K-4500 Service Program
K-4500 Service Program
K-4500 Service Program
Purpose:
Skips automatic cleaning during the start up and
enters the Ready state earlier
Purpose:
Runs the diaphragm pump and allows you to
check the sample dispensation volume.
Function:
One cycle of the diaphragm operation carries
out when this program is selected.
Purpose:
Checks operation of the Sampler sensors and SV
´s
Function:
Runs the Sampler SV´s
Displays the sensor ON/OFF status when the
corresponding SV is activated.
Purpose:
Checks the operation of the MCP sensors and
SV`s
Function:
Runs the Sampler SV´s
Displays the sensor ON/OFF status when the
corresponding SV is activated.
Purpose:
Prints the error history together with date and time.
Function
Up to the ten last errors can be recorded.
The error history can be cleared by setting PCB
3648 DIP SW 3 ON and turning ON the power
switch.
Only specific errors are printed in history.
Function:
Allows you to enter the % data for five parameters
to correct carryover differences between
instruments.
Three modes are available to enter correction
values.
Correction values in the Whole blood mode are
effective in all Modes.
When the Print key is pressed during setting, the
current setting values will be printed
Purpose:
Change analysis parameter labeling
Function:
Changes the WBC analysis labeling.
(W-SCR/LYMPH%)
Selects RDW (RDW-SD/ RDW-CD)
Selects PDW/ P-LCR output (output/do not output)
Purpose:
Checks Settings
Function:
When the print key is pressed, all settings
including Service Program Settings are printed
Purpose:
Initialize all the settings and values including those
in the service program
Caution:
COC settings are also reset to 100%. You
must print the values before initializing.
Function:
Checks the computer related functions (for
debugging
Caution:
Using this program is usually inhibited. Using
this menu causes the subsequent operation
to stop the system cannot recover until the
power switch is turned off and then on again.