Training Teknikal Servo-I Maquet

2016

Oleh : Engineer Servo-I


Servo-i
Bagian Utama Unit :
1. User Interface 
Control Setting &
Monitoring
Parameter

2. Patient Unit 
Pneumatic & Elektronik u/
Supply gas ke patient.
3. Trolley Termasuk juga Power
(Mobile Cart) Supply & Battery backup.
Servo-i
Semi Part/ Semi Consumable/
Accessories Utama :
1. Expiratory Cassette 3. Test Lung Adult 4. Battery Module 6. Nebulizer Aeroneb

2.a. O2 Cell Test lung 5. Test Tube Ultra Neb


Infant

2.b. O2 Sensor 7. Circuit Patient Adult Circuit Patient Infant 8. Servo Guard
(Bacterial Filter)
Servo-i
Set-up Breathing system
1. Servo Guard
2. Support Arm
3. Patient Tube
4. Nipple Connector for Nebulizer
5. Nebulizer patient unit
6. Tube Connection
7. Y-piece
8. Co2 Sensor
9. HME
10. Angled Connector
11. Water Trap
12. Nipple Connector for Chamber
13. Humidifier (F&P)
14. Heater Wire
15. Patient Tube with Probe Housing
Servo-i
Standar Pengoperasian/ Test Unit

1. Pre-Use Check Unit


2. Setting Kategori Patient, Mode Ventilasi &
Parameter
3. Test Unit denganTest Lung
4. Amati Nilai Parameter Monitor
Servo-i
Preventive Maintenance

a. Pengecheckan fungsi secara umum (per 6 bulan atau min 1 kali


setahun)
b. Penggantian Maintenance Kit 5000 H (per 5000 jam operasi)
c. Bersihkan Cooling fan Filter (per 4-6 bulan)
d. Bersihkan membrane expiratory cassette (per 4-6 bulan)
e. Charge Battery secara khusus jika unit tidak pernah/jarang terhubung
ke sumber listrik (1 bulan : 1 x 24 Jam)
f. Kalibrasi dengan Flow Analyzer 1 x per tahun.
Servo-i

Maintenance :
Maintenance
Kit 5000 H

Internal Fan
Membrane Expiratory
Filter
Cassette
Blok Diagram :
Servo-i
Trouble shooting : Pre-Use Check
Servo-i
Trouble shooting : Pre-Use Check
Servo-i
Trouble shooting : Pre-Use Check
Servo-i
Trouble shooting : Pre-Use Check
Servo-i
Trouble shooting : Monitoring
Servo-i
Trouble shooting : Monitoring
Servo-i

Trouble shooting : Monitoring


Servo-i

Disassembling User Interface


Servo-i
Disassembling Patient Unit
(Uncovering)
Terima Kasih