Download as ppt, pdf, or txt
Download as ppt, pdf, or txt
You are on page 1of 35

S/5 Aespire Anesthesia Machine

ABS Troubleshooting
Aespire Special Tools
Leak Test Tool 1407-7013-000 (1a thru f)

1c
1a
1d
1d

1b
1e-2ea 1f-2ea

Positive Pressure Leak Test Tool 1009-3119-000


Plug, Stopper
2 2900-0001-000
3

Vaporizer Manifold Test Tool 1006-3967-000

Aespire TRM page 8-4


7100 Special Tools
7100 Flow Orifice 1504-3016-000

Airway Pressure Sensing Tee 1504-3011-000

7100 TRM page 8-2


Isolating a Leak – Bag or Vent Mode?

• Perform a Low Pressure Leak Test for the machine.


• Breathing System Leak Test Aespire TRM p. 7-5 and 7-6
• - Check Valves
• - Ventilator Bellows
• - Service Mode Tests (Vent Mode)
• - Bag Circuit (Bag Mode)
• - APL Valve

• Based on the findings of the Bag Mode and Vent Mode tests, proceed to the
Flowcharts to isolate the leak.
Flowchart 1
TRM page 7-7
Flowchart 2
TRM page 7-8
Flowchart 3
TRM page 7-9 Pass
Flowchart 4
TRM page 7-10
Fail Pass
Flowchart 5
TRM page 7-11
Test 1: Integrity of Test Tool
Notice the direction of flow through the tool

Front View
Bulb Thru Port Fresh Gas Port
Pressure Sense Port

Alignment Post

Back View

Bulb Thru Port

Alignment Post
Test 2: Machine Low-Pressure Leak Test

Note the position of the adapter


Test 2 (cont)
Vacuum

X X

Circle or ACGO
Selector Switch
Test 3: Gauge, Port 1 & Port 3 U-Cup Seals

Note the position of the adapter


Test 3 (cont)
Fresh Gas Flow

Test Tool
Circle or ACGO
Selector Switch
Test 4: Bag Port, APL, B/V Switch & Neg.
Pressure Relief Valve
Test 4 (cont)
Fresh Gas Flow

Test Tool

Circle or ACGO
Selector Switch
Test 5: APL Diaphragm
Test 5 (cont)
Fresh Gas Flow

Test Tool
Circle or ACGO
Selector Switch
Test 6: Bellows Module & B/V Switch
Test 6 (cont)
Enter Service Mode, Breathing
System Leak Test

Peep Drive Gas


Fresh Gas Flow
Gas Trapped in
Bellows Housing

Test Tool
Circle or ACGO
Selector Switch
Test 7: Bellows & Pop-off, Bellows Base
Manifold and B/V Switch
Test 7 (cont)
Fresh Gas Flow
Gas Trapped in
Bellows Housing

Circle or ACGO
Selector Switch
Test 8: Bellows Assembly
Test 8 (cont)
Fresh Gas Flow
Gas Trapped in
Bellows Housing

Test Tool
Circle or ACGO
Selector Switch
Test 9: Flow Sensor and Circuit Modules &
Absorbent Canister

Plug
Test 9 (cont)
Fresh Gas Flow
Enter Service Mode,
Display A/D Channels:
Insp & Exp Actual Values To
- Should be Near Zero VMB

To VMB

Circle or ACGO
Selector Switch
Test 10: Circuit Module & Canister
Test 10 (cont)
Fresh Gas Flow

Circle or ACGO
Selector Switch
Test 11: Circuit Module
Test 11 (cont)
Fresh Gas Flow

Circle or ACGO
Selector Switch
Test 12: Inspiratory Limb of Circuit Module

Remove Plug Insp Port


Canister

Loop
Test 12 (cont)
Fresh Gas Flow

Circle or ACGO
Selector Switch
Test 13: Negative Pressure Relief
Fresh Gas Flow

Test Tool
Test 13: Negative Pressure Relief
Plug

Plug
Test 14 & Test 15
Test 14 Test 15

These tests are similar to the current tests on the Aestiva


Refer to the Service Manual.