Professional Documents
Culture Documents
Rehab Tech21
Rehab Tech21
PRE-SELECTION
SELECTION
ANSI VS ICRA
DIGI SPEECH
Interrupted composite noise. The pauses
are similar to that of natural speech
There are two Digital Speech signals
available
Probe tube calibration accounts for the acoustic effects the probe tube
introduces as sound travels through it.
In effect, calibration removes the acoustic effects the probe tube during real-ear
measurement.
One has to compare the output form the reference mic to the simultaneously
recorded response of the signal travelling thorugh the probe tube attached to
the test mic.
The difference between these two responses reflect the effect of probe tube.
In most equipments, the acoustical transmission effects of probe tube are stored
and hence the response from two mic should be same.
B. Otoscopic Examination
C. Location of speaker
D. PROBE TUBE PLACEMENT
METHODS
One method (visually-assisted positioning) involves inserting
the probe tube a constant insertion depth beyond the tragus or
inter-tragal notch. The guidelines regarding how far to insert
the probe tube can vary, depending on the age and gender of
the patient.
For children, insert the probe tube 20-25 mm past the inter-
tragal notch. Certainly normal anatomic variants will prohibit
the placement of the probe tube to these depths in some
patients, while in other patients these locations may not be
deep enough.
'geometrical positioning.‘
the ridge of the ear mold or hearing instrument
corresponding to the location of the inter tragal
notch is identified.
Lay the probe tube along the ridge identified above
with the open end of the probe tube extending 5
mm beyond the tip of the ear mold or hearing
instrument.
Mark the probe tube at the outer edge of the ear
mold or at the faceplate of the hearing aid and
then insert the probe tube into the ear canal until
the mark lies at the rim of the inter tragal notch.
Probe tube placement can also be assisted via acoustical
positioning procedures (ANSI, 1997; ISO 12124:2001). A simplified
method is through visualization and repositioning based on the
REUG curve, monitoring particularly the frequency region above
4000 Hz.
a. Insert the probe tube less than half way into the ear canal while
presenting a 65 dB pink noise signal or composite signal.
b. A notch in the gain curve above 4000 Hz is likely to be observed.
c. Gently insert the probe tube deeper while keeping an eye on the
notch which is moving towards higher frequencies.
d. The probe tube is located correctly as soon as the notch is no
longer dragging the gain curve down (-5 dB) in the high-
frequencies.
e. Once the measurement is stabilized move the probe tube marker
into position or to attach the probe tube to the probe tube support.
E. EQUALIZATION
It is the process of controlling the acoustic
signal at a specific point in space so that the
amplitude remains at the desired level across
frequencies.
There have been 2 commonly used methods
of sound field equalization
a. Substitution method
b. Modified pressure method (this term was
recently recommended by ANSI) / earlier it
was called as modified comparison method
(Preves, 1987; Preves and Sullivan, 1987).
a. Substitution method
The exact position in the room where the person will be
seated is identified
without the patient in the room, a mic is placed at the
location the person will occupy for measurements.
A signal is produced by the loud speaker, measured by the
mic and deviation from a flat free field is calculated.
After that the unoccluded ear testing will be conducted
with the patient in the exact position .The centre of the
patient’s head is placed in the precise location previously
occupied by microphone
microphone is located in the ear canal.
Then aided measurement conducted in the similar
manner as unaided testing. It is usually done in off line.
b. Modified pressure method
There are two mics, one that measures SPL in the
ear canal and one that is located some place in
the head and regulates the SPL being generated
by the loud speaker and maintain the signal at a
constant level. There are two major differences
between substitution method and this. First,
there is no equalization conducted with the
patient absent. Second, a second regulating mic
will be present for all the measurement. It can be
done either in online /offline.
Feed in thresholds
Select prescriptive formula
Select ear
Stimuli
Level
1. REUR (real ear unaided response)
SPL, as a function of frequency, at a
specified measurement point in the
unoccluded ear canal for a specified sound
field. This can expressed either in absolute
SPL (Response) or again in decibels relative
to the stimulus level (Gain).
They reflect the resonance characteristics of ear canal, concha,
and also the head and torso.
The average adult REUR has a primary peak around 2700 Hz of
about 17 dB, and a secondary peaks around 4000 Hz to 5000 Hz
region of 12 – 14 dB. It is the gain provided by the pinna and the
ear canal with consequent head diffraction effects as measured in
the ear canal.
Procedure:
1. Conduct otoscopic examination.
2. Place probe tube in the ear canal, with end of tube at
appropriate distance from the inter tragal notch (e.g., within 5
mm of the eardrum).
3. Place patient at appropriate distance/azimuth from the
loudspeaker.
4. Select desired input level.
5. Conduct the measurement.
Reference value for the calculation of
insertion gain.
Also reflects the abnormalities of ear canal
or the middle ear.
2. REAR (Real-Ear Aided Response)
SPL, as a function of frequency, between the
SPL at a specified measurement point in the
ear canal for a specified sound field, with
the hearing aid (and its acoustic coupling) in
place and turned on.
The gain of the hearing instrument across frequencies,
measured in the ear canal. It is a direct measurement how a
hearing aid will perform in a real ear.
Procedure:
1. Conduct otoscopic examination.
2. Seat the patient at the appropriate distance/azimuth from
the loudspeaker.
3. Place probe tube in the ear canal, with end of tube at
appropriate distance from the inter tragal notch (e.g., within
5 mm of the eardrum). NOTE: if the REAR/REAG is being used
to calculate insertion gain, be sure to position the probe tube
at the same location as the REUR/REUG measurement.
4. Insert the hearing instrument into the client's ear while
holding the probe tube so that its position in the ear canal is
not disturbed.
5. Turn the hearing instrument on and set the user gain
control to the desired setting.
6. Select desired input level.
7. Conduct the measurement.
Clinical applications .
Serves as a reference for insertion gain
calculation
To find out inter modulation distortions in
the hearing aid
Trouble shooting the hearing aids
Helps in finding out the maximum output in
the real ear of the hearing aid when it is in
saturation.
3. REIG (Real-Ear Insertion Gain)
The real ear insertion response (REIR)(ANSI S3.46-
1997): Difference in decibels, as a function of
frequency, between the REAR and the REUR,
obtained with the same measurement point in the
same sound field conditions.
The REIG is the value, in decibels of the REIR at a
specific frequency. The amount of gain provided by
the hearing instrument alone calculated by
subtracting the REUG from the REAG across
frequencies or by subtracting the REUR from the
REAR across frequencies.
REIG = REAG – REUG
Procedure :
Step 1: Conduct an REUR
Step 2 : Conduct an REAR, using the same sound field
conditions and measurement point as the REUR (i.e., probe
tube placement and signal level).
Step 3: Subtract the REUR from the REAR across frequencies
or subtract the REUG from the REAG across frequencies.
Step 4: Adjust hearing instrument characteristics so that the
REAR (REAG) and thus the subsequent calculation of REIG
provides the best match to the target REIG values across
frequencies.
REOR (Real era occluded response
SPL as a function of frequency, at a specified measurement
point in the ear canal for a specified sound field, with the
hearing aid (and its acoustic coupling) in place and turned
off. This can be expressed either in SPL or as gain in decibels
relative to the stimulus level.
Here the effect of placement of ear mould or hearing aid in
the ear easily can be measured. For open ear fitting REOR
may be very similar to REUR at some frequency region. For
most of the ear mould styles REOR will be substantially below
than REUR. In cases of venting REOR comes higher than that
of REUR at the region around 1500 Hz, because of the
resonance. REOR helps in the estimation of insertion loss.
Tight fitting hearing aid the REOR falls below the input level.
So it becomes very important in the measurement of REIR.
Procedure:
REUR measurement (even if it is not necessary)
Hearing aid will be placed. Make sure the hearing aid is turned off.
Place probe tube in the ear canal, with end of tube at appropriate
distance from the inter tragal notch (e.g., within 5 mm of the
eardrum).
Place patient at appropriate distance/azimuth from the
loudspeaker.
Select desired input level. (usually 60dB SPL)
Conduct the measurement
Clinical applications
Helps in determining appropriate vent size.
Indirect measure of occlusion effect.
Helps in selecting an acoustically appropriate sound delivery system
Real ear saturation response
Difference in decibels, as a function of
frequency, between the SPL at a specified
measurement point in the ear canal for a
specified sound field, with the hearing aid
(and its acoustic coupling) in place and
turned on. The measurement is obtained
with the stimulus levels sufficiently intense
as to operate the hearing aid at its maximum
output level.
It is very critical measurement for children
and non responsive patients when the
maximum output of the hearing aid must not
only comfortable but also safe.
Procedure:
Same as that of REAR, ensure that the
hearing aid is in saturation. This can be
accomplished by setting the input as 90dBSPL
and by adjusting the hearing aid volume
control to a just below the feedback.
Clinical applications
To measure the patient discomfort level
To make sure that the maximum output of
the hearing aid is both comfortable and safe.
Real ear coupler difference (RECD)
Difference in decibels, as a function of frequency, between the
outputs of the hearing aid measured in a real ear Vs a 2 cm3
coupler. This measure solves the problems of correcting
responses from the 2cm3 coupler to the real ear.
Procedure
Measure REAR
VCW at ½ or 2/3 rotation setting.
Remove the hearing aid from the ear and a 2cm3 coupler
response is obtained. ( VCW should be in the same position
2cm3 coupler values are subtracted from the REAR and the
difference is RECD.
The real ear analyzer measures the response near the ear drum with the
probe microphone. The difference between the SPL value measured by
real ear analyzer and the amplitude produced by the audiometer in dB
HL is the REDD at that frequency.
ADVANTAGES OF IGM
We obtain information across frequencies
It is not necessary to mask the other ear.
REG can be done for whom behavioral
responses are not available.
The effects of input level are assessed.
SOUND FIELD MEASUREMENTS
Sound field is any area in which sound waves
are present
Functional gain: is the dB difference
between aided and unaided behavioral or
neural thresholds
EQUIPMENT
The equipment used in sound field
measurements consists of a stimulus
generator, loudspeakers, and calibration
equipment.
A. Loudspeakers
The ideal loudspeaker for audio logic testing
should possess the following general
characteristics:
(a) Broad bandwidth (minimally 100–10,000
Hz);
(b) Constant output as a function of
frequency (c) low distortion;
(d) Capability of accurately transducing
transient as well as steady-state signals;
B.ENVIORNMENT
Sound field measurements are influenced by
the acoustic characteristics of the
environment in which auditory measures are
to be conducted.
When pure tones are introduced into the
sound field, the resonances of the room are
evidenced by standing-wave patterns with
resultant variation in measured SPL depending
on measurement location in the room.
Hence, warble tones
To achieve a reliable and repeatable
measurements in a sound field, it is
necessary to place the patient in an area
where ambient noise levels are controlled.
Less reverberant condition
Good absorbing material that dampens the
sound and reduces standing waves.
Patient placement in a sound field
SELECTION
VERIFICATION OF ELECTROACOUSTIC
CHARACTERISTICS
DIFFERENCES
DSL prescribes higher gain than NAL
DSL prescribes more low-frequency gain for flat loss
DSL prescribes more high-frequency gain for sloping loss
Behavioral measures:
Functional Gain is done: This is the difference, in
decibels between aided and unaided thresholds
cooperation from the child is required and thus
may require multiple sessions.
Speech tests:
Level of difficulty has to be appropriate to the age
and degree of hearing loss (either open or closed
set)
Ling sound test
Paired comparison:
This can be done in children aged six or older -two
alternative hearing aids in quick succession can be
given and compared
Evaluation of discomfort:
Face icons to represent different loudness categories
when a child’s LDL is measured.
Age of seven or older can be assessed using this.
Objective measures:
Insertion Gain Measures:
These are Individual and objective measures
If the child is not cooperative, RECD can be performed
6.7 to 17.3 years
Scollie et al, 2010
If RECD measurements cannot be performed,
applications of DSL v5, there are age appropriate
predicted RECD values.
And also LDLS can be predicted using Immittance since the ART
and loudness discomfort level for some types of acoustic stimuli
- especially speech - were at approximately the same levels.
OTOACOUSTIC EMISSION AND
COCHLEAR MICROPHONICS
Level at which OAE is recorded cannot be used to
predict behavioural thresholds from which to
prescribe amplification.
1) Threshold of wave V.
The hearing aids and/or settings which provide lowest threshold is
selected
Krebs (1976); Cox and Metz (1980) and Kilney (1982)
2) Latency of wave V.
The hearing aids and/or settings which provide shortest latency is
selected
L-I function
The observed patterns could be used to determine
what type of HA would best suit a HL.
that ASSR gain and REIG were highly correlated and there
was no significant difference at all test frequencies.
Amplitude Projection Procedure has been applied on
the ASSR data and ASSR could be used for selection
of gain and compression ratio.
LIMITATION OF ASSR
Hence, LLR can be used to ensure that the hearing aid is useful
or not.