Procedure On LAA Device Closure
Procedure On LAA Device Closure
FOR LAA
DEVICE
CLOSURE
Atrendynurse
DISCRIPTION
Left atrial appendage closure is a treatment strategy to reduce the
risk of left atrial appendage blood clots from entering the
bloodstream and causing a stroke in patients with non-valvar atrial
fibrillation.
ARTICLES NEEDED
Puncture needle-2(18 g or 20 g) for artery and vein Sheath introducer 11cm 10f
long Mullen sheath 8f
Guidewire Teflon 035 150 st tip
Catheter marker pigtail
Guidewire Teflon 035 150 j tip
Septal puncture needle Long sheath with dilator Device delivery cable
Cath lab kit with kidney tray set 3 way with pressure line
pressure injector and syringe or contrast in 50ml syringe
PREPARATION OF STERILE
TROLLY
Sterile Scrubbing, Gowning and Gloving Preparation of sterile
trolley Antimicrobial painting( From umbilicus to middle of thigh)
Draping the patient in such a way that inguinal fold lies diagonal to
the hole
Load 10 ml of xylocaine in BD syringe Load Heparin
5000iu(According to patient's weight) in 2ml syringe Flush the
puncture needle sheath, dilator catheters and guide wire
PROCEDURE
Puncture the femoral vein and artery with puncture needle (18 GX 7CM)
Introducing 0.035"Teflon introducer wire Removing puncture needle
Making a small incision with scalpel blade no 11
Introducing 10F sheath with dilator in the vein
Introducing 5F sheath with dilator in the artery Remove the dilators with wire
& flush the sheaths with heparinized saline.
Monitor the arterial pressure through the femoral arterial sheath.
PROCEDURE
Insert the long sheath with straight Teflon wire through the femoral
vein sheath to RA - Get an access to the LA by puncturing the
septum by using septal puncture needle with the help of echo and
fluoroscopy.
Position the long sheath in the LA and adequate flushing to be done
with heparinized saline and monitor LA pressure.
PROCEDURE
Heparin 5000 iu should be administered once the sheath positioned
in LA.
Advance marker pigtail through the long sheath with Teflon wire
and reach the LA appendage.
Take proper cine to make out the appendage size with 50ml of
contrast.
Exchange the pigtail catheter with a Long sheath which is
compactable with the Device by using stiff Teflon wire.
PROCEDURE
Preload the Device in the delivery cable.
Push the device through the long sheath and position the device in
the LA appendage.
Deploy the device with the guidance of echocardiogram(TEE is
recommended) and adequate fluoroscopy.
Remove the sheath and compress the site till bleeding stops.
Pressure dressing to be applied to avoid bleeding complications
AAA