Dräger-Tubes®

Short-Term Tubes
Dräger-Tube® gas detection
Short-term tubes

 Dräger-Tubes are measurement instruments for detection of


airborne pollutants
 Short-term tubes are intended for the measurement of
momentary concentrations. The measurement result is only
applicable to the location and time of the measurement
 The Dräger-Tube consists of a small glass tube. The tube tips
are fused on both ends. The tube contains a chemical
preparation that reacts with the substance to be measured
 To perform a measurement the tips of the tubes will be opened
and the tube will be inserted into a das detection pump.
Squeezing the pump the air to be measured will be sucked
through the tube. The chemical reagent discolored cause by the
substance to be measured
 We differentiate between scale tubes and color comparison
tubes

2 | 159 Title of presentation | Author of presentation | Date


3 | 159
Scale tube

Color
Types of tubes

comparison tube

Title of presentation | Author of presentation | Date


Qualitative
tube
Dräger-Tube® gas detection

Ampoule
tube

Tubes with
connecting hose

Combination
of 2 tubes

One-
Stroke-Tubes

Simultaneous
test set
Dräger-Tube® gas detection
Information on the tube

Quality control batch- /


serial-number

Number of strokes

Scale to evaluate
the concentration

Hermetically-sealed
glass tube housing

Order code

Substance to be
measured, Phosphine

Airflow directional arrow


(Arrow towards pump inlet)

4 | 159 Title of presentation | Author of presentation | Date


Dräger-Tube® gas detection
Information on the box

 Order code
 Batch- / serial-number
 Number of measurements (10, 9, 8, 5)
 Type of tube (name)
 Tube names in 4 languages
 Measuring range
 ISO 9001
 Storage temperature (25 °C )
 Expiry date

5 | 159 Title of presentation | Author of presentation | Date


Dräger-Tube® gas detection
Comparison to other detector tubes

Dräger-Tubes have a 7 mm diameter Tubes with a 5 mm diameter


 Less cross sensitivities  Higher cross sensitivities
PF:
PF:Please
Pleasecheck
checktext
text
 Short measuring time  Longer measuring times
– E.g., NH3 5 - 70 ppm, 1 minute – E.g., NH3 10 - 50 ppm, 3,5 minutes
 Clear length of stain  Shorter length of stain
– E.g., CO 2 - 60 ppm: 25 mm at TLV – E.g., CO 5 - 50 ppm: 11 mm at TLV AGW
 Long shelf life (2 years)  Ambiguous shelf life
 No cold storage required  Cold storage required
– 0 to 25 °C (0 to 77 °F) – 0 to 15 °C (0 to 59 °F)
 Advanced tube design  Pyrolyzer required
 Quality control batch testing  Batch testing
 ISO 9001 certification  ISO 9001 certification
 Detecting of more than 500 substances

6 | 159 Title of presentation | Author of presentation | Date