EVIDENCE

RECOMMENDED SAFETY PRACTICE


BASED:
FOR ELECTROSURGERY

Presented
Muhammad Hidayat Faisal
Product specialist and
National Service
Coordinator
Elec t r i cal
Basic Theory

• Selalu mencari hambatan yang paling


kecil

• Dan selalu mencari “grounding”


Komponen
Kelistrikan

Sumber
Tegangan

Sirkuit
Impedansi /
Hambatan

Arus
listrik
BASIC PRINCIPLES OF ELECTRICITY IN
SURGERY
PRO PERTIES O F ELECTRICITY
Properties1
Current
Flow of electrons during a period of
time, measured in amperes

Circuit
Pathway for the uninterrupted flow of
electrons (must be complete/closed to flow)

Impedance
Resistance to the flow of current
measured in ohms

Voltage
Force pushing current through the Electricity
resistance, measured in volts seeks the path
of
least resistance

4 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
Elec t rosur ger y
Basic Theory

Arus
listrik

Sumber Sirkuit
Tegangan

Impedansi /
Hambatan
Konsentrasi
Arus istrik
Pri nsi p Elec t rosur ger y
Monopolar
Pri nsi p Elec t rosur ger y
Bipolar
FREQUENCY
SPECTRUM
NO SHO CK
TO PATIENT

11 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
Pri nsi p Elec t rosur ger y
Panjang Gelombang
Elec t rosur ger y
Perkembangan Electrosurgery

2015
1926 RF 1968 – 1999 Tissue 2006 Tissue
1981 – REM 1995 IRT TISSUEFEC
ESU Solid state fusion Sensing
T
MONOPOLAR ELECTROSURGERY BASICS
VARIABLES THAT IMPACT
ELECTRO SURGERY

ELECTRODE WAVEFORMS GENERATOR POWER TIM THE


TYPE CAPABILITY SETTING E PATIEN
S T

14 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
MONOPOLAR ELECTROSURGERY
BASICS
ELECTRODE IMPACT O F CURRENT CO NCENTRATIO N
TYPE

15 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
MONOPOLAR ELECTROSURGERY
BASICS
WAVEFORMS WAVEFORMS

Lower Higher
Voltage Voltage

16 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
MONOPOLAR ELECTROSURGERY
BASICS
WAVEFORMS THERMAL SPREAD

RW
1

PURE CUT BLEND COAG

LOW HIGH

Thermal
Spread/Charring
Voltage

17 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019 US181098-[WF2441443]
Slide 17

RW1 Todd asked about the citation. These animations were created for us and come directly from approved document
MTS10309. These are straightforward concepts so it's my belief that no citation is necessary here Riseley, Will, 4/10/2019
MONOPOLAR ELECTROSURGERY BASICS
ACTIVATIO N O N/O FF TIMES 3
TIME

 Extended activations of the active


electrode should be avoided
 Per Valleylab™ patient return electrode
IFUs, a 25% duty cycle should be adhered
to (30 seconds off for every 10 seconds on)

18 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
MONOPOLAR ELECTROSURGERY
BASICS
GENERATOR
ADDITIO NAL MO DALITIES
CAPABILITY

Cut Electrosurgical
Modalities Technique

Coag
Modalities

19 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
MONOPOLAR ELECTROSURGERY BASICS
GENERATO R CAPABILITY – TISSUEFECT ™
GENERATOR
TECHNO LO GY
CAPABILITY

Monitors changes in tissue


impedance 434,000
times/sec 4

Adjusts energy output to deliver


appropriate amount of energy for desired
tissue effect 4

Results in enhanced (greater consistency and


efficiency) performance of monopolar
† As compared with ForceTriad™ Energy
devices4,5†
Platform

20 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
MONOPOLAR ELECTROSURGERY
BASICS
INFLUENCE O F THE
THE
PATIENT PATIENT

Each patient on the OR table is unique. It is important to be aware of how different body types may
impact the output and efficiency of surgical energy

21 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
§

INJURY
PREVENTION
40,000 burns per year
caused by Electrosurgery
7,†

$600MM/YR
22
LIABILITY 7,† .
© 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]

† In the United States, data from 1999 study


Electrosurgery
Sistem Grounding

“Menjawab semua kebutuhan di masa


depan. . .”
Cushing, 1927

Namun ...

Generator sistem grounding selalu mencari grounding


terdekat
Electrosurgery
Sistem Grounding

Generator sistem grounding tidak direkomendasikan untuk


digunakan
ECRI, 1995
Hindari
Cidera yang bisa terjadi dengan sistem grounding

Penggunaan generator dengan


sistem tersiolasi bisa
menghilangkan resiko cidera
AORN Recommended Practices 2008
Electrosurgery
Sistem Terisolasi

Power
Outlet
Hindari
Cidera yang bisa terjadi dengan sistem grounding

Sejarah mencatat, cedera yang


sering terjadi akibat pemakaian
ESU ada pada lokasi penempatan
“patient pad”

Penggunaan patient pad yang


memiliki teknologi return
electrode Contact Quality
Monitoring, dapat
meminimalisasi resiko cidera
pada pasien

AORN Recommended Practices 2008


Pri nsi p Elec t rosur ger y
Contact Quality Monitoring

Therapeutic
current

Interrogation
current
Ikuti Petunjuk Pemakaian dari Pabrik
SOP

• Generator dan
accessories harus
digunakan sesuai
petunjuk pemakaian.
• Setiap tipe generator
mempunyai petunjuk
pemakaian yang
berbeda
• Accessories harus
digunakan sesuai petunjuk
pemakaian
AORN Recommended Practices 2008
Elec t rosur ger y
Petunjuk Keselamatan

Bipolar Monopolla
r
Pri nsi p Elec t rosur ger y
Memilih Patient Pad
• Patien plate harus kompatibel dengan
generatornya.
– Jangan menggunakan patient plate yang
mematikan/melumpuhkan sistem contact
quality monitoring
(Generator valleylab mempunyai sistem
pengamanan, yang hanya berfungsi bila
menggunakan patient plate merk valeylab)
• Periksa Apakah:
– Patient plate masih utuh
– Conductive gel masih ada dan lembab
– Expire date masih jauh
• Patient Plate tidak boleh di reuse (1 kali pakai)
INJURY PREVENTION
SELECTING THE RIGHT
PRE Adult
Select patient electrode
according to patient
>30
weight
lbs
Infant
6– 30
lbs
Neonatal
1– 6 lbs

34 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
Elec t rosur ger y
REM Polyhesive ™II
Single Use Patient Return Electrodes
Technical Specs
NEONATAL INFANT ADULT ADULT
REM Polyhesive ™ REM Polyhesive ™ REM Polyhesive ™ NON-REM
Polyhesive ™

E7512 E7510-25 / E7510- E7507 / E7508 / E7506


Product Code 25DB E7509
Overall Dimensions 5,3 x 9,8 cm 9,5 x 12,1 cm 18,3 x 11,4 cm 18,3 x 11,4 cm
Conductive Surface Area 33 cm2 75 cm2 137 cm2 137 cm2
1~6lbs 6~30lbs
Weight Limits (0.45~2.7kg) (2.7~13.5kg)
0,71cm Including 2nd 0,95cm Including 2nd
liner over border when liner over border when
fluid seal is not fluid seal is not 1,11 cm 1,11cm
Border Adhesive required required

Backing Material Polyethylene foam


Cord Withstands 4000v (RMS) at 60Hz

35 |
Pri nsi p Elec t rosur ger y
Memilih Area Pemasangan Patient Pad

• Ikuti Petunjuk pemakaian dari Pabrik


• Sebaiknya penempatan patien plate:
– Kulit yang Bersih & kering ( non
Steril )
– Daerah yang luas, masa otot yang
bagus
– Dekat dengan area operasi
– Daerah yang Cembung
INJURY PREVENTION
PREVENTING PAD SITE
BURNS
Patient Return
Placement
Electrode

The PRE should be placed as close


to
the surgical site as possible while
adhering to other placement
guidelines found in product IFUs
and per accepted hospital
protocols3

37 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
INJURY PREVENTION
PREVENTING PAD SITE
BURNS
CAUTION: DO NOT 3

 Warm
 Cut or modify
 Overlap edges
 Apply after positioning
 Put sustained pressure
on pad site
 Prep site with
flammable
solutions
 Allow liquid to get
between pad and patient
*Where necessary, please remove hair per hospital guidelines. Razor in animation is for demonstration/representation purposes
only

© 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
INJURY PREVENTION
PREVENTING PAD SITE
BURNS
CAUTION:
AVOID 10

40 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]
Pri nsi p Elec t rosur ger y
Hindari Pemasangan Patient Pad Pada

Daerah yang ada protesa /


Daerah
Impal bertulang

Daerah Bekas Daerah Berambut


Luka
Ball,
2004
Pri nsi p Elec t rosur ger y
Persiapan Pemasangan Patient Pad

Ikuti Petunjuk Pemakaian Lindungi Patient Jangan gunakan cairan yang


untuk memotong, plate dari mudah terbakar
membersihkan dan masukanya cairan
mengeringkan area

AORN Recommended Practices


2008
Pri nsi p Elec t rosur ger y
Petunjuk Pemasangan Patient Pad

Jangan Jangan di potong atau di


Overlap modifikasi

AORN Recommended Practices


2008
Pri nsi p Elec t rosur ger y
Petunjuk Pemasangan Patient Pad

• Pilih patient plate sesuai dengan berat pasien


• Infant patient plate
– Power jangan melebihi 120 watts
– Gunakan aktivasi yang pendek-pendek

• Neonatal patient plate


– Hindari Aktivasi lebih dari 30 detik
– Untuk neonates < 0.45 kg,
gunakan bipolar

AORN Recommended Practices


2008
Pri nsi p Elec t rosur ger y
Petunjuk Pemasangan Patient Pad

• Patient plate jangan dihangatkan


• Pasang Patient plate setelah pasien diposisikan
• Letakan Patient plate dekat ke
tourniquet, Sebaiknya pada sisi tempat
operasi

AORN Recommended Practices


2008 Hotline News, 2003
Pri nsi p Elec t rosur ger y
Petunjuk Pemasangan 2 buah Patient Pad

• Pasien obesitas
• Pasien yang terlalu kurus
• Dan Prosedur yang
membutuhkan waktu lama

AORN Recommended Practices


2008
Pri nsi p Elec t rosur ger y
Petunjuk Penyimpanan Elektroda Aktif

Elektroda aktif harus dimasukkan


ke dalam sarung non-konduktif
(holster) pada saat tidak digunakan

AORN Recommended Practices


2008
Pri nsi p Elec t rosur ger y
Elektroda Aktif

Jangan gunakan karet merah atau material


lainnya sebagai pelindung elektroda aktif
– Karet atau plastik mudah terbakar
– Selalu gunakan tip terinsulasi yang
direkomendasikan pabrik
Pri nsi p Elec t rosur ger y
Kebocoran Arus

• Kabel kabel jangan di lilitkan ke


instrument
• Kabel pencil ,kabel patient pad dan
kabel dari alat alat lain jangan di
satukan(bundel) bersama sama.
Pri nsi p Elec t rosur ger y
Petunjuk Pembakaran Hemostat

– Jangan hemostat” dengan elctrode


bentuk jarum
– Jangan bersandar pada pasien, meja,
atau retraktor saat aktivasi
– Lebih baik pakai cut dari pada coag
– Pegang dengan benar sebelum aktifasi
– Pencil dibawah tangan jangan diatas
tangan
– Tempelkan pencil dulu baru aktivasi

Perhatian :
Sarung tangan tidak dapat menginsulasi arus
RF
Pri nsi p Elec t rosur ger y
Pasien dengan alat pacu jantung

• Konsultasikan dengan pabrikan alat


pacu jantung
• Gunakan bipolar
• Gunakan seting energi yang rendah
• Hindari arus melalui jantung dan
alat pacu jantung
• Jauhkan kabel dari alat pacu
jantung

AORN Recommended Practices


2008
Pri nsi p Elec t rosur ger y
Pasien Dengan Perhiasan

• Pastikan pasien sudah tidak


mengenakan perhiasan/logam pada
tubuhnya
AORN Recommended Practices
2008
Pri nsi p Elec t rosur ger y
Pasien dengan tubuh penuh tato

• Hidari penempatan patient plate pada tatto


• Tinta warna merah mengandung besi
bisa menimbulkan panas atau
konduktor listrik

Valleylab Clinical
Hotline
Pri nsi p Elec t rosur ger y
SOP Penggunanaan ESU

1. Nyalakan generator Pastikan alat bekerja dengan baik (Sebelum Patien masuk OK)
2. Posisikan pasien sesuai tindakan yg akan dilakukan
3. Pasang Patien pad, ke patien dan sambungkan ke alat
4. Pasang pincils ke alat dan tempatkan di tempat yg aman.
5. Setelah user siap untuk insisi setting power sesuai kebutuhan atau sesuai
yang di recomendasikan.
6. Jika di perlukan tambah atau kurangi powe sesuai permintaan user
7. Setelah oprasi selesai matikan generator terlebih dahulu dan cabut pincils dan patien
pad pada alat

Valleylab Clinical
Hotline
Pri nsi p Elec t rosur ger y
Surgical Smoke

Gas yang diproduksi dari hasil vaporisasi jaringan protein dan


lemak
Pri nsi p Elec t rosur ger y
Surgical Smoke
xylen
benzen
e e

cyanid
e
Pri nsi p Elec t rosur ger y
Surgical Smoke
Chemical Health Effects

Acetaldehyde Eye, skin, respiratory irritant; causes


erythema, coughing, pulmonary edema; may
be teratogenic

Acrolein Eye, skin, respiratory irritant; liver kidney


damage; increased clotting time

Acetonitrile Nose irritant, throat asphyxiant; liver and


kidney damage in animal models

Benzene Headache, fatigue, weakness; bone marrow


damage; trigger for leukemia

Styrene Respiratory irritant; vapors cause nose lining


damage in animal models

Toluene Eye, respiratory irritant; CNS function;


teratogenesis with high inhalation levels

Xylene Respiratory irritant; chronic exposure


associated with changes in blood cells
Pri nsi p Elec t rosur ger y
Surgical Smoke

• “The mutagenic effect created by thermal destruction of 1 g of tissue


is equivalent to that of three or six cigarettes for laser and
electrocautery smoke, respectively.” (W.L. Barrett, Surgical smoke: a
review of the literature, Surgical Endoscopy 17 (2003), pp. 979–
987).

• “The components of surgical smoke have been described as being, at the


very least, a nuisance and, at worst, carcinogenic.” (
http://findarticles.com/p/articles/mi_m0FSL/is_4_87/ai_n25374211/)
Valleylab E2100 Series Reusable
Pencil

With electrode
All Disposable Pencils

Rocker switch with arthroscopic


electrode

Suction Coagulator
Footor button activated

Push button

Rocker switch

All Pencils (Exc Foot activated) packed with blade & quiver
Disposable Active Electrodes – Stainless
steel
E1551-6 Extended blade electrode

E1552-6 Extended needle electrode

E1511 Arthroscopy electrode (Also: E1510, E1512, & E1513)

E1551G Blade (E1551X with Hex lock)

E1552 Needle electrode E1550 Ball electrode


ALSO:
E1504 / E1502 Electrode extensions – (Re-Usable)
Disposable Loop Electrodes – Stainless Steel
(LLETZ)

E1559

E1560

E1561

E1562
Standard EDGE™
Electrodes
E1450-6 & 4 Extended blade electrode

E1452-6 Extended needle electrode


E1452 Needle electrode

E1450G & X + E1475X Blade ( Use with


Accuvac)
Insulated EDGE™
E1455 Electrodes E1455-4

E1455B-4
E1455B

E1455-6

E1465 E1465-4

E1465B E1465B-4

E1465-6
ANATOMY OF
SUCTION
COAGULATORS
Shock Guard/ Bendable, Insulated
Suction Hole Shaft
Coag
Suction
Connection

Cable to Coagulating Tip


Generator
Re-Usable
Electrodes
E1000 Small ball
E1502 Electrode extension

E1001 Blade
E1020 Blade

E1002 Large ball

E1003 Needle

E1004 angled ball

E1005 Loop
Adaptors &
Cables

E0017

E0507-B

E0560 Re-Usable plate cable for E7509 plates


E0560E Plate cable (Erbe ESU’s)
OPTI 2 Solid Laparoscopic
Electrodes
E2770-28 Str Spat

E2771-28 Curved Spat

E2772-28 Wire J Hook

E2773-28 Wire L

Hook

E2774-28 Flat L Hook


OPTI 4 Hollow Laparoscopic
Electrodes
E2780-28 Str Spat E2781R-28ASP (Retractable electrode)

All OPTI 4 Electrodes are also available as


E2781-28 Curved Spat Retractable electrodes

E2782-28 Wire J Hook Electrode Fit

Connection to ESU E2750


Irrigation button
E2783-28 Wire L Hook

E2784-28 Flat L Hook


Irrigation
Aspiration Aspiration button ES Activation
E2786-28 Cylindrical tip
Bipolar
Forceps

E4058CT

E4062CT

E4060CT
FOOTSWITCH

Bi-Polar
E6009

Mono-Polar
E6008B
Pri nsi p Elec t rosur ger y
Surgical Smoke
REFERENCES
1. Ulmer, B. Electrosurgery Self-Study Guide. Boulder, CO: Covidien Energy-Based Devices, 2010, p. 5
2. Based on Covidien product verification report: "Ex vivo Monopolar Thermal Effect Evaluation on the Valleylab™ FT10 " January - February,
2015; RE0064439 Rev A Based on Covidien In-vivo GLP Acute report: "Verification – Report – GLP Acute Animal Lab – LigaSureTM Preclinical
Evaluation of Valleylab™ FT10." May 19, 2015: RE00005503. FT10 CM3 Document
3. Product E7507/E7507DB Product Instructions for Use. “Intended Use” Section, Page 2
4. McHenry J, Dunning J, Wagle K. Valleylab™ FT10 Energy Platform Tissuefect™ Technology. October 2015. Boulder, CO. Medtronic White
Paper, part number: US150755
5. Based on Covidien ex-vivo monopolar report: "Verification - report - Ex-vivo Monopolar Procedural Flow Evaluation on the Valleylab™ FT10
Orion." November, 2014; R0064443 Rev A. FT10 CM3 Document
6. Based on Covidien In-vivo GLP Acute report: "Verification – Report – GLP Acute Animal Lab – LigaSureTM Preclinical Evaluation of Valleylab™
FT10." May 19, 2015: RE00005503. FT10 CM3 Document
7. Jones, et al. Safe Energy Use in the Operating Room. 2012;52:447-468
8. Overbey D, Townsend N, Chapman B, et al. Surgical energy-based device injuries and fatalities reported to the Food and Drug
Administration. J Am Coll Surg. 2015;221(1):197-205
9. Bisinotto, FMB et al. Burns Related to Electrosurgery-Report of Two Cases. Sociedade Brasileira de Anestesiollogia. 2015:3
10. CSA Group: Standards Update Service Z387-13. “Safe Use of Electrosurgical Medical Devices and Systems in Healthcare” January 2013.
11. Alkatout et al. Principles and Safety Measures of Electrosurgery in Laparoscopy. JSLS (2012) 16: 130-139

75 © 2019 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands
are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 8/2019-US181098-[WF2441443]