Medilas H Solvo 30 & 35:

Service Training
Medilas H Solvo 30 & 35
Service Training
Medilas H Solvo 30 & H35
• DornierSolvoH35OperatingGuide
Eye Safety
• All personnel should wear wavelength specific laser safety glasses with side shields

• Patients should wear opaque eye protection or wavelength specific laser safety goggles. Use metal ocular
eye shields if the face is treated

• Safety filter should be attached when optical instruments are used

• Remove reflective objects from surgical field

• Never look directly into the laser handpiece or fiber, even with safty glasses in place

• Maintain the laser in standby position except when in use

Company confidential 4
Solid state laser

5 4
1 2
3
Laser Resonator
1 totally reflecting mirror
2 partly reflecting mirror
electronics cooling 3 flash lamp
4 laser active medium
5 laser beam
Laser Light: Technical Realisation
Components of a laser system:
• Laser-active medium – Ho-Yag laser rod
• Optical resonator with 80% outcoupling mirror and 100% refector mirror
• Flash lamp
• HV power supply
• Cooling Circult
• Aiming or Pilot laser
• Fiber Coupling Assembly and Objective lens
Laser Light: Technical Realisation
• Laser-active medium – Ho-Yag laser rod
Laser Light: Technical Realisation
• Optical resonator with 80% outcoupling mirror and 100% refector mirror
Laser Light: Technical Realisation
• Flash lamp
Laser Light: Technical Realisation
• HV Power Supply
Laser Light: Technical Realisation
• Cooling Circult
Laser Light: Technical Realisation
• Cooling Circult
Aiming Laser
• Aiming or pilot laser
Fiber Coupling
• Objectives lens
Fiber Coupling
• Fiber Coupling assembly
Fiber Delivery
• Fiber: Energy transmission
Service Tools
• Power meter:
o Ophir – Vega Gentec – Masstro
Service Tools
• Auto collimator K1009862
Service Tools
• Photo paper for alignment :
Operation
• Operation login
o 1234

• Service login
o 2365

• Dongle
o Configuration Dongle –Advance mode/ RFID 3/RFID 4
o Service Dongle K2013506
Installation & Preventive Maintenance

• Perform check as per installation and Preventive maintenance check list.


• Laser_H35_AR+PM+CM_K2015010-B.pdf

• Training Video
o http://servicedoc.dornier.com/
o Draining coolant.
o Filling the coolant (distill or De-ioned water), replace filter and ion exchange bag
o SolvoH35coolantcontainers.
Activation of Solvo H35
• Activation of Solvo H35
o Remove the top cover
Activation of Solvo H35
o Remove the left side panel
Activation of Solvo H35
o Fill the coolant according to page 33 to 35.
o Insert or replace Ion Exchnage bag
Activation of Solvo H35
o Activate the Advanced mode Dongles (if available)
o Connect the door interlock plug

o Pairing of wireless footswitch ( new replacement, SM Pg 66)


o Perform test and mesaurement according to the check list.