BC5000¡ BC5150 Service Training Material1.0

You might also like

Download as pptx, pdf, or txt
Download as pptx, pdf, or txt
You are on page 1of 180

BC-5000/BC5150 service training

© 2013 Mindray Confidential


Content

 Clinical
 Overview
 Principal
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

2 © 2013 Mindray Confidential


clinical—Blood Cells

 Blood consists of 40-50% blood cell and 50-60% plasma.


 Three main parts of blood cell:
red blood cell, white blood cell and platelet.

3 © 2013 Mindray Confidential


clinical—Blood Cells

Five groups of leukocyte

WBC 5-part proportion

 Lymphocyte 25 ~ 30%

 Monocyte 3 ~8%

 Neutrophil 50 ~70%

 Basophil 0 ~1%

 Eosinophil 0.5 ~ 3%

4 © 2013 Mindray Confidential


clinical—Blood Cells

 Morphology : size, shape, cytomembrane, cytoplasma, nucleus,


granule.

RBC WBC PLT


GRAN MONO LYMPH
NEU ESO BASO

5 © 2013 Mindray Confidential


Clinical—Blood Cells

6 © 2013 Mindray Confidential


Morphological

7 © 2013 Mindray Confidential


Clinical—WBC five part comparison

NEU EOS BASO LYMPH MONO


Diameter 10~14um 11~16um 10~12um 6~9um 12~20um

Cytoplasm Large more more less More than


amount LYM
karyon Multlobed, 2lobes mostly 3~4lobes, or Large round Irregular and
2~5 hard to indented
divide
Granule Neutral Eosinophilic basophilic None normally azurophilic
or
acidophili
c
Chromatin More lump More lump More lump Tightness loosen
evenly
Color of Reddish Reddish Reddish Baby blue Grey blue or
cytoplas little
m reddish

8 © 2013 Mindray Confidential


Clinical —granules in cytoplasm

Neutral Eosinophilic basophilic Azurophilic

Size Fine and even Thick and even Most thick and Thicker than
uneven NEU, uneven

Shape Fine particles Round or ellipse Differ in shape Differ in shape

Amount Much Much Less A few or


moderate

Color Pale violet red Orange red or dark Royal purple or Purple red
or reddish yellow dark purple

Distribution Even Even Uneven, coat the Uneven, coat the


nucleus normally nucleus
occasionally

9 © 2013 Mindray Confidential


Content

 Clinics
 Overview
 Principle
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

10 © 2013 Mindray Confidential


Overview—appearance

 BC-5150 BC-5000

11 © 2013 Mindray Confidential


Overview —Dimension

BC-5000
vs BC-2800

324mm 322mm

437mm
415mm

410mm 386mm

12 © 2013 Mindray Confidential


Overview —Report parameter
White Blood Cell count WBC Red Blood Cell count RBC
Basophils number Bas# Hemoglobin Concentration HGB
Basophils percentage Bas% Mean Corpuscular Volume MCV
Neutrophils number Neu# Mean Corpuscular Hemoglobin MCH
Neutrophils percentage Neu%

WBC Eosinophils number Eos# Mean Corpuscular Hemoglobin Concentration MCHC


11 RBC
Para Eosinophils percentage Eos% 8
Para Red Blood Cell Distribution Width Coefficient RDW-CV
Lymphocytes number Lym# of Variation
Lymphocytes percentage Lym% Red Blood Cell Distribution Width Standard RDW-SD
Deviation
Monocytes number Mon#

Monocytes percentage Mon% Hematocrit HCT

Platelet count PLT

Mean Platelet Volume MPV

PLT
Platelet Distribution Width PDW BC5150 only
6 Plateletcrit PCT
Para
Platelet-large cell ratio P-LCR

Platelet-large cell count P-LCC

13 © 2013 Mindray Confidential


Overview —RUO

BC5150 only

Name Abbr.

Abnormal lymphocytes number ALY#

Abnormal lymphocytes Percentage ALY%

Large immature cells number LIC#

Large immature cells percentage LIC%

14 © 2013 Mindray Confidential


Overview —appearance

1 、 Display screen
2 、 Power/status indicator
3 、 Probe wipe block
4 、 Sample probe
5 、 [ Aspirate] key

 Front of the analyzer

15 © 2013 Mindray Confidential


Overview —appearance

1 、 M-52D diluent inlet


2 、 Waste outlet
3 、 Waste sensor
4 、 AC input socket
5 、 Power switch
6 、 Air outlet for fan
7 、 Back panel

 Back of the analyzer


16 © 2013 Mindray Confidential
Overview —Main screen

17 © 2013 Mindray Confidential


Overview —difference between BC5000andBC5150

item BC-5000 BC-5150


Test speed 40Samples/h 60Samples/h

Report Para 23 25
RUO NONE 4Para
Sample storage 20000 份 40000 份
WBC linear 0.00×109/L ~ 100.00×109/L 0.00×109/L ~ 500.00×109/L
PLT linear 0×109/L ~ 1000×109/L 0×109/L ~ 5000×109/L
Front panel 043-002705-00 043-002706-00
Screen assembly 115-018417-00 115-018418-00
Key board 043-002703-00 801-3100-00194-00

18 © 2013 Mindray Confidential


Overview —Specification

 Measuring mode : CBC 、 CBC+DIFF


 Sample mode : WB Capillary W PD
 Sampling :
 WB Capillary W
 Sample volume of aspiration : CBC+DIFF—15ul
CBC—11.5ul
 OV-PD
 Sample volume of aspiration : 200ul
 Diluent dispensed : 480ul
 Blood collected : 20ul
 Scattergram : DIFF(Report:L-M) ,( RUO : L-W 、 M-W)
 Histogram : WBC 、 RBC 、 PLT

19 © 2013 Mindray Confidential


Overview —Specification

 Operating temperature : 10℃—30℃


 Screen : 10.4’LED , with touch screen
 Weight :≤ 25Kg
 Interface : 4USB, 1 ethemet network port
 Data transmission : HL7 supported
 Printer : external , PCL6
 printer : EPSON LQ-590K 、 HP Laser Jet P1505n 、 HP OfficeJet Pro
K5300 、 HP LaserJet P1606dn

20 © 2013 Mindray Confidential


Overview —reagent

Reagent Name Type

Diluent M-52D DILUENT M-52D

M-52DIFF LYSE M-52DIFF


Lyse
M-52LH LYSE M-52LH

Probe cleanser PROBE CLEANSER /

Control : Mindray made


Calibrator : R&D: S-CAL Mindray made : S50

21 © 2013 Mindray Confidential


目录

 Clinics
 Overview
 Principle
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

22 © 2013 Mindray Confidential


Principle —main technology

colorimetry
Laser Lambert
scattering –Beer
law
Impedance
Lyse
Flow
without
cytometry
cyanide

Cytochemical
staining

23 © 2013 Mindray Confidential


Principle—system block diagram

24 © 2013 Mindray Confidential


Principle—Flow cytometry
 Flow cytometry + Laser scattering—WBC 5DIFF
 Sheath fluid surrounds and forces blood cells suspended in dilution
to pass through the center of flow cell one by one.
Blood cells pass through the facula of flow cell with high speed

cell

Laser facula

Flow cell

sheath
sample

25 © 2013 Mindray Confidential


Principle—twice accelerating sheath flow

Flow cell Scatter light


Laser

sample

2ed Sheath flow 1st sheath flow

Blood sample

26 © 2013 Mindray Confidential


Principle—Multi-angle laser scattering

 Multi-angle laser scattering ;


 LAS—volume of the cells ;
 MAS—inner complexity
 WAS—granularity of the cells

27 © 2013 Mindray Confidential


Principle—cytochemical staining
 M-52 DIFF Lyse :
 Cationic surfactant To dissolve the RBC , special treatment to
WBC
 anionic groupMono
Lymph stain the
Nuegranular
Eos of the
BasoWBC


original

After adding
reagent

More difference in complexity

28 © 2013 Mindray Confidential


Principle—cytochemical staining

 M52-LH Lyse
 To beak down the RBC
 All the cells are naked (no membrane) except Baso, the Basophils are relatively
stable in size and inner structure, larger in volume than other kind of cells,
 To converts hemoglobin to a hemoglobin complex

original

After adding
reagent

Lymph Mono Nue Eos Baso

29 © 2013 Mindray Confidential


Principle—scattergram

Neu

Mon

Lym

Baso

Eos

30 © 2013 Mindray Confidential


Principle—L-M Scattergram ( report )

 DIFF : laser scattering+cytochemical staining

Monocyte
Neutrophil

Lymphocyte

Eosinophil

31 © 2013 Mindray Confidential


Principle—scattergram (RUO)

 L-W scattergram , M-W scattergram , special for Eos , more


accurate to large amount of EOS.

Eosinophil

32 © 2013 Mindray Confidential


Principle—histogram of Baso

 laser scattering + cytochemical staining scattergram


histogram

Basophil

33 © 2013 Mindray Confidential


Principle—position of cells

Scattergram
DIFF
Histogram
LAS

LIC

LIC

MON NEU

ALY EOS

LYM

Nrbc

BAS
GHO

MAS

34 © 2013 Mindray Confidential


Principle—HGB

 HGB measurement
 Lambert-beer law
 HGB(g/L) = Constant×Log 10 (Blank Photocurrent/Sample Photocurrent)

Fe 2+ 3+
Fe
+ Fe 2+ Fe
3+
Fe 2+ Fe
3+
Fe 2+ Fe
3+

Anionic surfactant

Lyse without cyanide


Special surfactant
ferrous iron is oxidized to ferric iron
35 © 2013 Mindray Confidential
Principle—RBC/PLT

 Impedance method

36 © 2013 Mindray Confidential


Principle—RBC/PLT

Time metering

 principle
 Volume metering : Use fixed volume of
metering tube to fix the quantity of each
test sample, Triggered by the optical
sensor 。
 Time metering : fixed test time to fix the
quantity of each test sample
volume=test time X velocity of the flow

Use large vacuum chamber(200ml)


to ensure stable negative pressure

37 © 2013 Mindray Confidential


Principle—RBC/PLT

 Test time
• Accurate hardware timing : RBC 9s

 How to ensure stable flow velocity


• Stable negative pressure
• Abnormal monitoring

1.Stable aperture voltage


2.Stable particle stream

Aperture voltage: 14.75-


23.75

38 © 2013 Mindray Confidential


Content

 Clinics
 Overview
 Principle
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

39 © 2013 Mindray Confidential


Installation—procedure

 Preparation
 Installation : check up, unpacking, emplacement, connection,
 Starting up and login
 Setting
 performance verification
 Background, reproducibility
 Quality control and calibration(if needed)
 Scattergram confirmation ( blood sample )
 Other setting : customized setting, maintenance, print.
 Customer training

40 © 2013 Mindray Confidential


Installation—preparation

 Tools :
 Blade or clipper
 USB flash disk
 Pipette (200ul)
 Cross-headed screwdriver
 Control (normal)
 Calibrator

41 © 2013 Mindray Confidential


Installation—preparation

 Environment :
 should be as free as possible from dust, mechanical vibrations, and
electrical interference;
 Far from brush-type motors, flickering fluorescent lights, and electrical
contacts that regularly open and close;
 Do not place the analyzer on a slope;
 Do not place the analyzer in direct sunlight or in front of a source of heat
or drafts;
 The environment shall be well ventilated

42 © 2013 Mindray Confidential


Installation—preparation

 Space : shown as in figure


enough room on or below the countertop to accommodate the
diluent and waste containers.
 Voltage : 100V-240V ~, 50Hz/60Hz
 Input power : ≤ 300VA
20cm
 Standard configuration :
 Analyzer, reagents
 Optional : printer , barcode scanner
50cm

43 © 2013 Mindray Confidential


Installation-unpacking

 Check :
 outer packing and apparatus
 Packing list and contents
 Apparatus arrangement

44 © 2013 Mindray Confidential


Installation—connection

Power switch
M-52D
diluent inlet

Waste outlet Waste sensor connector


Power socket

45 © 2013 Mindray Confidential


Installation—connection

46 © 2013 Mindray Confidential


Installation—connection

M-52DIFF lyse connector


M-52LH lyse connector

47 © 2013 Mindray Confidential


Installation—connection

 Peripheral devices :
Printer
Barcode scanner
 Computer
Network cable
Data management software

USB port and network port

48 © 2013 Mindray Confidential


Installation—Starting up

Hardware self test, liquid system self test and maintenance, background
test 。
49 © 2013 Mindray Confidential
Installation—Starting up

User ID : service
Password : Se s700 ( capital S, blank between Se and s700)

50 © 2013 Mindray Confidential


Installation—Starting up

51 © 2013 Mindray Confidential


Installation—Setting up

 reagent

press

press

52 © 2013 Mindray Confidential


Installation—Setting up

 reagent

press

Scan
or
input

53 © 2013 Mindray Confidential


Installation—Setting up

 Other settings

54 © 2013 Mindray Confidential


Installation—performance test

 Background Verification

Parameter Background/ blank count


requirements
WBC ≤ 0.20 109 / L
RBC ≤ 0.02 1012/ L
HGB ≤1 g/L
HCT ≤0.5 %
PLT ≤ 10  109 / L

55 © 2013 Mindray Confidential


Installation—performance test

 Background Verification

56 © 2013 Mindray Confidential


Installation—performance test

 Background Verification

57 © 2013 Mindray Confidential


Installation—performance test

 Reproducibility

58 © 2013 Mindray Confidential


Installation—performance test

 Reproducibility

59 © 2013 Mindray Confidential


Installation—quality control

press
60 © 2013 Mindray Confidential
Installation—quality control

 Click “By SD” to display the limits in the form of absolute value;
 Or click “By CV” to display the limits in the form of percentage
61 © 2013 Mindray Confidential
Content

 Clinics
 Overview
 Principle
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

62 © 2013 Mindray Confidential


Content

 Clinics
 Overview
 Principle
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

63 © 2013 Mindray Confidential


Software—calibration

old factor  reference value


new factor 
calculated mean

Transfer factor =CBC CBC+DIFF

Analysis Result = measurement result x factory’s factor x user’s factor x transfer


factor

64 © 2013 Mindray Confidential


Software—calibration

Factory
calibration
factor and
tansfer factor

with the engineer permission, factory’s Cal factor and transfer factor will
be modified, meanwhile the user’s factor will be adjusted to 100%

65 © 2013 Mindray Confidential


Software—calibration

User’s factor
only

with the administrator permission, user’s Cal factor is modified, it’s


recommend to be with “Admin” permission to login and calibrate

66 © 2013 Mindray Confidential


Software—gain

Target :4.5V

HGB gain can be modified manually, to modified the [set value] to


change the HGB blank voltage to around 4.5

67 © 2013 Mindray Confidential


Software—advanced toolbox

 Change language
 Choose the type of data saved
 Update
68 © 2013 Mindray Confidential
Software—one key export
U disk :
Kingston 8/16G
SanDisk 8/16G
Maxell 8/16G
FAT32
to reserve 4G
space.

Analyzer Info, software adjust Info, closed-reagent Info. Save to USB flash
disk, files named10000 和 10000_service, if the SN of the analyzer is 10000

69 © 2013 Mindray Confidential


Software—update

 Create an upgrade USB drive


Unzip update.rar and copy the unzipped “update” directory to the root directory
of the USB drive.
 Upgrade
 Insert the USB drive into the USB port on the analyzer. Enter the Advanced
Toolbox and press[update] to upgrade the software
 The upgrade process can be divided into two steps.
• Step 1: upgrade the guidance and operation system
• Step 2: upgrade the software components

When upgrading the boot and operation system, the operator will be prompted
to restart the analyzer between step 1 and step 2.
When upgrading the software components, the upgrade process will directly go
to step 2.

70 © 2013 Mindray Confidential


Software—communication

IP addresses of the PC and the analyzer shall be in the same


IP segment
Default IP of the analyzer : 10.0.0.2
71 © 2013 Mindray Confidential
Software—communication ( Mindray Hematology data
management Communication soft ware)

Port :5100
Analyzer’s IP

 Select the actual analyzer model


 the encoding mode of HL7 communication protocol is UTF-8,
and that of 15ID protocol is GBK

72 © 2013 Mindray Confidential


Software—communication ( LIS )

IP can
be
changed

contact your network administrator as to setting up IP address 、 subnet


mask and default gateway
73 © 2013 Mindray Confidential
Content

 Clinics
 Overview
 Principle
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

74 © 2013 Mindray Confidential


Hardware—boards

 Main board(digital control board)


 Analog drive board
 Power supply board
 Indicator board
 Touch screen control board
 Laser control board
 Pre-amplifier board
 MAS&LAS PD board
 WAS PD board
 Liquid sensor board

75 © 2013 Mindray Confidential


Hardware—system overview

Front panel Optical system


Touch
screen Laser PD
control
LED Screen
board
Touch screen Pre-
Laser drive
Indicator amplified
board
board board

Aurepert
Digital main board Analog driving board

HGB unit

Valve,
USB Motor floater
Power pump
supply
board Temp
Net Sensor hydraulic
control
Fluidic

76 © 2013 Mindray Confidential


Hardware—system control

Data Channel Master Control


Sample review Touch screen
Signal Signal
Sensor
preprocess sampling Result display TFT monitor
Signal
preprocess Result print USB printer

Channel Control Data storage

Data manager Network


interface
Bar-code
Input module
scanner
Drive/Detect
Control Start key

Motor Optical Alarm


Valve and
switch
pump control control control control

POWER Supervisor
POWER supervise system Indicator
board
State
supervise

77 © 2013 Mindray Confidential


Hardware—connection
Pinaster Master Analog&driver
Control Board Board
J1 Float
Ethernet J1
Communication J5 HGB
J81 J31
USB J2 J6 RBC
Optical
Analog Data J9 system
USB J3 J85 J4
Temperature
J10 sensor
BackLight J4 J11 Liq pressure
Supervise Signal sensor
LCD J86 J2 Optical
Data J16 J12 switch
Optical
Control Signal
J13 switch
Touch Screen J67 J77 J3
J14
Valve
Indicator J78 J15
J16 Pump
Start Key J99
J68 J17 Fan

P5V P12V J18 Fan


POWER Board
J19 Fan
D5V J1 Waste
J20
Pump
P12V/P24V J2 P5V
P12V J22 J21 Heat
A+12V/A-12V J3 P24V
J23 Motor
A+12V/A-12V
AC120V J4 J8 J24 Motor
AC120V

78 © 2013 Mindray Confidential


Hardware—Pinster main control board

 Composition : Digital section, analog section


 Function:
 A/D conversion
Converts the analog signals to digital signals which can be processed by FPGA or
CPU.
 Data processing
The digital signals obtained from A/D samples are digitally filtered and save by
FPGA, then is transmitted to CPU for further processing, output the result.
 Peripheral interface implementation
The CPU module provides a platform for system software, interfaces for peripherals
 Control interface extension
Provides control logic and interface for LCD, SD card and touchscreen, etc

79 © 2013 Mindray Confidential


Hardware—Pinster main control board

80 © 2013 Mindray Confidential


Hardware—analog drive board

 composition :
 Power module
 Control module
 monitoring module
 Power drive module

81 © 2013 Mindray Confidential


Hardware—analog drive board

 Power module
supplies power for all the electronic devices on the analog drive board, including both
input power from the analog drive board, and the power converted by the chip;
 Main control module
responsible for communicating with the control board and task scheduling;
 Monitoring module
collecting information such as temperature, pressure, voltage and fluid in the tubes;
 Power driving module
driving electric components such as motor, heater, fan, valves and pumps;
 RBC detection circuit
provides the RBC module with constant current source, RBC signal collection and
modulation;
 Modulation of optical signals and HGB signals.

82 © 2013 Mindray Confidential


Hardware—analog drive board

83 © 2013 Mindray Confidential


Hardware—Power supply board

Name Socket No. Description

PIN1 : GND
D5V J1
PIN2 、 PIN3 : 5V
PIN1 : GND
P12V PIN2 : 12V
J2
P24V
PIN3 : 24V
PIN1 、 PIN4 : GND

A+12V A- PIN2 、 PIN3 、 PIN5 : -


J3
12V 12V
PIN6 : 12V

AC120V J4 Alternating current

84 © 2013 Mindray Confidential


Hardware—Indicator board

 Indicate the current status of the analyzer by means of sound and


light

85 © 2013 Mindray Confidential


Hardware—status of indicator
Analyzer Status Indicator Comments

Ready Green light on Sequence is allowed

Running Green light flashing Sequence is being performed

Running with error Red light flashing An error is present and the
system is running
Stop with fault Red light on An error occurs and the
system is not running
No fault is present, but Yellow light on Initialization and standby
fluidics action is not status of sequence is not
allowed involved in the startup process
Enter/Exit standby status Yellow light flashing Enter/Exit standby status

86 © 2013 Mindray Confidential


Hardware—buzzer
Event Buzzer Prompts Comments
Startup completed One short beep Startup completed means the whole
startup process has been completed
and the analyzer is ready for operation
Open-vial aspiration completed Two short beeps

When count operation can not be One long beep If these screens have already
started in count related screens (including prompted, then the buzzer need not be
Sample Analysis, QC, Calibration, responded again.
Reproducibility, Carryover, Background,
Aging and Optical Gain Calibration
Count, etc.), press the aspirate key
Error Long intermittent Tap the touchscreen to stop the
beeps buzzer
Analyzer is ready One short beep Analyzer enters ready status from
other status
When the screen is black prompting stop beep If an error occurs during the
“Please power off the analyzer” shutdown process, the buzzer will
stop beeping when the screen is black

87 © 2013 Mindray Confidential


Hardware—touch screen drive board

 provides the interface between the touchscreen and the


control board
 converting the user's touch operation to identifiable signals
for main board.
 The touch screen drive board needs to be adjusted before
use.

88 © 2013 Mindray Confidential


Hardware—Liquid sensor board

 LED and Test point TP4


 output lower than 0.8V, the LED light is off, (have reagent)
 output higher than 2V, the LED light is on, (no reagent)

89 © 2013 Mindray Confidential


Hardware—optical boards

 Laser control board


 Pre-amplifier board
 MAS&LAS PD board
 WAS PD board

90 © 2013 Mindray Confidential


Hardware—Laser control board

 to provide stable and moderate-intensity driving current for laser


 Power modulation
 Laser driving current monitoring
 Laser power control.
 VR1:to adjust the power of the laser
Variable
Resistor

Control Laser driver Laser


(Closed-loop control) (HL6714G)

Voltage Laser current Power


supervise supervise PreProcess

A±12V
Laser Control
Board

91 © 2013 Mindray Confidential


Hardwar—Pre-amplifier board

 Power modulation
 I/V conversion
 Signal modulation.

92 © 2013 Mindray Confidential


Content

 Clinics
 Overview
 Principle
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

93 © 2013 Mindray Confidential


Optical system—introduction

 Simple structure, easy to maintain

94 © 2013 Mindray Confidential


Optical system—introduction
 Omit backward optical assembly, to receive scatter signal directly with a
large size PD instead, make it small and low cost

3106 optical

3107optical

95 © 2013 Mindray Confidential


Optical system—introduction

 Add a new signal sensor to collect the scatter at a wide angle, and the path
called WAS, three angles to collect scatter signal to for three-dimensional
scattergram that ensure the accurate category to the special sample ,
MAS
Collector Lens2

MAS Aperture
Collector Lens1

LD
Flow Cell Beam Splitter
Collimating Lens Cross Cylindrical
Lens
LAS Aperture
LAS

Flowcell Assembly
Light Source Assembly
Scattering Detection Assembly
96 © 2013 Mindray Confidential
Optical system—introduction

97 © 2013 Mindray Confidential


Optical system—principle

 To form an elliptic facula

Front view Focus point

Lasrer Collimating lens Cylindrical lens Light source assembly Flow cell

Focus point

Top view

98 © 2013 Mindray Confidential


Optical system—principle
Facula and signal : light intensity in horizontal direction
V

99 © 2013 Mindray Confidential 样本流


Optical system—principle

 Oval facula -Gaussian distribution


 Vertical direction: cover an entire blood cell
 Horizontal direction : wider than that of sample flow, even light
intensity in detection area.
cell

Laser facula

Flow cell

sheath
sample

100 © 2013 Mindray Confidential


Optical system—principle

 To collect the scatter light directly with PD assembly

101 © 2013 Mindray Confidential


Optical system—structure

 1-Light source assembly 5 4


6
 2 -Flow cell assembly 7
 3- base plate assembly
 4-Preamplifier and shield
shell
 5-Shield assembly
 6-Scattering detection
assembly
 7-Laser control board

3
2
1

102 © 2013 Mindray Confidential


Optical system—structure

103 © 2013 Mindray Confidential


Optical system—boards

 Laser control board


 Pre-amplifier board

Variable
Resistor

104 © 2013 Mindray Confidential


Optical system—temperature control

Item Temperature °C
Interior temperature 15~40 ( alarm )
Fan <25 , >30

Optical System
Temperature Testing Unit

J10
J10
J17
J17
Analog
Analog Drive
Drive J18
J18 Fan System
Board
Board J19
J19

 Heating : power supply board


 Heat dissipation : fan

105 © 2013 Mindray Confidential


Optical system—status confirmation

 To confirm the performance of optical system,


 CG position-intensity of the signal
 CV-signal stability
 Enter the “optical” screen, perform counting with 7um standard
particle solution (0.5ml deionized water+3drops standard particle )

Parameter Total CG Position CV


LAS 38~42 ≤6.50
MAS 1500~3000 107~158 ≤3.00
WAS 135~220 ≤8.00

106 © 2013 Mindray Confidential


Optical system—status confirmation

107 © 2013 Mindray Confidential


Optical system—maintenance

 Auto maintenance
 Clean the flow cell
Enter “maintenance]”–[clean] screen, press [flow cell],
 Probe cleanser soak
 Enter “maintenance]”–“maintenance]”–[overall]
 Perform the maintenance manually

108 © 2013 Mindray Confidential


Optical system—status confirmation

 Normal facular

109 © 2013 Mindray Confidential


Optical system—status confirmation

 Abnormal facula
 Use dust-free cloth with dehydrated alcohol to wipe the outer surface of
cylindrical lens ;
 Replace laser control board ;
 Replace the optical system ;
Abnormal
facula

110 © 2013 Mindray Confidential


Optical system—status confirmation

 If the exterior or interior of the flow cell is very bright, then the
exterior or interior may be stained.
 Use dust-free cloth with dehydrated alcohol to wipe the outer
surface of the floe cell.
 Auto or manual Before Cleaning After Cleaning

Clean the inner of the flow cell

111 © 2013 Mindray Confidential


Optical system—status confirmation

 Verify that the light beam is perpendicular to the flow cell


 Place an inner hexagon spanner in the path of the light. The output spot
and the spot reflected by the flow cell should overlap and strictly
coincide with each other

Output Spot

Reflection Spot

112 © 2013 Mindray Confidential


Optical system—replacement

 Shut down, disconnect the cables and fluidic tubes carefully.


 Unscrew
 Replace

 Drain out the residual fluid with syringe

113 © 2013 Mindray Confidential


Optical system—scattergram

Center of gravity is lower Scatter is dispersal and decentral


Low power

114 © 2013 Mindray Confidential


Optical system—scattergram

 Forward light assembly problem

115 © 2013 Mindray Confidential


Content

 Clinics
 Overview
 Principle
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

116 © 2013 Mindray Confidential


Fluidic system—overview

 2 Analysis channel : WBC&HGB channel, RBC&PLT channel


Sample

RBC/PLT WBC/HGB

M-52D diluent
M-52 DIFF lyse M-52 LH lyse

Impedance
Laser scattering Laser scattering
\colorimetry

RBC/PLT/MCV Lymph 、 Mon 、 Eos 、 N


eu WBC/BASO/HGB

117 © 2013 Mindray Confidential


Fluidic system—overview

Flowchart
( WB-CBC+DIFF )

118 © 2013 Mindray Confidential


Peripheral Blood
Sample 20μl

Fluidic system—overview
Diluent 480μl

Sample Diluent
Ratio 1:25 500μl
Outside the analyzer

Inside the analyzer Aspirate 200 μl of sample

Flowchart Diluent Background


358μl
WBC
bath

( PD-CBC+DIFF ) Aspirate 21.6 μl of


Sample Diluent secondary sample
Ratio 1:69.75

Variable temperature control

WBC
bath DIFF Lyse 1ml
Aspirate 380 μl of mixed fluid 36°

Sample WBC RBC


preparation LH Lyse 200μl Diluent 2.4ml
bath bath
tube

WBC DIFF Sample WBC BASO Sample RBC Sample Diluent


Diluent Ratio 1:199.8 Diluent Ratio 1:234.3 Ratio 1:7750

Flow cell WBC bath


DIFF HGB RBC
Measurement Measurement bath
RBC/PLT
Sample preparation tube, flow cell cleaning Measurement
Flow cell
119 © 2013 Mindray Confidential
WBC/BASO
Measurement
Fluidic system—reagent consumption

Diluent DIFF lyse(ml) LH lyse(ml) Probe


Mode (ml) cleanser
( ml )
CD 27.5 1 0.2 0
WB
CBC 23.2 0 0.2 0
CD 27.5 1 0.2 0
Capillary_W
CBC 23.2 0 0.2 0
CD 27.6 1 0.2 0
Prediluted
CBC 22.9 0 0.2 0
Shutdown 60.9 1 0.2 2
Normal startup 65 1 0.2 0
Wake-up 1 3.7 0 0 0
Wake-up2 16.9 0 0 0
Wake-up3 65 3 0.2 0

120 © 2013 Mindray Confidential


Fluidic system—fluidic diagram
1 2 3 4 5 6
MRSZ/R05N01.291.02(2.0) REV ECN DESCRIPTION DESIGN
Add the information of LF, T76 and tubes in the optical assembly, Revise T3,
2.0 CR-3107-042 T10, T11, T18, T20, T22, T32, T44, T57, T66, T70, T71, T75, C18, C19, Yang Yijie
C20, C21
3.0 CR-3107-047 Add T80,J32 and revise T48,J15 Yang Yijie

SV08 C17 LF C22 C23


T19 T76 T20

A C14 DILUENT
T25 T49
CV1
T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50
Preheat SV01
Bath T32

T26

T24
C27
J14-T48-J15-T80-J32 T47
T17

T77
SV02 DP2

J30-T78-J31
WBC&HGB channel T18
Case

LH Reagent Bottle
T79 SV07 C18
C19

T29

T71
J9
RBC channel
SV04
WBC RBC

T51
PV18
J25-T23-J26
B
C11 C10
J20-T63-J21 T34 T52
Case T33 C6

J22-T64-J23-T65-J24

J12-T45-J13

T72
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
C20
SV12

Isolation Chamber 1

Isolation Chamber 2
C21
T3

SV16 SV14 C12 C13

DIFF Reagent Bottle


T61 T62
SV06

T70
C5 SV13
T9

C4 T12 T13 T14 T37 T38


T39
C9

T40

T55
T15
T36
C3 SV11
T11 T16 C7
SV15
C26

T41
Hydraulic J5-T7-J6-T8
T10

C Sensor T53
C2 T46 C16
J4-T6 C8
Transducer
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66
1

Sampling/dispensing
T
SPB
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68 Confidential classification:
Confidential
SR CONFIDENTIAL
D
TITLE 3107 Pneumatic & Fluidics System Diagram
保密:此图及其全部知识产权(含著作权)归深圳迈瑞生物医疗电子股份有限公司所有。未经深圳迈瑞生物医疗电子股份有限公司预先书面许可,严禁出于任何目的,对 DOC No. A1-115-015670-00
此图的全部或部分内容(包括但不限于图中信息、数据、运算结果等)泄露、使用、拷贝或复制。
Classified documents, This set of drawing(s) and all it's intellectual property rights (including copyright) subsisting herein are property of DESIGN 杨一杰 REV. 3.0
Shenzhen Mindray Bio-medical Electronics Co.,Ltd. No disclosure,use,copies or reproductions should be made of this drawing or any part(s)
thereof for whatever purpose nor shall any information, data, calculations, or other contents contained in this drawing be disseminated without P.CODE 3107 SIZE A4
prior written permission of Shenzhen Mindray Bio-medical Electronics Co.,Ltd Software &Rev:
SHEET 1 OF 6
Microsoft visio 2003

121 © 2013 Mindray Confidential


Fluidic system—component
 Highly integrated air and hydraulic system
 17valves : 1 high pressure (2-way)(V3);1pinch valve(V18) 、 1LVM(V11)
 1 syringe assembly (2 syringes , 250ul 、 10ml )
 1 sampling assembly
 2 Electromagnetic Metering pumps ( 1ml 、 200ul )
 2 counting baths
 2 isolate chambers
 1 flow cell
 1 waster pump
 1 preheat bath assembly
 1 hydraulic pressure sensor
 1 check valve
 1 filter
 1 vacuum chamber

122 © 2013 Mindray Confidential


Fluidic system—component

 VALVE

123 © 2013 Mindray Confidential


Fluidic system—component

 Syringe Linkage

Syringe Linkage consist of a motor, a


driving assembly and two syringes

Name Specification Function


Small Used for quantitative aspiration, distribution and
Full scale
volume secondary aspiration of blood samples, and for injecting the
250μL
syringe sample into the flow cell for measurement
Used for quantitative addition of diluent to WBC and
Large RBC bath, wipe fluid supply, cleaning of interior and
Full scale
volume exterior of sample probe and reaction bath, forcing the
10mL
syringe sheath into the flow cell, cleaning of the flow cell and
sample preparation.

124 © 2013 Mindray Confidential


Fluidic system—component

 Electromagnetic  Hydraulic pressure sensor


Metering Pump

symbol
symbol

Function : Function:
DP1 : 1ml , DIFF lyse adding Monitoring fluid pressure, give a
DP2 : 200ul , LH lyse adding alert signal when abnormal or
beyond settings.
125 © 2013 Mindray Confidential
Fluidic system—component

 Preheat bath  Tem. Sensor for diluent

symbol symbol

To heat DIFF Monitor Temp of dluent,


lyse , maintain the Temp recommend a target Temp to
in DIFF reaction bath(WBC preheat bath
bath)
126 © 2013 Mindray Confidential
Fluidic system—component

 Fluidic temperature control

Item Preheat bath diluent


Target Refer to diluent
Temp
Alarm <target Temp -1.5 <10, >40
Temp >target Temp +3

Diluent
Temperature
Sensor

Reaction Bath
Temperature
J10 Sensor
PIN1 Temperature
Temperature
Analog Drive Switch
Switch
Preheat
Preheat
J21 assembly
assembly
Board PIN3

127 © 2013 Mindray Confidential


Fluidic system—component

 Waste pump  Check valve

drain the wipe block, WBC bath,


RBC bath and vacuum chamber, Define the flow direction of DIFF
create vacuum in the vacuum lyse in case of reverse flow
chamber Attention : direction !

128 © 2013 Mindray Confidential


Fluidic system—component  Wipe block

 Sample probe

Aspirate and dispense the blood sample


and probe cleanser.
Note: the sample probe is flat-tipped
Clean the sample probe and drain out the
with a side opening to ensure normal
waste fluid while cleaning the inner and
aspiration in case that the tip touches
outer of the probe
the bottom of the sample tube.

129 © 2013 Mindray Confidential


Fluidic system—component

 Count reaction bath


 New structure
 Easy to maintenance

 WBC bath : for sample mixing ,reacting, HGB measuring, provide


diluted and reacted sample for DIFF\BASO counting
 RBC bath :
 50um aperture(improve the accuracy of PLT)
 Special design to reduce bubble in back bath
 Consist of aperture, front bath and back bath
 RBC diluting and mix and counting

130 © 2013 Mindray Confidential


Fluidic system—component

 HGB assembly
 New design
 Convergent optical path
 Easy to Maintenance

131 © 2013 Mindray Confidential


Fluidic system—component

 Vacuum chamber : maintain negative pressure for RBC counting. Same


with that of BC3000

 WBC/RBC Isolation chamber :provide an isolation air room to cut off the
internal or external interference

132 © 2013 Mindray Confidential


Fluidic system—component
Cooperate with DP1 for DIFF Wipe block drainage , vacuum
SV01 SV10
lyse suction /drainage pressure release
Cooperate with DP2 for LH Control syringe switch to sampling or
SV02 SV11
lyse suction/ drainage sample preparation route
 Valve function

On/off between large and mall RBC bath to vacuum chamber on/off
SV03 SV12
syringe control for RBC counting
SV04 On/off the sheath flow SV13 RBC drainage
Control diluent adding to RBC Rinse sample preparation pipe ,RBC
SV05 SV14
bath bubbling
SV06 Bubbling for WBC bath SV15 WBC drainage
RBC bath rinse, cooperate with
SV07 SV16 Floe cell drainage
SV09 to rinse the wipe block
Control syringe suction of
diluent, cooperate with Pinch valve for sample preparation,
SV08 SV7/sv9 in dispensing diluent PV18 cooperate with syringe controlling
to RBC back bath and wipe sample preparation and rinse
block ,
cooperate with SV7 to rinse the
SV09
wipe block

133 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 Sampling/dispensing channel

SV08 C17 LF
T19 T76 T20

Function :
Diluent
 Sampling 、 HGB

T24
T17

SV07

C3
Fluidic
pressure
SV11
RBC
measurement

T51
C26

 Rinse the outside of the


J5-T7-J6-T8
T10

sensor T53
C2
J4-T6 SV10 T52

SP
C6
J3-T5

T57
2
T
J27-T22-J28

C1 SV03

1
T
SPB
T58 T66
 Rinse the inside of the
J1-T4-J2

SV09
T43 SP
T44
VC
 Suck and dispense
SR Waste
LP probe cleanser
T68

134 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 Sampling/dispensing channel
Function :
 Sampling 、 HGB
SV08 C17 LF
T19 T76 T20

diluent measurement

T24
T17

C3 RBC
SV07
 Rinse the outside of the
SP
SV11

T51
C26 J5-T7-J6-T8
T10

T53
C2
J4-T6 SV10 T52
C6
 Rinse the inside of the
SP
J3-T5

T57
2
T
J27-T22-J28

C1 SV03

1
T
SPB
T58 T66
 Suck and dispense
J1-T4-J2

SV09
T43 probe cleanser
VC
T44

LP
SR
T68

135 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 Sampling/dispensing channel
Function :
SV08
T19
C17
T76
LF
T20  Sampling 、 HGB
diluent measurement

T24
 Rinse the outside of
T17

SV07

C3 RBC
SV11
the SP

T51
C26 J5-T7-J6-T8
T10

 Rinse the inside of the


T53
C2
J4-T6 SV10 T52
C6

SP
J3-T5

T57
2
T
J27-T22-J28

C1 SV03

1
T
SPB
T66
 Suck and dispense
T58
J1-T4-J2

SV09
T43 probe cleanser
VC
T44

LP
SR
T68

136 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 Sampling/dispensing channel

SV08 C17 LF
T19 T76 T20

diluent

T24
Function :
T17

SV07

 Sampling 、 HGB
C3 RBC
SV11

T51
C26 J5-T7-J6-T8
T10

measurement
T53
C2
J4-T6 SV10 T52

 Rinse the outside of the


C6
J3-T5

T57
2
T
J27-T22-J28

C1 SV03

1
T
SPB
T66
SP
 Rinse the inside of the
T58
J1-T4-J2

T43
SV09

T44
VC SP
LP
 Suck and dispense
SR
T68 probe cleanser

137 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 WBC&HGB channel
SV08 C17 LF C14
T19 T76 T20 T25 T49
CV1
diluent J29 J7-T31-J8 T30
DIFF lyse
DP1
C27 Preheating T32
SV01
chamber C15
T77

J30-T78-J31
T50

T26
T9 LH lyse
J14-T48-J15-T80-J32 T47
T17

SV02 DP2
SV04
WBC
Function :

J9
PV18
J25-T23-J26

J20-T63-J21
C11

外套
T34
C10
T33
 Inject 240ul diluent
J12-T45-J13

T75
J10-T35-J11

 Dispense 15ul blood +


J22-T64-J23-T65-J24

J16-T59-J17
diluent 312ul
SV16 SV14 C12 C13
T61 T62
SV06
C5
T9

C4 T12 T13 T14 T37 T38


C3
T11
SV11
C9
 Suck diluted sample 12ul
T15

T36
C26
T10

 Add 1ml DIFF lyse


T53 T16 C7
C2 J5-T7-J6-T8
SV15
J4-T6
T41

 Bubble mixing
J3-T5

2
T T46
C8
J27-T22-J28

C1 SV03
T42

1
T
T66
SPB
J1-T4-J2

VC

Waste LP
SR T68
T69

138 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 WBC&HGB channel
SV08 C17 LF C14
T19 T76 T20 T25 T49
CV1
J29 J7-T31-J8 T30
diluent DIFF lyse
DP1
C27 Perheating SV01
T32
chamber
C15
T77

J30-T78-J31
T50

T26
T79
J14-T48-J15-T80-J32 T47 LH lyse
T17

SV02 DP2
SV04
WBC

J9
PV18
J25-T23-J26

J20-T63-J21
C11
T34
C10 Function :
T33

 Sample preparation 380ul


J12-T45-J13

T75
J10-T35-J11
J22-T64-J23-T65-J24

J16-T59-J17  Adjust the position in the


SV16 SV14 C12 C13
T61 T62
SV06
C5
T9

tube
C4 T12 T13 T14 T37 T38
C3 SV11
T11 C9
T15

T36
Fluidic
C26

 Add 200ul LH lyse


T10

pressure T53 T16 C7


C2 sensor J5-T7-J6-T8
SV15
J4-T6
T41

 Bubble mixing
J3-T5

2
T T46
C8
J27-T22-J28

C1 SV03
T42

1
T
T66
SPB
J1-T4-J2

VC

Waste LP
SR T68
T69

139 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 WBC&HGB channel
SV08 C17 LF C14
T19 T76 T20 T25 T49
CV1
diluent J29 J7-T31-J8 T30 DIFF lyse
DP1
C27 Preheating SV01
T32
chamber
C15
T77

J30-T78-J31
T50
T79

T26
LH lyse
J14-T48-J15-T80-J32 T47
T17

SV02 DP2
SV04
WBC
Function :

J9
PV18
J25-T23-J26

J20-T63-J21
C11

T75
T34
C10
T33
 To form the sheath flow
J12-T45-J13

J10-T35-J11

 To boost the sample


J22-T64-J23-T65-J24

J16-T59-J17
SV16 SV14 C12 C13

 DIFF count
T61 T62
SV06
C5
T9

C4 T12 T13 T14 T37 T38


C3
T11
SV11
C9
 HGB measurement
T15

T36
Fluidic
C26
T10

pressure T53 T16 C7


C2 sensor J5-T7-J6-T8
SV15
J4-T6
T41
J3-T5

2
T T46
C8
J27-T22-J28

C1 SV03
T42

1
T
T66
SPB
J1-T4-J2

VC

Waste LP
SR T68
T69

140 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 WBC&HGB channel
SV08 C17 LF C14
T19 T76 T20 T25 T49
CV1
diluent J29 J7-T31-J8 T30 DIFF lyse
DP1
C27 Preheating SV01
T32
chamber
C15
T77

J30-T78-J31
T50
外套

T26
T79 LH lyse
J14-T48-J15-T80-J32 T47
T17

SV02 DP2
SV04
WBC
Function :

J9
PV18
J25-T23-J26

J20-T63-J21
C11

外套
T34
C10
T33
 Suck 156.3ul sample
J12-T45-J13

from WBC bath


T75
J10-T35-J11
J22-T64-J23-T65-J24

J16-T59-J17
 Rinse the flow cell
SV16 SV14 C12 C13
T61 隔 隔 T62
SV06 离 离
C5 室 室
T9

C4 T12 T13 T14 1 2 T37 T38


C3
T11
SV11
C9
 Rinse the route of sample
T15

T36
C26
T10

preparation anf flush the


液压传感器 T53 T16 C7
C2 J5-T7-J6-T8
SV15
J4-T6
T41
J3-T5

2
T46

sample injector and flow


T
C8
J27-T22-J28

C1 SV03
T42

cell
1
T
T66
SPB
J1-T4-J2

VC

废液 LP
SR T68
T69

141 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 WBC&HGB channel
SV08
T19
C17
T76
LF
T20
CV1
T25
C14
T49 Function :
DIFF lyse
 Sample preparation
J29 J7-T31-J8 T30
diluent
Preheating DP1
C27 SV01
chamber T32
C15
800ul
T77

J30-T78-J31
T50
T79

T26
LH lyse
J14-T48-J15-T80-J32 T47
 To form sheath flow
T17

SV02 DP2
SV04

 To boost the sample


WBC

J9
PV18
J25-T23-J26

C11

 BASO counting
J20-T63-J21 C10
T34
T33
T75
J12-T45-J13

J10-T35-J11
J22-T64-J23-T65-J24

J16-T59-J17
SV16 SV14 C12 C13
 Rinse(same as DIFF)
 WBC bath drainage
T61 T62
SV06
C5
T9

C4 T12 T13 T14 T37 T38

 Inject diluent to WBC


C3 SV11
T11 C9
T15

T36
Fluidic
pressure C26
T10

T53 T16 C7

bath
sensor J5-T7-J6-T8
C2 SV15
J4-T6
T41
J3-T5

2
T T46
C8
J27-T22-J28

C1 SV03
T42

1
T
T66
SPB
J1-T4-J2

VC

Waste LP
SR T68
T69

142 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 WBC&HGB channel
SV08 C17 LF C14
T19 T76 T20

diluent J29
CV1
J7-T31-J8
T25

T30
T49

DIFF lyse
Function :
 Rinse (as DIFF)
DP1
C27 Reheating SV01
T32
chamber
C15
T77

J30-T78-J31
 Rinse and drain out
T50
T79

T26
LH lyse
J14-T48-J15-T80-J32 T47
T17

SV04
WBC
SV02 DP2
WBC bath
 Inject 2018ul diluent to
J9
PV18
J25-T23-J26

C11

WBC bath as base


J20-T63-J21 C10
T34
T33
T75
J12-T45-J13

J10-T35-J11
J22-T64-J23-T65-J24

J16-T59-J17
SV16 SV14 C12 C13
solutioon
T61 T62
SV06
C5
T9

C4 T12 T13 T14 T37 T38


C3 SV11
T11 C9
T15

T36
Fluidic
C26
T10

pressure T53 T16 C7


C2 sensor J5-T7-J6-T8
SV15
J4-T6
T41
J3-T5

2
T T46
C8
J27-T22-J28

C1 SV03
T42

1
T
T66
SPB
J1-T4-J2

VC

Waste LP
SR T68
T69

143 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 RBC&PLT channel SV08 LF


T19 C17

Fuction :
T76 T20
diluent

T21
SV05
 Inject diluent (1600ul)
T18

T24
SV07

 Add diluted sample

T29
T17

RBC

T51
(12ul) to RBC bath
 Bubble mixing
T52
C6
Temp. sensor J20-T63-J21 C11

T54
J18-T60-J19
 RBC 、 PLT counting
J10-T35-J11

SV12
J22-T64-J23-T65-J24
T3

C13
T62

SV14 SV13
C4 T37 T38
T39
C9

T40

T55
C3 SV11
T11 T36 C7
Fluidic T43
C26

T41
pressure J5-T7-J6-T8
T10

T53
sensor C16
C2
J4-T6 C8
Transducer
J3-T5

T56
T T42
J27-T22-J28

T67
C1 SV03 T66
1
T
SPB
J1-T4-J2

VC

SV09

T44 LP
Waste
T69 T68

SR

144 © 2013 Mindray Confidential


Fluidic system—functioning and sequence
 RBC&PLT channel
SV08 C17 LF
T19
T76 T20
diluent

T21
Function :
SV05

 Rinse the RBC


T18

T24
SV07

T29
front bath
T17

RBC

T51
T52
C6
Temp.sensor J20-T63-J21 C11

T54
J18-T60-J19
J10-T35-J11

SV12
J22-T64-J23-T65-J24
T3

C13
T62

SV14 SV13
C4 T37 T38
T39
C9

T55
T40
C3 SV11
T11 T36 C7
Fluidic T43
C26

T41
pressure J5-T7-J6-T8
T10

sensor T53
C2 C16
J4-T6 C8
Transducer
J3-T5

T56
T T42
J27-T22-J28

T67
C1 SV03 T66
1
T
SPB
J1-T4-J2

VC

SV09

T44 LP
Waste
T69 T68

SR

145 © 2013 Mindray Confidential


Fluidic system—functioning and sequence

 RBC&PLT channel

SV08 C17 LF
Function :
T19 T76 T20

diluent
 Rinse the RBC
back bath

T24
T17

SV07

C3
Fluidic
SV11
RBC
 Inject diluent

T51
(2560ul) to RBC
pressure C26 J5-T7-J6-T8
T10

sensor T53
C2
J4-T6 SV10 T52
C6
bath as base
J3-T5

T57
2
T
J27-T22-J28

C1 SV03

1
T
SPB
solution
T66

 Release vacuum
T58
J1-T4-J2

T43
SV09

VC
T44

LP
SR Waste
T68

146 © 2013 Mindray Confidential


Fluidic system—sequence in detail
 WB_CBC+DIFF
SV08 C17 LF C22 C23
T19 T76 T20

C14 DILUENT
T25 T49
CV1

Small syringe SV05


T21 J29 J7-T31-J8 T30

DP1
C15
T50
Preheat SV01

aspirate 15μL blood Bath T32

T26

T24
C27
J14-T48-J15-T80-J32 T47

sample, while large

T17

T77
SV02 DP2

J30-T78-J31
T18
Case

syringe aspirate 600ul

LH Reagent Bottle
T79 SV07 C18
C19

T29

T71
diluent,

J9
SV04
WBC RBC

T51
HGB background
PV18
J25-T23-J26

Empty WBC bath


C11 C10
J20-T63-J21 T34 T52
Case T33 C6

J22-T64-J23-T65-J24

J12-T45-J13

T72
Temperature Sensor T75

J10-T35-J11
Linkage SR aspirate

T54
J16-T59-J17

J18-T60-J19
C20

Isolation Chamber 2
Isolation Chamber 1
SV12
C21

DIFF Reagent Bottle


5360ul diluent
T3

SV16 SV14 C12 C13


T61 T62
SV06

T70
(SV03 、 PV18, SV11
C5 SV13
T9

C4 T12 T13 T14 T37 T38


T39
C9

T40

T55
T15
toT64 )
T36
C3 SV11
T11 T16 C7

Zap aperture of RBC


SV15
C26

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8
Transducer
T57
J3-T5

T58
2

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66
1
T
SPB
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

147 © 2013 Mindray Confidential


Fluidic system—sequence in detail
SV08 C17 LF C22 C23
T19 T76 T20

C14 DILUENT
T25 T49
CV1
T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50
Preheat SV01
T32

 Sample probe goes up


Bath

T26

T24
C27
J14-T48-J15-T80-J32 T47

T17
to upper position( the

T77
SV02 DP2

J30-T78-J31
T18
Case

LH Reagent Bottle
T79 SV07 C18

tip inside the wipe) C19

T29

T71
J9
SV04
WBC RBC

 Linkage SR rinse

T51
PV18
J25-T23-J26

exterior of SPB and J20-T63-J21


C11

Case
T34
C10
T33
T52
C6

J22-T64-J23-T65-J24

J12-T45-J13

T72
T75

RBC back bath, drain


Temperature Sensor

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
C20

Isolation Chamber 1

Isolation Chamber 2
SV12

by LP
C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13


T61 T62
SV06

T70
C5 SV13
T9

T12 T13 T14 T37 T38

 SPB to the position


C4
T39
C9

T40

T55
T15
T36

above the WBC bath


C3 SV11
T11 T16 C7
SV15
C26

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16

 Empty WBC bath for J4-T6 C8


Transducer
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

T67

the second time


C1 SV03 SV10
T43 T66
1
T
SPB
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

148 © 2013 Mindray Confidential


Fluidic system—sequence in detailv
SV08 C17 LF C22 C23
T19 T76 T20

C14 DILUENT
T25 T49
CV1
T21 J29 J7-T31-J8 T30

Linkage SR injects
SV05 C15
DP1 T50
Preheat SV01
Bath T32

T26

T24
240ul diluent into WBC
C27
J14-T48-J15-T80-J32 T47

T17

T77
SV02 DP2

J30-T78-J31
bath as base solution

T18
Case
T79

LH Reagent Bottle
SV07 C18

Empty RBC bath


C19

T29

T71
J9
SV04

SPB goes down


WBC RBC

T51
PV18
J25-T23-J26

into WBC bath J20-T63-J21


C11

Case
T34
C10
T33
T52
C6

J22-T64-J23-T65-J24
Small syringe

J12-T45-J13

T72
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
C20

Isolation Chamber 1

Isolation Chamber 2
SV12

dispenses 18 sample,
C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13


T61 T62
SV06

T70
large SR rinse RBC C5 SV13
T9

C4 T12 T13 T14 T37 T38


T39

inlet with 720ul diluent


C9

T40

T55
T15
T36
C3 SV11
T11 T16 C7

Linkage SR inject C26


SV15

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16

312ul rinse interior of


J4-T6 C8
Transducer
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

T67

SPB(SV03,
C1 SV03 SV10
T43 T66
1
T
SPB

SV11),sample diluted
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

149 © 2013 Mindray Confidential


Fluidic system—sequence in detail

 WBC bubble mixing, SV08


T19
C17
T76
LF
T20
C22 C23

linkage SR T21 J29


CV1
J7-T31-J8
T25

T30
C14
T49
DILUENT

SV06(SV03, SV11, SV05


Preheat
Bath T32
SV01
DP1
C15
T50

T26
PV18)

T24
C27
J14-T48-J15-T80-J32 T47

T17

T77
SV02 DP2

J30-T78-J31
T18
 SPB goes up (inside the
Case
T79

LH Reagent Bottle
SV07 C18
C19

T29

T71
wipe), LP drain the waste

J9
SV04
WBC RBC

T51
PV18

of wipe block
J25-T23-J26

C11 C10
J20-T63-J21 T34 T52
Case T33 C6

 Rinse interior of SPB, ),

J22-T64-J23-T65-J24

J12-T45-J13

T72
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
C20

LP drain the waste of


SV12

Isolation Chamber 2
Isolation Chamber 1
C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13


T61 T62

T70
SV06

wipe block C5 SV13


T9

C4 T12 T13 T14 T37 T38


T39
C9

T40

T55
T15
T36

 Linkage SR aspirate C3
T11

C26
SV11

SV15
T16 C7

T41
Hydraulic J5-T7-J6-T8
T10

1580ul diluent. Sensor T53


C2 T46 C16
J4-T6 C8
Transducer
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

 SPB down into WBC


SV10 T67
C1 SV03
T43 T66
1
T
SPB

bath
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

 DP1 suck DIFF lyse T44 Waste


Tank
WASTE
C25 C24
LP

T69 T68

SR

150 © 2013 Mindray Confidential


Fluidic system—sequence in detail

 Empty RBC bath


SV08 C17 LF C22 C23
T19 T76 T20

C14 DILUENT
T25 T49

 Small SPB aspirate


CV1
T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50

12ul diluted sample


Preheat SV01
Bath T32

T26

T24
C27
J14-T48-J15-T80-J32 T47

from WBC bath, large


T17

T77
SV02 DP2

J30-T78-J31
T18
Case

SPB for 480 diluent T79 SV07 C18

LH Reagent Bottle
C19

T29

T71
(SV03 off)

J9
SV04
WBC RBC

T51
PV18
J25-T23-J26

 SPB goes up (inside J20-T63-J21


C11
T34
C10
T52
Case T33 C6

J22-T64-J23-T65-J24
the wipe),rinse exterior

J12-T45-J13

T72
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
C20

of SPB LP drain the


SV12

Isolation Chamber 1

Isolation Chamber 1
C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13


T61 T62

waste of wipe block,


SV06

T70
C5 SV13
T9

C4 T12 T13 T14 T37 T38


T39

then turn to RBC bath.


C9

T40

T55
T15
T36
C3 SV11
T11 T16 C7
SV15

 DP1 add DIFF lyse to


C26

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8
Transducer
WBC bath
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66

 Bubble mix WBC via


1
T
SPB
J1-T4-J2

Diluent Tank
T73
VC

SV06.

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

151 © 2013 Mindray Confidential


Fluidic system—sequence in detail
SV08 C17 LF C22 C23
T19 T76 T20

C14 DILUENT
T25 T49
CV1
T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50
Preheat SV01
Bath T32

T26

T24
C27
J14-T48-J15-T80-J32 T47

T17
 Linkage SR add1600ul

T77
SV02 DP2

J30-T78-J31
T18
Case

diluent via SV05 to


T79

LH Reagent Bottle
SV07 C18
C19

T29

T71
J9
RBC bath as base SV04
WBC RBC

T51
PV18

solution
J25-T23-J26

C11 C10
J20-T63-J21 T52
T34

 SPB down to RBC


Case T33 C6

J12-T45-J13

T72
J22-T64-J23-T65-J24
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
C20
SV12

Isolation Chamber 1

Isolation Chamber 2
 SPB dispense 12ul
C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13


T61 T62
SV06

T70
diluted sample + 800ul C5 SV13
T9

C4 T12 T13 T14 T37 T38


T39
C9

diluent (rinse interior of

T40

T55
T15
T36
C3 SV11
T11 T16 C7

SPB) C26
SV15

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8
Transducer
SPB to upper position LP
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66

drain waste of wipe block 1


T
SPB
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

152 © 2013 Mindray Confidential


Fluidic system—sequence in detail
SV08 C17 LF C22 C23
T19 T76 T20

C14 DILUENT
T25 T49
CV1
T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50
Preheat SV01
Bath T32

T26

T24
 Linkage SR aspirate C27
J14-T48-J15-T80-J32 T47

T17

T77
SV02 DP2

J30-T78-J31
diluent from diluent tank

T18
Case
T79 C18

LH Reagent Bottle
SV07
C19

T29
 SPB to the up position of

T71
J9
SV04
WBC RBC

T51
sampling
PV18
J25-T23-J26

C11 C10
J20-T63-J21
T34 T52

 Linkage SR aspirate Case T33 C6

J12-T45-J13

T72
J22-T64-J23-T65-J24
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
380ul DIFF sample into SV12
C20

Isolation Chamber 1

Isolation Chamber 2
C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13

sample preparation
T61 T62

T70
SV06
C5 SV13
T9

C4 T12 T13 T14 T37 T38

pipe(T65-T63), C9
T39

T40

T55
T15
T36
C3 SV11
T11 T16 C7

 DP2 suck LH lyse C26


SV15

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8

 Aspirate 200ul via C11-


Transducer
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66

T35 –SV14-ISC1, adjust 1


T
SPB
J1-T4-J2

sample position

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

153 © 2013 Mindray Confidential


Fluidic system—sequence in detail

 Add LH lyse to WBC bath SV08


T19
C17
T76
LF
T20
C22 C23

DILUENT
 WBC bubble mixing
C14
T25 T49
CV1
T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50

 Large SR inject diluent to


Preheat SV01
Bath T32

T26

T24
C27
J14-T48-J15-T80-J32 T47

form sheath flow via


T17

T77
SV02 DP2

J30-T78-J31
T18
Case

SV04,

LH Reagent Bottle
T79 SV07 C18
C19

T29

T71
J9
 At the same time, SV03 SV04
WBC RBC

T51
PV18
J25-T23-J26

open 0.2s, linkage SR J20-T63-J21


C11
T34
C10
T52

boost sample to flow cell


Case T33 C6

J12-T45-J13

T72
J22-T64-J23-T65-J24
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
C20

Isolation Chamber 1

Isolation Chamber 2
SV12

 SV03 closed, small SR


C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13


T61 T62
SV06

T70
push the sample pass C5 SV13
T9

C4 T12 T13 T14 T37 T38


T39
C9

through the flow cell,

T55
T40
T15
T36
C3 SV11
T11 T16 C7

sheath flow remains. C26


SV15

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8
Transducer
 DIFF detection for 9.1s
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66

 HGB measurement
1
T
SPB
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

154 © 2013 Mindray Confidential


Fluidic system—sequence in detail
SV08 C17 LF C22 C23

 Linkage SR pull down, C14


T19 T76 T20

DILUENT

large SR aspirate diluent,


T25 T49
CV1
T21 J29 J7-T31-J8 T30
SV05 C15

small SR aspirate 156.3ul Preheat


Bath T32
SV01
DP1 T50

T26

T24
sample from WBC bath
C27
J14-T48-J15-T80-J32 T47

T17

T77
SV02 DP2

J30-T78-J31
T18
Case

 Create negative pressure T79 SV07 C18

LH Reagent Bottle
C19

T29
in VC

T71
J9
SV04
WBC RBC

T51
PV18
J25-T23-J26

 Rinse flow cell via SV04- J20-T63-J21


C11 C10
T34 T52

SV16. rinse injector via Case T33 C6

J12-T45-J13

T72
J22-T64-J23-T65-J24
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
SV04-sv14. rinse sample C20

Isolation Chamber 1

Isolation Chamber 2
SV12
C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13

preparation route via SV06


T61 T62

T70
C5 SV13
T9

SV11-SV14-isolation
C4 T12 T13 T14 T37 T38
T39
C9

T55
T40
T15
T36

chamber2 to bubble mix C3


T11
SV11

SV15
T16 C7

the sample in RBC


C26

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8
Transducer
J3-T5

T57

 The constant current


T58
2

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66

source of RBC bath turns


1
T
SPB
J1-T4-J2

Diluent Tank
ON

T73
VC

T74
SV09
WASTE
 RBC counting(SV12 on)
T44 LP
Waste
Tank C25 C24
T69 T68

SR

155 © 2013 Mindray Confidential


Fluidic system—sequence in detail
SV08 C17 LF C22 C23
T19 T76 T20

Aspirate rest sample into CV1


T25
C14
T49
DILUENT

preparation pipe(T63-T65) SV05


T21

Preheat
J29 J7-T31-J8 T30

SV01
DP1
C15
T50

from WBC Bath T32

T26

T24
C27
J14-T48-J15-T80-J32 T47

T17

T77
 Large SR inject diluent to
SV02 DP2

J30-T78-J31
T18
Case

LH Reagent Bottle
T79 SV07 C18

form sheath flow via SV04, C19

T29

T71
J9
SV04
RBC

 At the same time, SV03


WBC

T51
PV18
J25-T23-J26

open 0.2s, linkage SR boost J20-T63-J21


C11

Case
T34
C10
T33
T52
C6

sample to flow cell

J12-T45-J13

T72
J22-T64-J23-T65-J24
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
C20

Isolation Chamber 1

Isolation Chamber 2
SV12
C21

DIFF Reagent Bottle


 SV03 closed, small SR
T3

SV16 SV14 C12 C13


T61 T62
SV06

T70
C5 SV13

push the sample pass


T9

C4 T12 T13 T14 T37 T38


T39
C9

T40

T55
T15
through the flow cell,
T36
C3 SV11
T11 T16 C7

sheath flow remains


SV15
C26

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8
Transducer
 WBC /BASO counting
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66
1
T
SPB
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

156 © 2013 Mindray Confidential


39.8~48.8s RBC 、 PLT 计数
Fluidic system—sequence in detail
SV08 C17 LF C22 C23
T19 T76 T20

C14 DILUENT
T25 T49
CV1
T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50
Preheat SV01
Bath T32

T26

T24
C27
J14-T48-J15-T80-J32 T47

T17

T77
SV02 DP2

J30-T78-J31
T18
Case

 Empty RBC bath T79 SV07 C18

LH Reagent Bottle
C19

T29

T71
J9
 Rinse sample preparation SV04
WBC RBC

T51
PV18
J25-T23-J26

pipe via (PV18 power on) J20-T63-J21


C11
T34
C10
T52

SV03-SV11-C11- Case T33 C6

J12-T45-J13

T72
J22-T64-J23-T65-J24
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
SV14 、 rinse flow cell

Isolation Chamber 1

Isolation Chamber 2
C20
SV12

DIFF Reagent Bottle


C21
T3

SV16 SV14 C12 C13

via sv04-sv16 SV06


T61 T62

T70
C5 SV13
T9

C4 T12 T13 T14 T37 T38


T39

 Linkage SR aspirate C9

T40

T55
T15
T36
C3 SV11

diluent.
T11 T16 C7
SV15
C26

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8
Transducer
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66
1
T
SPB
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

157 © 2013 Mindray Confidential


Fluidic system—sequence in detail
SV08 C17 LF C22 C23
T19 T76 T20

C14 DILUENT
T25 T49
CV1
T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50
Preheat SV01
Bath T32

T26

T24
C27
J14-T48-J15-T80-J32 T47

T17

T77
SV02 DP2

J30-T78-J31
 Empty WBC bath

T18
Case

LH Reagent Bottle
T79 SV07 C18
C19

T29

T71
 Flush injector via SV04-

J9
SV04
WBC RBC

T51
PV18

C11-SV14-ISC J25-T23-J26

C11 C10
J20-T63-J21 T34 T52

 Rinse pinch tube to WBC Case T33 C6

J12-T45-J13

T72
J22-T64-J23-T65-J24
Temperature Sensor T75

J10-T35-J11

T54
J16-T59-J17

J18-T60-J19
(PV18 power off) C20

Isolation Chamber 1

Isolation Chamber 2
SV12
C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13


T61 T62

 Empty WBC bath


SV06

T70
C5 SV13
T9

C4 T12 T13 T14 T37 T38


T39
C9

T40

T55
T15
T36
C3 SV11
T11 T16 C7
SV15
C26

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8
Transducer
J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66
1
T
SPB
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

158 © 2013 Mindray Confidential


Fluidic system—sequence in detail
SV08 C17 LF C22 C23
T19 T76 T20

C14 DILUENT
T25 T49

 Add 2560ul diluent to RBC


CV1
T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50

bath
Preheat SV01
Bath T32

T26

T24
C27
J14-T48-J15-T80-J32 T47

T17
 Add 2018ul diluent to WBC

T77
SV02 DP2

J30-T78-J31
T18
Case
T79

LH Reagent Bottle
SV07 C18

bath C19

T29

T71
J9
SV04
RBC

 Turn on the LP rinse wipe


WBC

T51
PV18
J25-T23-J26

block J20-T63-J21
C11

Case
T34
C10
T33
T52
C6

J12-T45-J13

T72
J22-T64-J23-T65-J24
Temperature Sensor T75

J10-T35-J11

T54
 SPB to ready position

J16-T59-J17

J18-T60-J19
C20
SV12

Isolation Chamber 1

Isolation Chamber 2
C21

DIFF Reagent Bottle


T3

SV16 SV14 C12 C13


T61 T62

 Small SR aspirate 4ul air to SV06

T70
C5 SV13
T9

C4 T12 T13 T14 T37 T38

create an isolation bubble


T39
C9

T40

T55
T15
T36
C3 SV11
T11 T16 C7

 Release vacuum pressure


SV15
C26

T41
Hydraulic J5-T7-J6-T8
T10

Sensor T53
C2 T46 C16
J4-T6 C8

(SV10 on, LP off) Transducer


J3-T5

2 T57 T58

T56
T42
T
J27-T22-J28

SV10 T67
C1 SV03
T43 T66

 Ready for next test. 1


T
SPB
J1-T4-J2

Diluent Tank
T73
VC

T74
SV09

T44
WASTE LP
Waste
Tank C25 C24
T69 T68

SR

159 © 2013 Mindray Confidential


Fluidic system—sequence in detail

 Measurement sequence in PD-CBC+DIFF mode


 Measurement sequence in PD-CBC+DIFF mode is basically the same
with the sequence in WB-CBC+DIFF mode, except that:
 Blood sample has been prediluted outside the analyzer,
 200μL of prediluted blood is aspirated in the PD mode
 Measurement sequence in CBC mode
 Compared with the sequence in CBC+DIFF mode, the measurement
sequence in CBC mode is basically the same except that:
 Sample volume: 15μL in the WB-CBC+DIFF mode, 11.7μL in the WB-
CBC mode.(but for PD mode, CBC+DIFF and CBC mode are the same )
 There is no actions related to DIFF measurement, including addition of
DIFF reagent, preparation of DIFF samples, DIFF measurement, and
cleaning process of the tubes and the flow cell after the DIFF measurement.

160 © 2013 Mindray Confidential


Fluidic system—sequence for maintenance
Maintenance with probe cleanser(shut down ) SV08 C17 LF C22 C23
T19 T76 T20

The orange lines are the CV1


T25
C14
T49
DILUENT

locations soaked by the


T21 J29 J7-T31-J8 T30
SV05 C15
DP1 T50
Preheat SV01
T32

probe cleanser,
Bath

T26

T24
C27
J14-T48-J15-T80-J32 T47

T17

T77
SV02 DP2

J30-T78-J31
the blue lines are the

T18
Case

LH Reagent Bottle
T79 SV07 C18
C19

locations passed by the

T29

T71
J9
SV04

probe cleanser
WBC RBC

T51
PV18
J25-T23-J26

C11

enhanced probe cleanser J20-T63-J21 C10


T34 T52
Case T33 C6

J22-T64-J23-T65-J24

J12-T45-J13

T72
Temperature Sensor T75

J10-T35-J11

T54
maintenance sequence

J16-T59-J17

J18-T60-J19
C20

Isolation Chamber 1

Isolation Chamber 1
SV12
C21

DIFF Reagent Bottle


will be called every 300
T3

SV16 SV14 C12 C13


T61 T62
SV06

T70
times of sample
C5 SV13
T9

C4 T12 T13 T14 T37 T38


T39
C9

T40

T55
T15
measurement.
T36
C3 SV11
T11 T16 C7

The soaking time of


SV15
C26

T41
Hydraulic
T10

Sensor T53
C2 T46 C16
J5-T7-J6-T8 C8

enhanced maintenance is
J4-T6
T2 Transducer
J3-T5

T57 T58

T56
T42
J27-T22-J28

SV10 T67

6.67S
C1 SV03
T43 T66
T1
SPB

In the Maintenance

Diluent Tank
J1-T4-J2

T73
VC

screen, the probe

T74
SV09

T44
WASTE LP

cleanser maintenance is
Waste
Tank C25 C24
T69 T68

defaulted as enhanced SR

probe cleanser
maintenance.
161 © 2013 Mindray Confidential
Fluidic system—sequence in detail
Starting up rinse
 The fluidic actions on startup can be divided into three procedures: :
1. Initialization of fluidics components: initialization of sampling assembly and
syringe assembly; 1 metering pump action (without consumption of reagents);
creating vacuum and releasing vacuum.
2. Cleanup: including cleaning of all the tubes and components on the analyzer;
removing bubbles from the flow cell; discarding 1 ml of DIFF reagent and 0.2 ml
of LH reagent.
3. Background: with WB-CBC+DIFF measurement sequence

If the background fails, perform the cleanup again and measure the
background.
For startup after an abnormal shutdown, perform the cleanup twice. 。
Initialization of fluidics components, cleaning up of the interior and exterior
surface of the sample probe while starting up after updating

162 © 2013 Mindray Confidential


Fluidic system—standby

 The instrument will enter standby status after idling for 15 to 30 minutes (defaulted as
15 minutes). After entering standby status, operations without fluidic actions can be
performed from the screen.
 Exit standby status 1: Standby for less than 1 hour.
Clean the exterior of the sample probe and the WBC bath and rebuild the
isolation bubble without consumption of lyse.
 Exit standby status 2: Standby for more than 1 hour and less than 5 hour.
 Clean the interior and exterior of the sample probe, the WBC bath, the RBC
bath, the sample supply tube and the flow cell, and rebuild the isolation bubble
without consumption of lyse.
 Exit standby status 3: Standby for not less than 5 hour.
The same as cleanup on startup, including cleaning of all the tubes and
components on the analyzer; removing bubbles from the flow cell; discarding 3 ml of
DIFF reagent to eliminate the effects of bubbles in the preheat bath tube; discarding
0.2 ml of LH reagent to eliminate the effects of crystal and bubbles.
163 © 2013 Mindray Confidential

164 © 2013 Mindray Confidential
目录

 Clinics
 Overview
 Principle
 Installation
 Software
 Hardware
 Optical system
 Liquid system
 Mechanical structure

165 © 2013 Mindray Confidential


Mechanical system—overall layout

① 2-way valve
② 3-way valve
③ Syringe assemly
④ 1ml metering pump
⑤ microswitch
⑥ [aspirate] key
⑦ 200ulmetering pump
⑧ Sampling assembly
⑨ Fluidic pressure sensor

 Front view of the main unit (cover opened)


166 © 2013 Mindray Confidential
Mechanical system—overall layout

① Analog drive board


② Main control board
③ Fan assembly

 Back view of the main unit (internal structure)


167 © 2013 Mindray Confidential
Mechanical system—overall layout

① Optical system
② Front cover
③ Reagent baffle
④ LH lyse
⑤ DIFF lyse
⑥ Net 、 USB interface
⑦ Power supply assembly

 Left view of the main unit (left


door removed)
168 © 2013 Mindray Confidential
Mechanical system—overall layout

① Fluid inlet/outlet
② Waste pump
③ Vacuum chamber
④ valve
⑤ LVM valve
⑥ Diluent Temp. sensor
⑦ Sampling assembly
⑧ Front cover
⑨ RBC bath
⑩ WBC bath

 Right view of the main unit (right door removed)


169 © 2013 Mindray Confidential
Mechanical system—overall layout

① Bath cap
② Pan head screw M3X6
③ Bath holder
④ Bath adjust holder
⑤ RBC bath
⑥ Aperture (D50um)
⑦ Front bath washer
⑧ RBC back bath

 RBC bath subassembly


170 © 2013 Mindray Confidential
Mechanical system—overall layout

① Bath cap
② Pan head screw M3X6
③ Bath holder
④ Bath adjust holder
⑤ WBC bath
⑥ HGB assembly

 WBC bath assembly

171 © 2013 Mindray Confidential


Mechanical system—disassembling

RBC bath maintenance assembly

172 © 2013 Mindray Confidential


Mechanical system—disassembling

pipes are fixed to the holder


with cable tie

Winding pipe and


adapter should
be tied up with
cable tie
Replace with specific pipe and
surround with winding pipe as
illustration in the figure

173 © 2013 Mindray Confidential


Mechanical system—disassembling

pipes are fixed to the holder


with cable tie

Winding pipe and


adapter should
be tied up with
cable tie
Replace with specific pipe and
surround with winding pipe as
illustration in the figure

174 © 2013 Mindray Confidential


Mechanical system—disassembling

Disassembly rear plate of power supply


assembly

175 © 2013 Mindray Confidential


Mechanical system—adjustment of bath position

176 © 2013 Mindray Confidential


Mechanical system—adjustment of bath position
Quality Control
Operation
Operation content
Procedure QC Point QC Standard
Click on “Service → Sample Probe Debug”. Enter the
1.1
Debug screen as shown in Figure 1.1.
First, click on the “Initial position” so that the sample
1.2 probe is above the RBC bath, as shown in Figure 1.2.
Click on the “WBC bath”(or “RBC bath”) button shown in Fine adjust The sample
Figure 1.2. Click the “Middle position” button. Check if the the relative probe is
sample probe is basically in the center of the bath. If there position of basically in the
1.3 is any deviation, loosen the four screws shown in Figure the bath center of the
1.3 and slightly adjust the bath in left/right and front/back bath
directions so that the sample probe is in the center of the
bath. Then tighten one screw in each direction
Click on the “Initial position” button, then click “WBC Check if The sample
bath”(“RBC bath”) →“Down position” button. Check there the bath probe can not
is no interference with the bath as the sample probe position is contact with the
1.4 enters the bath. If the interference occurs, readjust the successfully bath 。
bath position in accordance with step 1-3 and repeat this adjusted
step to check. Finally, tighten the 4 screws.
Note:
In Figure 1.1, all the buttons except the “Initialization” button will be grey before clicking on Initialization, the
three buttons in the “Sample probe horizontal position” column will be illuminated and available for use Only
after clicking on Initialization. The “Middle position” and “Down position” buttons will not be available until
clicking on the “WBC bath” or “RBC bath” button.

177 © 2013 Mindray Confidential


Preheating temperature calibration and verification

Fill in the sum of the machine


measurement value and the FRU
value

Machine
measurement
value
Quality Control
Operation
Operation Content
Procedure QC Point QC Standard
Click the Start Menu bar, then click
“Service→ Temp.& Press.
Calibration” to enter the
2.1 “Temperature calibration” screen.
Click on the “Preheat bath
temperature calibration” button to
enter the screen shown in Figure 2.1
Record the machine measurement Measure the
value, and enter the sum of the preheat bath
2.2 machine measurement value and the temperature
value on the service FRU as the calibration factor
meter measurement value

178 © 2013 Mindray Confidential


Mechanical system—matters need attention

 All parts of the analyzer are potential infectious, safety protection measures
should be taken during operation and maintenance
 Reagents may irritate eyes, skin and mucous membrane. Operator who
contact such items as reagent in the laboratory, should comply with the
laboratory safety operating rules, and wear personal protective outfit (such as
laboratory protective clothing, gloves, etc.)
 Once the skin is in contact with Reagent, wash immediately with plenty of
water, if necessary, please go to see a doctor; Once the reagent is splashed
into your eyes, rinse immediately with plenty of water, and treated by doctors
when removing some component with electrostatic sensitive sign, electrostatic
ring or static gloves and other protective equipment need to be equipped to
avoid damage to the parts
 Choose the correct specifications and models of tube and screw
 keep the original tube length. Any prolongation and shortening are strictly
forbidden.
 Pay attention that the tube and cable could not bend or squeeze, when reinstall, to
prevent block or short circuit

179 © 2013 Mindray Confidential


Thanks !

180 © 2013 Mindray Confidential