Preventing Accidental Exposures From New External Beam Radiation Therapy Technologies
Preventing Accidental Exposures From New External Beam Radiation Therapy Technologies
Yes, indeed
Beam calibration:
independent
verification of absorbed
dose at the reference
point
115
patients
severely
affected
INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION ——————————————————————————————————————
Lessons from conventional techniques
valid for new technologies?
Need for commissioning of
treatment planning systems
(TPS)
Validation of procedures and
of their modifications,
particularly TPS
For individual patients: 28 patients severely
check of the dose to a point, overdosed in Panama
independently fromTPS 1045 patients
calculations underdosed in the UK
INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION ——————————————————————————————————————
Lessons from conventional techniques
valid for new technologies?
27
patients
severely
affected in
Spain
Test of interlocks at
equipment acceptance
INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION ——————————————————————————————————————
Tolerances and interlocks
• Misadjustment of the tolerance limits of a dose-
rate interlock occurred on a tomotherapy unit
• Important: understanding interlocks and
ensuring their testing at equipment acceptance
Avoid “believes or assumptions” and base the treatment on proper understanding and verification!
• Left-right error
• Distorsion of images when transferring them from the
TPS to the “record and verify”
• Potential problems of image artefacts and wrong tissue
density
• Wiith increased use of different imaging modalities,
consistency in imaging identification and image labelling
becomes more critical
???
40 (mm)