Professional Documents
Culture Documents
e Learning檔案 2.1醫學影像 2.1.7 Principleof Computed Tomography
e Learning檔案 2.1醫學影像 2.1.7 Principleof Computed Tomography
e Learning檔案 2.1醫學影像 2.1.7 Principleof Computed Tomography
tomography
嘉義長庚放射科 廖書柏
Introduction
In 1950 , Allan M. Cormack develop the theoretical and
mathematical methods used to reconstruct CT images.
ionization
Electrical signal
Solid-state
Solid state
Ceramic or crystal scintillatior
Photon capture
Light
Photo-diode
Electrical signal
Collimators
Pre-patient collimator- control slice thickness
Pre-detector collimator-reduce scattered radiation
History of CT
Variations in scanner design based on :
X-ray tube and detector movement
Detector arrangement
Rotating mechanism
First-generation ~1972
single X-ray tube and one
detector element
Pencil beam
about 5 minutes per slice
from 180 degrees rotation.
Translate-rotate movement
EMI CT1000
Second-generation ~1975
Translate-rotate movement
Third-generation ~1975
Single X-ray tube, rotating
movement
Multiple detectors in
curvilinear design,
rotating movement
Fan beam(~30 。 )
Several seconds per slice
Rotate-rotate movement
Fourth-generation ~1976
Single X-ray tube, rotating
movement
Fixed ring as many as
8000 detectors inside of
gantry
1-s scan time
Avoiding ring artifact
problem of 3rd generation
scanner
Rotate-stationary movement
Fifth-generation ~1984
four semicircular tungsten target rings spanning 210
degrees about the patient
Multiple detectors of two banks, fixed inside of the
gantry
no mechanical movement
By using four target rings and two detector banks,
eight slices of the patient may be imaged without
moving patient.
EBCT( electron beam CT)
I0 x I
I = I0 e -μx
I0 x1 x2 x3 I
I = I0 e -(μ1x1+μ2x2)
I0 x1 xn I
n
I = I0 e Σ
- i=1μixi
CSF 0.207
Water 0.206
Fat 0.185
Air 0.0004
Linear attenuation coefficient of various body tissues for 60 keV x-ray
The hounsfield scale of CT numbers
Image reconstruction
Image reconstruction
The image is reconstructed from projections by a
process called Filtered Backprojection .
"Filtered" refers to use digital algorithms called
convolution to improve image quality or change certain
image quality characteristics, such as detail and noise
"Backprojection" is the actual process used to produce or
"reconstruct" the image.
The filtered backprojection process involves the
following steps:
generating a sinogram from a set of N projections
filtering the data to compensate for blurring
Backprojecting the data
.
Projection and sinogram
Ray: the X-ray read by every one detector within a short
time interval.
Projection:yall rays sum in a direction
Sinogram: all projections
P(t)
t
x
μ(x,y)
t
X-rays Sinogram
P(t) =
Filter
a de-blurring function is combined (convolved) with the
projection data to remove most of the blurring before the
data are backprojected.
A high-frequency filter reduces noise and makes the
image appear “smoother.”
A low-frequency filter enhances edges and makes the
image “shaper.”
中央切面投影理論
(Central Slice Projection Theorem, CSPT) :
If a 1D Fourier Transform is performed on a projection of
an object of some angle, the result will be identical to one
line on 2D Fourier Transform of that object and at that
angle.
Central Slice Projection Theorem
y ky
P(t)
t
F[P(t)]
x kx
μ(x,y) F(kx,ky)
CSPT can relate the Fourier transform of the projection to one line in the 2D K
space formed by the 2D Fourier transform of μ(x,y)
y
ky
F-1[F(kx,ky)] v
x u
kx
u(x,y) F(kx,ky)
Inverse 2D-FT
object 1D-FT
image
Inverse 2D-FT
Backprojection
Filtered backprojection
Slice thickness 5 mm 5 mm
overlaping 50%
RI=2 mm
overlaping (5-
2)/5=60%
Image quality
Spatial resolution in CT
focal spot size
detector dimensions
Slice thickness
Pixel size Pixel size= FOV/matrix size
Pitch
artifact
Image Artifact in CT
Image Artifact
Artifacts are any discrepancy between the CT numbers
represented in the image and the expected CT numbers
Common artifacts
Beam hardening
Partial volume effect
bad detector(3th scanner)
Metal
Patient motion
Beam hardening effect
Linear attenuation coefficients vary with photon energy.
After passing through a given thickness of tissue , lower-energy
x-rays are attenuated to a greater extent than high-energy x-rays
are.
artifacts such as a reduced attenuation toward the center of
tissue (cupping) and streaks that connect tissues with strong
attenuation.
polychromatic spectrums
Cupping Artifact
Means for suppressing beam hardening effect
pre-filtering X-rays
avoiding high X-ray absorbing regions if
possible
applying appropriate algorithms
Partial volume effect
Partial volume artifacts are the result of a variety of
different tissue types being contained within a single voxel
Measured attenuation coefficient are averaged by all
components
use thinner slice to reduce
bad detector(3th scanner)
96 24 12
the number of views (projections)
Effect of reducing rays
200 50 25