Topic Assigned: Submitted To:: Online Gas Chromatograph Excel Technological Services (Sir Wajeeh Abbas) Akasha Noor
Topic Assigned: Submitted To:: Online Gas Chromatograph Excel Technological Services (Sir Wajeeh Abbas) Akasha Noor
Comparison
Packed Column Capillary Column
Parameter
Glass or stainless
Material Fused Quartz
steel
Length 1 to 3 meters 10 to 100 meters
Internal Diameter 2 to 4mm 0.1 to 1mm
Higher efficiency,
Produce broad peak,
Advantage handle large sample faster, better for
complex mixtures
volumes
Detectors
Several types of detectors are available for gas chromatographs, including
Detectors Selectivity Sensitivity
Thermal Conductivity Detector Universal 10^-9 g
(TCD)
Flame photometric Detector Sulphur & 10^-5 g
(FPD) phosphorus
Flame Ionization Detector (FID) Organic compounds 10^-11 g
Selectivity:
It determines the type of compounds that the detector will detect.
Sensitivity:
It defines the concentration levels the detector will detect.
Codes & Standards
Meter type (Orifice ,Turbine,
Sr# Standard PD, Coriolis Purpose
& Ultrasonic )
1 AGA8 Applies to all types Density , Compressibility factor (z) ,super compressibility factor
Natural Gas energy measuerement (Heating Value) for custody
2 AGA5 Applies to all types
transfer
3 API14.1 Applies to all types Collecting and handling of Natural gas samples for custody tranfer
4 GPA 2166-05 Applies to all types Techniques for natural gas spot sampling
7 GPA 2145-03 Applies to all types Table of Physical Constants for Hydrocarbons &
Other Compounds of Interest to the Natural Gas Industry
Selection, Preparation, Validation, Storage of Natural Gas &
8 GPA 2198-03 Applies to all types Natural Gas Liquids Reference Standard Blends
Calculation of Gross Heating Value, Specific Gravity,
9 GPA 2172-96 Applies to all types & Compressibility of Natural Gas Mixtures from Compositional
Analysis
Converting Mass of Natural Gas Liquids & Vapors to Equivalent Liquid
10 API 14.4 Applies to all types
Volumes
Difference b/w NIST & ISO 17025
ISO 17025 (Internationally recognized
NIST
standard)
Does not take into account the Certifies the competence of the
monitoring of the environment in laboratory that produces the
which the calibration was informed calibration gas
(Two remote serial digital communications ports, One USB Host and one
Ethernet port)
• Display (Built in/ separate unit) 1/4 VGA display with multiple
screen access