Installation & Basic Operations: Medcaptain Service Dept
Installation & Basic Operations: Medcaptain Service Dept
www.medcaptain.com 成为医疗健康领域领航者
Contents
Installation Procedure
Unboxing/check/accessories-assembly/
Setting/calibration/training/acceptance/
Upload installation information
Overall Introduction
Parts/functionality introduction
Operation Introduction
Common clinical operations
Q&A
内部资料,请勿外传
Installation Procedure
Training
accessories- Acceptance/up-load
Unboxing/check
assembly installation info
Calibration
内部资料,请勿外传
Installation Procedure - Unboxing/check
Check the out-looking of the overall package, record and report to MedCaptain when damage or
abnormal was found;
Check the configuration list after unboxing, record and report to MedCaptain when the following
was/were found:
① Damage on the main unit & accessories;
② Not matching the configuration list.
内部资料,请勿外传
Installation Procedure – Assembly/settings
Assembly/settings
Install pole clamp / handle and accessories, MP-80 need to combine cabinet and cabinet cover(Pumps
need to be installed inside when connecting to the workstation)
Make sure the power cord is well-connected;
Boot up the product;
Install syringe or consumable on the corresponding pump;
Set the settings properly according to the need of the client;
Check the performance and fill in the checking list.
Checking list
内部资料,请勿外传
Installation Procedure – Calibration
Calibration
Check whether the consumable provided by the client was already built-in.
① If yes, need to verify whether the accuracy is within the allowable range, calibration is
need when the accuracy is out of the range;
② If no, Customize calibration is needed;
Allowable range:
± 5% for infusion pumps
± 2% for syringe pumps
Contact MedCaptain if the accuracy cannot reach the range.
内部资料,请勿外传
Installation Procedure – Training
内部资料,请勿外传
Installation Procedure – Acceptance/
Upload installation information
After the product was successfully delivered, upload the installation information to
MedCaptain for 6 months extend warranty(Standard warranty:18 months).
Link: https://t3jjrct42u.jiandaoyun.com/f/5eb92af35e9ae600063f4876?ext=MEDCAPTAIN
内部资料,请勿外传
Overall Apparance
MP-30--Front view 1 – Syringe clamp
Used to clamp the syringe and identify the syringe
size
2 – Slit
3 – Drive head
Used to secure the syringe plunger and drive the
syringe plunger to move
4 – Clutch
Used to open or close the syringe claw and control
the movement of the drive head.
7 – Shell
8 – Alarm indicator
内部资料,请勿外传
Overall Apparance
MP-30--Back view
1 – Water-proof cover 2 – Battery cover
9 – External inlet 1
Used to connect with drop sensor and barcode scanner
10 – External inlet 2
Used to connect with UI upgrade cable for UI upgrade
11 – External inlet 3
Used to connect with nurse call & connect with drive upgrade
cable for drive upgrade
内部资料,请勿外传
Overall Apparance
MP-60--Front view
1 – Alarm indicator
3 – [HOME] key
4 – [ON/OFF] key
5 – [OPEN] key
Used to open the pump door
内部资料,请勿外传
Overall Apparance
MP-60--Inner view 1 – Lighting Lamp
To provide lighting in a dim environment, so as to
install and check the
infusion tube
2 – Depressor
4 – AC power inlet
5 – Multi-stackable lock
6 – External inlet 1
Used to connect with drop sensor and barcode
scanner
7 – External inlet 2
Used to connect with UI upgrade cable for UI
upgrade
8 – External inlet 3
Used to connect with nurse call & connect with
drive upgrade cable for drive upgrade
9 – Speaker 10 – Shell
内部资料,请勿外传
Overall Apparance
MP-80--Controller
1 – Buzzer
2 – Handle lock
3,4,5 – USB1/USB2/USB3
USB 1 is used to connect with nurse call
USB 2 is used to connect with upgrade cable for
upgrade
1 – Alarm indicator USB 3 is used to connect with barcode scanner
Visual expressions of alarms at three levels(low-level,
mid-level, and high-level alarms) 6 – Combination socket
2 – Handle slot
7 – Handle slot
3 – Pump plug(3)
To connect the USB socket with pump
4 – Combination plug
Used to connect to the combination socket of controller
or cabinet body
6 – IV tube clamp 2
内部资料,请勿外传
Overall Apparance
MP-80--Back view
1 –Combination AC power inlet
2 –Combination button
4 – IV tube clamp 2
6 – Combination inlet
Used to connect to the combination plug
of another cabinet body
7 – IV tube clamp 1
内部资料,请勿外传
Overall Apparance
Accessories
内部资料,请勿外传
Operation introduction-MP-30
内部资料,请勿外传
Operation introduction-MP-60
Door open/Install
Boot up Set Speed Start infusion
consumable
内部资料,请勿外传
Operation introduction
Change brand
Pic_1 Pic_2
内部资料,请勿外传
Operation introduction
Set infusion mode
Choose 【 home 】 -- Choose 【 Infusion Set 】 -- Choose 【 Infusion mode 】 -- Choose proper mode
内部资料,请勿外传
Operation introduction
Bolus
During infusion, choose 【 Bolus 】 -- 【 Bolus VTBI 】【 Bolus Rate 】 -- 【 Bolus Start 】
Choose 【 Bolus Stop 】 to stop Bolus
内部资料,请勿外传
Q&A
Apparance ?
Operation ?
Features ?
Others ?
MP Series
内部资料,请勿外传