Professional Documents
Culture Documents
Preanalytical Errors
Preanalytical Errors
By Dr Sulakshmi B Kurlekar
INTRODUCTION
• Historical Perspective
• Laboratory and Diagnostic Errors
• Brain to Brain Loop– The Lundberg Concept
• The Total Testing Process
• Introduction
Diagnostics in the clinical chemistry laboratory is a pivotal part of clinical
decision-making but is not exempt from ‘human errors’.
Scientific innovations such as automation and electronic order test requesting
have contributed to substantial improvements in the field of laboratory
science, but errors still occur.
One major example of such failing is connected to the prevalence of errors
occurring in pre-analytical phase of the Total Testing Process(TTP).
Pre- analytical errors can occur at the time of patient assessment, test order
entry, patient identification, sample collection, sample transport, or sample
receipt in the laboratory. Previous work and clinical insights suggest that
most errors in the TTP are extra-laboratory (i.e. they occur before the
samples reach the laboratory for analysis). Such errors are frequently the
results of human mistakes during phlebotomy practice. Therefore to reduce
these errors the pre-analytical phase of the TTP must be prioritized.
Researchers observed that about two-thirds of errors in the
clinical chemistry laboratory occur in the pre-analytical phase,
when compared to the analytical or post- analytical phase. This is
plausible because the pre-analytical stage requires more human
intervention when compared to the analytical and post-analytical
phases of the TTP.
Significant advances in automation; robotics; laboratory
information management systems (LIMS) and precision engineering
have remarkably simplified many laborious and cumbersome
procedures in the clinical chemistry laboratory. Therefore,
analytical errors are no longer the main factors influencing the
reliability of clinical application of laboratory diagnostics.
Historical perspectives
In the 1970s a new term “pre-analytical phase” was introduced to the laboratory
medicine lexicon (Statland and Winkel, 1977). In the 1980s terminologies such as
interfering factors were introduced to professional training programmes,
culminating in the publishing of the first book on pre-analytical variables in 1987
(Einer and Zawta, 1987); and these have become part of the terminologies used in
clinical laboratory sciences vocabulary (Dybkaer, 1997) and international guidelines
(CLSI, 2004).
In 1981, the Clinical Science Laboratory Institute (CSLI) introduced pre- analytical
standards to examine pre-analytical procedures, and the European Committee on
Clinical Laboratory standards (ECCLS) followed this movement. The actions of
these bodies meant that the term ‘pre-analytical phase’ was included in training
and teaching programmes in clinical chemistry and laboratory medicine (Guder
and Wahlefeld, 1983; Hagemann, 2005; ISO, 2007).
Historical examples of pre-analytical errors may help us understand
how ‘seemingly normal’ extra-laboratory routines may impact on
the analytical phase and lead to spurious results. For example in the
early 1960s, clinicians requested urinary amylase measurements in
order to exclude pancreatitis as the cause of acute abdominal pain.
Surprisingly most of the results showed increased amylase activity
even though some of the patients showed no signs of
‘pancreatitis’. It was soon discovered that because the urine
samples were collected in open vessels, they apparently became
contaminated by salivary amylase from drops of spittle from the
nursing staff as they held discussions over the urine samples when
transporting them to the laboratory for analysis (Guder, 2014). The
cause of the pre-analytical error was eventually eliminated by
collecting the samples in closed containers (Guder, 2014). Fluid or
blood amylase samples have now increasingly become the samples
of choice for pancreatic amylase activity investigation.
•Laboratory and diagnostic errors
A laboratory error is defined as any defect, from test ordering to reporting test
results and appropriately interpreting and reacting on these results.
Conventionally medical errors are grouped into 4 types namely: errors of diagnosis,
errors of treatment, errors of prevention and errors of miscellaneous origin. Errors
in the TTP are associated with all of the four error groupings mentioned, albeit most
medical errors are closely linked to ‘errors of diagnosis’.
According to recent data, diagnostic errors rank as the most common source, most
costly and most precarious of medical mistakes for inpatients as well as patients for
attending outpatient clinics or departments. Despite the ubiquitous nature of
diagnostic errors that often lead to avoidable disability or death in some cases,
diagnostic errors still remain a relatively unmeasured subject area of patient safety.
Medical errors can result in physical and emotional suffering for the patient and
frequently lead to a number of mortalities annually, along with excess economic
burden and litigations.
Medical errors that put patients at risk have generated a lot of
media attention in recent years, so much so that publications
such as that of the Institute of Medicine (IOM), estimate that
preventable errors leading to about 1.5 million adverse events
occur in the United States annually and that between 44,000 to
98,000 deaths occur annually as a result of medical errors,
excluding unreported events. About 55% of missed or delayed
diagnoses in the ambulatory setting and about 58% in accident and
emergency departments arise from failure to request appropriate
diagnostic and laboratory tests in the pre-analytical phase of the
TTP.
• Brain-to-brain loop: The Lundberg concept
It was Lundberg (1981, 1999), who first proposed the brain-to brain loop concept
(Figure 1.1) to describe the total testing cycle (Figure 1.2). According to
Lundberg, the activity loop begins outside the walls of the clinical chemistry
laboratory with a clinical question in the physician’s mind. This leads to test
requesting or ordering, collection of the patient’s sample and identification,
transportation of sample to the laboratory, sample separation (by centrifugation
to make it suitable for analysis), sample aliquoting and sorting into batches.
Next, the sample is delivered for analysis on automated platforms. In the final
step results are generated and reported, and action taken (interpretation and
decision making) by the physician. Each stage in this complex loop, when
performed correctly guarantees quality procedures in the clinical chemistry
laboratory, thus ensuring valuable medical decision making and effective
patient management.
It can be appreciated that several transitional steps are involved in
the loop, some of which are pre-analytic (before laboratory testing
of samples); some are analytic (relating to the actual laboratory
testing of samples) and a post-analytic step that involve the
transmission and interpretation of test results into the health
records. The introduction of the brain-to-brain model resulted in an
identification and classification of errors related to laboratory
testing. Previous studies (Plebani and Carraro, 1997; Carraro and
Plebani, 2007; Plebani, 2006, 2007, 2009, 2011) suggested that
most errors in the ‘Lundberg loop’ do not occur within the analytical
stage, nor do they frequently fall within the ‘pre-analytical’ or ‘post-
analytical’ stages under the control of biomedical scientists.
Figure 1.1: Brain-to-Brain Loop Concept for Laboratory Testing.
Adapted from Plebani et. al., (2011). Am J Clin Pathol 2011;
136:829-833.
Sepulveda (2013) summarized the Lundberg loop in seven steps shown by the coloured
numbered boxes:
1. The right question was asked from the patient by the clinician or physician
2. The right test was ordered by the physician
3. The right sample was collected on the right patient, at the correct time, with appropriate
patient preparation.
4. The right technique was used collecting the sample to avoid contamination with
intravenous fluids, tissue damage, prolonged venous stasis, or hemolysis.
5. The sample was properly transported to the laboratory, stored at the right temperature,
processed for analysis, and analysed in a manner that avoids artifactual changes in the
measured analyte levels.
6. The analytical assay measured the concentration of the analyte corresponding to its
“true” level (compared to a “gold standard” measurement) within a clinically acceptable
margin of error (the total acceptable analytical error (TAAE).
7. The report reaching the clinician contained the right result, together with interpretative
information, such as a reference range and other comments, aiding clinicians in the
decision-making process.
It has been established that most of the errors occur before
and after laboratory analysis (Plebani, 2010), and that the pre-
and the post analytical phase are responsible for up to 96% of
total Turn around Time (TAT) anomalies (Manor, 1999;
Rodriquez- Borja et al., 2014 Rodriquez-Borja, 2015).
Therefore incorrect interpretation of laboratory or diagnostic
tests in the final stages of the brain-to-brain loop (B-T-B) also
causes a large proportion of errors in the ambulatory and
emergency settings (Kachalia et al., 2007).
• The total testing process (TTP)
The total testing process (TTP) is the sequence of events starting with the
ordering of a test by a clinician and ending with the interpretation of the test
result by the clinician (Figure 1.2). The TTP concept is similar to the previously
described Lundberg’s loop, the difference being that it occurs in phases. To fully
appreciate the processes involved in generating a result for a patient’s sample in
a clinical chemistry laboratory it is imperative to elucidate the three phases of
the TTP: Any analytical process in the clinical chemistry laboratory involves
these three major stages:
1. The pre-analytical stage, which involves patient specimen acquisition and
preparation.
2. The analytical stage, which is the measurement of the test analyte.
3. The post-analytical step, which consists of reporting results.
Figure 1.2. Total Testing Process (TTP): Adapted from Boone (2007): Presentation at the Institute on
Critical Issues in Health Laboratory Practice: Managing For Better Health, September 23–26, 2007
Atlanta, GA, USA Centers for Disease Control and Prevention.
Laboratory testing in clinical chemistry is a significant source of
medical errors affecting patient management and safety. Errors
generated in the pre-analytical phase decisively influence the total
error and consequently the diagnostic accuracy. The pre-analytical
phase is probably the most important step in the TTP in the clinical
chemistry laboratory and indeed in laboratory medicine, if total
laboratory quality is to be achieved. This assertion is true because
the pre-analytical phase is conceivably the most complex and highly
vulnerable to uncertainties (such as biological variations) and
untoward incidents. Recent studies carried out by a number of
investigators have shown that approximately 93% of errors
encountered within the TTP is due to the lack of standardised
operating protocols for sample collection, which includes patient
preparation, phlebotomy, sample handling/transportation and
storage.
Errors relating to pre-analytical stages of the TTP are arduous to
control and therefore require the adoption of drastic but suitable
approaches for the prevention (or reduction) of these errors. A
number of suitable strategies for error prevention have been
examined and streamlined into the TTP. The major challenge for
biomedical scientists and other laboratory staff has been the absence of
an effective communication system with which to engage with the end-
users of the clinical chemistry laboratory service (i.e. physicians, nurses,
consultants, phlebotomists and other healthcare professionals). A
good communication system is essential for prevention or reduction
of pre-analytical errors.
Currently there has been increased focus on the reduction of pre-
analytical errors (by healthcare professionals working in the clinical
chemistry laboratory) and the possible impact on patient
management by development of quality improvement ‘tools’ such as
1 .THE PRE-ANALYTICAL PHASE
The medical needs and clinical questions usually determine the source of sample
and time of sampling.
Besides medical knowledge about the physiology of analytes intended to be
measured, criteria such as practicability of sampling and risk of patient harm may
influence the decision
Most commonly used sample type used is Venous blood, followed by capillary
blood, which , because of the lower amount of sample needed, became the
preferred sample in newborns and young children. Arterial blood on the other
hand, is mainly used for analysing blood gases and acid base state. Other sample
types are urine, saliva and other body fluids.
CONTROLLABLE VARIABLES
Time of sampling
Influence of Circadian rhythm
Menstrual cycle
Influence of Diagnostic and Therapeutic Procedures
Diet
Effects of Fluid Intake Before Sampling
Smoking Tobacco
Body position and Tourniquet
Muscular activity
Preparing for Blood sampling
INFLUENCE OF CIRCADIAN RHYTHM Several analytes tend to fluctuate in terms of their
plasma concentration over the course of a day
TIMING OF COLLECTION
For cortisol, between 7 and 9 a.m.
Before interfering diagnostic and therapeutic procedures
In drug monitoring: consideration of the peak after administration and the steady state phase
before the next dose.
Documentation of the exact time of sampling is very important.
DIURNAL VARIATION
OF SELECTED ANALYTES
MENSTRUAL CYCLE
Prior to blood sampling, the confounding influences of foods and fluid intake should be
excluded.
The activity of some enzymes ( ALP, AST, ALT) increase upto 20% following a meal.
Triglyceride and glucose concentrations increase during the absorptive phase. It is the
turbidity of the plasma/serum samples, caused by chylomicrons following absorption of
lipids which interfere with various measurement procedures and causes apparent changes
in analyte concentrations.
To avoid any misinterpretation , it is therefore recommended that the blood sampling be
done after 12 hours of fasting and reduced activity ( bed rest).
In response to a meal, hydrochloric acid secretion in the parietal cells of the stomach is
coupled with extraction of chloride from and release of bicarbonate into plasma in order to
maintain electrical neutrality. Thus the venous blood leaving the stomach enriched with
bicarbonates, and this phenomenon is responsible for a mild postprandial metabolic alkalosis
( i.e the alkaline tide ) with concomitant increase of pCO2 and a subsequent reduction of
ionized calcium by 0.2 mg/dL.
Effect of Fluid Intake Before Sampling
Caffeine containing drinks such as tea, coffee and cola drinks stimulate the
adrenal cortex and medulla, leading to the subsequent
increase of the concentration of catecholamines and their metabolites
free cortisol
11-hydroxycorticoids, and
5-HIAA in serum. They inturn increase the plasma glucose concentration
Plasma renin activity may also be increased following caffeine ingestion.
As caffeine is a diuretic inhibits the reabsorption of electrolytes, thus causing
transient increase in excretion ( calcium, magnesium, sodium, chloride,
potassium)
Ingestion of coffee increases the rate of lipid catabolism, thus leading to the
increase of plasma lipids, free fatty acids, glycerol, and lipoproteins.
ALCOHOL
Acute effects occur within 2-4 hours of ethanol consumption, causing
• decrease in plasma glucose and increase of lactate
• increases hepatic formation of uric acid, inhibits renal urea excretion, thus causing
an increase of uric acid in plasma.
• acetate decreases plasma bicarbonate, resulting in mild to severe metabolic
acidosis, depending on the amount of ingested alcohol.
• increases activity of serum GGT and some other enzymes ( eg, isocitrate
dehydrogenase, ornithine carbamoyltransferase)
They have the ability to interfere with the analytical procedure and alter the test
results. Their effect depends on the method-that is, the same interferent may not
necessarily affect two different methods used to measure the same analyte.
The most common interference factors being
Hemolysis
Lipemia
Icterus
Drugs
Paraprotiens and
Various contaminants such as gels, tube additives and fibrin clots
Interfering factors are considered clinically relevant when the
bias caused by their interference is greater than the maximum
allowable deviation of a measurement procedure
They can be endogenous and exogenous. Endogenous
interferences originate from the substances present in the
patient sample, whereas exogenous interferences relate to the
effects of various substances added to the patient sample, such as
separator gels, anticoagulants, surfactants, and so on.
HEMOLYSIS
Release of the Cell Components Into the Sample: Some components are
present in blood cells in concentrations that are several times higher than
those in the extracellular space.
RATIO BETWEEN INTRACELLULAR AND EXTRACELLULAR CONCENTRATION OF
VARIOUS PARAMETERS IN RED CELLS
ANALYTE INTRACELLULAR CONCENTRATION
( Compared to extracellular )
Lactate dehydrogenase 160
Inorganic phosphate 100
Potassium 40
Aspartate aminotransferase 40
Folic acid 30
Alanine aminotransferase 7
Magnesium 3
SAMPLE DILUTION
Various blood cell components may affect the analyte measurement procedure by
directly or indirectly competing for molecules in the reagents, inhibiting indicator
reactions or modifying the analyte by complexation, proteolysis, or precipitation.
One such effect is caused by the adenylate kinase, which is present in both
erythrocyte and platelets. Adenylate kinase is an enzyme that catalyzes the
reversible conversion of ATP and AMP to two ADP molecules and maintains the
adenine nucleotide cell content. When released from the cells during hemolysis,
adenylate kinase may compete for ADP with creatine kinase in a creatine kinase
assay if inhibitors are not supplied in the reaction mixture.
Hemoglobin released from erythrocytes during hemolysis may interfere with
various assays through its pseudo-peroxidase activity. Pseudo-peroxidase activity
of free haemoglobin released from erythrocytes interferes in the assay for
measurement of bilirubin concentration through the inhibition of the formation of
diazonium salt.
Hemolysis may cause a clinically significant interferences on a wide
range of analytes in immunochemistry assays. This interference is
caused by modifying the reaction analytes ( antigens and
antibodies) by the proteolytic action of cathepsin E, the major
proteolytic enzyme in mature erythrocytes. Proteolytic enzymes
released from erythrocytes may mask or potentially enhance epitope
recognition in various immunoassays. Interference caused by
proteolytic activity may cause measurement bias of various
degrees and various directions, depending on the assays.
Plasma in healthy individuals in the fasting state consists of only minor portion of
lipids (<10% of total plasma volume). The rest of the plasma is water. The increase
in the concentration of lipoprotein particles leads to an increase in the plasma
volume occupied by lipids. Particles that are not lipid soluble are displaced by the
lipids to the to the water part of the plasma. Therefore lipemia leads to a false
decrease in the concentration of the measured analyte in all methods in which
concentration of respective analyte is measured in the plasma volume.
One example of interference caused by the volume depletion effect is the bias in
electrolyte measurement, leading to so-called pseudo-hyponatremia. This type
of interference affects electrolytes only if measured by flame photometery and
by indirect measurement using ion-selective electrodes but not in direct
potentiometry. However , it must be noted that the volume displacement effect
of the lipemic sample will affect the electrolyte measurement only in grossly
lipemic samples with concentrations of triglycerides greater than 17 mmol/L
( 1504 mg/dL )
INTERFERENCE CAUSED BY PARTITIONING OF THE SAMPLE
Hemolysis becomes visible at the concentration of 0.3 to o.5 g/L of free haemoglobin, and the
intensity of the red colour of the serum or plasma further increases with the increase in
concentration of free serum haemoglobin.
Lipemia causes sample turbidity, which approximately corresponds to the concentration of
serum triglycerides.
Increased concentrations of serum bilirubin lead to yellow to orange colouration of the
serum, and the change of the colour correlates to the increasing concentration of bilirubin in
the serum.
Free hemoglobin, triglycerides, and bilirubin have characteristic absorption peaks in a wide
wavelength range of 300 to 600 nm. This is also the range where sample absorbance is
measured in spectrophotometric methods, and that is why hemolysis, icterus, an lipemia
cause spectral interferences
AUTOMATED SERUM INDICES
Most mainstream chemistry analyzers can detect serum indices by the use of
semiquantitative, spectrophotometric measurement and grading the
interfering substances into categories.
MANAGEMENT OF HEMOLYTIC , ICTERIC, AND LIPEMIC
SAMPLES
There is large discrepancy among the ways results are reported from samples with
interferences.
When interferences from hemolysis, icterus, and lipemia are causing unacceptable
bias and results are clinically inaccurate, such results should not be reported and
sample redraw should be requested. Such a test report should always be
accompanied with comments informing the clinical staff about the reasons for not
reporting the originally requested test results. It is also important that a laboratory
notify the staff when sample appearance ( colour , turbidity) deviates from a
normal state by including a comment on a test report ( eg sample hemolysed,
icteric, lipemic or turbid. Comments should also indicate if the sample has been
treated in any way to minimize the effect of interfering substances ( eg
delipidation)
DRUG INFLUENCES ( DRUG EFFECTS) AND INTERFERENCES
Mechanisms of Interferences
According to the mechanism of action, drug interferences on laboratory tests can be
categorized as either BIOLOGICAL or CHEMICAL
INDUCTION OF HEPATIC ENZYMES – Phenytoin raises levels of GGT through this mechanism
ENZYME INHIBITION – Finasteride and dutasteride cause a decrease in PSA concentration by
inhibition of 5α-reductase
All these drug actions occur in vivo, and thus the changes in parameters reflect a true state
in human body, thus ruling out analytical error.
Analytical ( chemical ) interference is caused when the parent drug or metabolite has
structural similarity to the analyte therefore interfering in the immunometric or photometric
methods.
Reaction interference can occur when the compound or its metabolite catalyse or inhibit the
steps of chemical reaction.In above cases Manufacturer Claims Regarding Drug Interferences,
can be of great use. And also the patients drug history.
Creatinine is one of the most commonly measured parameters in a clinical chemistry
laboratory, the laboratories worldwide use either the Jaffe colorimetric assay with alkaline
picrate or an enzymatic assay for creatinine measurement.
Some commonly prescribed antibiotics and analgesic drugs can interfere with creatinine
measurement. Cephalosporin antibiotics cause falsely increased creatinine in sera shortly
drawn after intravenous administration of cefpirome.
Also subtherapeutic, therapeutic and toxic concentrations of acetaminophen, acetylsalicylic
acid and metamizole cause falsely raised creatinine when measured by jaffes method
High concentrations of catecholamines also show strong negative interferences with
creatinine enzymatic assay.
Ethamsylate, a hemostatic drug causes a significant decrease in creatinine when
measured by enzymatic technique. Also evidences suggest ethamsylate causes
significant false decline of cholesterol, triglycerides and uric acid.
The active metabolite of leflunomide, a immunosuppressant with antiviral
property shows interference with ionized calcium by falsely decreasing the
results. Commonly seen in kidney transplant patients. Similar interference is seen
with sodium perchlorate used as a oral drug in treatment of hyperthyroidism.
Carbohydrate-deficient transferrin (CDT) is a biomarker for long term alcohol
consumption. The sensitivity of the test can be affected by number of drugs like
amlodipine, perindopril, atorvastatin, isosorbide mononitrate.
Thiopental is a barbiturate used for treatment of elevated intracranial pressure,
falsely elevated sodium concentrations were seen when V-LYTE Integrated
multisensory technology system was used for electrolyte measurement
Supportive Medical Therapy
Iodine-based compounds ( iohexol, iodixanol, and ioversol) are mostly used in
x-ray methods can interfere with chemical reaction in several ways. Iodine
molecules have high density and can prevent proper formation of the barrier in
the serum gel separator tubes. Iopromide is used as contrast media agent in
coronary angiography. A false increase in troponin I concentration measured
by Opus Magnum reagent is detected if the sample is taken immediately after
the procedure. This interference is reagent specific.
Gadolinium contrast agents act as chelators, and that seems to be the main
interference mechanisms. Therefore calcium measurement is mostly affected
by the presence of these compounds. Other clinical chemistry assays are also
influenced by several different gadolinium contrast agents: angiotensin-
converting enzyme, total iron-binding capacity, zinc, magnesium and
creatinine.
Natural Preparations
Chinese medicine like Chan Su, can contain some physiologically highly active
molecules like bufadienolides. The structural similarity of bufadienolides and
digoxin is responsible for both cardiotoxicity and interference in the
immunochemistry method. Ingestion of this medicine can cause false-positive
and false-negative results for the digoxin measurement depending on assay
format. Herbal supplements that are used widely worldwide, like ginseng, can
also interfere with digoxin measurement. Chinese herbal remedies may
contain steroids, with the potential to interfere with some assays and to
cause suppression of the hypothalamic-pituitary-adrenal axis.
Similarly, it is also well documented that some Ayurvedic herbal medicine
products may be contaminated with lead, with the potential to cause toxicity.
Accidental Exposure and Poisoning
In case of accidental poisonings with herbs, household cleaning products, or any
other exogenous compounds, interferences on laboratory test results can
potentially prolong the diagnostic procedures in acute patient care and cause
harm to the patient.
Cardiotonic glycosides like Convallaria majalis and Nerium oleander, due to their
structural similarities to digoxin, can interfere with digoxin measurement.
Accidental intoxications often occur as a result of children ingesting cleaning
products. Miniature racing cars run on nitromethane, which has been shown to
interfere with creatinine measurement, producing falsely elevated
concentrations.
Extreme creatinine concentration without evident renal failure has been
observed in suicide attempt in which Blue Thunder fuel containing nitromethane
was ingested. In a jaffe method, nitromethane also reacts with alkaline picrate
and forms a red chromophore, thus causing falsely elevated creatinine.
Sample Contaminant
Several components of the blood collection tubes can interfere with the measurement and
influence laboratory results.
Lubricants like silicone oils and glycerol facilitate insertion and removal of the stoppers. Glycerol
should not be used for the lubrication of stoppers in tubes that will be used for triglyceride
measurement because glycerol interferes with most triglyceride assay.
Silicone-based lubricants are less likely to interfere with assays, although silicone can falsely
elevate ionized magnesium and triiodothyronine levels.
Some clot activator based silica particles affect the measurement of some analytes like lithium
and testosterone.
Silicone surfactants used to decrease nonspecific adsorption of components on tube walls may
interfere with measurement of vitamin B12 and cancer antigen 15-3
Antibiotics are commonly added into reagents and buffer solutions to prevent microbial
growth.
Carryover from reagents containing gentamicin may cause spuriously high gentamicin results.
Gentamicin is present in some diagnostic reagents ( Glucose, urate, direct bilirubin, CK ALT, AST,
beta-hydroxybutyrate) as an antibacterial additive and is known to cause spuriously high
gentamicin results on the Beckman Coulter AU-480 analyser due to reagent carryover.
INTERFERENCES IN IMMUNOASSAY
Potential exogenous sources of preanalytical interferences include
• Sample type (type of additive)
• Underfilling the tube
• Sample stability during transport and storage
• Hemolysis
• Lipemia or sample turbidity
• Icterus (hyperbilirubinemia)
• Fibrin clots
• Carryover
• Administration of radioactive or flourescent compounds
• Drugs
• Herbal medicine
• Nutritional supplements
TUBE ADDITIVES AS POTENTIAL INTERFERING FACTORS IN
IMMUNOASSAY
Under optimal clotting conditions, serum is considered to be free of fibrin, fibrinogen, and cells, and it is
a preferred matrix for most immunoassays.
To allow complete clot formation , serum tubes should be allowed to clot for a minimum of 30 minutes.
However, with new tube types, containing clot activators ( thrombin-based clotting agent), serum
clotting time is substantially reduced ( on average < 2.5 minutes) without compromising the sample
quality and stability for most chemistry analytes.
Blood from patients who are receiving heparin therapy may require a longer time to completely clot, and
there is a greater likelihood for latent postcentrifugation clot formation. Insoluble fibrin , fibrin strands,
and microclots may affect instrument performance and cause interferences. If fibrin is aspirated for
analysis and goes undetected, there is high likelihood of getting false-positive results for that sample.
This is manifested by duplicate measurement errors ( i.e unacceptable deviations in two measurements
on the same sample)
Such discrepancies has been repeatedly reported in cardiac troponin measurements, a possiblre
approach is to recentrifuge all positive troponin samples at high speed ( 6700 g for 5 min) and then
repeat the analysis
Interference Caused by Separator Gels
Separator gels are used to ensure rapid and prolonged separation of serum/plasma
from clotted blood and cells, respectively. Separation of the sample is enabled due to
the specific gravity of the gel, its ability to undergo a temporary change in viscosity
during centrifugation, and its ability to lodge between the packed cells and the top
serum/plasma layer.
Hydrophobic compounds may bind to the gel, which is why tubes containing
separator gels are not appropriate for some hydrophobic drugs and hormones such as
–
DRUGS : phenytoin, phenobarbital, carbamazepine, tricyclic antidepressants,
quinidine, lidocaine
HORMONES : testosterone, estradiol, cortisol, free thyroxine, total triiodothyronine
At times when kept under improper storage conditions (time and temperature) the
gel may degrade and release small particles or globules into the supernatant.
Interferences Caused by Sample Carryover
Sample carryover in automated analyzers occurs due to the inefficient probe and cuvette
washing and subsequent inability of an instrument to successfully remove any remnants of
sample or reagent. Due to improper washing, certain amount of reagent or analyte can be
transferred ( carried ) by the measuring system from one assay reaction to a subsequent
reaction thereby erroneously affecting test results. Additional probe and cuvette washing
may be performed after the extremely elevated results with an increased risk for carryover.
According to International Union of Pure and Applied Chemistry ( IUPAC) , sample carryover
testing is performed by running one sample with high concentration of an analyte at least
two times, followed by at least three runs of a sample with low concentration of the
analyte
According to CLSI, carryover testing for immunoassays is done by running four consecutive
analyses of two samples at different levels ( samples with extremely high analyte
concentration (A) followed by samples with very low (B) concentrations for the same
analyte)
Interferences Caused by Paraprotiens
Paraprotien interferences are not uncommon. The frequency of paraprotien interferences has
been estimated to be as high as 3% to 4% in hospitals .
GROUP OF ANALYTES MOLECULE
Enzymes Alkaline phosphatase
GGT , LDH
Electrolytes, minerals, and Calcium
microelements Inorganic phosphorous , Iron
Metabolites Bilirubin , Cholesterol, Creatinine,
Glucose, Urea , Uric acid
Protiens C-reactive protiens , IgA , IgG
Hormones TSH , HCG
Drugs Gentamicin , Vancomycin , Valproic acid
Phenytoin
Cardiac markers Troponin I
Tumor markers Α-fetoprotein
CA-125
Problems with sample preparation, storage and
transportation
The length of time of sample preparation, including the speed and temperature of centrifugation, exposure
to light and preparing aliquots, are important factors that must be considered before analysis. Proper
sample storage condition is paramount (this relates to the length of storage, temperature, freezing and
thawing). However during transport, a sample may be exposed to shaking, changes in light conditions, and
changes in temperature. Transport delays to the laboratory can give rise to clinically important errors if
transport conditions are not optimized.
The National Committee for Clinical Laboratory Standards (NCCLS) H5-A3 1994 recommended a maximum of
2 hours for transporting blood samples at a temperature of 10–22 °C. Delay in transporting the specimen to
the laboratory where processing begins, not only prolongs the time until the attending physician receives a
test result but also impacts specimen integrity that may lead to a false-negative or false-positive result of
analysis and a misleading answer. A delay in transporting a blood sample to the laboratory for measurement
of ‘blood sugar’ is likely to cause a ‘falsely reduced’ blood glucose result. Such errors that occur at this stage
cannot be corrected later.
These mistakes would require collection of fresh blood samples, which
could be quite distressing for the patient. The sample preparation steps
contribute to approximately 19% of the overall cost of analysing a single
sample and are time-consuming (37% of time spent in producing result).
Clinical laboratory informatics
Conventionally healthcare providers have always ordered laboratory tests using manual requisitions
(paper-based orders). Healthcare providers would place a tick or write the name of the requested
test on a pre-printed hospital laboratory order form. The sample would then be collected, labeled by
a phlebotomist and placed in a specimen bag along with the requisition, and delivered to the clinical
chemistry laboratory for analysis. However due to increasing pre-analytical errors, complexities,
challenges, and resource constraints of modern healthcare systems, paper orders are becoming
increasingly unsuitable. Manual order requests are gradually being replaced by electronic order
entry systems, which allow direct healthcare provider input of diagnostic testing orders into
requests into a computer system are known as Computerized Provider Order Entry (CPOE) systems.
Implementation of laboratory CPOE systems may offer healthcare centres many benefits, including
reduced test turnaround time, improved test utilization. Electronic requesting should improve the
accuracy of patient demographic information (Yorkshire Centre for Health Informatics, 2014); allow
better tracking of results and allow laboratories to report the results of tests they send away for
analysis via their normal reporting systems.
Approaches for reducing pre-analytical errors in clinical
chemistry
Performance and outcome measures can significantly improve the quality of patient
care (Plebani, 2011). Such measurements of improvement can be achieved through the
development and monitoring of specific quality indicators (QI) in the TTP. The pre-
analytical phases in clinical chemistry are more prone to errors than the analytical and
post-analytical phase. Evidence abounds that before the turn of the century, reliable
QIs and quality specifications have been developed and introduced for effective
management of analytical processes. The development of these specific QIs were an
extension to the Internal Quality Control (IQC) systems coupled with availability of
different External Quality Assessment (EQA)1 schemes already in place, which have
made it possible for the clinical chemistry laboratory to measure, monitor and improve
their analytical performance on a monthly basis, although retrospectively
QIs have been defined as vital tools (IOM, 2000) for enabling clinical chemistry
laboratory staff to quantify the quality of a selected aspect of patient care by comparing
it against a defined criteria (ISO 15189: 2012). A QI objectively measures and evaluates all
of 6 critical domains as outlined by the IOM, namely:
a) Patient safety, b) Patient-centeredness, c) Effectiveness, d) Efficiency, e) Equity and f)
Timeliness. Assessment of QI is therefore based on the evidence associated with the
aforementioned domains and can be implemented in a constant and comparable
manner across settings and over a period (IOM, 2000); the identification of a reliable QI
represents an essential stage in the schemes targeting evaluation and improvement of
the quality of patient care. Although many authors are in agreement that QIs are
important in providing information on continued improvement it has two major
shortcomings: firstly, it is often difficult to compare data reported in the literature
because of differences in QIs us and the methods of collecting these data. Secondly,
while the most common errors in the pre- and post- analytical phases are considered,
available QIs do not include appropriate choice and selection of test as well as
appropriate interpretation and utilization of laboratory results at the right time.
Improved communication between clinical chemistry
laboratory staff and other healthcare professionals.
Many mistakes generated in the TTP may be due to poor communication, action taken by others involved
in the testing process (e.g., physicians, nurses and phlebotomists), or poorly designed processes, all of
which are beyond the laboratory's control (Plebani, 2006). Inter-professional working to improve service
delivery involves several healthcare professionals with expertise; knowledge and skill bases and
experience being drawn together in a structure to provide services. These different professionals make
adjustments and allowances in their responsibilities to take account and relate with the roles of others.
Many workers value inter-professional team working as a source of mutual support in the face of internal
and external pressures (Payne, 2000) and particularly the demands of working in the NHS and spending
cuts. Failure to initiate and maintain communication between the laboratory and the users of its service
may be responsible for causing certain medical errors. Clinical communication is highly complex and
prone to errors especially during transitions of patient care and emergency situations. Standardised
approaches and tools may provide potential solutions to improve the quality of communication and
prevent subsequent patient harm.
Quality indicators in the pre-analytic phase
IOM (2000) defined quality as: “The degree to which health services for individuals and
populations increase the likelihood of the desired health outcomes and are consistent
with current professional knowledge”
As stated above, the pre-analytical phase should be subdivided into an extra-
laboratory pre-analytical phase and a ‘true’ pre-analytical phase, which is undertaken
within the laboratory after sample reception. The former phase, which comprises
initial procedures usually performed neither in the clinical laboratory nor undertaken,
at least in part, under the control of laboratory personnel, includes test requesting,
patient and sample identification and sample collection. The latter phase involves the
steps required to prepare samples for analysis (order entry into computer interface,
centrifugation and aliquotting into secondary tubes etc.). In a patient-centred
scenario, QIs should be designed to cover all steps of the pre-analytical phase,
including the appropriateness of test selection, which is a key issue in projects aiming
to ensure clinical effectiveness. The IFCC Working Group in a recent study have
identified 16 Quality Indicators to monitor the pre-analytical phase of TTP
Q I-1 : A p p ro p ria te n e ss o f te st re q u e st N u m b e r o f re q u e sts w ith c lin ic a l q u e stio n (% )
Q I-8 : S a m p le s N u m b e r o f sa m p le s lo s t/n o t re c e iv e d (% )
Q I-1 0 : S a m p le s N u m b e r o f sa m p le s h a e m o ly s e d (h a e m a to lo g y , c h e m istr y ) (% )
Q I-1 4 : S a m p le s N u m b e r o f sa m p le s d a m a g e d in tra n s p o rt (% )
Q I-1 6 : S a m p le s N u m b e r o f im p ro p e rly s to re d sa m p le s ( % )
The lack of patient identification or patient misidentification have serious
consequences for reaching the final conclusion and clinical decision, as well as for
patient safety, thus this is one of the key indicators in the process. Errors in
patient identification may also occur during the procedures of sample
preparation. Ordering inappropriate tests is another pre-analytical variable with a
negative impact on patient safety. It is the cause of unnecessary test repetitions
in up to 30% of cases (Plebani, 2006). Biomedical scientists in conjunction with
clinical scientists should acquaint the clinician with the importance of biological
variations, potentially significant in the pre- analytical process (test selection) but
also in the post-analytical process (interpretation of test results). It is necessary
for the clinician to understand the concept of biological variations that include
intra-individual biological variations (deviation of results in relation to the
person’s homeostasis) and inter-individual biological variations (variations
between different persons in relation to the established homeostatic value).
Based on such knowledge it is possible to select between two tests, the best one
with the highest diagnostic value.
In the pre-analytical phase, major sources of variability - which includes biological
variability, environmental conditions, postural changes, patient identification and
sample labelling tourniquet time, type of container (e.g., vacutainer tube),
phlebotomy procedure (order of sample draw, contamination, mixing); sample
transportation (time, pneumatic tube transport systems from wards and clinics);
sample preparation for analysis (length, temperature of centrifugation,
preparing aliquots, can occur during patient preparation for phlebotomy. Errors
that may occur in this process often become obvious in the analytical and post-
analytical phase as well. For instance, the effects of interferences may be
discovered during analysis or the clinical interpretation of results (for example a
falsely elevated potassium result is observed in ‘hemolysed’ samples; similarly a
sample with a high degree of lipemia or protein concentration will adversely
affect the ‘true’ sodium result). For these reasons identification of quality
indicators is necessary in order to avoid potential errors in the pre-analytical
phase.