Professional Documents
Culture Documents
HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM (Autosaved)
HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM (Autosaved)
SYSTEMS
PARTS OF Building Supply
Pipe from the source of supply to a building or structure.
MEDICAL GAS Critical Care Area
AND VACUUM Special care units such as intensive care units, coronary
care units, cardiac catheterization, delivery rooms,
SYSTEMS FOR operating rooms and other units which are connected to
line operated, patient care related electrical appliances
MEDICAL Manifold
Device for connecting outlets of one or more gas
FACILITIES cylinders to the central piping system for the specific gas
Medical Air
Air which is supplied from cylinders, bulk containers, medical air
PARTS OF compressors or air which is reconstituted from oxygen USP.
According to the NFPA 99.3.3.106 and NFPA 5.1.3.5.1, medical
MEDICAL GAS
should be oil free, has dry nitrogen NF, must gave no detachable
hydrocarbons, have less than 25 ppm hydrocarbons and have equal
or less than 5 of permanent particulates which are sized 1 micron or
AND VACUUM larger at normal atmospheric pressure.
Medical Gas
SYSTEMS FOR Gas used in a medical facility including oxygen, nitrous oxide,
carbon dioxide, helium, medical air, and mixtures of these gases.
MEDICAL Standards of purity apply.
Medical Gas System
SYSTEMS FOR A valve that isolates one piece of equipment from the other
Riser Valve
MEDICAL A valve at the base of a vertical riser that isolates that riser
Service Valve
FACILITIES - A valve serving horizontal piping extending from a riser to a
station outlet or inlet.
PARTS OF
MEDICAL GAS Source Valve
A single valve at the source that controls the number of
AND VACUUM units that make up the source.
FACILITIES
PARTS OF MEDICAL GAS AND VACUUM SYSTEMS FOR MEDICAL
FACILITIES
Medical air compressors shall be installed in a well lit, ventilated, and clean
location and shall be accessible. The medical air compressor area shall be
located separately from the medical gas cylinder system sources, and shall be
readily accessible for maintenance.
Medical air compressors shall be sufficient to serve the peak calculated demand
with the largest single compressor out of service. In no case there be fewer than
2 compressors. ( NFPA 995 1.3.5.11.2)
Medical air compressor systems shall comprise of: an automatic means to
shutoff backflow, manual shutoff valve, intake filter mufflers, pressure relief
valves, shutoff valve that is compatible with oxygen that does not contribute to
contamination levels and lastly components that comply with NFPA
99.5.1.3.5.4, 99.5.1.3.5.10 and 99.5.1.3.5.11)
MEDICAL AIR
RECEIVERS