Professional Documents
Culture Documents
Imperfections in SOLIDS ( ) : Dr. Aneela Wakeel
Imperfections in SOLIDS ( ) : Dr. Aneela Wakeel
Imperfections in SOLIDS ( ) : Dr. Aneela Wakeel
IMPERFECTIONS IN
SOLIDS(Ⅱ)
Dr. Aneela Wakeel
2
Characteristics of dislocations
3
4
5
6
7
8
Beam
splitter Reflected light
Olympus
BX51
Research
Microscope
Cutaway
Diagram
Transmitted light
http://micro.magnet.fsu.edu/primer/java/microassembly/index.html
Objective Lens
dmin = 0.61l/NA
Objective specifications Anatomy of an objective
rical
ture
(Diaphragm)
M=(L/fo)(25/fe)
Eyepieces (Oculars) work in combination with microscope
objectives to further magnify the intermediate image
http://micro.magnet.fsu.edu/primer/anatomy/oculars.html
http://www.birdwatching.com/optics/diopter_set.html
Common Modes of Analysis
Depending on the nature of samples, different illumination
methods must be used
• Transmitted OM - transparent specimens
thin section of rocks, minerals and single crystals
• Reflected OM - opaque specimens
most metals, ceramics, semiconductors
Specialized Microscopy Techniques
• Polarized LM - specimens with anisotropic optical
character http://www.youtube.com/watch?v=ulNZ3u7_J5I to ~1:05
Resolution (Lateral)
The smallest distance between two specimen points
that can still be distinguished as two separate entities
To achieve high
resolution, We
should use dmin = 0.61l/NA NA=nsin()
Shortest wavelength
Larger numerical
aperture
https://www.youtube.com/watch?v=n2asdncMYMo at~5:35-6:00
What is resolution?
Resolution describes the minimal distance of two points that can be distinguished.
α2
α1
Objective lens
Oil (n = 1.5)
Air (n = 1.0)
Coverslip (n = 1.5)
Glass slide (n = 1.5)
1.1
1.0
0.9
Normalized Intensity
0.8 1.4 NA
0.7
0.6
0.5
0.4
0.7 NA
0.3
0.2
0.1
0.0
-1.10
-0.96
-0.81
-0.66
-0.51
-0.37
-0.22
-0.07
0.37
0.66
0.96
0.07
0.22
0.51
0.81
1.10
Microns
Increasing NA
b) a)
NA f F
0.1 0.13 15.5
0.4 3.8 5.8
.95 80.0 0.19
The distance above and below The axial range through which
geometric image plane within an object can be focused without
which the image is in focus any appreciable change in image
sharpness
M NA f F
F is determined by NA.
M NA f F
http://www.matter.org.uk/tem/depth_of_field.htm
http://www.youtube.com/watch?v=FvC2WLUqEug at~3:40
Brightness
Brightness refers to the intensity of light.
In a transmission light microscope the brightness is related to
the numerical aperture (NA) and magnification (M).
http://micro.magnet.fsu.edu/primer/techniques/contrast.html
Contrast
Contrast is defined as the difference in light intensity
between the specimen and the adjacent background
relative to the overall background intensity.
Sspecimen-Sbackgroud S
C= =
Sspecimen SA
Objective Lens
x100
Reflected OM
Grain Growth - Reflected OM
5m 30mm
m
Scanning electron
microscopy (SEM)
Dr Aneela Wakeel
39
Probe scanning
• Probe scanning is operated by a beam deflection system incorporated within
the objective lens in an SEM. The deflection system moves the probe over
the specimen surface along a line and then displaces the probe to a position
on the next line for scanning, so that a rectangular raster is generated on the
specimen surface.
• The signal electrons emitted from the specimen are collected by a detector,
amplified, and used to reconstruct an image, according to one-to-one
correlation between scanning points on the specimen and picture points on a
screen of a cathode ray tube(CRT) or liquid crystal display.
• The deflection system of the electron probe is controlled by two pairs of
electromagnetic coils (scan coils). The first pair of coils bends the beam off
the optical axis of the microscope. The second pair of coils bends the beam
back onto the axis at the pivot point of a scan.
43
Detector
Image
44
Electron guns
• We want many electrons per time unit per
area (high current density) and as small
electron spot as possible
• Traditional guns: thermionic electron gun
(electrons are emitted when a solid is
heated)
• W-wire, LaB6-crystal
Electron guns
• A field emission gun is 1000×brighter than a tungsten thermionic gun and
100×brighter than a LaB6 thermionic gun. The brightness is proportional to
the acceleration voltage (Vo) of the gun.
• With field emission guns we get a smaller spot and higher current densities
compared to thermionic guns
• Vacuum requirements are tougher for a field emission guns
Signal detection
When high energy electrons strike a specimen, they produce either elastic or
inelastic scattering.
Elastic scattering produces the backscattered electrons (BSEs), which are
incident electrons scattered by atoms in the specimen. BSEs are typically
deflected from the specimen at large angles and with little energy loss; they
typically retain 60–80% of the energy of incident electrons.
Backscattered electron
detector:
(Solid-State Detector)
Contrast formation
X-rays
Sample
52
BSE vs SE
X-rays
• Photons not electrons
• Each element has a fingerprint X-ray
signal
• Relatively few X-ray signals are
emitted and the detector is inefficient
relatively long signal collecting times are
needed
Depth of focus
Optical microscopy vs SEM
Topographic contrast
Topographic contrast in an SEM refers to variation in signal levels that corresponds
to variation in geometric features on the specimen surface. An SEM image with
topographic contrast often has the stereoscopic appearance of a rough specimen
surface.
Topographic contrast occurs because signal electrons arise from two effects: the
trajectory effect and the electron number effect.
The trajectory effect arises from variations in how the specimen surface is
oriented with respect to the detector.
60
Topographic contrast
In an SEM image, the electron number effect will create bright areas in the image
that do not correspond to surface contours on the specimen. Figure 4.13 illustrates
the electron number effect. When the electron probe hits a surface at an angle,
more electrons can escape from the specimen than when the probe hits a flat
surface directly. Thus, certain areas of the specimen (such as edges of spherical
particles, raised areas and cavities) will appear bright in an SEM image. Figure 4.14
shows the example of topographic contrast from the electron number effect
Compositional Contrast
Compositional contrast refers to the variation in gray levels in an SEM image that
correspond to variation in chemical composition in a specimen.
An image formed by BSEs exhibits very useful compositional contrast if the
specimen consists of more than one chemical element.
The origin of compositional contrast arises because the capability of BSEs to
escape from the specimen depends on the atomic numbers of the specimen atoms
Compositional contrast
Thus, any area in a specimen containing chemical elements with higher atomic
number will generate more BSEs. The difference in the number of BSEs collected
by a detector will appear as differences in gray levels in a black and white image;
that is, an area with atoms of higher atomic numbers will appear brighter. Thus, a
BSE image shows the atomic number contrast or compositional contrast as
demonstrated in Figure 4.16
63
In-situ imaging
• A modern SEM can be equipped with various
accessories, e.g. a hot stage
64
Summary
• The scanning electron microscope is a versatile
instrument that can be used for many purposes
and can be equipped with various accessories
• An electron probe is scanned across the surface
of the sample and detectors interpret the signal
as a function of time
• A resolution of 1 – 2 nm can be obtained when
operated in a high resolution setup
• The introduction of ESEM and the field emission
gun have simplified the imaging of challenging
samples
66
Summary
Signals:
•
• Secondary electrons (SE): mainly topography
• Low energy electrons, high resolution
• Surface signal dependent on curvature
• Backscattered electrons (BSE): mainly chemistry
• High energy electrons
• “Bulk” signal dependent on atomic number
• X-rays: chemistry
• Longer recording times are needed
67
Electron Back
Scattered Diffraction
Introduction to EBSD
EBSD = Electron Backscatter Diffraction
EBSD is a technique that allows:
• Crystallographic Orientations
• Misorientations
• Texture trends
• Grain size and boundary types
• Phases
Deformed silica
(quartz)
• Nickel
• Austenitic
Stainless
Steel
Introduction to EBSD - Color Key for COMs
• COM with
Inverse Pole
Figure color key
for cubic
material
• Red = 100
• Green = 110
• Blue = 111
planes parallel
to the surface
72
TEM
74
75
76
77
78
79
80
81
82
83
84
85
86
87
http://spm.phy.bris.ac.uk/
90
Operation
• Scanning probe
microscopes operate
by detecting the
deflection in the
cantilever
• Modern scanning
probe microscopes
use a split photo
diode to detect the
deflection http://spm.phy.bris.ac.uk/
92