Professional Documents
Culture Documents
Qualification of UV - Visible Spectrophotometer
Qualification of UV - Visible Spectrophotometer
VISIBLE
SPECTROPHOTOMETER
By
Student Name-Gitanjali Sahu
Roll No-21ph204a03
Branch- Pharamaceutical Analysis
Under the guidance of
Dr G.Kiran
CONTENT
1. Qualification
2. Instrumentation of UV Visible
spectrophotometer
3. Level II installation procedure
4. Level III Performance qualification
parameters
5. Level IV in use instrument checks
QUALIFICATION-
It is an act or process to assure something complies with some
conditions, standards or specific conditions.
Types
Design Qualification- document which shows that plant
design agrees with the design specifications of the customer.
Installation Qualification-documented evidence is given that
all parts of equipment are installed according to the
equipment supplier's and purchase specifications.
Operational Qualification-documented evidence which
shows that all parts of plant and equipment work with in their
specifications and process parameters.
Performance Qualification-document evidence that all parts
of plant and other processes produce products of specified
quality under conditions of normal production for a
longer period of time.
INSTRUMENTATION OF UV
VISIBLE
SPECTROPHOTOMETER
LEVEL II INSTALLATION
PROCEDURE
When the UV instrument was shipped after the precise adjustment and inspection at the factory,
it is recommended to install according to procedure to provide optimum performance.
Installation site-
Room temperature during use of 15 to 30°c
Out of direct sunlight
No strong vibration or continuous weak vibration
No strong magnetic fields
No corrosive gases or organic or inorganic gases with absorptivity in the UV region
Small amount of dust
Humidity of 45 to 80%
LEVEL III PERFORMANCE
QUALIFICATION
PARAMETERS
Parameters to be checked
Typical acceptance limits
• Absorbance accuracy
• Photometric linearity
4. Start scan and examine the trace for a peak around 656.1 nm.
5. Measure spectral band width which should be with in ± 10%
6. Check the calibration at a silt width of 0.2nm If silt width is too small –instrument have more photometric
noise.
WAVELENGTH PRECISION
It is defined as deviation of wavelength reading at an absorption band and emission band from the
wavelength band.
Materials -
Holonium perchlorate solution R -40g/L of Holonium oxide in 14g/L solution of perchloric acid.
Mercury lamp
Certified filters
Method - Carry out 6 measurements of absorption maxima.
Limit -
Repeatability - The RSD of absorption maxima should satisfy manufacture specifications.
the difference between 6 individual absorption maxima should be < 0.5nm.
BASELINE NOISE
The flat baseline test shows the ability of instrument to normalize light intensity measurements
and spectral output at different wavelength throughout spectral range.
Method
Make 61 absorbance measurements with an integration time of one second at wavelength of
500nm, with no sample in the sample chamber and calculate the mean.
Acceptance limit is <0.01 AU.
PHOTOMETRIC DRIFT
It should be checked in both UV and Visible region ,at appropriate wavelength.
It is determined by comparing difference between measured absorbance of reference material
and established value.
Method-
As routine test the drift is measured at 250 nm over a period of 2 hours by using the time scan
mode of instrument, with no sample in the chamber.
Limit- ±0.001AU
LEVEL IV IN USE
INSTRUMENT CHECKS
Parameter to be checked Typical acceptance limits