PROCEDURE CALIBRATION INSTRUMENT MIQ


MAINTENANCE METHOD
FOTO EQUIPMENT TOOLS
TASK
Procedure Calibration Pressure
Gauge MIQ
PEAPLE TIME Wrench Calibrator
Spanner
Notice Clean cloths Pressure
Padlock
55”
Board
01
gauge
Reserve

Step 1 Step 2 Step 3


-Set the Humidity and Temperature -Check calibration schedule daily. -Open the software QMAN calibration
(Taken from the QMAN) -worksheet, print reports.

20' 3”
10´

MANUTENÇÃO PLANEJADA - GRAVATAÍ


PIRELLI - PNEUS S.A. PIRELLI USO INTERNO

Step 4 Step 5 Step 6


-Print calibration worksheet of instruments. -Move up the machine to remove the -Install PCEP device and padlock
instrument for calibration
-Check locking points and take the
line pressure.

3” 5”
10´

Step 7 Step 8 Step 9


-Install the notice board -Remove the pressure gauge with the -Install pressure gauge reserve
“Warning, do not touch”. necessary tools

5” 5”
10´

MANUTENÇÃO PLANEJADA - GRAVATAÍ


PIRELLI - PNEUS S.A. PIRELLI USO INTERNO

Step 10 Step 11 Step 12


-Remove the PCEP device. - Leave the instrument in acclimatization in -Note on the instrument
the laboratory for 30 min. calibration record the time,
temperature and humidity

30' 30” 10'

Step 13 Step 14 Step 15


-Mount the pressure gauge in the - Calibrate the pressure gauge by comparison, -Enter in the software QMAN
Calibrator together with the standard written the calibration record results . the values found in
pressure gauge. calibration.

5” 8”

MANUTENÇÃO PLANEJADA - GRAVATAÍ


PIRELLI - PNEUS S.A. PIRELLI USO INTERNO

Step 16 Step 17 Step 18


- Date of the next generation for calibration -Make the MIQ label with current date and -Cleaning the instrument and
the next calibration date. placing the MIQ label.

1” 3”
50’

Step 19 Step 20 Step 21


-Close the cuts flow and remove the line - Install the device PCEP -Install the notice board
pressure. “Warning, do not touch”.

40’ 10'
1”

MANUTENÇÃO PLANEJADA - GRAVATAÍ


PIRELLI - PNEUS S.A. PIRELLI USO INTERNO

Step 22 Step 23 Step 24


- After the end of the service make
a Check-Up to inspect the
-Install pressure gauge. - Remove the device PCEP condition of the equipment.

10' 2”
1”

Step 25

- Release the machine.

MANUTENÇÃO PLANEJADA - GRAVATAÍ