• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
Application:
- Intensive care wards,
ICU
- Recovery rooms and generally for hospital use
- Patient request long term ventilation
- During transport within the hospital
A&E SV300 PACU
Patient Type:
- Adult to Pediatric
Trans
port
Price
-turbine based ventilator -High pressure gas source
-transport capacity (HPGS)
-NIV -support critical ill patient e.g
-excellent cost benefit ratio ARDS
-extensive functions
E5
E3 SV600/800
SV100 SV300
Performance
Blower
Constant speed Dynamic speed
SV 300 C2
Savina 300
V200
Non turbine based
Piston
PB 760
Alarm indicator
AC power &
battery indicator
Control knob
Expiration port
Nebulizer connector
DC power inlet
support up to
HPO (High Pressure O2) inlet
2 E-size
cylinder
LPO (Low Pressure O2) inlet
Monitoring
- Up to 35 monitored patient parameters
- Tve/IBW
- Weaning indicator: P0.1, RSBI, NIF
- 72 hrs trends data
- 5,000 event log
- Data export (Screen short, History, configuration)
Optional feature
- SPO2 monitoring
- ETCO2 monitoring preconfigured (built-in)
- Mainstream with (Volumetric CO2 )
- Side stream
Mobility
Integrated Turbine
- Ultra quiet noise level down to 45 db (a)
Especially good for patient require long term
ventilation
- High performance drive up to 210 L/min
fulfill most of the acute patient needs
- 20,000 hours before maintenance needed
Low maintainance cost
up to 4 waveforms displayed
complete picture of patient condition
Pop-up relevant
parameters value
suggestion
current active
alarms
current alarm setting
current patient value
better
understanding of the
current situation and
hence better
judgment
most recent alarm
(not active right
now) Maximum 9
• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
Tools List
Name Qty
Allen Key Set 1set
Philips Screwdrive 1set
Maintain Passwords
User password 1234
Service password 1118
1 2 3
4 5 6
Test lung
Watertrap
- Water traps should be at the lowest point of the
limb. Humidifier
- Humidifier should be in inspiratory limb.
- Nebulizer should be in the inspiratory limb, best
45cm away from patient.
- Keep the nebulizer vertical.
Inspiratory limb
To
Ventilator patient
Adapter
Thermometer
Adapter
Chamber
Humidifier
082-001927-00
Pressure regulator 14Mpa (for cylinder),
can be ordered from Mindray
Item
Test Purpose Schedule
No.
1. Set O2 supply type
System 2. O2% calibration if O2 sensor is installed After
1
settings 3. For high altitude area, altitude compensation is installation
necessary
1. Check turbine, air/O2/expirtory flow sensor,
After
System pressure sensor, valves and O2 sensor works well.
2 installation /
check 2. Check system leakage 、 compliance and airway
repair / visit
resistance.
1. To confirm ventilator works normally, and no alarm. After
Ventilation
3 2. To confirm the measured pressure or tidal volume installation /
check
are consistent with settings repair / visit
6
4
After altitude is updated,
user flow calibration
should be done.
(P-A/C)
Steps
1. Make sure that the supply pressure is
normal and that the breathing circuit are
correctly connected.
2. Connect a 2L bag as the test lung to the Y
piece in the patient circuit.
Pass Criteria
• The difference between the TVi
reading and set value does not
exceed 10ml + 10% setting value
Example
• If you set TV=600ml.
• Max accepted tolerance will be
10+600*10%=70ml, i.e. acceptable
measured TVi is 530~670ml.
Steps
1. Make sure that the supply pressure is
normal and that the breathing circuit are
correctly connected.
2. Connect a 2L bag as the test lung to the Y
piece in the patient circuit.
6
3. Select P-A/C as the ventilation mode. 6
3
5. Set others to the defaults.
4 4
6. Record the displayed Peak and PEEP
values in each setting stabilized status.
Pass criteria:
Difference between the displayed
Difference between the displayed
Ppeak and the set Pinsp
PEEP value and the set value
(△Pinsp+PEEP) value does not
does not exceed 2cmH2O+5% of
exceed 2cmH2O+5% of the set
the set value.
value.
Example:
− If you set △Pinsp=15cmH2O, PEEP=5cmH2O.
Do not use antistatic or conductive masks or breathing hoses. They can cause
burns if working with high frequency electrosurgical equipment.
Re-do system check each time when accessories or components like hose,
humidifier, and filter are replaced.
Humidifier connector must be lower than the ventilator’s breathing connectors and
the patient.
To reduce the risk of fire, do not use a low-pressure O2 supply that delivers a flow
greater than 15L/min.
• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
Settings
1
1. Patient type
- Adu / Ped
2 - Last patient
5 2. Patient settings
- Gender
- Height / IBW (Ideal
body weight)
- Invasive for intubated
4 patient
- Non-invasive for
patient wearing mask
3
3. Ventilation mode
4. Start Ventilation
5. Other functions
- System Check
- O2 Therapy
46 © 2018 Mindray Confidential
Ventilation Modes
Volume • V-A/C
controlled • V-SIMV
Pressure • P-A/C
controlled • P-SIMV
Spontaneou • CPAP/PSV
s
• PRVC
Advanced • APRV
• Duolevel
During the inspiration phase, when the delivered volume exceeds the
upper alarm limit of tidal volume, the system switches to expiratory
phase immediately and airway pressure is fully released.
End users are most familiar with, and used most often. All of them support
synchronous trigger or spontaneous breath.
V-A/C
• In this mode, mechanical ventilation is synchronous
P-A/C with patient spontaneous effort. It can improve man-
machine coordination, reduce sedative and muscular
relaxants consumption, expedite weaning process.
PEEPi O2 Therapy
---Inspiratory valve is closed, and flow turns to zero. Then there is no airway resistance. At
this moment, Pplat is generated by lung capacity and compliance
Function
1. Optimize parameters with the
measurement of Cstat;
2. Improve ventilation efficiency;
3. Ensure patient safety.
--- Expiratory valve closed. Flow is zero and pressure in the lung and ventilator
circuit equilibrate. The static airway pressure represents total PEEP.
Expiration hold
Suction
Suction
Typical ventilation curves with static P-V Typical static P-V loop
tool function
UIP: Upper
inflection point
There are clinical evidence proven that if the PEEP value is set at
LIP+2cmH2O, it will significant reduce the mortality rate of ARDS patients
62 © 2018 Mindray Confidential
Special Functions
P-V tool
3
1
2
ATRC
---Automatic Tube Resistance Compensation
ATRC algorithm
ΔPtube = Paw - Ptrach=Comp.*Ktube*Flow2
ATRC Settings
1. ATRC type
- Choose type
ET tube; Trach Tube; Disable ATRC
1 2 3
2. Diameter
- Set tube diameter 4
Adult: 5.0-12.0mm; Pediatric :2.5-8.0mm
3. Compensation ratio
- set compensation ratio
range:1%-100% , default setting: 80%
• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
Name Type
Autoclavable
OR
Items Operations
Data export Export log data from service menu via USB stick
System check Check if machine pass all system check items and
record leakage value
Check before
maintenance Ventilation Check if machine pass ventilation test and no alarm
check
Items Operations
System check Check if machine pass all system check items and
record leakage value
Parts replacement
048-004133-00 045-001333-01
(Dustproof net) (HEPA filter assembly)
045-001298-00
(Air-intake dust filter)
Parts replacement
049-000778-00
(Expiration valve membrane)
115-028051-00
(Blower box FRU)
79 © 2018 Mindray Confidential
Maintenance - Turbine Blower