STP 120 TISSUE PROCESSOR

Janine Marcos Sacramento


Nicole Ivy del Rosario
Ralph Joshua Castro
STP 120 TISSUE PROCESSOR
Tissue processing is a
technique than uses
alcohols to remove
water from tissues and
replace it with a
medium that allows
sectioning of tissue.
Several methods are
used to achieve this.
▪ Immediate and delayed start processing modes
▪ Programmable spinning speed of either 60 or 70
The STP 120 has a
rpm; spinning can also be programmed off
patented and unique
technique that ▪ Programmable immersion time in each station (from
combines several
movements for the 1 minute to 99 hours: 59 minutes)
tissue to achieve perfect ▪ Basket capacity of 120 or 240 cassettes
infiltration results.
▪ Battery backup system in case of power failure
▪ Reagent vessel tops and charcoal-enhanced
Performance ventilation help control processing vapors
▪ Up to 10 different processing programs maintained
▪ Cassette baskets spin counterclockwise and
clockwise within reagent container to improve
processing
▪ Reagent carryover is reduced through gentle
centrifugal spinning of basket above reagent vessel
▪ Each reagent vessel holds 1.8 L of processing
reagent
Paraffin Vessels
▪ Connection marked with the number 10 only if a third paraffin
bath has been installed in position 10
▪ Connection marked with the number 11 only if the paraffin
bath has been installed in position 11
▪ Connection marked with the number 12 only if the paraffin
bath has been installed in position 12
Physical Specifications:

The Epredia STP 120 spin tissue processor


Electrical Requirements: 100-
240 V, 50/60 Hz can easily adapt to your lab’s specimen
Diameter: 85 cm (33.5 in.) processing needs with programmable, gentle
centrifugal force to augment the normal
Height: 50 cm opening to 70 cm
(19.6 in. opening to 27.5 in.) vertical agitation process associated with
Weight: 154 lb. (70 kg) carousel processors.
Features

ROTATIONA ▪ The basket with the cassettes is immersed into the


L reagent vessel. In this position, the basket turns at

AGITATION 60 rpm and changes the rotational direction every


60 seconds.
Features

SHAKING ▪ This movement can be optionally activated on the


control panel
Features

▪ This function is activated as soon as the infiltration


CENTRIFUGING time comes to an end. The basket rises above the
reagent’s level but rests inside the vessel
Features

Ergonomic ▪ The buttons of the control panel are arranged


control ergonomically for easy handling.

panel
Features

▪ The instrument has the capacity to store up to 10


Easy different programs that can be freely set up by the
handling user.
Maximum ▪ The individual cover for each one of the vessels
reduces the emission of vapors to a minimum. The
safety STP 120-2 and STP 120-3 versions incorporate a
standards for fume extraction system that – through a fan and an
the user and active charcoal filter – cleanse the vapors before

the specimen being discharged into the air.


▪ If the specimens remain inside a station for a longer

Alarms time than the programmed time, due to a power


failure, the display will show a message indicating
during the the station number and the overtime spent in that

process station compared with the initially programmed


time.
CONTROL PANEL

The control panel at


the front of the
carrousel contains all
the keys
required to handle and
program the carrousel.
Entering the Program The 'programmed value' is a digit that appears on the
programming display.
It has the following meaning: 0 = without rotation stirring 1 =
rotation stirring at 60 rpm. 2 = rotation stirring at 70 rpm

▪ The vessels contain the reagents indicated on the list.


▪ The basket holder is located at station 1
▪ Press the UPWARD or DOWNWARD key to lower the basket
holder into reagent vessel 1 and prevent the reagents from
evaporating while carrying out the programming
If the date, time hour and language are correct,
proceed as follows.

PROGRAM key will light and the screen will display


the following message

If program number 01 is selected, the screen will


appear as shown.
BASKET
Default Values
• Paraffin bath: 62°C
For changing reagent vessel
• Vertical Shake (only if
this option is selected):
every 10 minutes Centrifugation

• Spin Time: 60 seconds


Immersion
▪ 1. Raise the basket holder.
▪ 2. Place the basket with the specimen holders in the
basket holder.
Starting the Process
▪ 3. Press STOP to exit manual operating mode
▪ 4. Press the SHAKE key if vertical shake is required,
the key indicator lights up.
▪ 5. Press the X2 key if working with a double basket
the pilot beside the key will light.
▪ 6. Press the START key to begin the program.
▪ 7. If a program other than 01 is to be used, press the
UP or DOWN keys to modify the option.
▪ 8. Start the program immediately by pressing START
again. The program will begin to execute.
END OF THE PROCESS Paraffin Temperature Display
The temperature of the paraffin
The program will stop in baths can be viewed at any time
paraffin bath 12, and during the process by pressing
issue a warning beep the DOWN key.
every five seconds. If the
program being run was
P 01, the screen will End of the Process Display
display the following The expected time for
message completing the process can be
consulted at any time by
pressing the UP key.
ADVANTAGES

Customize ▪ Immediate and delayed-start modes allow you to


to Your process tissues when convenient

Needs
ADVANTAGES

Save on
Steps and ▪ The STP 120’s microprocessor unit can maintain up

Cost to ten processing programs


ADVANTAGES

Ensure ▪ Reagent vessel covers and charcoal-enhanced


ventilation help control processing vapors for
Your Safety increased user safety.