Professional Documents
Culture Documents
Pacemaker Malfunction
Pacemaker Malfunction
• Failure to output
• Failure to capture
• Undersensing
• Pseudomalfunction
– Crosstalk (dual-chamber PM)
– when the monitor system does not display the pacemaker stimulus
artifact when it is really present (frequently with bipolar pacing)
Oversensing
• Unexpected sensing of an intracardiac or extracardiac signal.
• Intermittent or constant
• Electrical signals that may cause oversensing include myopotentials, T waves, and P
waves.
• Pseudomalfunction –
– Inappropriately low voltage-amplitude and pulse-duration settings
– A PM artifact occurring within the myocardial refractory period
Intermittent ventricular failure to capture in a patient with a dual-chamber
pacemaker
DDDR pacemaker. All but one ventricular pacing artifact fail to result in
ventricular depolarization, that is, failure to capture
Undersensing
• Intermittent or total
• Can result in an unwanted rhythm (eg. atrial pacing that competes with NSR may
result in AF).
• Competition in the ventricle is possible but is almost never a problem except when
the fibrillation threshold has been altered by ischemia, electrolyte imbalance, or
some other metabolic abnormality.
• Pseudomalfunction
– Magnet application
– Environmental electrical noise
– When a P wave or QRS complex falls within the RP of PM
– Monitor artifact
• In dual-chamber PMs, apparent undersensing may occur during the initial portion of
the AV interval (blanking period).
During this interval, the ventricular channel of the pacemaker is refractory to avoid
sensing of the atrial stimulus and depolarization. If an intrinsic ventricular event
occurs during the blanking period, it is not sensed.
This problem can be corrected by shortening the blanking period.
ventricular PM programmed to 50 bpm. The second and third pacing artifacts
occur inappropriately close to the preceding QRS complex (that is, at less than
1200 ms the programmed rate of the PM)
• Pseudomalfunction
Sensing abnormalities
– Tracking of atrial fibrillatory or flutter waves
– PM re-entrant tachycardia (seen in dual-chamber PMs; occurs when sensing of a
retrograde atrial depolarization initiates ventricular pacing, which in turn leads
to retrograde conduction and repetition of the cycle).
• Transthoracic Defibrillation
– Can cause reversion to back-up mode, transient increases in capture threshold
and loss of capture as well as destruction of the PM generator and circuitry
– Damage is minimized by positioning paddles anteroposteriorly and as far from
the pacemaker or lead as possible (ideally 15cm).
• Therapeutic Radiation
– Failure of various battery components or accelerated battery depletion
– Changes in sensing capability, failure of telemetry function, runaway function
and complete shut down may all occur
– No specific prediction relative to dose can be made.
– Particularly patients undergoing radiation for thoracic / chest wall malignancy.
– Precautions - Position the field of radiation at an angle oblique to the PM, total
accumulated dosage limit of 2 rad, shielding of the PM with a 1cm margin may
be required.
– If this is not possible, the PM should be explanted and moved to another site .
• RFA
– PM should be checked before and after ablation.
– Rate response function should be turned off.
– RF applications should be as brief as possible and remote from the electrode tip.
– If the patient is not dependent, the pacemaker can be programmed to OOO or
VVI at a lower rate than the intrinsic heart rate.
– If the patient is dependent, the PM should be programmed to VOO mode
– Patients who are PM dependent should use an analog type cellular phone
system. Carrying the phone on the same side of the body as the implanted
PM may cause interference. When using the phone, it should be held at
least 15 cm away from the PM and on the opposite ear.
N Eng J Med 1997; 336: 1473-9
• Potentially significant restrictions exist for a small subset of patients
who work in environments with equipment capable of causing
significant electromagnetic interference— eg. internal combustion
engines, arc welding equipment, degaussing equipment and induction
ovens.
The use of bipolar leads can minimize or eliminate the problem.
• Triboelectric signals are usually wider and more irregular than pacemaker
stimuli
• Defined as the symptoms associated with right ventricular pacing relieved with the
return of A-V and V-V synchrony.
• Additional symptoms include easy fatigability, malaise, headache, and the sensation
of fullness and pulsations in the head and neck