New Fluoro Taper Calibrations: Peter Varhegyi
New Fluoro Taper Calibrations: Peter Varhegyi
Peter Varhegyi
Fluoro Taper Calibration
feature content service impact
• Calibration can be executed entirely • Manual protocol selection is
from SUIF including acquisitions needed in the beginning of the
• Software guided calibration process calibration
(dynamic checks of system settings) • The calibration can be successful
• A check phase follows the calibration only if the check is successful too
to verify it • Failed check forces to go back to
• Possibility to reset the calibration the Fluoro Taper Calibration
• Dose values stored in xml file (iteration).
“IqInspecLog.xml” on DL
2
Prerequisites
Settings are displayed
here and dynamic
checks will be
performed on PA, SID
and DLB.
Protocol selection on
DL console is enabled
during this page (but
only during this page!)
WARNING:
You have to
reselect protocol!
Large pole
Small pole
During dose
measurements verify
that the dosimeter
chamber is in the center
of the image!
4
Setup for Innova 2100
Z shape holder
Image not on SUIF.
During dose
measurements verify
that the dosimeter
chamber is in the center
of the image!
5
Phases
Calibration Phase: Check Phase:
This sets the new Fluoro Taper This verifies the Fluoro Taper
2 dose measurements are needed for the After the calibration there is a mandatory
calibration phase at different SID. Follow check phase with 3 dose measurement and
the instructions to measure dose, set new similar measurement SUIF pages.
SID and measure dose again.
6
Dose measurements
Make two acquisitions and measure the
dose in the second!
2a
cqu
isit
ion
Select unit and accuracy on every s
page
7
Reset and iterate Reset: ”big problem; start from default”
The acquired data depends on the initial calibration status of
Calibration phase the system. If for some reason the system is far away from
proper calibration, then it is possible that the acquired dose
data is rejected in calibration. Restoring the default
calibration status can solve this problem. The "Reset" button
appears on some error cases for this reason. By pressing the
"Reset" button the default calibration status is restored and the
entire calibration has to be repeated.
8
Trouble shooting
Q: I press a button and nothing happens
A: SUIF window can hide the popup messages of the DL console. Check that there is no popup on DL by minimizing
or moving the SUIF window.
Q: I measure about 300 mR/min and dose limit is displayed as 87600000 mGy/min
A: You forgot to make a protocol selection during the prerequisite page. Abort the current calibration section, restart it
and make a protocol selection!
A2: If calibration still not converging your tube can be weak. In that case there is no simple solution, inform us about
the problem.