Life Pack 12

 Biphasic technology.

 Compatible with QUIK-COMBO™


pacing/ defibrillation/
ECG electrodes.
Life Pack 12
What can it do?
 ECG MONITORING

 ACQUIRING A 12-LEAD ECG

 PULSE OXIMETRY

 AUTOMATED EXTERNAL DEFIBRILLATION

 MANUAL DEFIBRILLATION

 SYNCHRONIZED CARDIOVERSION

 NONINVASIVE PACING

 NIBP MONITORING

 THERAPY WARNINGS
CONTROLS AND INDICATORS
Controls and Indicators - Area 1
 ON Button: Switches power on / off

 Energy Select: Selects energy levels in


manual shock. Press the up or down button.

 Charge: When the defibrillator is charging a
charging bar appears and a ramping tone
sounds.

 Shock: Discharges the shock to the patient.


If shock is not delivered in 60 sec, the stored
energy is removed. If you change the
charge energy, the energy is removed and
you will need to recharge.

 Sync: Use for cardioversion.

MO-1 DMAT
Controls and Indicators - Area 1
 Battery Charge: Indicates that at least one
battery is installed and being charged by an
AC/DC power adapter

 Service: Indicates that the internal self-


check failed

 Advisory: Activates continuous patient


surveillance system (CPSS) which monitors
the patient for potential shockable rhythm

 Analyze: Activates the Shock Advisory
System (SAS). SAS is a ECG analysis system
that will advise the user if it detects a
shockable or nonshockable rhythm

MO-1 DMAT
Controls and Indicators - Area 2

 Pacer: Switches pacer on and off indicated


by light. Once the ECG confirm a triangle
sensor marker on the top of the QRS, if not
change leads or adjust.

 Rate: Press the button (10ppm) or rotate


the dial (5ppm)to the desired rate

 Current: Press the button for 10 increments


or rotate the dial for 5 increments until
capture occurs. Pacer indicator will flash
with each pacing stimulus.

 Pause: Push and hold this button to pause


pacing. Pacing will continue at 25% of set
rate. Release button and pacing will
continue at set rate
Controls and Indicators - Area 3

 Lead: Select or change ECG leads. Press


the lead button again or rotate the dial to
the desired lead. Display will be on the right
side of the screen.

 Size: Press the button a display will appear.


Either press again or rotate the dial to the
desired ECG size.

 NIBP: Press the button to activate. Cuff


should inflate 30mmHg higher then pt
systolic pressure.

 Alarms: Can be configured on or off.


Alarms are set to a predetermined limits.
They can be adjusted.
Controls and Indicators – Area 3

 Options: Press the button and a screen will


display Patient, Pacing, Date/Time, Alarm
volume, Archives, Print and User Test.

 Events: Press the event button and scroll


thru the screen selecting the choice from
Adenosine, ASA, Atropine, Bicarb, CPR,
Dopamine, Epinephrine, Intubation for
example. By pressing the button a
timestamp will appear on the code summary.

 Home Screen: Pressing this button will


return you to from any overlay screen to the
home screen EXCEPT during AED analysis or
manual defibrillation charging or shocking.
Controls and Indicators - Area 4

 CO2 tubing connector: Open the connection door and


connect the Capnography tubing next connect to the
patient.

 SPO2 Cable Connector: When plugged in performs a 17


sec calibration and 3 sec self test. Measures SpO2
between 50% and 100%.

 NIBP Tubing Connector: When connect there is a 3 sec


self test. Typically takes pressure within 40 sec. but no
longer 120 sec.

 ECG Cable Connector: Connect the trunk lead and the


limb leads and/or the 12 leads are connected the trunk
lead.
Controls and Indicators - Area 4

 Speaker

 Printer

 Therapy Cable Connector: Connect cable to therapy electrodes or


to the paddles.

MO-1 DMAT
Controls and Indicators - Area 5

 12 Lead: Press to initiate acquisition of a 12


lead ECG.

 Transmit: Capability to transmit patient data.

 Code Summary: Pressing button will print the


code summary with critical events that were
recorded.

 Print: Press the button to start and to stop the


printer.
Controls and Indicators - Area 6

MO-1 DMAT
ECG MONITORING

ECG Monitoring - Patient ECG cable

ECG Monitoring – Combo Electrode

Selecting ECG leas and size

Adjusting systole volume

Alarm settings
SpO2 MONITORING

 Adjust Sensitivity normal/high

 Average Sensing time 4,8,12 and 16 Sec

 Hi light and select SpO2 volume

 Cover sensor if excessive ambient light present

 SpO2 waveform displays on channel 2


AUTOMATED EXTERNAL DEFIBRILLATION

 AED – Do not place electrodes in


Anterior and posterior

 AED – Not for Pediatric less than 8Yrs

 Press Advisory – (CPSS On)

 Press Analyze – (SAS On)

 AED – Motion detector Auto on


AED SAMPLE VIDEO

AED SAMPLE VIDEO.MSWMM


MANUAL DEFIBRILLATION

 Press Energy Select or Rotate energy


select dial

 Press charge – Press shock

 Shock not pressed in 60sec. Auto


discharge internally
SYNCHRONIZED CARDIOVERSION

 ECG through Therapy Electrode, Select Paddles

 Press sync – LED blinks each QRS

 Confirm triangle sense marker

 Press Energy Select or Rotate energy select dial

 Press charge – Confirm ECG rhythm and


available energy

 Press and hold shock button until discharge


occurs with next QRS complex
NONINVASIVE PACING

 Demand and Nondemand Pacing

 Place ECG and Therapy electrodes

 Press pacer. Confirm LED illuminates

 Press rate or rotate selector set PPM

 Press current or rotate selector – electrical capture


occurred

 View intrinsic rhythm press and hold pass.(25%)


NONINVASIVE BLOOD PRESSURE
 Initial Inflation pressure

 Manual single measurement

 Timer controlled measurement

 Pulse not displayed when spo2 / ECG active


THERAPY WARNINGS
 Do not clean with ketones or other flammable agents. Do
not autoclave or sterilize this device accessories

 Do not use this device in the presence of flammable gases


or anesthetics. Use care when operating this device close
to oxygen sources (such as bag-valve-mask devices or
ventilator tubing). Turn off gas source or move source
away from patient during defibrillation.

 Avoid operating device near cauterizers, diathermy


equipment, or cellular phones. Maintain equipment
separation of at least 1.2 mtrs.
CODE SUMMARY REPORT
THANK YOU