Professional Documents
Culture Documents
Quality Control 2023 +CLIA 2024
Quality Control 2023 +CLIA 2024
in Laboratory Medicine
by
Bashira Khalidy, MPH
Outline
Introduction
Consequences of Poor quality
Phases of Laboratory Quality
What is quality control
Objective of quality Control
Basis Statistics
Accuracy/Precision/Trends/Shifts
Random and Systematic Errors
Proficiency testing
Allowable Total Error
Westgard Rules
References
2
Introduction
From the first handling of test specimens and
paperwork, to the reporting of results, quality
should be monitored.
Bashira Khalidy 3
Laboratory Testing
Impacts Nearly Everyone
Bashira Khalidy 4
5
6
7
Phases of Laboratory Quality
1. Pre-Analytical Activities
Bashira Khalidy 8
Pre-Analytical
Where Quality Originates
Bashira Khalidy 9
Analytical Activities
Proceeding with Utmost Care
Instrument maintenance and operation
Test reagents
Supplies
Personnel
Actual test performance (Test method used)
Policies and procedures followed
Internal Quality control procedures
External Quality control /proficiency testing
Bashira Khalidy 10
Post-Analytical Activities
Reporting with Quality
Bashira Khalidy 11
The Quality Assurance Cycle
Patient/Client Prep
Sample Collection
Personnel Competency
Reporting Test Evaluations
Data and
Lab
Management
Safety
Customer Sample Receipt
Service and Accessioning
Record
Keeping
Sample Transport
Quality Control
Testing
Bashira Khalidy 12
Which Errors are Worse?
Bashira Khalidy 15
Analytical Phase
Reagents
Bashira Khalidy 16
Reagents Cont’
Bashira Khalidy 17
Analytical Phase Cont’
Bashira Khalidy 18
Analytical Phase Cont’
Daily QC is a process that is established to detect any
systematic errors that may interfere with reporting
patient samples.
Bashira Khalidy 19
What is Quality Control?
20
What is Quality Control?Cont’
QC results are used to validate whether the
instrument is operating within pre-defined
specifications, inferring that patient test results are
reliable.
21
What is Quality Control? Cont’
Good laboratory practice requires testing normal and
abnormal controls for each test at least daily to
monitor the analytical process.
22
Objective of Quality Control
23
Understanding Quality control
In order to better understand quality control one must
become familiar with certain statistical terms,
concepts, and calculations.
Bashira Khalidy 24
Basic Statistics
Mean: sum of data
# of data points
e.g. Calcium levels: mg/dl
1. 9.2
2. 9.3
3. 9.1
4. 9.2
5. 9.3
6. 9.9
Bashira Khalidy 25
Basic Statistics Cont’
Mean & SD
Bashira Khalidy 26
Standard Deviation
27
Basic Statistics Cont’
SD
SD for calcium levels:
√ (9.2 – 9.2) 2 + (9.3 – 9.2)2 + ….
6 -1
= 0.2 mg/dl
SD can compare a set of data measured in the
same units.
The SD often increases as the concentration
increases, therefore it is often useful to express the
SD as a percentage of the mean concentration.
Bashira Khalidy 28
Basic Statistics Cont’
SD
SD: Most laboratories use the acceptable parameter of
a 2 SD range, which means that 5 % of values will fall
outside this range
Bashira Khalidy 29
Normal Distribution
30
Basic Statistics Cont’
SD
Bashira Khalidy 31
Basic Statistics Cont’
Accuracy (Bias)
Bashira Khalidy 32
33
Bias
34
Precision
Precision tests the reproducibility of an assay method.
Bashira Khalidy 35
Accuracy & Precision
Bashira Khalidy 36
Accuracy and Precision
37
Basic Statistics
Coefficient of Variation (CV)
Can compare variability between different units or
different populations: Kg,m ; male, female
Because the CV reflects a ratio of the SD to the
concentration, it often provides a better estimate
of method performance over a range of
concentration
= SD * 100
mean
CV (calcium levels) = 0.2 mg/dl * 100
9.2 mg/dl
= 2.17
Bashira Khalidy 38
Coefficient of Variation (CV)
The smaller the CV,
the more the values tend to cluster around the mean
and the more reproducible or reliable is the result
e.g.:
Cholesterol levels
Set A (10 values): Mean: 145, SD: 10
Set B (10 values): Mean: 80, SD: 10
Bashira Khalidy 39
Coefficient of Variation (CV)
Bashira Khalidy 40
Types of Analytical Errors
Systemati
c Errors
Random
Errors
41
42
43
Random Error
Bashira Khalidy 44
45
Systematic Errors
Systematic error, or inaccuracy is
an error that is always in one
direction.
Bashira Khalidy 46
Causes of Systematic Errors
A. change in reagent or calibrator lot
B. wrong calibrator values
C. improperly prepared reagents
D. deterioration of reagents, calibrators, or
photometric light source
E. inadequate storage of reagents or calibrators
F. change in sample or reagent volumes due to
pipette maladjustments
G. change in temperature of incubators and
reaction blocks
H. change in procedure from one operator to
another.
Bashira Khalidy 47
48
Inaccuracy: Trend & Shift
Control results should be distributed on both sides of the
mean value.
Bashira Khalidy 49
Trend
A trend is said to occur when there is a consistent
decrease or increase in control values.
50
Trend
Trends may be caused by:
Deterioration of the instrument light source
Gradual accumulation of debris in sample/reagent
tubing
Gradual accumulation of debris on electrode surfaces
Aging of reagents
Gradual deterioration of control or calibration
materials
Gradual deterioration of incubation chamber
temperature
51
Shift
A shift is found when 5 successive values fall above
or below the mean.
52
Shift
Shifts may be caused by:
Sudden failure or change in the light source
Change of reagent lot
Major instrument maintenance
Sudden change in incubation temperature
Change in room temperature or humidity
Failure in the sampling system
Failure in reagent dispense system
Inaccurate calibration/recalibration
53
Perfect control
No trends
No shifts
Bashira Khalidy 54
Shift
Trend
Bashira Khalidy 55
56
57
58
59
60
61
Proficiency Testing
(EQC)
62
63
Proficiency Testing Cont’
64
Proficiency Testing Cont’
The proficiency agency provides a summary
report that contains summary data of all the
participating laboratories along with an accuracy
grading report.
65
EQC normal distribution charts
66
EQC normal distribution charts
67
EQC Histogram
68
How do proficiency testing schemes evaluate the accuracy of individual laboratories?
‘z-score’ (SDI)
This score reflects the actual accuracy achieved (i.e.,
the difference between the participant’s result and the
accepted true value.
69
Z- Score
Measures your Lab accuracy in comparison with the
other participating Labs
Z score = xi – μ
s
70
How should z-scores be interpreted?
71
Standard Deviation Index [SDI]
72
SDI acceptance criteria
The target SDI is 0.0 which indicates a perfect
comparison with the peer group.
The following guidelines may be used with SDI:
•• 1.25 or less is considered acceptable.
•• 1.25 – 1.49 is considered acceptable to marginal
performance.
•• 1.5 – 1.99 is considered marginal performance
•• 2.0 or greater is generally considered to be
unacceptable performance
73
74
75
76
77
78
79
WHY Proficiency Testing???
Help evaluate existing test methods by comparison to
other peer labs.
Help to assure accurate test analysis under laboratory
operating conditions.
Help complement test kit acceptance criteria
Provide added confidence
Comply with regulatory guidelines
Test competency of personnel who are performing the
analysis.
80
List of International Proficiency Agencies (in our Lab)
85
Allowable Total Error Cont’
Allowable Total Error (TEa): states the laboratory’s policy for
how much error is medically acceptable. Regulatory
requirement (e.g. CLIA’88) represent an upper limit.
e.g.: SGPT : ± 20%
Amylase: ± 30%
Calcium: ± 1.0 mg/dL
Hemoglobin: ± 7%
Platelets: ± 25%
WBC: ± 15%
86
87
88
89
90
91
92
CLIA TEa: old and new (2024)
New Old
93
94
How to measure Sigma for analytes
95
96
97
98
99
100
101
102
Sten Westgard
103
104
Westgard QC Rules
First, a non-technical description.
When my daughter Kristin was young and living at
home, she liked to party.
One day when she told me she was again intending to
be out late, I felt the need to exert some parental
control over her hours.
So I told her that if she was out once after three,
twice after two, or four times after one, she was in
big trouble. That's multirule control.
105
Westgard Multirule QC
Multirule QC uses a combination of decision criteria, or control
rules, to decide whether an analytical run is in-control or out-
of-control.
107
108
Westgard QC Rules Cont’
Some alternative control rules are more suitable when
three control materials are analyzed.
109
110
111
112
113
Westgard Rule: 10X
114
115
116
Westgard rules applications
If one measurement is made on each of two different control materials in an
analytical run, control rules can be applied as follows:
The two control results "within a run" can be inspected by applying a 13s rule to
each material, as well as the 22s and R4s rules "across materials."
The 22s rule can also be applied to the last two measurements "within a
material and across runs."
The 41s rule can be applied to the two control measurements in the current
run and the two measurements in the previous run, i.e., the rule can be
applied "across materials and across runs".
The 41s rule can also be applied to the last four measurements "within a
material and across runs," which now requires the control results from the
three previous runs.
The 10x rule can be applied to both control measurement in a run for the last
five runs, or to the measurements on just one material for the last ten runs.
117
118
119
Westgard Rules Cont’
Random (sudden, unpredictable)
1-3S, R4-S
120
121
Every lab result might include an error.
Our job is to limit that error to a tolerable amount that is not clinically
significant for patient management.
122
Quality is everyone’s job
If it’s not documented, it’s not done
Say what you do; do what you say (Walk the Talk)
Quality first, production second
Inaccurate results are never timely
Do things right the first time
125