Professional Documents
Culture Documents
Filters and Beam Restrictors.
Filters and Beam Restrictors.
BEAM RESTRICTORS
Moderator-Dr.Bhagyavathi.M.k
Presentor-Dr.Bhagyashree Sajjan
FILTERS
INSULATING OIL
Filters thickness
Two millimeters of aluminum absorb nearly all photons
with energies less than 20kev.
National council on radiation protection and measurement
has recommended the following total filtration for
diagnostic radiology the figures include both inherent and
added filtration
Comprised of lead, attached to the collimator
Ideal geometric configuration
But flare of the cone is larger than flare of xray
beam
Only the base plate that attaches device to tube
housing is part that restricts beam
CYLINDERS
Comprised of lead, attached to the collimator
Beam restriction takes place at the far end of
barrel, so less penumbra
Disadvantage :Limitation of available field sizes
COLLIMATORS
• Best all round x-ray beam restrictor device.
ADVANTAGES:
• Provides an infinite variety of rectangular xray fields.
• Light beam shows the centre and the exact
configuration of the x-ray field.
• Helps to limit penumbra.
• Provides adjustable rectangular or square field.
• Reduce scatter radiation that increase image
contrast.
•It reduces patient dose.
STRUCTURE OF A COLLIMATOR
• Two sets of shutters to control the dimension.
- each shutter contains 4 or more lead plates, which move in
independent pairs.
- when the shutters are closed, they meet at the centre.
• Light beam is from a light bulb in the collimator.
Precautions while using collimator
X ray field is illuminated by light
beam from bulb
Light beam is deflected by mirror
at 45 degree
Target of x-ray tube and bulb
placed at same distance from mirror
and both should coincide at the
second shutter
If too far from x-ray target, x ray
beam will be smaller than light
beam
• A collimator can also identify
the centre of the x-ray field
which is accomplished by
painting a cross line on a thin
sheet of plexiglass mounted on
the end of the collimator.
MATERIALS REQUIRED:
• Four L shaped wires.
• 17 x 14 inch x-ray film
• Lead letter “R”
Procedure
• Place the film on the top of the x-ray table
The dark centre shows the position of the xray beam and
the wires indicate the position of the light beam.
HOW TO ADJUST A MISALIGNED MIRROR?