrode housing in the
instrument
Air Bubbles
• Valve Module consists of two valves, valve 1 and valve 2 respectively. Valve 1 is used to
separate CAL A & CAL B flow, Valve 2 is used to take Air and Valve 1 output.
• First remove CAL-A & CAL-B tubes from thermal system. Take out whole valve tubing from
valves (V1 & V2) slots. Take syringe with DI water. Flush whole tubing with DI water in
order to clear all the salt formation in flow path. Finally flush whole tubing with air and
then reinsert in respective positions at valves flow path tube slots.
• Note: Pinch valve tubing should be totally inserted in the valve towards
inner side. Partial or incomplete inserted tube can let two solutions mix
up.
Thank You