BC-6000 - Air Dynamic and Liquid System - V3.0 - EN
BC-6000 - Air Dynamic and Liquid System - V3.0 - EN
Three Way
Valve
③ ① ②
① is NC
② is COM
③ is NO)
Bottom
Release vacuum to
get accurate -30KPa
940
920
900
880
90kPa 860
840
820
800
200 250 300 350 400 450 500
315
310
305
N30kPa 300
295
290
200 250 300 350 400 450 500
DR LD LN LH
Air pump 1
WC2
WC1
New design, diluent and LH lyse goes in from the bottom side
to make circle flow for mixing. Needs less bubble for mixing.
Warm up sheath fluid, HGB diluent, RBC diluent and diluent cleaning the
sample preparation tube. Target is 32 ℃
35 © 2018 Mindray Confidential
Components – WNB and DIFF reaction bath
assembly
Not functional
RET
ON
ON
OFF
ON
OFF
Improvement: ON
OFF
No bubble in sample aspiration tube.
No blood stain on test tube capOFF
100,000 times of piercing ON
No movement
Atmosphere
W
HG
DI
RE
NB
FF
B
T
RB
C
Primed by syringe
Troubleshooting:
The sample flow is visible if put a white paper behind the joint
below flow cell.