Professional Documents
Culture Documents
Fisika Usg
Fisika Usg
Fisika Usg
ULTRASOUND
Trijayanti P.
19708251022
INTRODUCTION AND OBJECTIVES
1 Is sound a transverse or longitudinal vibration?
where ψ(x, t) gives the displacement of an element of fluid initially at point x and
where
c is the speed of the wave, B the bulk modulus of the medium and ρ the density.
In an ideal gas, where (PV )γ is constant, c is given by
ρ will be less for helium than for air and so, from the above equation, c will
be higher than in air. Now for any transmitted wave c = fλ, where f is the
frequency and λ the wavelength. The wavelength is fixed by the size of the
vocal cavity and therefore frequency will increase in proportion to velocity.
Radiation pressure
A sound wave travelling through a medium exerts a static pressure on any
interface across which there is a decrease in intensity along the direction
of propagation. The mechanism is disputed, but the result can be used to
measure the power output of ultrasound transducers. Power = force ×
velocity, so the transducer power can be found by using a force balance.
A force of 6.7 × 10−4 N is generated by the complete absorption of the
beam from a 1 W transducer in water, where the ultrasound velocity is
1500 m s−1.
7.3. GENERATION OF ULTRASOUND
7.3.1. Radiation from a plane circular piston
We start by assuming that a piston (i.e. a rigid circular plate) of radius a is mounted on
a rigid infinite baffle (figure 7.1), and use the notation of Kinsler et al. The piston is
moving normal to the baffle (i.e. In the Z direction) at a speed U0 exp(jωt). Each
infinitesimal element, of area dS, acts as a point source on the piston. It can be shown
that the pressure generated by a point source at a distance r with respect to time t is
given by
where U0 dS is the strength of the source, k is the wavenumber 2π/λ, and ρ0 is the
density of the undisturbed medium. The total pressure generated by the piston is
therefore given by
The pressure amplitude on the axis is given by the magnitude of p(p, 0, t),
which after some tedious manipulation can be shown to be:
This is plotted in figure 7.2 for a piston radius of five wavelengths
The integral has a real and imaginary part. The imaginary part vanishes, by
symmetry, and the real part is a first-order Bessel function:
thus giving:
We can relate this to the previous result for the asymptotic case. If θ = 0, the
bracketed term is unity. The exponential term gives the time variation of the pressure,
and the remaining term is the same as the asymptotic expression for the axial case. All
of the angular dependence is in the bracketed term. If we examine a table of first-order
Bessel functions, we find that J1(x) = 0 for x = 3.83, 7.02, 10.17, etc. In other words,
the beam energy is confined to lobes, with zeros at angles defined by the values of x
for which J1(x) = 0. Conventionally, for an ultrasound transducer, the side lobes are
ignored, and the axial spread of the beam is taken as the angular size of the main lobe.
The centre lobe reduces to zero for angles given by θ = sin−1(3.83/ka) =
sin−1(0.61λ/a). For a transducer that is 10 wavelengths in diameter, the beam diverges
in the far field with a half-angle of 7.0◦ (in the near field, the beamcan be considered to
be parallel).
If a/λ < 0.61 (i.e. the transducer is less than 1.22 wavelengths in diameter),
J1(x) = 0 for any real values of θ, so there are no side lobes. In the limit as ka → 0
the transducer radiates spherical waves. Figure 7.4 shows the beam shape for a
piston that is 10 wavelengths in diameter.
where lx and ly are the lengths of the sides of the plate, and θ and ϕ are the angular
deviations from the axis in the planes parallel to lx and ly , respectively.
7.3.2. Ultrasound transducers
Table 7.1 gives the velocity of sound in some biological media, from which it can be
calculated that the wavelength at 1 MHz is of the order of 1.5 mm in soft tissue.
Acoustic impedance is the product of the density and velocity of the sound for the
mediumand is therefore very low for air.
Figure 7.5. Simplified cross-section of an ultrasound transducer
7.4. INTERACTION OF ULTRASOUND WITH MATERIALS
7.4.1. Reflection and refraction
In the current context, for audible sound, we are usually only interested in the
transmission through air. Environmental and architectural acoustics, which deal
with noise propagation in the world in general and in buildings in particular, are
important and interesting fields in their own right, but we refer interested readers
to Kinsler et al (1982).
Of more importance to us is the interaction of ultrasound with biological tissue,
because this interaction is the basis of both ultrasound imaging techniques and
Doppler blood velocity measurement. The interaction of ultrasound with tissue is
complex, and what follows is a summary.
Aplane wave (see figure 7.6) in medium1(acoustic impedance Z1 = ρ1c1) is incident
on the boundary (at x = 0) with medium 2 (acoustic impedance Z2 = ρ2c2). The
complex pressure amplitudes of the incident, transmitted and reflected waves are
Pi,Pt and Pr, respectively. We define the pressure transmission and reflection
coefficients Tp and Rp as Tp = Pt/Pi and Rp = Pr/Pi. We have shown in section
3.4.1 that the intensity of a plane progressive wave is given by I = p2/2ρc, so the
intensity transmission and reflection coefficients are real and given by
The power transmitted by the beam is the product of the intensity and the cross-
sectional area. The reflected beam always has the same cross-sectional area as
the incident beam, whereas the cross-sectional area of the transmitted beam is,
in general, different. The power transmission and reflection coefficients are
therefore given by
Note the reversal of the sign of x for the reflected wave, and the different wave
numbers k1 and k2 in the two media, as a result of the different phase speeds c1
and c2. We can now state the boundary conditions which have to be satisfied. The
acoustic pressures on both sides of the boundary are equal (i.e. there is no net
force acting to displace the boundary), and the particle velocities normal to the
boundary are equal (i.e. The two fluids do not separate). The boundary conditions
are
Dividing pressure by velocity gives
A plane wave has p/u = ±ρc, so this is a statement of the continuity of the normal
specific acoustic impedance across the boundary. The incident, reflected and trans-
mitted waves must all satisfy the relationship pressure/velocity = acoustic impedance,
therefore
For normal incidence, all the beams have the same cross-sectional area, so
the power coefficients are the same as the intensity coefficients.
Figure 7.7. Reflection and transmission of an oblique incident
beam.
The
The incident, reflected and transmitted waves now subtend angles of θi, θr and θt,
respectively, at the x-axis. The equations for the incident, reflected and transmitted
waves now include the angles θ
These will be recognized from optics, the first states that the angles of incidence
and reflection are equal, and the second is Snell’s law (see section 3.5.1). As the
exponents are equal, continuity of pressure at the boundary gives the previous
expression,
This is referred to as the Rayleigh reflection coefficient. From Snell’s law we can
find the angle of transmission in terms of the velocity of sound:
If the media can support shear waves, the boundary conditions must be satisfied
parallel to the boundary, as well as perpendicular to the boundary. The resulting
analysis is complex, particularly for anisotropic media such as tissue, and will not
be pursued here. The shear wave velocity is always less than the longitudinal
wave velocity, so that a shear wave can still be propagated into the second medium
when the angle of incidence is equal to the critical angle. As the angle of incidence
is further increased, the shear wave will approach the boundary. It is thus possible
to propagate a shear wave along the boundary by a suitable choice of angle of
incidence. This has been used as a means of measuring the mechanical properties
of the skin.
7.4.2. Absorption and scattering
A pressure wave travelling through a medium will decrease in intensity with
distance. The loss of intensity is due to the divergence of the beam, scattering which
is not specular, mode conversion at interfaces and absorption. Beam divergence and
scattering reduce the intensity but not the energy of the beam, whereas mode
conversion and absorption both result in the beam energy being converted to another
form.
The most important contribution to absorption comes from relaxation processes, in
which the ultrasound energy is converted into, for instance, vibrational energy of the
molecules of the tissue. This is not a one-way process-there is an exchange of energy.
If an increase in pressure compresses the tissue, work is done on the tissue which is
recovered as the pressure drops and the tissue expands
In general, the returned energy is out of phase with the travelling wave, so the
beam intensity is decreased. The lost energy is converted to heat. The absorption is
frequency dependent and varies for different materials. As the absorption usually
increases with frequency, it is normally quoted in terms of dB cm−1 MHz−1, but this
should not be taken to mean that the absorption is strictly linear with frequency.
Typical values are given in table 7.2. The very high absorption in bone and lung
tissue means that they are effectively opaque to ultrasound, and structures which are
behind them will be hidden. As the absorption rises with frequency, there will be a
maximum depth for detecting echoes with ultrasound of a particular frequency.
Frequencies of 5–10 MHz can be used for scanning the eye, but the upper limit for
the abdomen is 2–3 MHz.
Thank you
Insert the title of your subtitle Here