2.washer Peristaltic Pump Soak Abnormally-20161104
2.washer Peristaltic Pump Soak Abnormally-20161104
2.washer Peristaltic Pump Soak Abnormally-20161104
Version : 20160930
FOCUS MAKES PROFESSION P1
CONTENTS
01
PART ONE Principle
02
PART TWO Problem & Solution
Possible reasons:
White wire connected with True Message.
injection needle
1.The peristaltic pump is damaged.
Black wire connected with
2.Blockage inside the suction
suction needle needle or the tube connected to
waste pump.
Three clamps on black line are 3.Damage of signal wire.
conducted to one another , same for 4.False alarm.
Injection needle white line But the black and white line
are not conducted to each other.
Suction needle
Step 1:
A. Observe the peristaltic pump
carefully to check it is vibrating or
not. If not, check the mechanical part
of the pump.
B. If the peristaltic pump can actually
vibrate but still cannot aspire the
liquid outside, tighten or loosen the
pump tubes by adjusting the
pressure lever. Especially the part
rubber tube rub with peristaltic
pump rod. Replace the tubes if they
have been abraded.
3. Damage or wrong
connection of signal
wire:
Solution :
A. After you follow the procedures
to maintain washer but still
showed error ”Washer peristaltic
pump soak abnormally”, and
there exist no liquid inside the
channel after priming. Check
whether there remains liquid
drop hanging between injection
needle and suction needle. Liquid remain
between needles Wrong connetion
If so: (1)make sure system liquid is
sufficient;
(2)clean the surface of needles
with 75% alcohol;
(3)prime 5 times;
(4)contact with Snibe engineer if
it still not fixed after maintenance.