Installation Guide
1)Environment
Working Temperature: 10℃ ~ 35 ℃
Humidity:≦ 80%(No Dewing)
Air Pressure: 75 kPa ~ 106 kPa
2)Space
The installation space should be:
Distance of left/right side from the wall ≥ 50 cm
Distance of rear side from the wall ≥ 50 cm
•Power Supply
Voltage : A.C.100 ~ 240V ;
Frequency : 50/60 Hz ;
Input Power : 120VA
Reagent A/B/C/L
Unsrew the knob of filter, and put the Filter in. Then tight the knob.
Pull the holder on top right corner of printer, and open printer door.
Load thermal paper according to the direction in Picture , and leave a segment of paper
outside the exit.
System will do Self testing after switching on, after Self Testing, Press “No” to avoid
System Cleaning.
After that, it will jump to Menu Interface.
H9 SE
reagent
,
version
:A0
Then press “System Maintenance”, enter password “661166”. Enter Hardware Test
interface.
If Ok, go to Calibration.