Professional Documents
Culture Documents
Mass Spectrometry
Mass Spectrometry
Mass Spectrometry
Mass spectrometry
1. Sample inlet
Which bring the sample from the laboratory environment.
2. Ion source
Where the sample molecule is transform into gas phase ions.
3. Accleralator
Once the sample has been ionize the beam of ion is accelerated by electric
field and then pass into analyzer.
4. Mass analyzer
The region of the mass spectrometer where the ions are separated
according to m/z.
5. Detector
The ions are counted by the detector.
6. Data system
The signal is recorded and process by the data system by pc.
Sample introduction
A repelar plate
Which carries a positive electric potential directs the newly created ions toward a series of accelerating plates
Accelerating plates
A large potential difference ranging from 1-10 k volts applied across the accelerating plate produces a beam of rapidly
traveling positive ions.
B.Chemical ionization
The sample molecules are combine with reagent gas .
When the sample molecule collide with the pre-ionized reagent gas
some sample molecules are ionized by various mechanisms proton transfer, electron transfer
Volatile liquid or gas can be use as reagent gas.
Common reagent gas are methane, ammonia gas, iso-butane and methanol
The choice of reagent gas should be made carefully to match the proton affinity of the reagent with that of
sample.
Mass Analyses
The region of the mass spectrometer where ions are separated according to their m/z .
There are several types of mass analyzer are
Magnetic sector Mass analyzer
Double focusing Mass analyzer
Quadruple Mass analyzer
Time of flight Mass analyzer
Detection
The detector of a typical mass spectrometer consist of a counter that produces a current that is proportional to
the number of ions that strike it.
Through the use of electron multiplier circuits, this current can be measured accurately.
Current cause by just one ion striking the detector can be measured.
When an ions strike the surface of the electron multiplier 2 electrons are ejected.
The approximately 2 k volts potential difference between the opening and end of the detector.
This draws the electron further into the electron multiplier.
Each electron strike the surface again.
This process continues until the end of electron multiplier is reached.
Electrical current analyzed and detected.
termination of molecular weight
The user observe the m/z that correspond to the molecular ion to determine the molecular weight of the sample molecule.
The masses that are observed for molecular ions are the masses of the molecules in which every atom is present .
One method of confirming that a particular peak corresponds to a molecular ion is to vary the energy of the ionizing electron beam
If the energy of the beam is lowered, the tendency of the molecular ion to fragment lessens.
The peak must correspond to the ion of highest mass in the spectrum, excluding isotopic peaks that occur at higher masses. The iso
peaks are usually of much lower intensity then the molecular ion peak.
The ion must have odd number of electrons.
The ion must be capable of forming the important fragment ions in the spectrum, particularly the fragments of relatively high mass,
of logical neutral fragments.