Download as ppt, pdf, or txt
Download as ppt, pdf, or txt
You are on page 1of 39

System Overview

01-OCT-09 CONFIDENTIAL
The Story of Access

 1967 - Kallestad Diagnostics


 Focused on high quality immunoassay reagents
 Entered instrumentation market in 1987
 1989 - Purchased by Sanofi Diagnostics Pasteur
 Acquired Genetic Systems Corp in 1990
 1997 - Access product line (Chaska site) purchased by
Beckman Instruments
 Hybritech technology becomes available
 Cooperative partnership with BioRad remains
 1997 - Beckman acquires Coulter
 Immunotech joins immunodiagnostics group

22-APR-10 CONFIDENTIAL
The Evolution of the Access Family

UniCel DxI®
800

LXi®

“Classic” Access®

Access® 2

2004

1990 1991 1992


1993 1993 1994 1995 1996 1997 1998 1999 2001
2000 2001
2002 2002
2003 2003
1990
22-APR-10 CONFIDENTIAL
Immuno Platform Strategy

A technology continuum DxC 660i


Access 2 and identical
DxC 600ireagents across
multiple testing platforms guarantees
equivalent patient results
Universal
CTA

DxI 800 DxC 880i


DxI 600
2006 2007 - 2008

22-APR-10 CONFIDENTIAL
What are Reagents?

Individual
calibrators
for PSA
Calibrator
kit
PSA
Reagent Pack
PSA
Calibrator
card
(comes with
calibrator kit)

22-APR-10 CONFIDENTIAL
The Reagent Pack

 The Access Reagent Pack is


a 5 compartment pack with
an elastomeric seal
 Average on board stability
of 28 days
 Reagent usage tracked by
software:
 Pre-run warning for supply
depletion or expiration
 Superior Calibration Curve
stability

22-APR-10 CONFIDENTIAL
Reagent Pack
Well 0 Well 1 Well 2 Well 3 Well 4

Magnetic Miscellaneous Chemicals Alkaline


Particles Phosphatase

22-APR-10 CONFIDENTIAL
Calibrators
Calibrators are tested like samples, except we already know their
concentration

22-APR-10 CONFIDENTIAL
Generic Reagents
 Used for all assays (not assay-specific)
 Located on fluidics module to the left of the system

Substrate Wash Buffer

22-APR-10 CONFIDENTIAL
Substrate

 Added to all reaction vessels

 Reacts with a chemical from the


reagent pack to produce light

 Light is measured by the luminometer

22-APR-10 CONFIDENTIAL
Wash Buffer

 Multiple Uses
 Rinses away unwanted reaction components
 Cleans system

22-APR-10 CONFIDENTIAL
Good Instrument Maintenance

22-APR-10 CONFIDENTIAL
Instrument Maintenance

Weekly Probe Clean Daily Probe Wash

Preventive Maintenance (6
Weekly System Check
months)

22-APR-10 CONFIDENTIAL
Access Process Flow

Supplies loaded
Samples loaded
Pipettor transfers
fluids to reaction
vessel
Vessel incubates
Vessel is washed
Vessel incubates
and is read
Vessel is sent to
waste

22-APR-10 CONFIDENTIAL
Modular Instrument
 Processing Functions
 Carousel Module
 Main Pipettor Module
 Analytical Module
 Fluidics Module
 Control Functions
 Electronics Module
 Software Module
 Interface Functions
 Base Plate Module
 Peripheral Module

22-APR-10 CONFIDENTIAL
Modular Components

22-APR-10 CONFIDENTIAL
Carousel Module

Reagent Carousel

Internal bar code reader

Tube detector

Sample carousel

22-APR-10 CONFIDENTIAL
Carousel Module

 On Board capacity of 24
Reagent Packs
 Reagent packs are
cooled to 2-10°C
 6 Sample Racks/Trays
of 10 samples each with
continuous load and
continuous STAT load

22-APR-10 CONFIDENTIAL
Carousel Module

 The carousel holds 6


sample racks/trays -
The Sample Rack/Tray 10 samples each
 Primary bar coded
tubes loaded through
host query or
broadcast download
 Continuous Load
 Continuous STAT
access

22-APR-10 CONFIDENTIAL
Main Pipettor Module

Pipettor gantry

Ultrasonic transducer

Primary probe

22-APR-10 CONFIDENTIAL
Main Pipettor Module

 Titanium Primary
Pipettor with multi-
tasking capabilities
 Aspirate, Dispense,
Level Sense, and Mix
 Ultrasonic mixing of
Paramagnetic
Particles and
Reactants
22-APR-10 CONFIDENTIAL
Analytical Module

Wash arm
Luminometer
Wash carousel
Shuttle

Waste chute

Rake

Incubator belt RV storage area

22-APR-10 CONFIDENTIAL
Analytical Module

Immunoincubation
 Asynchronous
Incubator Belt:
 Accommodates
multiple assay
formats with no
interruption to
workflow
 Facilitates a smooth
flow of routine and
STAT samples
22-APR-10 CONFIDENTIAL
Analytical Module

Wash/Read
 Paramagnetic
Particle Separation
 3 Washes prior to
substrate addition
 Continuous Glow
Chemiluminescence

22-APR-10 CONFIDENTIAL
Analytical Module

 Wash arm
Dispense probes
 3 dispense probes
 3 aspirate probes
 Substrate probe
 Wash carousel
 Moves RVs through wash
process, substrate incubation
and signal reading

Aspirate probes

Substrate probe

22-APR-10 CONFIDENTIAL
Fluidics Module
Precision pump

Vacuum pump Vacuum jar

Precision valve

Wash valve
Wash pump

Wash tower

Peristaltic pump Substrate pump


Substrate heater

22-APR-10 CONFIDENTIAL
Electronics Module
Power supply
Card cage

Hard disc drive

Reset button

Floppy drive

22-APR-10 CONFIDENTIAL
Chassis Base Module

Top cover

Carousel door
Front panel

Supply cover

Base panel

22-APR-10 CONFIDENTIAL
Peripherals

 Access  Access 2
 Monitor  Flat panel touch screen
– Monitor arm – Monitor arm
 Keyboard  Keyboard
 Printer  Printer
 Bar code wand  Bar code wand
plus
 External computer
 Mouse
 Hub
22-APR-10 CONFIDENTIAL
External Bar Code Reader

 Used when configuring


 New lots of calibrator

 Used when loading


 Reagent packs
 Substrate (Access 2 only)
 Optional for racks and patient samples

22-APR-10 CONFIDENTIAL
Access Feature Summary

 Onboard capacity
• 60 samples
• 24 on-board assays/1200 tests
• 3 hours on-board supplies/consumables
 Continuous access
• Liquid reagent, 24 hour refrigeration
• Can load samples and most supplies while
running
• QC - 2 levels, once/day

22-APR-10 CONFIDENTIAL
Access Feature Summary (cont’)

 Clot detection (on Access 2)


 Throughput
 100 tests/hour (1-step assays)
 50 tests/hour (2-step assays)
 Same Reagents for
 Access/Access2
 LXi
 DxI

22-APR-10 CONFIDENTIAL
Access 2 WorkGroups
CLIENT SYSTEMS
 Access 2 Workgroups
 Up to four systems can
be networked together

SERVER SYSTEM

22-APR-10 CONFIDENTIAL
Server/Client Communication*
Server PC

DB
 Server PC supports
HD its AU and all clients
Server AU
UI by providing a
UI location for the DB

 Communication flows
UI between AU and DB
Client AU HD through UI

Client PC
*Note this is not how the cabling is set up
22-APR-10 CONFIDENTIAL
Access 2 Glossary
 NT (Microsoft Windows Operating System)
XP (for new systems as of Feb 20100

 RTOS (OS9000 Real Time Operating System)

 UI (user interface)

 AU (analytical unit)
 Access 2 instrument without PC and monitor
 HD with OS9000 RTOS
 Has no UI or database

 PC (console)
 Access 2 PC
 Has UI (and may have database if a server)

22-APR-10 CONFIDENTIAL
Access 2 Glossary
(continued)
 Ethernet card
 Network PCB in Access 2 (NIC)
 Hub
 Multi-port connection box for Ethernet (10Base-T)
 Crossover cable (single system cable)
 Used in older systems to connect PC to AU for single system
configuration (yellow cable), and troubleshooting
 Network cable
 Used to connect PC and AU to Hub (blue = 14 feet, gray = 50
feet)
– Required for workgroup
– Provided, and required for new single system installs,
eliminating need to have PC powered on before AU

22-APR-10 CONFIDENTIAL
System
Status
ACCESS 2 MAIN MENU Buttons

Status
Mode

Help

Completion time clock on all screens Supplies Required


when in Running and Paused modes
(includes utility assay) Work Pending

System clock
Status
on all screens
Line
401422

22-APR-10 CONFIDENTIAL
Interlock Switch
 Location
 Under front panel hinge on
right upper corner of system
 Engaged by flag on hinge
when panel is closed
 Function
 Cuts off power to pipettor
gantry when front panel is
opened
 Purpose of cover
 Safety
 Keeps out room light
 Keeps instrument interior free from dust

22-APR-10 CONFIDENTIAL
Lab Exercise
“Getting Started”

 Objectives
 Use safe practices, including the use of PPE
(personal protective equipment) and the
interlock switch
 Observe an assay being processed through
the system with all the covers open

22-APR-10 CONFIDENTIAL