BX Series

SOFTWARE
SPH

Slide 1
KEY OPERATION

Operates the analyzer using the computer keyboard and mouse. Functions for
each key are as follows:

1. [F1]: Start measurement


• Press this key to start or restart measurement.

2. [F2]: Stop sampling


• Press this key to stop dispensing of sample or reagent.
• However, measurement is continued only for the samples after
dispensing.

3. [F3]: Emergency sample addition (STAT)


• Press this key to add an emergency sample.

4. [F4]: Alarm
• Press this key to display the alarm screen.
KEY OPERATION

5. [F5]: Run (F5)


• Press this key to switch to the [Run (F5)] menu.

6. [F6]: Parameter (F6)


• Press this key to switch to the [Parameter (F6)] menu.

7. [F7]: Calibration (F7)


• Press this key to switch to the [Calibration (F7)] menu.

8. [F8]: QC (F8)
• Press this key to switch to the [QC (F8)] menu.

9. [F9]: System (F9)


• Press this key to switch to the [System (F9)] menu.
KEY OPERATION

10. [F10]: Maintenance (F10)


• Press this key to switch to the [Maintenance (F10)] menu.

11. [F11]: Switch to tab menu (upward) within the selected menu.
• Press this key to switch the tab menu upward.
KEY OPERATION

12. [F12]: Switch to tab menu (downward) within the selected menu.
• Press this key to switch the tab menu downward.

13. [Shift] + [F1]: Start orderless measurement


• Press the above keys at the same time to start orderless
measurement.

14. [Shift] + [Esc]: Alarm sound stop


• Press the above keys at the same time to stop alarm sound.

15. [Ctrl] + [F2]: Emergency stop


• Press the above keys at the same time to immediately stop
operation.

16. [Ctrl] + [F5]: Print screen


• This function is not available during measurement process.
OPERATION SCREEN

A. Menu Icons
1. Run (F5)

Monitor

Round

Test Selection

Results
OPERATION SCREEN

A. Menu Icons
1. Run (F5)

Inventory

Wakeup Setup

Auto Template

Patient Information

Daily Maintenance
OPERATION SCREEN

2. Parameter (F6)
• Press the icon switch to the [Parameter (F6)] menu.

Chemistry Parameters 1

Chemistry Parameters 2

Chemistry Parameters for ISE

Serum Indices
OPERATION SCREEN

2. Parameter (F6)
• Press the icon switch to the [Parameter (F6)] menu.

Calculated Test

Profile

Order

Wash Program
OPERATION SCREEN

3. Calibration (F7)
• Press the icon to switch to the [Calibration (F7)] menu.

Calibration Registration

Diluted Calibration

Multi-Calibrator

Result of ISE Calibration


OPERATION SCREEN

4. QC (F8)
• Press the icon to switch to the [QC (F8)] menu.

QC Graph

QC Results (Detail)

QC Results (Daily)

QC Results (Cumulative)

QC Settings

QC Sample Registration
OPERATION SCREEN

5. System (F9)
• Press the icon to switch to the [System (F9)] menu.

System Setup 1

Define Report Format

Reagent Registration

Versions
OPERATION SCREEN

5. System (F9)
• Press the icon to switch to the [System (F9)] menu.

Backup Procedures

System Setup 2

Definition

Normal Range
OPERATION SCREEN

7. Shutdown
• Pop-up message appears for system shutdown.

8. Sleep
• Displays the window for sleep mode.
• In the operational mode: Before sleep setting window appears.
• In the sleep mode: Confirmation window to wake-up from sleep
mode appears.

9. Alarm (F4)
• Press this icon to switch to the [Alarm] screen.
OPERATION SCREEN

6. Maintenance (F10)
• Press the icon to switch to the [Maintenance (F10)] menu.

Sequence

Water Blank

Working Hour Counters

Performance

Sensor

Volume Adjustment (for maintenance only)

DTR Positioning
OPERATION SCREEN

B. Operational Icons

1. Save
2. Cancel
3. Delete
4. Delete all
5. Search
6. Print
7. Execute
8. Close
ANALYZER STATUS

The pop-up message indicating the system status appears in the upper left of
the screen.

1. Not ready
• Not ready for connection between the analyzer and the host computer. Or
the analyzer is OFF.
2. Pre-ready
• The analyzer is ON, however, preparatory operation such as dispensing
into cuvettes and priming has not been completed.
3. Ready
• It is ready for starting measurement.
4. Measurement
• Measurement is performed.
ANALYZER STATUS

5. STAT-measurement
• Measurement for an emergency sample is performed as a priority.
• Measurement for normal samples and QC samples is not performed.
6. MSStopping
• Dispensing is stopped due to sample addition Or dispensing is completed.
7. ESStopping
• Since an error occurs, dispensing is stopped.
8. Sleeping
• The analyzer is in sleep mode.
9. Maintenance
• Maintenance procedures are performed.
10. Emergency Stop
• The analyzer is emergency stopped due to errors or user’s operation.
ANALYZER STATUS

11. Sampling stop in progress


• Continuous dispensing is not performed when this error occurs.
12. Sampling stop complete
• The measurement has been completed after the dispensing error
(interruption).
13. Maintenance complete
• Maintenance procedures are completed.
14. Em Stop in progress
• Emergency stop is executed.
15. Em Stop Complete in progress
• Ending procedure for Emergency stop is performed.
DATA SAVE

Data to be saved from operational software is as follows.


• System parameter
System parameter indicates data related to measurement conditions.
This is not included test orders and results information.
• Database
Database indicates data related to the analyzer such as measurement
conditions and results information.

The data can be saved to the hard disk (HD) or an external memory medium
(such as a USB memory).
It is also available to return the designated operational status by loading the
previous data from the HD or external memory medium.
DATA SAVE

System (F9) > Backup Procedures


DATA SAVE

A. System parameter
1. Save Parameters
2. Load Parameters
3. [...] Select the directory to save or load the system parameters.

B. Database
1. Save Database
2. Load Database
3. [...] Select the directory to save or load the database.

C. Load previous database


It is recommended to periodically save the system parameters and database
for security.
Reboot your operational PC after loading the system parameters or database.
Do not turn the analyzer or the operational PC OFF during backup process.
Data corruption may occur if the power is OFF during backup process.