Flexible endoscopes :
Basic construction.
Inspection.
Description of most common problems.
Repair schemes.
Summery.
ENDOSCOPY
FLEXIBLE ENDOSCOPY
SURGICAL ENDOSCOPY
Basic construction.
Inspection.
Care and maintenance.
RIGID ENDOSCOPE
Basic Construction of Rigid
Endoscopes
The basic construction of a rigid endoscope : The lenses
of the optical system are arranged in the inner tube, and
held in place with distance pieces. The distal region of
the inner tube is closed by the deflector prism (that
determines the direction of view) and with the eyepiece
at the proximal part. The inner tube is enclosed by the
jacket tube containing optical fiber cable and it’s
connector.
The jacket tube has a covering lens at the distal part and
an ocular at the proximal part. One of its significant
attributes of the jacket tube is the ability to seal against
invading substances.
Construction of a rigid endoscope
3- Measuring of Transmission
This is a method determining light transmission of the endoscope.
A luxmeter is adapted to the eye piece of the endoscope. At the distal part
light is directed through an adapter, the proximal output represents the
data.
4-Test method using Monocular Tube
LUXMETER
Testing Method with the
Monocular Tube
The endoscope is inserted into a jacket
tube containing an integrated
magnifying glass at the proximal end.
A distance piece, gliding over the
jacket tube canters the telescope
within this construction.
By varying the distance between
magnifying glass and the eyepiece,
different depths within the endoscope
can be focused. Variation in distance is
achieved by sliding the endoscope
within the construction.
CARE OF RIGID ENDOSCOPE
VIDEOSCOPE FIBERSCOPE
Image Evaluation
I. Perfect scope
II. Scope with 24 broken fibers
III. Leaky and stained scope
II III
MOST COMMON PROBLEMS WITH
FLEXIBLE ENDOSCOPES
Leakage in General.
BUCKLED INSERTION TUBE
BUCKLED INSERTION TUBE
BUCKLED BIOPSY CHANNEL
BIOPSY CHANNEL DAMAGE
Kinked, damaged or open biopsy forceps .
Buckling of the insertion tube can also
cause buckles in the channel.
Forcing instrumentation through the
channel can cause wear or tears in the
channel material. This frequently occurs in
the bending section when resistance is
met when the scope is angulated.
Using the noncorrect size cleaning brush
LEAKAGE PROBLEM
BENDING SECTION RUBBER LEAK
REPAIR SCHEMES
TECHNICAL SUPPORT
PREVENTIVE MAINTENANCE MINOR REPAIR MAJOR REPAIR
LENSES
. . . And many other repair schemes .
ANGULATION REPAIR
AIR / WATER CHANNEL REPLACEMENT
SUCTION SYSTEM REPAIR
BIOPSY CHANNEL REPLACEMENT
DISTAL COVER
CCD RETERMINATION
REMOTE SWITCHS REPAIR
GOOD HANDLING
Leakage Test
Endoscopes must be checked against any leak or damage before
use and processing to ensure its efficiency and avoid
instrument malfunction during endoscopy.
Leakage tester is an instrument
which can be attached to endoscope
and blows certain pressure of air-set
by the manufacturer- inside it then
endoscope is immersed in a water
basin and checked against any
leak,if any leak is seen endoscope
must be immediately transferred
for repair and must not been used.
LG PLUG REPAIR
INSERTION TUBE REPAIR
SUMMARY STATEMENT
This educational activity has
been developed in response to GI
nurses and technicians who have
expressed a need to increasetheir
knowledge base of the endoscope
construction, video system
components, care and
maintenance of flexible
endoscopes and repair reduction
strategies. Attendees will be given
updated information that will
reduce repair costs and improve
unit operations and patient care.