PRINCIPLE OF MRI IMAGING

Nuclei with an odd number of protons and neutrons


possess a property called spin.
 In quantum mechanics spin is represented by a magnetic
spin quantum number.
Spin can be visualised as a rotating motion of the nucleus
about its own axis.
As atomic nuclei are charged, the spinning motion causes
a magnetic moment in the direction of the spin axis.
The strength of the magnetic moment is a property of the
type of nucleus.
Hydrogen nuclei (1H), as well as possessing the strongest
magnetic moment, are in high abundance in biological
material.
Consequently hydrogen imaging is the most widely used
MRI procedure.
A charged, spinning nucleus creates a magnetic
moment which acts like a bar magnet (dipole)
SPINNING OF 1H NUCLEI
Consider a collection of 1H nuclei (spinning
protons).
In the absence of an externally applied magnetic
field, the magnetic moments have random
orientations.
However, if an externally supplied magnetic field
B0 is imposed, the magnetic moments have a
tendency to align with the external field .
Nuclei with an odd number of protons and
neutrons possess a property called spin.
(A)A collection of 1H nuclei (spinning protons) in the
absence of an externally applied magnetic field. The
magnetic moments have random orientations.

(b) An external magnetic field B0 is applied which causes


the nuclei to align themselves in one of two orientations

with respect to B0 (denoted parallel and anti-parallel).


PARALLEL AND ANTI PARALLEL
ORIENTATION
The magnetic moments or spins are constrained to adopt
one of two orientations with respect to B0, denoted parallel
and anti-parallel.
The angles subtended by these orientations and the direction
of B0 are labelled theta.
The spin axes are not exactly aligned with B0, they precess
around B0 with a characteristic frequency.
This is analogous to the motion of a spinning top precessing
in the earth's gravitational field.
Atomic nuclei with the same magnetic spin quantum number
as 1H will exhibit the same effects - spins adopt one of two
orientations in an externally applied magnetic field.
Elements whose nuclei have the same magnetic spin
quantum number include 13C, 19E and 31P. Nuclei with higher
magnetic spin quantum number will adopt more than two
orientations.
(a) In the presence of an externally applied magnetic
field, B0, nuclei are constrained to adopt one of two
orientations with respect to B0. As the nuclei possess
spin, these orientations are not exactly at 0 and 180
degrees to B0.
(b) A magnetic moment precessing around B0. Its
path describes the surface of a cone.
 The Larmor equation expresses the relationship
between the strength of a magnetic field, B0,
and the precessional frequency, F, of an
individual spin.

 The proportionality constant to the left of B0 is


known as the gyromagnetic ratio of the nucleus.
 The precessional frequency, F, is also known as
the Larmor frequency. For a hydrogen nucleus,
the gyromagnetic ratio is 4257 Hz/Gauss. Thus
at 1.5 Tesla (15,000 Gauss), F = 63.855
Megahertz.
Radiofrequency field and MR
signal
 For a collection of 1H nuclei, let the number of spins
adopting the parallel and anti-parallel states be P1 and P2
respectively, with corresponding energy levels E1 and E2.
 E2 is greater than E1 causing P1 to be greater than P2.
 An obvious question is why do spins adopt the higher
energy anti-parallel state?
 The answer is that spins of P2 may move to P1 if the
exact amount of energy, delta(E) = E2 - E1 is supplied to
the system.
 If the temperature of the system were absolute zero, all
spins would adopt the parallel orientation.
 Thermal energy will cause P2 to be populated.
 At room temperature in a 1.5 Tesla magnetic field, there
will typically be a population ratio P2:P1 equal to
100,000:100,006.
 Consider a collection of 1H nuclei (spinning protons).
Radiofrequency field and MR
signal
 At any given instant, the magnetic moments of a
collection of 1H nuclei can be represented as
vectors.
 Every vector can be described by its components
perpendicular to and parallel to B0.
 For a large enough number of spins distributed on
the surface of the cone, individual components
perpendicular to B0 cancel, leaving only
components in the direction parallel to B0.
 As most spins adopt the parallel rather than the
antiparallel state, the net magnetisation M is in the
direction of the B0 field.
A collection of spins at any given instant in an external
magnetic field, B0. A small net magnetisation, M, is
detectable in the direction of B0.
 Suppose the direction of B0 is aligned with the z-axis of
Euclidean 3-space.
 The plane perpendicular to B0 contains the x and y-axes.
In order to detect a signal from 1H nuclei, radio
frequency (RF) energy must be applied.
 RF energy at the Larmor frequency causes nuclear spins
to swap between parallel and anti-parallel states.
 This has an oscillatory effect on the component of M
parallel to the z-axis. RF energy, like all electromagnetic
radiation, has electric and magnetic field components.
 Suppose the magnetic field component is represented by
B1 and lies in the x-y plane.
 The x-y components of M will be made coherent by the
B1 field giving a net x-y component to M and hence
effectively cause M to tilt from the z direction into the x-y
plane.
 The angle through which M has rotated away from the z-axis is
known as the flip angle.
 The strength and duration of B1 determine the amount of energy
available to achieve spin transitions between parallel and anti-
parallel states.
 Thus, the flip angle is proportional to the strength and duration of
B1.
 After pulses of 90 degrees and 270 degrees, M has no z component
and the population ratio P2:P1 is exactly one.
 A pulse of 180 degrees rotates M into a position directly opposite to
B0, with greater numbers of spins adopting anti-parallel (rather than
parallel) states.
 If the B1 field is applied indefinitely, M tilts away from the z-axis,
through the x-y plane towards the negative z direction, and finally
back towards the x-y plane and z-axis (where the process begins
again).
The effect of RF radiation on the net magnetisation M is to
produce a second magnetic field Mx-y. M is tilted from its
original longitudinal z-axis orientation, along the direction of
the external magnetic field B0, into the transverse x-y plane.
An illustration of flip angle, which is the angle through which
M has rotated away from the z-axis.
Figure shows the situation after an RF pulse is applied that
causes the net magnetisation vector M to flip by 90
degrees.
M lies in the x-y plane and begins to precess about the B0
axis. M will induce an electromotive force in a receiver coil
according to Faraday's law of magnetic induction.
This is the principle of NMR signal detection.
It is from this received RF signal that an MR image can be
constructed. Figure shows a graph of the voltage or signal
induced in a receiver coil verses time.
Such a graph, or waveform, is termed a free induction
decay (FID). The magnitude of the generated signal
depends on the number of nuclei contributing to produce
the transverse magnetisation and on the relaxation times
(see next section).
(a) After a 90 degrees RF pulse, M lies in the x-y plane and
rotates about the z-axis. The component of M in the x-y
plane decays over time. An alternating current, shown in
Figure (b), is induced in the receiver coil.
RELAXATION PROCESSES
 The return of M to its equilibrium state (the
direction of the z-axis) is known as relaxation.
 There are three factors that influence the decay
of M:
 Magnetic field inhomogeneity.
 Longitudinal T1 relaxation (Spin-Lattice
relaxation) is the realignment of spins (and so
of M) with the external magnetic field B0 (z-
axis).
 Transverse T2 relaxation (T2 decay, Transverse
relaxation or Spin-Spin relaxation) is the
decrease in the x-y component of
magnetisation.
MAGNET INHOMOGENEITY
 It is virtually impossible to construct an NMR magnet with
perfectly uniform magnetic field strength, B0.
 Much additional hardware is supplied with NMR machines to
assist in normalizing the B0 field.
 However, it is inevitable that an NMR sample will experience
different B0's across its body so that nuclei comprising the
sample (that exhibit spin) will have different precessional
frequencies (according to the Larmor equation).
 Immediately following a 90 degree pulse, a sample will have
Mx-y coherent.
 As time goes on, phase differences at various points across
the sample will occur due to nuclei precessing at different
frequencies.
 These phase differences will increase with time and the
vector addition of these phases will reduce Mx-y with time.
T1 RELAXATION
 Following termination of an RF pulse, nuclei will dissipate their excess
energy as heat to the surrounding environment (or lattice) and revert to
their equilibrium position.
 Realignment of the nuclei along B0, through a process known as recovery,
leads to a gradual increase in the longitudinal magnetisation.
 The time taken for a nucleus to relax back to its equilibrium state depends
on the rate that excess energy is dissipated to the lattice.
 Let M-0-long be the amount of magnetisation parallel with B0 before an RF
pulse is applied. Let M-long be the z component of M at time t, following a
90 degree pulse at time t = 0. It can be shown that the process of
equilibrium restoration is described by the equation

where T1 is the time taken for approximately 63% of the longitudinal


magnetisation to be restored following a 90 degree pulse.
T2 RELAXATION
 While nuclei dissipate their excess energy to the lattice
following an RF pulse, the magnetic moments interact with
each other causing a decrease in transverse magnetisation.
 This effect is similar to that produced by magnet
inhomogeneity, but on a smaller scale.
 The decrease in transverse magnetisation (which does not
involve the emission of energy) is called decay.
 The rate of decay is described by a time constant, T2*, that
is the time it takes for the transverse magnetisation to decay
to 37% of its original magnitude.
 T2* characterises dephasing due to both B0 inhomogeneity
and transverse relaxation. Let M-0-trans be the amount of
transverse magnetisation (Mx-y) immediately following an RF
pulse.
 Let M-trans be the amount of transverse magnetisation at
time t, following a 90 degree pulse at time t = 0. It can be
shownthat