Professional Documents
Culture Documents
Poster
Poster
Impaired Individuals
Kenneth Lee and Kayla Seto
There is a need to provide an alternative A volunteer test participant was recruited to
Specification Unit Description
method for visually impaired individuals to test Safe N/A As safe as current methods verify and validate our design.
blood glucose. Capable of being sterilized or consist of parts that
To verify and validate our prototype design, Mr. Raymond
N/A can be disposed and replaced
Only one percent of the blind population is born without N/A Non-perishable for up to 1 year Bishop agreed to test our device. We had him follow our
sight. The vast majority of people lose their vision later in life Accurate transfer/collection Enough blood transfered and collected from 1 µL
instructions to first become accustomed to the device. We
due to illnesses such as macular degeneration, glaucoma and of blood µL blood droplet to make get a reading then had him run through the entire set up and glucose testing
diabetes. Diabetic retinopathy is a complication from diabetes Accurate as current Must not decrease the accuracy of current scenario with a simulated prick. We also tested accuracy by
that occurs when diabetes damages the tiny blood vessels inside methods of glucose testing N/A methods dropping testing solution where the blood droplet would
User friendly N/A surface and then had Mr. Bishop let the magnets guide his
the retina. While there are some treatments to delay these Considers people with low strength, low dexterity,
conditions, there is no cure. Within the United States, there are low touch, and low blood circulation to the finger to the test strip. Finally, Mr. Bishop answered a survey
fingertips regarding user friendliness of the device.
approximately 4.1 million adults, aged 18 or older, reporting sec Assembly must take less than 20 sec
visual impairment with diagnosed diabetes. This is projected to
Method must include less than 5 parts excluding
be 7.2 million by 2020, and could reach 16 million by 2050. parts glucometer/test strip together
The difficulty of the current method for those with extreme of Portable N/A Largest piece must fit into quart sized plastic bag
complete vision loss has not been adequately addressed, even oz Weigh less than 4 ounces
though a few solutions exist. Fig. 1 shows the accuracy needed to Inexpensive $/yr Less than $475/yr
transfer blood to the glucometer test strip.
Acknowledgments
The prototype employs magnetic guidance to draw Thanks to our client, Suzanne Erb, for providing us with the idea for this
the finger to the test strip. project and guidance along the way. Also thank you to Raymond Bishop for
volunteering to test our device.
The device is made of a glucometer frame and a finger attachment. The
frame is a trestle design with folding legs and holds the glucometer with References
the test strip hanging from the bottom. A ring magnet surrounds the test
Healthline.com. 19 July, 2014,
strip. So the test strip is centered The finger attachment slides onto the https://www.healthline.com/diabetesmine/ask-dmine-the-worst-intro-
finger and centers the lancing device. Another magnet in the finger to-blood-sugar-testing-ever#1.
attachment pulls the finger toward the test strip, thus sight is not Park et al. “Haptic glucometer guide.” Google Patents. Jun 2016,
Figure 2. Haptic Glucometer Guide. needed. Fig. 3 shows how the device is intended to be used. https://www.google.com/patents/US9364183.