03-Recent and Pending Changes Concerning Airborne Particle Counting ISO 14644-1, Annex 1, ISO 21501
03-Recent and Pending Changes Concerning Airborne Particle Counting ISO 14644-1, Annex 1, ISO 21501
-
Theory
-
Guidelines
-
Monitoring
-
Liquid Particle counters
1
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
• Integration and Automation – Getting It All Together: Industry Trends & Case
Studies
3
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
– ISO 14644-1
• Classification of air cleanliness
– ISO 14644-2
• Specifications for testing and monitoring
to prove continued compliance with ISO
14644-1
– ISO 14644-3
• Guidance on instrumentation to be used
for testing for compliance with ISO 14644-
1999 1
ISO 14644 4
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
5
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
Minimum number of 1
whichever is
locations greatest
with at least 3 samples
total
6
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
ISO 1 10 2
ISO 2 100 24 10 4
7
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
20 particles
Volume = X 1000 liters
3520 particles
8
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
Time required (in minutes) at 1 cfm (28.3 lpm) flow rate with 1-minute limit imposed
9
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
Maximum Time
Test Parameter Class Test Procedure
Interval
Particle Count
≤ ISO 5
All 24
6 Months
ISO 14644-1
Test classes
> ISO 5 12months
Months
Annex A
Important
Air Pressure note: If Particle counts and Differential
ISO 14644-1
All Classes 12 Months
Pressure
Difference are monitored continuously (forAnnex B5
example, by a Facility Monitoring System) then
ISO 14644-1
the Maximum Time Interval for re-certification
Airflow All Classes 12 Months
Annex B4
can be extended
10
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
Vial
Washing Lyo 1
Station 5m
Lyo 2
8m
5m
Calculations for Number of Points:
Area of clean zone = 80 m²
Lyo 3
Take the SQRT (80) = 8.94
Rounding up to next integer = 9 sample positions
4m
11
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
Vial
Washing Lyo 1
Station
1 2 3 4 5 6
Lyo 2
12
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
Vial
Freeze
Washing
Dryer 1
System
Freeze
Dryer 2
• Need to adjust for equipment in room.
• Under ISO 14644-1, if you sample at 10 or more
positions, you can avoid the added calculation of
the UCL (Upper Confidence Limit). Calculation of
the UCL is only mandated when the number of
positions used is between 2 and 9. Freeze
• Best to sample near potential problem spots which Dryer 3
are near entrances and exits and near operator
positions.
13
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
9
1
Vial 10 Freeze
Washing
8 Dryer 1
System
2
11
3 4 5 6 7
12 Freeze
Dryer 2
• Need to adjust for equipment in room.
• Under ISO 14644-1, if you sample at 10 or more 13
positions, you can avoid the added calculation of
the UCL (Upper Confidence Limit). Calculation of
the UCL is only mandated when the number of
positions used is between 2 and 9. Freeze
14
• Best to sample near potential problem spots Dryer 3
which are near entrances and exits and near
operator positions.
14
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
If the number of points sampled is more than 1 but less than 10,
then the UCL factor must be applied:
a) Calculate the standard deviation
b) Use Student’s T-factor from tables
c) Calculate UCL
d) Compare to classification limit; UCL must
not exceed the applicable limit
15
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
Proposed:
• For all zones, every 12 months
16
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
17
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
18
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
20
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
Normative reference
The following normative document contains provisions, which, through reference in
this text, constitute provisions of this part of ISO 14644. For dated references,
subsequent amendments to, or revisions of, any of these publications do not apply.
However, parties to agreements based on this part of ISO 14644 are encouraged to
investigate the possibility of applying the most recent editions of the normative
documents indicated below. For undated references, the latest edition of the
normative document referred to applies. Members of ISO and IEC maintain registers
of currently valid International Standards.
21
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
22
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
23
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
Maximum particle Particles per cubic Particles per cubic Particles per cubic 10% Coincidence loss (calculated)
concentration centimeter centimeter meter (minimum size) On screen calculator
Volumetric flow rate
MFR standard MFR standard
Sample flow rate +/- 5% If unit has flow rate control system
uncertainty uncertainty
Include flow rate uncertainty (all)
Sample time +/- 1% +/- 1% +/- 1% On screen calculator
Sample volume +/- 5% +/- 5% N/A On screen calculator
10 min at concentration limit (minimum size),
Response rate N/A N/A 0.5% Sample (<60 sec) then filter (10 sec), then
sample (< 60 sec)
25
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications
26
Market Leader in
ANATEL TOC Analysis
Global Innovator in
HIAC Fluid Particle Applications